Owner’s Guide
Decade 1011/1021 Chair ®
85.2635.00
ALPHABETICAL EQUIVALENT TO THE NUMERAL OF THE MONTH MANUFACTURED A
January
B
February
C
March
D
April
R
E
May
F
June
G
July
H
August
I
September
J
October
K
November
L
December
2601 CRESTVIEW DRIVE NEWBERG, OREGON 97132
Designated EU Representative: A-dec Dental U.K Austin House, 11 Liberty Way, Attleborough Fi Nuneaton, Warwickshire, England CV116RZ Tele: (44) 1203-350901
SN: J828287
SERIAL NUMBER
REF
MODEL NUMBER
YEAR MANUFACTURED
Serial Number Identification Decade 1011/1021 Chair Serial Number Location: • Under the seat/toeboard For service information contact your local authorized A-dec dealer. Check with local codes and A.D.A. (Americans with Disabilities Act) Requirements for Installation of this product.
Warranty
Warranty: A-dec warrants all products in this owner's guide against defects in material or workmanship for one year from time of delivery. A-dec's sole obligation under the warranty is to provide parts for the repair, or at its option, to provide the replacement product (excluding labor). The buyer shall have no other remedy. All special, incidental, and coincidental damages are excluded. Written notice of breach of warranty must be given to A-dec within the warranty period. The warranty does not cover damage resulting from improper installation or maintenance, accident or misuse. The warranty does not cover damage resulting from the use of cleaning, disinfecting or sterilization chemicals and processes. The warranty also does not cover light bulbs. Failure to follow instructions provided in the A-dec owner's guide (operation and maintenance instructions) may void the warranty. A-dec warrants A-dec dental chair cylinders, both lift and tilt, for ten years from the date of purchase of the chair or the cylinder. This warranty is retroactive to A-dec chair cylinders already in the field. The warranty covers chair cylinders A-dec finds to have manufacturing related irregularities. Stool cylinders are covered under A-dec's one-year warranty. NO OTHER WARRANTIES AS TO MERCHANTABILITY OR OTHERWISE ARE MADE.
Decade 1011/1021 Chair Return Policy: U.S. and Canadian dealers wishing to return overstock (unopened) merchandise to A-dec for credit consideration must include a copy of the original invoice number. A return authorization form from an A-dec Territory Manager must be included with serial numbered equipment or A-dec/W&H handpieces. A 15% restocking fee will be assessed. Merchandise that cannot be returned for credit includes parts assembled to the dental unit, chair, light, or dental furniture; obsolete parts; and specials. Dental furniture cannot be returned for credit. Standard color upholstery ordered for obsolete chairs or stools cannot be returned for credit. In the case of a defective warranty item, a copy of the replacement invoice, serial number of the unit under which it was replaced, and a description of the symptoms of the defect must be returned with the part to: A-dec Inc. 2601 Crestview Drive, Newberg, Oregon 97132, USA.
Decade 1011/1021 Chair Equipment Alterations Policy: Certain modifications or alterations of A-dec equipment which expand the use of A-dec equipment beyond its design and intent, or which override any safety features of A-dec equipment may jeopardize doctor, patient or staff safety. Field modifications that alter the electrical and/or mechanical safety of A-dec dental devices are in conflict with Underwriters Laboratory (UL) construction file requirements and are not sanctioned by A-dec. Examples of field modifications that diminish safety design include, but are not limited to, rendering access to the line voltage without the use of tools, modification of supporting elements that increase or shift loading characteristics, and the addition of any powered device that exceeds the design limits of the dental system. The use of accessory equipment not complying with the equivalent safety requirements of A-dec equipment may lead to a reduced level of safety of the resulting system. It is the responsibility of the equipment distributor and the installer to assure that the installation complies with all building code requirements. The responsibility to determine whether a modification or alteration of A-dec equipment falls within these constraints is with the person(s) who initiates, approves and/or performs such modification or alteration. A-dec will not respond to inquiries on an individual basis. This person(s) will be deemed to have assumed all associated risks with such alteration or modification and will hold A-dec harmless from resulting claims, including product liability claims. Additionally, such modification or alteration voids A-dec's warranty and may invalidate UL or other regulatory agency approval.
All product names used in this document are trademarks or registered trademarks of their respective holders.
Printed in U.S.A. • Copyright © 2006 • All Rights Reserved
v
Decade 1011/1021 Chair
Decade 1021 Chair
Decade 1011 Chair
Decade 1011/1021 Chair
CONTENTS Serial number location, service information, and warranty information are located on the inside front cover and front page. About Your Decade Chair ... 2 Chair LED ... 2 Chair Stop Plate... 2 8-Function Footswitch ... 3 Touch Pad ... 3 Programming the Chair ...5 Folding Armrest ... 7 Swivel Brake ... 8 Double-Articulating Headrest ... 9 Positioning For Wheel Chair ... 10 Headrest Drift Adjustment ... 11 Left/Right Conversion ... 12 Care Instructions ... 14 Upholstery Replacement ... 15 Backrest Upholstery ... 15 Seat Upholstery ... 16 Armrest Upholstery ... 17 Double-Articulating Headrest Upholstery ... 18 Maintenance ... 19 Adjustments and Specifications ... 19 Transport the Decade dental unit ... 20 Identification of Symbols ... 21 Classification of Equipment (EN 60601-1) ... 21 1
Decade 1011/1021 Chair
About Your Decade Chair Your Decade chair is an electronically controlled, hydraulically powered dental chair. Chair functions are controlled by the 8-function footswitch or touch pad (see Figure 2 or 2a on
CHAIR BACK / TOEBOARD TILT CHAIR STOP PLATE
BASE LIFT
LED
Figure 1. Lift, Tilt, and Stop Plate
Chair LED page 3). The chair LED indicates the status of the chair: ON: Normal operation. SLOW BLINK: The cuspidor or stop plate limit
Chair Stop Plate switches have been activated. Remove any obstructing object. The chair stop plate (see Figure 1) stops the chair immediately when any part of it is pressed. Should anything inadvertently become lodged under the chair, press Base UP on the footswitch or touch pad 2
Decade 1011/1021 Chair
8-Function Footswitch to raise the chair so the object can be removed. As long as pressure is applied to the stop plate, the chair base will not go down any further. The 8-function footswitch (see Figure 2) or the chair touch pad (see Figure 2a) gives you both manual and programmed control of chair PROGRAM BUTTON PROGRAMMABLE POSITION (1) PROGRAMMABLE ENTRY/EXIT (0)
PROGRAMMABLE POSITION (2)
2
1
3
0
CUSPIDOR/R ETURN (3) BASE UP
BACK DOWN
BACK UP
BASE DOWN
Figure 2. 8-Function Footswitch
Touch Pad PROGRAM BUTTON
PROGRAMMABLE POSITION (2)
PROGRAMMABLE POSITION (1) PROGRAMMABLE ENTRY/EXIT (0)
1 0
2 3
CUSPIDOR/R ETURN (3) BASE UP
BACK DOWN
BACK UP
BASE DOWN
Figure 2a. Chair Touch Pad
3
Decade 1011/1021 Chair positioning. The arrows on the footswitch and touch pad manually control chair back/toeboard tilt and chair lift. The numbered buttons are for Entry/ Exit and programmed chair positions. Manual Control The Base Up/Down (lift) function controls the chair’s lift, or vertical movement. To raise the chair, push the up arrow on the footswitch or touch pad. To lower the chair, push the down arrow on the footswitch or touch pad. Push the button until the chair reaches the desired height, then release it. The Back Up/Down (tilt) function controls the chair back/toeboard tilt. To raise the chair back, push the right arrow on the footswitch or touch pad. To lower the chair back, push the left arrow on the footswitch or touch pad. Push the button until the chair back reaches the desired position, then release it. Program Button The program button (located on the top middle of the footswitch, or in between the arrows on the touch pad) is used to save the settings for Entry/ Exit (0), programmable positions (1 and 2), and Cuspidor/Return (3).
4
Decade 1011/1021 Chair
Programming the Chair NOTE When 1 or 2 are pressed on the footswitch or touch pad, the chair base and back go to the programmed position. To stop the chair at any point, press any button on the footswitch (or press any button on the touch pad). Programmable Positions 1 and 2 To send the chair to a programmed operating position, push either 1 or 2 on the footswitch (press 1 or 2 on the touch pad). Positions 1 and 2 are programmed at the factory to move the chair to the same position. To change a programmable position, find the program button on the footswitch or touch pad (see Figures 2 or 2a on page 3). 1. Using the manual arrows on the footswitch or touch pad, set the chair to the operating position that you prefer. 2. Press and release the program button. An audible tone will be emitted. Then, within four seconds, push the button for 1 or 2 to store that position. You will hear an audible tone confirming that the programmable position function has been reprogrammed. 3. Check the programmed position by manually moving the chair to another position. Then push and release the button programmed in Step 2. The chair should move automatically to the position set in Step 1.
5
Decade 1011/1021 Chair Optional Program Functions Position 3 is factory set in the Cuspidor/Return mode. In this mode, the chair back will rise to a pre-programmed upright position providing the patient access to the cuspidor. Pressing position 3 a second time lowers the chair back to its previous operating position. Position3 may also be used as a third Pre-Position or as a last position recall. Contact an authorized A-dec Dealer to have Position 3 reconfigured to a third Pre-Postion or as a last position recall.
6
Decade 1011/1021 Chair Entry/ Exit (0) To send the chair to a preset entry/exit position, push the position 0 button (see Figure 2 or 2a on page 3).
NOTE When pushed, Entry/ Exit (0) will cause the chair base and back to go to the preset entry/exit position. To stop the chair at any point, push any button on the footswitch or touch pad. If you want to change the preset entry/exit position, first locate the program button on the footswitch or touch pad (see Figure 2 or 2a on page 3). 1. Using the manual arrows, set the chair to the desired patient entry/exit position. 2. Press and release the program button. An audible tone will be emitted. Within four seconds, press the Entry/ Exit (0) button on the footswitch or touch pad. An audible tone will confirm that the chair has been reprogrammed. 3. Check the Entry/ Exit (0) function by manually moving the chair to another position. Push the Entry/ Exit (0) button. The chair will move to the position you set in Step 1.
Folding Armrest The armrests fold down and out of the way to facilitate patient entry/exit. When lifted, the armrests lock securely in place for patient comfort during procedures. To lower an armrest, press the button on the side of the arm. Lock the armrest by lifting it until it locks in place.
7
Decade 1011/1021 Chair
BUTTON
Figure 4. Folding Armrest
Swivel Brake LOCK
SWIVEL BRAKE LEVER
RELEASE
Figure 5. Swivel Brake
When engaged, the chair swivel brake restricts rotation of the chair. With the brake released, you can rotate the chair to any position within approximately 30° either side of center. To release the swivel brake, push the brake lever to the right. To engage the swivel brake, push the brake lever to the left. If the chair swivels left or right with the brake engaged, or if it is difficult to move with the brake disengaged, the swivel brake tension must be adjusted. Using a 3¼16-inch hex key, adjust the swivel brake tension. Turn the adjusting screw clockwise to increase brake friction. Turn the screw counterclockwise to decrease brake friction. 8
Decade 1011/1021 Chair
Double-Articulating Headrest
Figure 6. Double-Articulating Headrest
The double-articulating headrest offers the widest range of headrest positions possible. A convenient locking knob allows you to easily accommodate most patients and procedures. To position the headrest, release the locking knob by turning the knob out (counterclockwise), then adjust the headrest as necessary to fit the head and neck. Lock the headrest in the desired position by turning the knob in (clockwise). To move the headrest higher or lower, simply pull up or push down on the headrest until it is at the desired height. (Refer to for Headrest Drift Adjustment on page 12.)
9
Decade 1011/1021 Chair
Positioning For Wheel Chair
Figure 7. Positioning For Wheel Chair Usage
The headrest can be used to accommodate wheelchair patients. Slide the headrest up until it is free from the chair, turn it 180°, then slide it back into the backrest and push it all the way down. Run the chair to its full Back Up position. Adjust headrest height by moving the chair up or down (using the Base Up function on the footswitch), then position the headrest as desired.
10
Decade 1011/1021 Chair
Headrest Drift Adjustment
HEADREST GLIDE BAR
Figure 8. Headrest Drift Adjustment
NOTE The back cover does not have to be removed to adjust glide bar tension. If the headrest drifts downward, or if you find the headrest difficult to move up or down, the glide bar tension must be adjusted (see Figure 8). Remove the headrest glide bar to access the adjusting screw. Using a Phillips screwdriver, adjust the glide bar tension. Turn the adjusting screw clockwise three to four revolutions to increase friction and hold the headrest more securely. Turn the screw counterclockwise three to four revolutions to decrease friction and allow the headrest to move up and down more freely. Reinstall the headrest and recheck glide bar tension. 11
Decade 1011/1021 Chair
Left/Right Conversion When combined with a conversion-compatible A-dec chair-mounted handpiece control system, your Decade chair allows you to convert for either left- or right-hand delivery. If you have a Radius® Delivery System, go to page 15. 1. Raise the chair back, then disconnect the chair from the power source to prevent injury caused by accidental lowering of the chair. 2. Remove the chair swivel brake cover. 3. Remove the upholstery pin on both sides of the chair casting to remove the seat upholstery (see Figure 9).
UPHOLSTERY PIN
REMOVE SCREWS
Figure 9. Seat Upholstery Removal
4. Place the armrests in their full down position. 5. Move the master on/off toggle to the OFF position. If the unit is equipped with a selfcontained water system, use the syringe to bleed the system of air and water pressure, then remove the bottle. 6. Position the control head over the post box and the assistant’s arm in front of the chair. 12
Decade 1011/1021 Chair 7. Unscrew and remove the hex-head bolt from the chair adapter (see Figure 10). Pivot the chair adapter post and the post box to the other side of the chair. Align the hole in the chair casting with the hole in the adapter post. 8. Reinstall the hex-head bolt in the chair adapter and tighten it firmly. 9. Position the post box: POST BOX
HEX BOLT
CHAIR ADAPTER
Figure 10. Left/Right Chair-Mount Conversion
NOTE Avoid rotating the post box in a complete circle. Doing so will twist the lines inside the floor box.
Cascade Delivery System Remove the post box covers. Use a 15¼16-inch wrench to loosen the bolt located in the center of the post box frame. Rotate the post box to the desired position. Then tighten the bolt and replace the covers. 13
Decade 1011/1021 Chair 10. Replace the self-contained water bottle. 11. Plug in the chair, and turn the system on. Check the post box connections for leakage or pinched tubing. 12. Return the chair to the entry position and move the foot control to the other side of the chair.
Radius Delivery System Simply rotate the handpiece control and light arms to the desired position.
Care Instructions The unique Radius bearing mount allows the delivery system to rotate around the toeboard in one continuous motion, eliminating the need for tools or fine adjustments. For surface cleaning and disinfection instructions, refer to your Equipment Asepsis Owner’s Guide, A-dec Publication No. 85.0696.00.