Owner’s Guide
CASCADE 1040 Chair ®
85.2605.00
85260500.QXD
3/13/06
9:53 AM
Page ii
ALPHABETICAL EQUIVALENT TO THE NUMERAL OF THE MONTH MANUFACTURED A
January
B
February
C
March
D
April
E
May
F
June
G
July
H
August
I
September
J
October
K
November
L
December
2601 CRESTVIEW DRIVE NEWBERG, OREGON 97132 USA Designated EU Representative: A-dec Dental U.K., Ltd. Austin House, 11 Liberty Way, Attleborough Fields, Nuneaton, Warwickshire, England CV116RZ Tele: (44) 24 7635 0901
SN: J828287
REF: 2122
SERIAL NUMBER
MODEL NUMBER
MADE IN USA
1999
YEAR MANUFACTURED
Serial Number Identification Cascade Chair Serial Number Location: • On the upper structure, under the upholstery
For service information contact your local authorized A-dec dealer. Check with local codes and A.D.A. (Americans with Disabilities Act) Requirements for Installation of this product.
85260500.QXD
3/13/06
9:53 AM
Page iii
Warranty
A-dec warrants all products in this catalog against defects in material or workmanship for one year from time of delivery. A-dec’s sole obligation under the warranty is to provide parts for the repair, or at its option, to provide the replacement product (excluding labor). The buyer shall have no other remedy. All special, incidental, and coincidental damages are excluded. Written notice of breach of warranty must be given to A-dec within the warranty period. The warranty does not cover damage resulting from improper installation or maintenance, accident or misuse. The warranty does not cover damage resulting from the use of cleaning, disinfecting or sterilization chemicals and processes. The warranty also does not cover light bulbs. Failure to follow instructions provided in the A-dec owner’s guide (operation and maintenance instructions) may void the warranty. A-dec warrants A-dec dental chair cylinders, both lift and tilt, for ten years from the date of purchase of the chair or the cylinder. This warranty is retroactive to A-dec chair cylinders already in the field. The warranty covers chair cylinders A-dec finds to have manufacturing related irregularities. Stool cylinders are covered under A-dec‘s one-year warranty. No other warranties as to merchantability or otherwise are made.
All product names used in this document are trademarks or registered trademarks of their respective holders.
Printed in U.S.A. • Copyright © 2006 • All Rights Reserved
85260500.QXD
3/13/06
9:53 AM
Page 1
Cascade 1040 Chair
CONTENTS Serial number location, service information, and warranty information are located on the inside front cover and front page. About Your Cascade Chair ... 2 Chair LED... 2 Chair Stop Plate ... 2 8-Function Footswitch ... 3 Touch Pad ... 3 Programming the Chair ...5 Swivel Brake ... 8 Double-Articulating Headrest ... 9 Headrest Positioning for Wheelchair ... 10 Headrest Drift Adjustment ... 11 Upholstery Replacement ... 12 Back Upholstery ... 12 Double-Articulating Headrest Upholstery ... 13 Seat Upholstery ... 14 Armrest Upholstery ... 15 Care Instructions ... 16 Adjustments and Specifications ... 17 Maintenance ... 17 Safety Considerations for Accessory Equipment ... 18 Transporting the Dental Unit ... 18 Identification of Symbols ... 19 Classification of Equipment (EN 60601-1) ... 19
1
85260500.QXD
3/13/06
9:53 AM
Page 2
Cascade 1040 Chair
About Your Cascade Chair Your Cascade chair is an electronically controlled, hydraulically powered dental chair (see Figure 1). Chair functions are controlled by the 8-function footswitch or touch pad (see Figure 2 or 2a on page 3).
CHAIR BACK / TOEBOARD TILT BASE LIFT CHAIR STOP PLATE
LED
Figure 1. Lift, Tilt, and Stop Plate
Chair LED The chair LED indicates the status of the chair: ON: Normal operation. SLOW BLINK: The cuspidor or stop plate limit switches have been activated. Remove any obstructing object.
Chair Stop Plate The chair stop plate (see Figure 1) stops the chair immediately when any part of it is pressed. Should anything inadvertently become lodged under the chair, press Base UP on the footswitch or touch pad to raise the chair so the object can be removed. As long as pressure is applied to the stop plate, the chair base will not go down any further. 2
85260500.QXD
3/13/06
9:53 AM
Page 3
Cascade 1040 Chair
8-Function Footswitch The 8-function footswitch (see Figure 2) or the chair touch pad (see Figure 2a) gives you both manual and programmed control of chair positioning. The arrows on the footswitch and touch pad manually control chair back/toeboard tilt and chair lift. The numbered buttons are for Entry/ Exit and programmed chair positions. PROGRAM BUTTON PROGRAMMABLE POSITION (1) PROGRAMMABLE ENTRY/EXIT (0)
PROGRAMMABLE POSITION (2)
2
1
3
0
CUSPIDOR/RETURN (3)
BASE UP BACK DOWN
BACK UP
BASE DOWN
Figure 2. 8-Function Footswitch
Touch Pad
PROGRAMMABLE POSITION (2)
PROGRAMMABLE POSITION (1) PROGRAMMABLE ENTRY/EXIT (0)
1 0
CUSPIDOR/RETURN (3)
2 3
BASE UP BACK DOWN
BACK UP PROGRAM BUTTON BASE DOWN
Figure 2a. Chair Touch Pad
3
85260500.QXD
3/13/06
9:53 AM
Page 4
Cascade 1040 Chair Manual Controls The Base Up/Down (lift) function controls chair lift, or vertical movement. To raise the chair, push the up arrow on the footswitch or touch pad. To lower the chair, push the down button on the footswitch or touch pad. Push the button until the chair reaches the desired height, then release it. The Back Up/Down (tilt) function controls the chair back/toeboard tilt. To raise the chair back, push the right arrow on the footswitch or touch pad. To lower the chair back, push the left arrow on the footswitch or touch pad. Push the button until the chair back reaches the desired position, then release it. Program Button The Program Button (located on the top middle of the footswitch, or in between the arrows on the touch pad) is used to save the settings for Entry/ Exit (0), programmable positions (1 and 2), and Cuspidor/Return (3).
4
85260500.QXD
3/13/06
9:53 AM
Page 5
Cascade 1040 Chair
Programming the Chair Programmable Positions 1 and 2
NOTE When 1 or 2 are pressed on the footswitch or touch pad, the chair base and back go to the preset position. To stop the chair at any point, press any button on the footswitch (or press any button on the touch pad).
To send the chair to a programmed operating position, push either 1 or 2 (or press 1 or 2 on the touch pad). Positions 1 and 2 are programmed at the factory to move the chair to the same position. To change a programmable position, locate the program button on the footswitch or touch pad (see Figures 2 or 2a on page 4). 1. Using the manual arrows on the footswitch or touch pad, set the chair to the operating position that you prefer. 2. Press and release the program button. An audible tone will be emitted. Then, within four seconds, push the button for 1or 2 to store that position. You will hear an audible tone confirming that the programmable function has been reprogrammed. 3. Check the programmed position by manually moving the chair to another position. Then push either 1 or 2 programmed in Step 2. The chair should move automatically to the position set in Step 1.
5
85260500.QXD
3/13/06
9:53 AM
Page 6
Cascade 1040 Chair
Optional Program Functions Position 3 is factory set in the Cuspidor/Return mode. In this mode, the chair back will rise to a pre-programmed upright position providing the patient access to the cuspidor. Pressing position 3 a second time lowers the chair back to its previous operating position. Position 3 may also be used as a third Pre-Position or as a last recall. Contact an authorized A-dec Dealer to have Position 3 reconfigured to a third Pre-Position or as a last position recall.
6
85260500.QXD
3/13/06
9:53 AM
Page 7
Cascade 1040 Chair
Entry/ Exit (0) To send the chair to a preset entry/exit position, push the position 0 button (see Figure 2 or 2a on page 3).
NOTE When pushed, Entry/ Exit (0) will cause the chair base and back to go to the preset entry/exit position. To stop the chair at any point, push any button on the footswitch or touch pad. If you want to change the preset entry/exit position, first locate the program button on the footswitch or touch pad (see Figure 2 or 2a on page 3). 1. Using the manual arrows, set the chair to the desired patient entry/exit position. 2. Press and release the program button. An audible tone will be emitted. Then, within four seconds, press the Entry/Exit (0) button on the footswitch or touch pad. You will hear an audible tone confirming that the chair has been reprogrammed. This stores your preferred patient entry/exit position. 3. Check the Entry/ Exit (0) function by manually moving the chair to another position. Push the Entry/ Exit (0) button. The chair should move automatically to the position you set in Step 1.
7
85260500.QXD
3/13/06
9:53 AM
Page 8
Cascade 1040 Chair
Swivel Brake When engaged, the chair swivel brake restricts rotation of the chair. With the brake released, you can rotate the chair to any position within approximately 30° either side of center. To unlock the swivel brake, push the brake lever to the right. To lock the swivel brake, push the brake lever to the left. ADJUSTING SCREW
LOCKED
UNLOCKED
Figure 4. Swivel Brake Adjustment
If the chair swivels left or right with the brake engaged, or if it is difficult to move with the brake disengaged, the swivel brake tension must be adjusted. Using a 3⁄16-inch hex key, adjust the swivel brake tension. Turn the adjusting screw clockwise to increase brake friction. Turn the screw counterclockwise to decrease brake friction.
8
85260500.QXD
3/13/06
9:53 AM
Page 9
Cascade 1040 Chair
Double-Articulating Headrest
LOCKING KNOB
Figure 5. Double-Articulating Headrest
The locking knob allows you to easily adjust the headrest for a full range of positions. To position the headrest, release the locking knob by turning the knob out (counterclockwise), then adjust the headrest as necessary to fit the head and neck. Lock the headrest in the desired position by turning the knob in (clockwise). To move the headrest higher or lower, simply pull up or push down on the headrest until it is at the desired height.
9
85260500.QXD
3/13/06
9:53 AM
Page 10
Cascade 1040 Chair
Headrest Positioning for Wheelchair
Figure 6. Headrest Positioning for Wheelchair Usage
The headrest can be used to accommodate wheelchair patients. Slide the headrest up until it is free from the chair, turn it 180°, then slide it back into the backrest and push it all the way down. Run the chair to its full Back Up position. Adjust headrest height by moving the chair up or down (using the Base Up function on the footswitch or touch pad), then position the headrest as desired.
10
85260500.QXD
3/13/06
9:53 AM
Page 11
Cascade 1040 Chair
Headrest Drift Adjustment
HEADREST GLIDE BAR
Figure 7. Headrest Drift Adjustment
If the headrest drifts downward, or if it is difficult to move up or down, the glide bar tension must be adjusted. Remove the headrest glide bar to access the adjusting screw. Using a Phillips screwdriver, adjust the glide bar tension. Turn the adjusting screw clockwise three to four revolutions to increase friction and hold the headrest more securely. Turn the screw counterclockwise three to four revolutions to decrease friction and allow the headrest to move up and down more freely. Reinstall the headrest and recheck glide bar tension.
11
85260500.QXD
3/13/06
9:53 AM
Page 12
Cascade 1040 Chair
Upholstery Replacement A-dec’s unique formed upholstery makes replacing upholstery quick and easy. The upholstery on your Cascade 1040 Chair is installed in four sections: back, headrest, seat, and armrests. Each section is easily removed and replaced.
Back Upholstery KEY SLOT IN BACK UPHOLSTERY
LARGE HEAD FASTENER
Figure 8. Replacing the Back Upholstery
To remove back upholstery firmly grasp the bottom edge of the cushion and lift upward, approximately 1-inch, to release the four large, flat-headed fasteners from the cushion key-slots. To install back upholstery place the large fastener heads in the cushion keyhole pockets then push in and down in one motion.
12
85260500.QXD
3/13/06
9:53 AM
Page 13
Cascade 1040 Chair
Double-Articulating Headrest Upholstery
MOUNTING SCREWS (one hidden)
HEADREST BACK-PLATE
TOP MOUNTING SCREW (hidden)
Figure 9. Replacing the Headrest Upholstery
Use the following procedure to replace the headrest upholstery. 1. Loosen the headrest knob and rotate the headrest to a full upright position. 2. Remove the top Phillips mounting screw from the headrest back-plate, located just above the knob assembly. 3. Rotate the headrest back 45° to expose the two Phillips mounting screws on the lower headrest back-plate (see Figure 9). Remove the two screws and the headrest cushion. 4. To attach the replacement upholstery, reverse this procedure.
13
85260500.QXD
3/13/06
9:53 AM
Page 14
Cascade 1040 Chair
Seat Upholstery SEAT UPHOLSTERY
SCREWS
3/16" HEX KEY
SEAT FRAME
Figure 10. Replacing the Seat Upholstery
Use the following procedure to replace the seat upholstery. 1. Lower the chair back, lift the toeboard, and hold it by the frame. 2. Detach and remove the clear vinyl toeboard cover. 3. Using a 3/16-inch hex key, loosen the four screws that secure the seat/toeboard upholstery assembly to the seat frame. 4. Lower the toeboard and lift off the seat/toeboard upholstery assembly. 5. Place the new seat/toeboard upholstery assembly on the seat frame. 6. Lift the toeboard and tighten the four screws, then lower the toeboard. 7. Attach the new clear vinyl toeboard cover.
14
85260500.QXD
3/13/06
9:53 AM
Page 15
Cascade 1040 Chair
Armrest Upholstery ARMREST UPHOLSTERY ASSEMBLY
RETAINING SCREW
Figure 11. Replacing the Armrest Upholstery
Use the following procedure to replace the armrest upholstery assembly. 1. Lift the armrest away from the seat pocket. Use a Phillips screwdriver and remove the retaining screw from the underside of the armrest. 2. Push the armrest upholstery assembly toward the toe end of the chair, approximately 1⁄2-inch, then lift the armrest upholstery away from the arm. 3. Place the new armrest on the arm and slide it toward the chair back. Replace the retaining screw on underside.
15
85260500.QXD
3/13/06
9:53 AM
Page 16
Cascade 1040 Chair
Care Instructions For recommended asepsis instructions, refer to the Asepsis section of the Owner’s Guide (A-dec Publication No. 85.0696.00).
16
85260500.QXD
3/13/06
9:53 AM
Page 17
Cascade 1040 Chair
Adjustments and Specifications Programming the Chair ... page 5 Swivel Brake ... page 8 Double-Articulating Headrest ... page 9 Headrest Drift Adjustment ... page 11 Chair Capacity: Patient Load: 300 lbs. (135 kg) maximum. Accessory Load: 150 lbs. (67.5 kg) maximum. Maximum Load from Accessories: Chair Adapter: 200 lbs. (90 kg) weight combined with a moment of 250 ft. lbs. (339 N·m) Radius Front Mount: 180 lbs. (81 kg) weight combined with a moment of 525 ft. lbs. (711.8 N·m) Radius Rear Mount: 70 lbs. (31.5 kg) weight combined with a moment of 160 ft. lbs. (216.9 N·m)
Maintenance Upholstery Replacement ... page 12 Back Upholstery ... page 12 Double-Articulating Headrest Upholstery ... page 13 Seat Upholstery ... page 14 Armrest Upholstery... page 15 Care Instructions ... page 16
Equipment Asepsis Owner’s Guide...85.0696.00 Maintenance Parts Owner’s Guide ...85.2634.00 Specifications are subject to change without notice.
17