Inspection Testing
1 Page
Preview
Page 1
Customer Name: ___________________________________________________________
1733 W Parkside Lane Phoenix, AZ 85027 Ph: 623-780-2020 website: www.accutron-inc.com
Address: __________________________________________________________________
Step
Item / Function
1
Physical Check
2
Tubing and Rubber Goods
3,4
Leak Check 1 (outspout)
5,6
Leak Check 2 (air override, tubes, joints)
7
Non-rebreathing Replace non-rebreathing (outspout) leaf. No testing valve required.
9
Oxygen Flush
10
Failsafe Test
11
O2 Flow Point 1
12
O2 Flow Point 2
15,16 13,14
SN:
N2O Flow Point 1
17
Model#:
Percent N2O Control
Ball style (PC,DC,CM only) - O2 at 6 lpm and N2O at 6 lpm. Verify Nitrous % knob at 50% per procedure. (non-critical if flow tubes pass previous 4 test steps)
[ ] Return to Accutron for repair [ ] Customer chooses to keep unit and rely on flow tubes since %knob is only an aid. Customer Signature:
18
Pass (P) Fail (F)
Meter Desc:
O2 Resuscitator (100 lpm min)
Plug fixture into resuscitator outlet. Verify min flow in fixture
[ ] Return to Accutron for repair [ ] Customer chooses to keep unit, believes current flow meets emergency needs. Customer Signature:
N2O Flow Point 2
Final Status:
Requirement / Test Actions
Possible Corrective Actions if Test Fails
Check for damaged components (list parts here)
[ ] Non-critical (Cosmetic) damage, use as is [ ] Return to Accutron
Check for tears, cracks, holes, etc (list parts here)
[ ] Order parts (Remove from service until new parts arrive, no need for follow-up) ______________________________________________________________________
With gas supply connected and unit off, soap outspout
[ ] Return to Accutron for repair [ ] Customer keeping unit, is aware it might increase ambient levels of N2O in room and cause some gas loss. Customer Signature:
With gas supply connected and unit flowing 4 lpm of each gas, soap override in two places per procedure, (PC/DC only) test joints per procedure
[ ] Return to Accutron for repair [ ] Customer keeping unit, is aware it might increase ambient levels of N2O in room and cause some gas loss. Customer Signature: [ ] Non-rebreathing leaf replaced
Minimum flow of 20 lpm
[ ] Return to Accutron for repair [ ] Customer accepts Flush as is. Customer Signature:
Both gases to 4 lpm flow. Shut oxygen supply at source. Verify all flow stops (digital gives Failure Oxygen alarm)
[ ] Stop using and return to Accutron for repair
Ball style meters - O2 3 lpm. Verify fixture 3 lpm. Digital - Total Flow 3 lpm %N2O at 0. Verify fixture 3 lpm.
[ ] 1st test failed. Gas supply was low. After correcting supply unit passed retest. [ ] Proper flow could not be achieved, return to Accutron for repair
Ball style meters - O2 6 lpm. Verify fixture 6 lpm. Digital - Total Flow 6 lpm %N2O at 0. Verify fixture 6 lpm.
[ ] 1st test failed. Gas supply was low. After correcting supply unit passed retest. [ ] Proper flow could not be achieved, return to Accutron for repair
Ball Style - O2 3 lpm, N2O 3 lpm. Verify fixture 6 lpm. Digital - Total flow 6 lpm, %N2O 50%. Verify fixture 6 lpm. Ball Style - O2 6 lpm, N2O 6 lpm. Verify fixture 12 lpm. Digital - Total flow 9.9 lpm, %N2O 70%. Verify fixture 9.9 lpm.
[ ] 1st test failed. Gas supply was low. After correcting supply unit passed retest. [ ] Proper flow could not be achieved, return to Accutron for repair
[ ] Meets minimum requirements
Tested by: __________________________________ Date: _______________
[ ] 1st test failed. Gas supply was low. After correcting supply unit passed retest. [ ] Proper flow could not be achieved, return to Accutron for repair
[ ] Fails minimum requirements, return to Accutron for repair Test fixture # _________________ Fixture Calibration Due Date: __________________
Comments:
F 1224 rev02
CN#:CN-0646
Field Certification Test Sheet
Accutron, Inc. MASTER