a-dec
371 Dental Light Instructions for Use Rev A Dec 2008
Instructions for Use
24 Pages
Preview
Page 1
INSTRUCTIONS FOR USE A-dec® 371 Dental Light
A-DEC 371 DENTAL LIGHT
INSTRUCTIONS FOR USE
Copyright © 2008 A-dec Inc. All rights reserved. A-dec Inc. makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. A-dec Inc. shall not be held liable for any errors contained herein or any consequential or other damages concerning the furnishing, performance or use of this material. The information in this document is subject to change without notice. If you find any problems in the documentation, please report them to us in writing. A-dec Inc. does not warrant that this document is error-free. No part of this document may be copied, reproduced, altered, or transmitted in any form or by any means, electronic or mechanical, including photocopying, recording, or by any information storage and retrieval system, without prior written permission from A-dec Inc. This document is intended to be used in conjunction with the equipment.
Trademarks and Additional Intellectual Property Rights A-dec, the A-dec logo, A-dec 500, Cascade, Cascade Master Series, Century Plus, Continental, Decade, ICX, ICV, Performer, Preference, Preference Collection, Preference ICC, and Radius are registered trademarks in the U.S. Patent and Trademarks office. Preference Slimline and A-dec 300 are also trademarks of A-dec Inc. Certain touchpad symbols are proprietary to A-dec Inc. Any use of these symbols, in whole or in part, without the express written consent of A-dec Inc. is strictly prohibited.
Comments and Feedback If you have any feedback or comments about this document, contact us at: A-dec Inc. Technical Communications Department 2601 Crestview Drive Newberg OR 97132 USA Tel: 1.800.547.1883 Within USA/Canada Tel: 1.503.538.7478 Outside USA/Canada E-mail: [email protected] Website: www.a-dec.com
Product Service For service information, contact your local authorized A-dec dealer. To find your local dealer, go to www.a-dec.com.
A-dec 371 Dental Light Instructions For Use
CONTENTS 1 Introduction ... 1 About Your Dental Light ... 1 Serial Number ... 3 2 Basic Operation ... 5 Power On/Off Button ... 5 Light Intensities... 6 Auto On/Off Feature ... 6 Circuit Breaker Locations ... 7 3 Adjustments ... 9 Focus Setting... 9 Dental Light Positioning... 9 Rotation Adjustments... 10 Bulb Replacement ... 11 Maintenance Parts ... 13 Care & Asepsis Instructions ... 14 Touch and Transfer Surfaces... 14 Appendix A: Warranty and Specifications ... 15 Warranty ... 15 Specifications ... 16
86.0220.00 Rev A
iii
A-dec 371 Dental Light Instructions For Use
iv
86.0220.00 Rev A
A-dec 371 Dental Light Instructions For Use
INTRODUCTION
1
Thank you for choosing A-dec equipment. This guide contains basic operating instructions, adjustments, and maintenance for your A-dec 371 dental light.
About Your Dental Light The A-dec 371 dental light includes a number of features designed to make your job easier. Balanced light gives a clear view of tooth surfaces, the precision-balanced arm allows for smooth positioning, and touchpad controls make operation convenient for you and your team. Figure 1. A-dec 371 Dental Light
86.0220.00 Rev A
1
A-dec 371 Dental Light Instructions For Use
Figure 2. A-dec 371 Dental Light Features
C
B
A
D
E
2
Item
Description
A
Light Handle
B
Pivot Housing
C
Flexarm End Cap and Spare Bulb Holder (inside)
D
Bulb Cap Assembly
E
Light Shield
86.0220.00 Rev A
Introduction
Serial Number Whenever inquiring about service for the A-dec 371 dental light, please provide the serial number. The serial number is located under the flexarm (see Figure 3). Figure 3. Serial Number Location
Figure 4. Serial Number Label Example
A B
(A) The REF number is the model number. (B) The first letter of the serial number indicates the month the product was manufactured. The first digit of the serial number is the year of manufacture (for example, L8 = December 2008). Table 1. Key to Month of Manufacture
86.0220.00 Rev A
Letter
Month
Letter
Month
A
January
G
July
B
February
H
August
C
March
I
September
D
April
J
October
E
May
K
November
F
June
L
December
3
A-dec 371 Dental Light Instructions For Use
4
86.0220.00 Rev A
A-dec 371 Dental Light Instructions For Use
2
B A S I C O P E RA T I O N Light operations are controlled through the touchpad.
Power On/Off Button The A-dec 371 dental light can be operated from the assistant-side or doctor-side touchpad. To turn the light on, press the dental light button on the touchpad (see Figure 5). To turn the light off, press and hold the light button. Figure 5. Dental Light Touchpad Controls A-dec 300TM Standard and Deluxe Touchpads F
A A
p
E E D
B C
B
C
D
A) Dental Light Button
D) X-Ray/Rinse Position Button
B) Entry/Exit Button
E) Program Button
C) Treatment Position Button(s)
F) Dental Light Indicator (LED)
Note: Touchpad symbols are proprietary to A-dec Inc.
86.0220.00 Rev A
5
A-dec 371 Dental Light Instructions For Use
Light Intensities The A-dec 371 dental light features two light intensities: high and composite. After turning the light on, press the dental light button to toggle between the two intensity settings. The LED indicator on the touchpad illuminates when the dental light is in the high intensity setting. When the dental light is in the composite setting, the LED indicator flashes.
Auto On/Off Feature The dental light includes a convenient auto on/off feature. Press any button that moves the chair into an operating position (see Figure 5), and the dental light turns on when the chair reaches that position. Press the Entry/Exit or X-ray/Rinse buttons and the dental light automatically turns off. To activate/deactivate the auto on/off feature: • Press and hold the program and dental light buttons simultaneously for three seconds. One beep confirms the auto on/off is off. • Press and hold the program and dental light buttons simultaneously for three seconds. Three beeps confirm that the auto on/off is on. For further instructions on programming the touchpad, please refer to the A-dec 300 Systems Instructions For Use (P/N 86.0092.00). NOTE
6
86.0220.00 Rev A
Basic Operation
Circuit Breaker Locations Under abnormal conditions, the circuit breaker interrupts the flow of electricity. If the circuit breaker should trip, reset it by pushing the circuit breaker. The circuit breaker for your dental light is located on the power supply, whose location varies by chair (see Figure 6). Figure 6. Circuit Breaker Locations
86.0220.00 Rev A
Chair Model
Power Supply Location
A-dec 311
Chair motor pump area
A-dec 500®
Chair motor pump area
Cascade,® Decade,® Performer,® and non-A-dec chairs
Utility floor box
Circuit Breaker Location
10 A M P
10 A M P
7
A-dec 371 Dental Light Instructions For Use
8
86.0220.00 Rev A
A-dec 371 Dental Light Instructions For Use
ADJUSTMENTS
3
Focus Setting The A-dec 371 dental light is factory set for optimum illumination between 12 - 27 inches (300 - 700 mm) from the oral cavity. This setting is fixed and cannot be adjusted.
Dental Light Positioning Properly adjusted, your dental light should rotate and pivot to provide smooth positioning (see Figure 7). If the light head is difficult to position, moves too easily, or tends to slip out of position, you can adjust the horizontal and vertical tension (see Figure 8 and Figure 9). Figure 7. Light Head Rotation
86.0220.00 Rev A
9
A-dec 371 Dental Light Instructions For Use
Rotation Adjustments 1. For horizontal rotation, use a hex key to turn the adjustment screws, beginning with the screw at the top of the switch housing (see Figure 8). 2. If the light head moves too easily, or tends to drift out of position, increase the tension by turning the screws right. 3. If the light is difficult to move, loosen the tension by turning the screws left. Use the same procedure to adjust vertical rotation (see Figure 9).
Figure 8. Horizontal Rotation Adjustments
A
(A) Horizontal Rotation Tension Screws Figure 9. Vertical Rotation Adjustments
A
(A) Vertical Tension Screw
10
86.0220.00 Rev A
Adjustments
Bulb Replacement A spare bulb is attached to the inside of the light flexarm end cap. To remove the end cap, pull from the bottom first, then straight out, being careful not to damage the bulb (see Figure 10). Figure 10. Remove Flexarm End Cap A
(A) Flexarm End Cap
For your convenience, the spare bulb part number is etched on the inside of the end cap (see Figure 11). Use this number when ordering replacement bulbs. Figure 11. End Cap With Spare Bulb A
B
(A) Bulb in Wrapper; (B) Replacement Bulb Part Number (041.709.00)
86.0220.00 Rev A
11
A-dec 371 Dental Light Instructions For Use
CAUTION Do not remove the outer wrapper when removing the bulb from the end cap. Finger oils can affect light performance and severely limit bulb life. If you should inadvertently touch the bulb, gently clean it with cotton dampened with isopropyl or ethyl alcohol.
To avoid burning your fingers, allow the bulb to cool before removing. Never operate the light with the light shield removed. The clear shield contains UV blocking additives. The light shield is also your protection in the unlikely event that the bulb shatters. WARNING
Follow these steps when replacing the bulb: 1. Turn OFF the light and allow it to cool. 2. Gently pull the light shield tab out on one side while gripping the top portion of the shield to remove, then set it aside (see Figure 12). Figure 12. Remove Light Shield
A
C
B
(A) Tabs; (B) Bulb Cap; (C) Shield
3. Remove the bulb cap assembly by rotating the cap assembly counter clockwise, then set aside. 4. Using a gauze pad or cloth to protect your fingers, carefully pull the old bulb from its socket. Discard the bulb.
12
86.0220.00 Rev A
Adjustments
5. Holding the new bulb in its outer wrapper with the pins away from you, carefully insert the bulb pins into the socket (see Figure 13). The bulb base is fragile and can break under excess pressure. Figure 13. Insert New Bulb
A
B
(A) Outer Wrapper; (B) New Bulb
6. Remove and discard the outer wrapper, then reinstall the bulb cap assembly and light shield. 7. Verify the operation of the light by turning it on and operating it at each intensity setting.
Maintenance Parts Table 2. Maintenance Parts List
86.0220.00 Rev A
Description
Part Number
Bulb
041.709.00
Bulb Cap Assembly
28.1635.00
Shield
26.1634.00
13
A-dec 371 Dental Light Instructions For Use
Care & Asepsis Instructions Touch and Transfer Surfaces Touch surfaces are those areas that require contact and become potential crosscontamination points during dental procedures. The minimum touch surface locations with A-dec 371 lights include touchpads and the dental light handle. Transfer surfaces are those surfaces that are contaminated by contact with instruments and other inanimate objects. There are no transfer surfaces on the A-dec 371 dental light. A-dec recommends barrier protection for all applicable touch and transfer surfaces. When used, barriers must be FDA market-cleared barrier plastic. Cover-All™ barrier film or any other FDA market-cleared barrier film is suitable for this application. Refer to your national regulatory authorities for barrier recommendations specific to your locale. Barrier plastics should be removed and discarded after each patient treatment. For recommendations on cleaning and chemical disinfection of touch and transfer surfaces (where barrier protection is not applicable or when barriers are compromised), please refer to the A-dec Equipment Asepsis Guide (P/N 85.0696.00).
14
86.0220.00 Rev A
A-dec 371 Dental Light Instructions For Use
A P P E N D I X A : WA R RA N T Y AND SPECIFICATIONS
A
Warranty A-dec warrants all products against defects in materials or workmanship for one year from time of delivery. A-dec’s sole obligation under the warranty is to provide parts for the repair, or at its option, to provide the replacement product (excluding labor). The buyer shall have no other remedy. All special, incidental, and coincidental damages are excluded. Written notice of breach of warranty must be given to A-dec within the warranty period. The warranty does not cover damage resulting from improper installation or maintenance, accident or misuse. The warranty does not cover damage resulting from the use of cleaning, disinfecting or sterilization chemicals and processes. The warranty also does not cover light bulbs. Failure to follow instructions provided in the A-dec Instructions For Use (operation and maintenance instructions) may void the warranty. A-dec warrants A-dec dental chair cylinders, both lift and tilt for ten years from the date of purchase of the chair or the cylinder. This warranty is retroactive to A-dec chair cylinders already in the field. The warranty covers chair cylinders A-dec finds to have manufacturing irregularities. Stool cylinders are covered under A-dec’s oneyear warranty. No other warranties as to merchantability or otherwise are made. For service information, contact your local authorized A-dec dealer. Check with local codes and ADA (Americans with Disabilities Act) requirements for installation of this product. CAUTION Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner licensed by the law of the state in which he or she practices to use or order the use of this device.
86.0220.00 Rev A
15