Calibration and Test instruction fabian +nCPAP evolution (SN prefix AN-)
Table of contents: 1) Configuration ... 4 2) Visual inspection ... 5 3) Start-up / Self-test ... 5 4) Blender / purge flow ... 6 4.1 Blender tightness ... 6 4.2 Tightness calibration valves... 6 4.3 Purge flow ... 7 5) Calibration ... 8 5.1 Set date and time ... 8 5.2 Pressure regulator calibration ... 8 5.3 Blender calibration... 9 5.3.1 Calibration AIR valve / Flow sensor ... 10 5.3.2 Calibration O2 valve / Flow sensor ... 10 5.4 Inspiration sensor (p insp.)... 11 5.4.1 Pinsp 0 mbar calibration ... 11 5.4.2 Pinsp 60 mbar calibration ... 12 5.5 Proximal sensor ... 13 5.5.1 Pprox 0 mbar calibration ... 13 5.5.2 Pprox 60 mbar calibration ... 13 5.6 Touch screen calibration ... 15 5.7 Exhalation valve calibration ... 15 5.8 Pressure relief valve calibration ... 16 5.9 O2 Sensor calibration ... 17 5.10 Flow Sensor Calibration ... 17 6) Functional test ... 18 7) Trigger Check ... 19 8) Volume guarantee Check (if licence installed)... 19 9) O2 / Air blender ... 19 10) Patient alarms ... 20 10.1 Tube leak ... 20 10.2 ET tube check ... 20 10.3 Pressure high ... 20 11) System Alarms ... 21 12) CPAP Mode... 21 13) DUOPAP Mode (if licence installed) ... 21 14) SpO2 functional test (if module installed) ... 22 15) Maintenance ... 23 16) Setting check ... 23 17) Electrical Safety Test ... 23 18) Battery calibration ... 24 TS-AA-03e V08 TI fabian +nCPAP evolution
Page 2 of 24
Used Test equipment: Calibrated tester for: • Flow measurement • Pressure measurement • Oxygen measurement Test-Set for intubated ventilation modes Part No. 7398 / 111291 Test Set1
Test-Set for non-intubated ventilation modes Part No. 7398.N / 111292 Test Set2
Test lung Part No. 1116 / 151901
Test lung small Part No. 7204 / 151900
Flowsensor Part No. 1031 / 151120
Flowsensor cable Part No. 1016 / 151100.xx
SpO2 Tester, Part. No. 152402
(if module installed)
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 3 of 24
1)
Configuration
Adjust the Altitude settings in the Service-mode Calibration screen. Set here the proper altitude for the flow compensation. One step is 100m. After this calibrate the flowsensor before going ahead with this instruction. Info Screen: From Main screen press 2x INFO • Device ID • Version Software • Monitor PIC • Conductor PIC • Blender PIC • Kernel Version • Language Version • Hours device, battery
then
In Service mode – open Accuboard screen • Accu PIC
All this information’s are also available when you save the “Device-Info” in the menu TOOLS. Menu point Content of the file
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 4 of 24
2)
Visual inspection
Inspect the device for general condition. 3)
Start-up / Self-test
Use the ON / OFF button on fabian to power on the equipment. When the device is powered on the start-up screen will appear and the equipment will perform a self-test. The SOFTWARE REVISION, the CHECK SUMS and the status of the interfaces will be checked. (ok ≙ no error)
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 5 of 24
4)
Blender / purge flow
Gas standardisation: 20/1013 Gas type: Air 4.1
Blender tightness
• • • • •
4.2
Connect fabian to gas supply. fabian is switched OFF Connect a tube to the ”patient gas” outlet from the blender (red) and block the “purge flow” tube (blue) of the blender. See picture below. Place the other end in a drinking glass filled with water OR connect to your flow meter or Respirator tester. With water it is more accurate. NO bubbles coming out of the tube OR 0.00 l/min on your Respirator tester.
Tightness calibration valves
• • • • •
•
Connect fabian to gas supply. Turn the device upside down to gain access to the blender fabian is switched OFF Connect a tube to the Oxygen sensor calibration outlet. ( see picture below ) Place the other end in a drinking glass filled with water OR connect to your flow meter or Respirator tester NO bubbles coming out of the tube OR 0.00 l/min on your Respirator tester.
Reconnect the tube from the blender back to its position, see white finger in picture above
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 6 of 24
4.3
Purge flow
• • •
Connect fabian to gas supply. Connect a tube to the proximal sensor port. (see picture below) Switch ON fabian
Settings: Mode I - Flow E - Flow PEEP P Insp Frequency I - Time Oxygen •
•
IPPV 8 l/min 5 l/min 5 mbar 30 mbar 30 bpm 1.00 sec. 21 Vol.%
Connect your flow meter or respirator tester to the prox. port. ( see picture below ) Start manual inspiration by pressing: Reading on the tester between 150 and 350 ml
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 7 of 24
5)
Calibration
5.1
Set date and time
• 5.2
In menu date / time Pressure regulator calibration
• •
• • • • •
•
Connect Air and O2 supply Start fabian: o In O2 Therapy mode with a setting of 3l/min of flow, if not available use o IPPV mode with I-Flow 3l/min. Set. O2 value to 60 Vol.% Service menu -> Monitor PIC -> compare values P-AIR and P Oxy. ( Fig. 1 ) Adjust pressure regulator to 2400 ± 200 (Fig. 2). “Old” version. Lift the cap from the pressure regulator to adjust the value. “New” version. Loosen the ring and adjust with a flat screwdriver and secure with the ring. OLD Type New Type
Check the values after 5 minutes again to be sure the regulator is working properly. O2
O2
AIR Figure 1
•
Figure 2
If adjusted secure the regulators.
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 8 of 24
5.3
Blender calibration
Calibration must be done by a calibrated mass flow meter. Power OFF the device. The measurement unit must be set to the following setting: Gas standardisation: 20/1013 Gas type: Air
Supply Air and O2 connection with compressed air!
• • • •
Connect flow meter to inspiration port Switch ON fabian Wait until the calibration of the blender is done! Let the fabian run for 5 minutes with an O2 saturation of 60%
Settings: Mode
IPPV
O2 Therapy
I - Flow
10 l/min
10 l/min
E - Flow
5 l/min
PEEP
5 mbar
P Insp
30 mbar
Frequency
10 bpm
I - Time
2.00 sec.
O2
21 Vol.%
21 Vol.%
O2-Flush
100 Vol.%
100 Vol.%
Set the manual Breath time to the maximum of 30 Seconds. See Settings menu on the left side or in the Ventilation menu. •
Switch OFF all Alarms by pressing Simultaneously
and
hold for approx. 5 sec.
In the upper right corner, the symbol will appear. From now on ALL alarms are disabled until restart of the device
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 9 of 24
Before adjusting the potentiometers try to restart the device 2-3 times and normally the value is coming into the range. If this is not the case, you then have to adjust the potentiometers as it is described now 5.3.1
Calibration AIR valve / Flow sensor
Set O2: 21 % Reading: 9.5 ± 0.1 l/min Adjust with potentiometer 21% (Fig. 3) For IPPV-Mode you have to press the button “manual breath”
5.3.2
Calibration O2 valve / Flow sensor
Activate the O2 Flush (100Vol.%) Reading: 9.5 ± 0.1 l/min Adjust with potentiometer 100% (Fig. 3) For IPPV-Mode you have to press the button “manual breath”
-
+
Figure 3
• • • •
Switch OFF fabian Switch ON fabian Compare the values 9.5 l/min for AIR and O2 Possibly recalibrate ( sometimes need 2 or 3 restarts )
When the calibration is done, power OFF the device and connect AIR / O 2 as in normal use. Restart the device
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 10 of 24
5.4
Inspiration sensor (p insp.)
5.4.1 Pinsp 0 mbar calibration • fabian Service Mode
•
Monitor PIC
•
Disconnect the tube from the pressure sensor (Fig. 5, mark 3)
• No patient tubing connected. Value p-insp. / p-mix 570 ± 10 ( Fig. 4 ) • Adjust with potentiometer ( Fig. 5, mark 1 )
Figure 4
Figure 5
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 11 of 24
5.4.2
Pinsp 60 mbar calibration
• Apply 60mbar to the pressure sensor (Fig.7 mark 3) directly • Value p-insp. / p-mix. 2400 ± 10 ( Fig. 6 ) • Adjust with potentiometer ( Fig. 7, mark 2 )
2400
Figure 6
Figure 7
•
Connect the tube again to pressure sensor ( Fig 7, tube from mark 3 to mark 7 )
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 12 of 24
5.5
Proximal sensor Before you start the calibration, do the following:
-
Connect gas supply Ventilation Mode: select CPAP Setting: CPAP = 5mbar, Flow = 4l/min
Figure 8
5.5.1
• • • •
Pprox 0 mbar calibration Service menu -> calibration Prox. Port is open / no tube connected Press – Calibrate Confirm 0 mbar
5.5.2 Pprox 60 mbar calibration There are two ways to perform the 60 mbar calibration. Option A is the recommended way once the device is open. In cases when the device is not open, the 60 mbar calibration can be done without opening it by following Option B.
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 13 of 24
5.5.2.1 Option A
• • • •
Connect calibration set on prox. port. ( Fig. 8 ) Set 60 mbar on prox. port Confirm 60 mbar button Sensor is calibrated
For the blender V2, you have to block the tube (blue), outlet from blender to manifold.
5.5.2.2 Option B
• • • •
• • • •
Connect “Test Set 1” and test lung „ small „ Connect measurement device with T Piece (Prox. port) Press 60 mbar Cal. Confirm the next message
Keep the “manual breath” button pushed Check the external pressure Release the “manual breath” button Adjust the difference with +/-, you can use the buttons or the knob. You have to confirm the value. As long the value is “green” it is not taken, it has to be dark blue.
• • • • •
Redo a “manual breath” Check the external pressure Adjust the difference with +/Redo a “manual breath” When the pressure is okay on the external device, push the SAVE button during the “manual breath” • Now the calibration is stored. Redo the whole procedure until you get a stable value with a maximal difference of 0.2mbar.
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 14 of 24
5.6
Touch screen calibration
• • • •
5.7
Service menu Display Calibration Follow messages on the screen (possibly you must perform the calibration more than one time!)
Exhalation valve calibration
• •
Service mode - > calibration Connect “Test Set 1” as for ventilating a patient
•
Exhalation valve calibration
•
Follow messages on the screen Be sure to have the test lung removed from the Y Piece. Hold Y Piece tight
•
When the values shows p peak / PIP 50 ±1 mbar and PEEP 5 ±0.5 push the “Home button” to store the calibration data.
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 15 of 24
5.8
Pressure relief valve calibration
• • • •
Start fabian in nCPAP mode Set CPAP 5 mbar Set P manual 15 mbar Connect pressure gauge (range 0 - 1 bar) directly to the CPAP port. ( Fig. 9 )
CPAP
Figure 9
• • •
Apply manual breath Safety valve opens between 450 to 600 mbar Adjust safety valve by turning the calibration screw (Allen key 8 mm). Clockwise will increase pressure ( Fig. 10 ) REMARK: The screw can be very strong, because it is secured with Loctite 480 After adjusting secure the screw with Loctite 480 or similar
Figure 10
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 16 of 24
5.9
O2 Sensor calibration
Access the calibration menu directly via the calibration key: Start the calibration with pressing 21 + 100%
5.10 Flow Sensor Calibration
First you have to go into the “IPPV” mode. Go into the calibration menu directly via the calibration key: Ensure no gas will flow through the flow sensor during calibration. The best way is to close the sensor on both sides, with the fingers. Following successful calibration, the date of the last calibration will be displayed in the corresponding field.
Close the device now. All the tests have to be done with a closed device to be sure that no cables or tubes are kinked.
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 17 of 24
6)
Functional test All tests have to be done with the Gas Standard ATP Activate also “enable O2 comp.” in “Service-Menu / Calibration” and also to have the correction factors for flow sensors to 0.960
We have adapted the gas-standard from BTPS to ATP as in the field it is easier to work without a humidifier in the system. Connect “Test Set1” and test lung „ small „
Trigger settings for the measurement device, e.g. Flow Analyser PF-300 are:
0.2
-0.2
The “Test Set1” with small lung has to be connected in the “low flow” line of the tester The test settings are then shown in the “TS-FB-21e V08 TPS fabian +nCPAP evolution” protocol TS-AA-03e V08 TI fabian +nCPAP evolution
Page 18 of 24
7)
Trigger Check
Use the following settings for the test: Mode
S - IPPV
Trigger
5
I - Flow
8 l/min
E - Flow
5 l/min
PEEP
5 mbar
P Insp
15 mbar
Frequency
15 bpm
I - Time
2.00 sec.
Oxygen
21 Vol.%
At Expiration phase pull the test lung small lightly to initiate a trigger. If a trigger is applied the breathing curve will turn green. 8)
Volume guarantee Check
(if licence installed)
Connect Test Set1and test lung „ small „ Set the Patient range to Neonatal (Ventilation Mode) Important: The leakage must be 0, otherwise the measurements will be different Use the following settings for the test: Before the measurement, let the device run with 60%O2 for 5 minutes. Mode
S - IPPV
I - Flow
6 l/min
E - Flow
6 l/min
PEEP
5 mbar
P Insp
25 mbar
Frequency
30 bpm
I - Time
1.00 sec.
If you do not pass this test, try it with another flow sensor. Most of the time the flow sensor is causing the problem. At the end of this test, set VG off 9)
O2 / Air blender
Use the following settings for the test: Mode
IPPV
O2 Therapy if installed
I - Flow
5 l/min
5 l/min
E - Flow
5 l/min
PEEP
5 mbar
CPAP
-
P Insp
20 mbar
P duo I - Time
0.5 sec.
Frequency
60 bpm
In O2 Therapy you can connect only the flow tube. For IPPV connect “Test Set1” without lung direct to the insp. port of the fabian.
TS-AA-03e V08 TI fabian +nCPAP evolution
Page 19 of 24
10) Patient alarms Connect “Test Set1” and test lung. Settings are as follows: Mode
S - IMV
I - Flow
10 l/min
E - Flow
5 l/min
PEEP
5 mbar
P Insp
20 mbar
Trigger
1
Frequency
10 bpm
I - Time
1.00 sec.
Oxygen
21 Vol.%
10.1 Tube leak
Set leak alarm to 20% and Apnea to 2 seconds Set the red t-piece to the position below:
Alarm messages: “High tube leak” and “Apnea” will switch from one to the other If you breathe between with the test lung, then you only get the leak alarm Test lung back to normal position
10.2 ET tube check
With the same settings, close now the yellow t-piece
Alarm message : “check ET tube “ After this test disable leak and apnea alarm settings 10.3 Pressure high
Squeeze the test lung during the inspiration phase
TS-AA-03e V08 TI fabian +nCPAP evolution