INSTALLATION INSTRUCTIONS
Radius® Cuspidor ®
®
(for Cascade 1040 or Decade 1011/1021 Chairs)
You Will Need: • Hex key set • Level • Large flat head screwdriver • Pliers • 1/2" and 9/16" wrench • Cutters Chair footswitch
Before Beginning this Install 1. Install the air and water manual shutoff valves. Refer to the Floor Box Template (A-dec Publication No. 85.0472.00). Purge the air and water supplies of debris.
Figure 1 Touchpad
2. Install the dental chair. Refer to Cascade 1040 chair Installation Instructions (A-dec Publication No. 86.0157.00) or Decade 1011/1021 Chair Installation Instructions (A-dec Publication No. 86.0156.00). WARNING The chair must be anchored to the floor to meet agency requirements. Refer to the Radius Floor Box Installation Instructions (A-dec Publication No. 85.2885.00) or the Radius 2122/2132 Delivery System Installation Instructions (A-dec Publication No. 85.2877.00) for chair anchoring instructions.
Figure 2
ELECTRICAL WARNING 3. If a delivery system is being installed, do so now. Refer to Radius Delivery System Installation Instructions (A-dec Publication No. 85.2877.00).
Failure to disconnect the chair from its power source can result in serious injury.
Raise the Chair • Raise the chair base completely up using the footswitch (see Figure 1) or the touchpad on the control head (see Figure 2). • Unplug the chair. 2006-04 Rev G (05190)
Page 1
86.0104.00
Remove the Chair Covers
Install the Arm Mount
1. Remove the chair stop plate (see Figure 3).
1. Remove the arm mount from the shipping carton. Loosen the bolt almost all the way (see Figure 6).
• If you have a non-Radius chair, discard the stop plate. Go to step 2. • If you have a Radius chair, remove and discard the cover on the stop plate. Go to step 2.
2. Place a small amount of blue Loctite® on the threads of the bolt (see Figure 6). Loctite here (place on bolt threads)
Chair stop plate, non-Radius chair (discard) Bolt (loosen)
Chair stop plate, Radius chair (keep) Stop plate cover, Radius chair (discard)
Figure 3 2. Remove and discard the chair upper lift arm cover (see Figure 4).
Arm mount
Figure 6 3. Move the brake handle to the right. Place the arm mount onto the chair swivel casting. Center the arm mount within the grooves of the chair swivel casting and under the lip of the swivel casting (see Figure 7). Be sure that the limit switch cable is beneath the arm mount. CAUTION
Chair upper lift arm cover (discard)
Be careful not to damage the limit switch cable while installing the arm mount.
Chair swivel brake handle
Figure 4
Chair swivel casting
Screws (discard)
Arm mount
3. Remove the support bracket using a large screwdriver (see Figure 5). Set it aside.
Limit switch cable
Figure 7 Screwdriver Support bracket (keep)
Figure 5 2006-04 Rev G (05190)
4. Tighten the bolt securing the mounting block to the swivel casting (see Figure 6 and Figure 8). Page 2
86.0104.00
Screw (place blue Loctite on threads) Thrust washers Thrust bearing
Link arm
Bolt (tighten)
Washer Cuspidor arm
Figure 8
Install the Mounting Arm 1. Remove the link arm from the shipping carton. 2. Install the link arm to the arm mount as shown in Figure 9. Screw (place blue Loctite on threads)
Figure 10
Thrust bearing
Thrust washers Link arm
Rotation stop cover (remove to install mount cover, then reinstall.)
Washer Arm mount cover Arm mount cover screw
Arm mount
Figure 9 3. Install the Radius cuspidor arm as shown in Figure 10. Prop the packaging insert under the cuspidor to help support it while installing. 2006-04 Rev G (05190)
Page 3
86.0104.00
2. Connect the cuspidor limit switch terminal to the chair circuit board.
Connect the Cuspidor Umbilical Position the cuspidor to the right or left of the chair according to the doctor’s requests before connecting the tubings.
Connect the Limit Switch and Touchpad Cable 1. Remove the jumper from the cuspidor limit switch terminal. Tape the jumper inside the chair stop plate in case of any future need.
3. Connect the cuspidor touchpad cable to the Y-adapter circuit board (see Figure 11). If a delivery system is not present, connect the cuspidor touchpad cable to the chair circuit board (see Figure 11). 4. Peel the covering from the Y-adapter circuit board and adhere it to the power supply.
Connect the Tubing Municipal Water Connections 1. Go to Install the Utilities, if you are installing municipal water connections.
Connect the Self-Contained Water 1. Cut the red tubing from the floor box to the junction box. Inline the red cuspidor tubing to the
Radius Cuspidor Umbilical Connections (For clarity, delivery system tubing has been omitted.) Ground wire (connect to the chair structure)
Inline cuspidor tubing to junction box tubing.
Red tubing (1/8" OD)
(Cut)
Optional low voltage water heater cable
Cuspidor umbilical
Junction box tubing
To the delivery system (if present) Cuspidor limit switch and touchpad cable
Delivery system touchpad cable
Yellow with green dashes tubing (1/8" OD) Yellow tubing (1/8" OD) To the cuspidor
Blue with white dashes tubing (3/8" OD)
Vacuum tubing
Cuspidor limit switch terminal connects to chair circuit board
Cuspidor drain tubing
Y-adapter circuit board
Floor box side
Tubing clamp
To the floor box
To chair circuit board
Cuspidor side
2006-04 Rev G (05190)
Page 4
86.0104.00
(see Figure 15).
tubing going to the junction box (see Figure 11). Use the tubing clamps provided to secure the cuspidor drain and vacuum drain tubing. NOTE:
5. Connect the three lengths of red tubing (1/8" OD) using a tee and uni-clamps (see Figure 14). Push
The self-contained water option is not available on Radius cuspidors without a delivery system.
Bracket
Delivery system post
2. Connect the red tubing (1/8" OD) to the matching red tubing in the junction box using an inline barb and uni-clamps (see Figure 12)
Bracket mounting screw and spacer
3. Remove the water bottle bracket from the Radius delivery system post. Cut the red tubing (1/8" Red tubing
Cascade Radius junction box Red tubing (1/8" OD)
Tee
Red tubing
Figure 14 the tee into the water bottle bracket. 6. Reinstall the bracket to the delivery system post. 7. Remove the thumbscrew underneath the cuspidor and slide the cover down (see Figure 15).
Decade Radius junction box
Figure 12 OD) connected to the bracket (see Figure 13). 4. Pull out the end of the red tubing (1/8" OD) Bracket Delivery system post
Tubing hole
Cuspidor cover
Bracket mounting screw and spacer
Red tubing (1/8" OD) cut
Thumbscrew
Red tubing (1/8" OD) coming from chairmount end of post
Figure 15
Figure 13 coming from the chair-mount end of the delivery system post (see Figure 14) and route the tubing through the tubing hole in the bracket 2006-04 Rev G (05190)
Page 5
86.0104.00
8. Disconnect the red syringe tubing from the water manifold. Install a barb plug onto the barb and secure with a uni-clamp (see Figure 16). CAUTION If the water manifold barb is not plugged, flooding will occur.
Red tubing (1/8" OD) In-line barb
Sleeve
Red syringe tubing Barb cap
Figure 16 9. Connect the red syringe tubing to the red tubing using an in-line barb and uni-clamps (see Figure 16). 10.Slide up the cover and tighten the thumb screw (see Figure 17). Pull on the saliva ejector and syringe tubing while sliding up the cover to make sure they do not get pinched.
Figure 17
2006-04 Rev G (05190)
Page 6
86.0104.00
Install the Utilities
Test the Cuspidor
Refer to the appropriate Plumbing Diagram (Figure 18 or 19). Also refer to Radius Floor Box Installation Instructions (A-dec Publication No. 85.2885.00) or Radius 2122/2132 Delivery System Installation Instructions (A-dec Publication No. 85.2877.00) for utility connections.
1. Plug in the chair. 2. Open the manual air and water shutoff valves. For air and water specifications, refer to the Floor Box Owner’s Guide (A-dec Publication No. 85.2611.00). 3. Plug in the power supply if applicable.
Install barb with nylon washer for red tubing (remove the hex plug) Red tubing (oral water, 1/8" OD) Tubing from the umbilical
Blue with white dashes tubing (water supply, 3/8" OD) Yellow with green dashes tubing (40psi regulated air, 1/8" OD)
Water filter/ regulator assembly (black body)
40 psi pre-regulator (non-adjustable)
Yellow tubing (unregulated air, 1/8" OD)
40 psi regulator
Radius Cuspidor (without Radius Delivery System) Figure 18
Air filter/ regulator assembly Yellow tubing (air supply, 1/4" OD) Tubing from the umbilical
Yellow with red stripe tubing (1/8" OD) Yellow tubing (1/8" OD)
Yellow tubing (1/8" OD) Yellow with red dashes tubing (1/8" OD)
Water filter/ regulator assembly
Red tubing (oral water, 1/8" OD) Do not connect Blue w/white dashes tubing (water supply, 3/8" OD) Yellow with green dashes tubing (1/8" OD) 40 psi preset regulator (non-adjustable) Yellow with green dashes tubing (1/8" OD) 40 psi regulator
Radius Cuspidor Utility Plumbing Diagram (with Radius Delivery System) Figure 19 2006-04 Rev G (05190)
Page 7
86.0104.00
• Secure the cuspidor umbilical to the arm mount with a tie wrap (see Figure 22).
4. Move the master on/off toggle on the delivery system or floor box to the ON position. 5. Check for air and water leakage at all connections while operating the syringe (see Figure 20). Air
Water
Arm mount
Spray
Tie wrap
Umbilical
Figure 20 6. Check for vacuum leaks while operating the HVE.
Figure 22
To operate the HVE, simply turn the control valve.
Install the Support Bracket
7. Test the cuspidor and check for water leakage at all connections. For operating instructions, refer to the Cascade 7284 and Radius 7285 Cuspidor Owner’s Guide (A-dec Publication No. 85.2609.00).
Reinstall the support bracket by snapping the bracket onto the chair link arms (see Figure 23).
8. Test the touchpad operation. For operating instructions, refer to the Decade 1011/ 1021 Chair Owner’s Guide (A-dec Publication No. 85.2635.00) or the Cascade 1040 Chair Owner’s Guide (A-dec Publication No. 85.2605.00). For adjustments to the cuspidor and its components, refer to the appropriate Owner’s Guide.
Adjust the Tension • Adjust the tension of the cuspidor arm to the desired resistance (see Figure 21). The cuspidor arm should rotate easily but not drift. Install caps over the screws after the tension has been adjusted.
Support bracket
Figure 23
Tension adjustment
Before Leaving the Facility
Cuspidor arm Caps
Figure 21
• Reinstall the floor box cover and stop plate if you have a Radius chair. If not, install the new stop plate onto the chair. Test the operation of the stop plate. • Distribute the Owner’s Guides and instruct the dental team on the operation and maintenance of the equipment.
2006-04 Rev G (05190)
Page 8
86.0104.00
2006-04 Rev G (05190)
Page 9
86.0104.00
Identification of Symbols SYMBOL
Classification of Equipment (EN-60601-1)
DESCRIPTION Recognized by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1 (2601-1) and under mutual recognition agreement with CAN/CSA C22.2, No. 601.1. Classified by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1 (2601-1) and under mutual recognition agreement with CAN/CSA C22.2, No. 601.1. UL listed to UL 61010A-1, BS EN 61010-2-010 and Canadian (CAN/CSA C22.2, No. 1010.1-92) safety standards. Conforms to applicable European Directives (refer to Declaration of Conformity). Protective earth (ground).
Functional earth (ground). Attention, consult accompanying documents. No user serviceable parts. Attention, line voltage. Only licensed electrician should remove cover. Type B applied part.
TYPES/MODE
CLASSIFICATION
Types of shock protection
CLASS I EQUIPMENT: Dental chairs, dental lights, and power supplies CLASS II EQUIPMENT: Chair, wall, and cart-mounted delivery systems
Degree of shock protection
TYPE B APPLIED PART: Delivery systems only
Degree of protection against water ingress
ORDINARY EQUIPMENT: All products
Mode of operation
CONTINUOUS OPERATION: All models except dental chairs CONTINUOUS OPERATION WITH INTERMITTENT LOADING: Dental chairs - 5% duty cycle
Flammable gasses:
Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide, where such gasses may accumulate in concentration (closed space).
Class II equipment.
Caution: Metal surfaces can be hot during and following the dry cycle.
Environmental Specifications TEMPERATURE/ HUMIDITY
SPECIFICATION
Storage/ Transportation
Temperature:
-40C to 70×C (-40×F to 158×F)
Relative humidity:
80% for up to 31×C, decreasing linearly to 50% at 40×C.
Termperature:
10×C to 40×C (40×F to 104×F)
Relative humidity:
80% for up to 31×C, decreasing linearly to 50% at 40×C.
Operating
Indoor Use:
Electrical Rating TYPE
SPECIFICATION
Volts:
100/110-120/220-240 VAC
Frequency:
50-60 Hz
Current:
As configured and specified in equipment manual (products labeled 15A or greater require dedicated circuit, identified in distribution panel).
Altitude up to 2,000M (6,563 ft.), installation category II, pollution degree 2. (UL 61010A-1 and CAN/CSA C22.2, No. 1010.1-92 only)
Printed in USA. Copyright © 2006, All Rights Reserved. 2006-04 Rev G (05190)