Acutronic Medical Systems
fabian Ventilator Systems
fabian +nCPAP evolution Service Manual Ref 122010.02 Rev 2.0 April 2019
Service Manual
99 Pages
Preview
Page 1
fabian +nCPAP evolution Ref: 122001 Prefix: AN
Service Manual SM Rev. 2.0
fabian +nCPAP evolution | SM-Rev.2.0 Ref: 122010.02EN / Date: 24.04.2019
Sales and Service Information
Sales and Service Information The ACUTRONIC Medical Systems AG sales and service centre network is world-wide. For the address of your local distributor, contact your nearest ACUTRONIC Medical Systems AG subsidiary. In case of difficulty a complete list of all distributors and subsidiaries is provided on our internet site: http://www.acutronicmedical.ch. Sales information can also be obtained from: [email protected]
Address Headquarter
ACUTRONIC Medical Systems AG Fabrik im Schiffli 8816 Hirzel / Switzerland Phone: +41 44 729 70 80 Fax: +41 44 729 70 81 E-Mail: [email protected] www.acutronic-medical.ch
2 / 99
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 24.04.2019
Table of Contents
Table of Contents Sales and Service Information ... 2 Table of Contents... 3 1
Preface ... 5
1.1
Document Structure ...5
1.2 1.3
List of abbreviations...6 Service manual revision history...6
2
Hardware information... 7
2.1 2.2
Serial number description...7 Hardware Revision History ...7
3
Important preliminary remarks ... 8
3.1 3.2
ESD Precautions ...8 Warranty ...8
3.3
Liability for functionality damages...8
3.4
Service, Repair, Maintenance, Calibration & Testing ...9
3.5
Contact ...9
4
System overview ...10
4.1 4.2
Manifold user side ... 11 Rear panel ... 12
4.2.1 4.2.2
External alarm connector (Nurse call) ... 13 Connector for 12 - 24 VDC ... 13
4.3
Hardware components ... 14
4.3.1 4.3.2 4.3.3 4.3.4 4.3.5 4.3.6
Mainboard... 14 NetDCU ... 16 Power module ... 17 Manifold... 20 O2 sensor ... 22 Proportional gas blender ... 24
4.4
Internal tubing ... 27
4.5
PDMS/ Network/ acuLink connection ... 29
5
Failure analysis ...30
5.1
Alarms messages on device ... 30
5.2
Power PCB - status LEDs and fuses ... 32
5.3 5.4
Mainboard - watchdog LEDs ... 33 Software watchdog ... 34
5.5
Alarm messages about the Proportional Gas Blender ... 34
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 29.09.2018
3 / 99
Table of Contents
4 / 99
5.6
Trends, Log files, Device info... 36
6
Calibration and Service information ...41
6.1 6.2
System and service information ... 42 Calibration submenu ... 46
6.3
Blenderboard submenu ... 46
6.4
MonitorPIC submenu ... 47
6.4.1 6.4.2 6.4.3
Inspiratory pressure (P-insp or P-MIX) calibration to 0mbar ... Error! Bookmark not defined. Inspiratory pressure (P-insp or P-MIX) calibration to 60mbar ... Error! Bookmark not defined. Input pressure regulator calibration ... Error! Bookmark not defined.
6.5
Valves submenu... 48
6.6
Display submenu ... 49
6.7
Settings submenu ... 51
6.8 6.9
Licensing submenu and License handling ... 53 Maintenance submenu ... 56
7
Software update ...58
8
Servicing ...59
8.1
Cleaning and disinfection ... 59
8.2
Transportation ... 59
8.3 8.4
List of the test equipment and special tools... 59 Alarms off... 60
8.5
Service-Password off... 60
8.6
Spare parts list ... 60
8.7
Handling of the used components ... 60
9
Repair and various replacements/ Disassembly and assembly ..61
9.2
Housing removal ... 65
9.2.2 9.2.3
Remove the Loudspeaker and Fan module ... 75 Complete Front Panel removal ... 91
9.3
Device check ... 98
9.3.1
Battery calibration ... 98
10
Technical Specifications ...99
10.1
Ambient Conditions ... 99
10.2
Dimensions / Weight ... 99
10.3
Ratings ... 99
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 24.04.2019
1. Preface
1 Preface This service manual is intended for use by ACUTRONIC Medical Systems AG trained and authorized service personnel. ACUTRONIC Medical Systems AG does not condone or approve of service activity on its products by other than ACUTRONIC Medical Systems AG trained or authorized personnel and ACUTRONIC Medical Systems AG is not responsible for any unauthorized repairs or modifications, or any repairs or modifications made by unauthorized procedures. Use of the incorrect part, or failure to exercise due care in the installation, removal, servicing, checkout or calibration of parts and equipment or the use of unauthorized accessories, may result in damage to the equipment which may in turn result in damage in property and injury (including death) to persons. The purchaser and installer of these parts shall bear full responsibility and liability for the above. All maintenance performed within the applicable warranty period must be authorized in advance by an ACUTRONIC Medical Systems AG service representative in order to retain the warranty status of the subject unit.
1.1 Document Structure Please carefully read the Service manual before using or servicing the ventilator. The table of contents will assist you in quickly locating the respective topic.
fabian +nCPAP evolution | SM-Rev.2.0 Ref: 122010.02EN / Date: 24.04.2019
1. Preface
1.2 List of abbreviations Abbreviations which are not generally known, but are used in this manual, are listed below. CPAP DIO ESD GUI HL7 IfU IVOI O2 P PCB PDMS PIF RS232 SpO2
Continuous Positive Airway Pressure Digital Input Output Interface Module Electrostatic Discharge Graphical User Interface Health Level 7 Instruction for Use Protocol from Philips for Intellibridge/Vuelink Oxygen Pressure Printed Circuit Board Patient Data Management System Parallel interface Serial communication transmission of data Peripheral Oxygen Saturation
1.3 Service manual revision history Version
Change
1.00
•
First edition
1.01
•
Connection for nurse call
• • • • • • • • • • • • • • • • • •
Chapter – Calibration removed. Test and calibration now in separate document 6.1 Serial number description 7.1 Information maintenance internal battery 12.1.5 replacing internal battery Minor changes Chapter “SW update” revised Battery board now called “power board” Chapter 10.4 renamed in: Processor port connections Delete chapter 10.4.4 – 10-4.6 PIC processor upgrade See new document: “Processor update instruction” 8.5 RS232 not for CO2 8.5.2 power connector / corrected 8.6.2 NEW mainboard 1.3 8.8.2 blender V2 8.9 Internal tubing / NEW 10.2 removed. See new document: fabian update instruction 10.3.4 Removed. See new document: Processor update instruction 12.2 NEW. Error codes.
1.06
•
Chapter 12.2 revised
1.07
•
New Chapter 10.3 revised
2.00
• • •
Complete revision Chapter varying Content modification
1.02
1.03
1.04
1.05
6 / 99
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 24.04.2019
2. Hardware information
2 Hardware information 2.1 Serial number description Example 1: Example 2:
AN10AA-0001 AN-03002
A: N:
only for internal use Model
10:
HW Revision
AA:
Date of manufacture
01:
Sequential number
(Before 1st of January 2016) (From 1st of January 2016) N – fabian +nCPAP evolution T – fabian Therapy evolution H – fabian HFO I – fabin HFOi K – fabian HFO without HFO Module 10 – 1.0 11 – 1.1 etc. A – January A - 2012 B – February etc. B – 2013 etc.
From 1st of January 2016 the serial number have been changed to AN, for the device and xxxxx is a sequential number: AN-xxxxx
2.2 Hardware Revision History Serial no..
Rev.
Change
AN10xx-xxxx
1.0
• Without external 12 – 24 VDC power capability. Connector is already mounted but not connected inside the unit.
AN11xx-xxxx
1.1
≥ AN11BB-0219
1.1
≥ AN11CC-0497
1.1
≥ AN11FC-0597
1.1
≥ AN11KC-0727
1.1
≥ AN11ED-1119
1.1
≥ AN11FD-1177
1.1
AN12xx-xxxx
1.2
≥ AN12KD-1365
1.2
≥ AN-01657
-
≥ AN-01804
-
≥ AN-01948
-
≥ AN-02135
-
≥ AN-03000
-
≥ AN03014
-
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 29.09.2018
• External power connector with integrated filter • Changing from “FAS” to “MIM” valves • Colour coded internal tubing • Adaptions in fan holder • New Mainboard 2.0 • SD-card inserted • New blender (new pressure input regulators Kuhnke) • New Mainboard 2.1 • New air filter (blue) build in in upper and lower cover • Changed the rotary knob to plastic • New front foil without edge around the display • New manifold with adapted fittings for the membrane holder • Blue Inspiratory port • New housing type (Hakama) • New front foil 7 / 99
3. Important preliminary remarks
3 Important preliminary remarks 3.1 ESD Precautions Always take standard ESD precautions when working on fabian ventilator systems. Ensure the ventilator is disconnected from the AC power supply before performing service or repairs or maintenance.
3.2 Warranty We warranty that this product is free from defects in material and workmanship and, when properly used, will perform in accordance with applicable specifications. If within two years after the original shipment it is found not to meet this standard, it will be repaired at Acutronic Medical Systems AG, or replaced at no charge when returned to an Acutronic Medical Systems AG service facility. Any modification to the equipment, rework or repair by other than Acutronic Medical Systems AG approved personnel will render this warranty null and void. Acutronic Medical Systems AG shall not be liable for any indirect, special or consequential damages, even if notice has been given of the possibility of such damages. THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. The policy of ACUTRONIC Medical Systems AG is to maintain product repair capability for a period of five years after the official sales stop. The two years warranty can only be guaranteed when the yearly maintenance with the preventive maintenance kit was done. All maintenance performed within the applicable warranty period must be authorized in advance by an ACUTRONIC Medical Systems AG service representative in order to retain the warranty status of the subject unit.
3.3 Liability for functionality damages The liability for proper function of the apparatus is irrevocably transferred to the owner or operator to the extent the apparatus has been serviced or repaired by personnel not employed or authorised by ACUTRONIC Medical Systems AG Service or when the apparatus was used in a manner not conforming to its intended use. ACUTRONIC Medical Systems AG cannot be held responsible for damage caused by non-compliance with the above-mentioned recommendations. The warranty and liability provisions of the terms of sale and delivery of ACUTRONIC Medical Systems AG are likewise not modified by the recommendations mentioned above.
Description of: Service: Repair: Maintenance: 8 / 99
measures to maintain desired condition to restore desired condition inspection, service and if applicable, repair fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 24.04.2019
3. Important preliminary remarks
3.4 Service, Repair, Maintenance, Calibration & Testing The Device is a ventilator classifies as Class IIb according to the European directive, as such: • Inspection as per manufacturer specifications is required every 12 months • Maintenance must be performed by Acutronic Medical System AG trained experts with access to appropriate test and measuring equipment. We strongly recommend Acutronic Medical System AG exclusive representatives for service agreements and repairs. Use only original Acutronic Medical System AG parts for repair. Always clean and disinfect equipment or equipment components prior to any maintenance – including when returning the equipment for repair.
Every 12 months Perform the following work: • Maintenance and safety inspections according to the following documents: o TS-AA-17e, Calibration and Test instruction, See Software Release Package, website o TS-FB-27e, Test protocol for service & maintenance, See Software Release Package, website • Preventive Maintenance kit for the device, see “Spare Part Price List” on the website.
Every 4 years Perform the following work: • Maintenance and safety inspections according to the following documents: o TS-AA-17e, Calibration and Test instruction, See Software Release Package, website o TS-FB-27e, Test protocol for service & maintenance, See Software Release Package, website • Preventive Maintenance kit for the device, see “Spare Part Price List” on the website.
Maintenance and the safety check may only be carried out by specialists trained by Acutronic Medical System AG , who have suitable measuring equipment and test facilities. Original packaging should be kept for the possible return of the unit. Only original packaging ensures optimal protection of the equipment during transport. If it is necessary to return the equipment during the warranty period, the manufacturer is not liable for damages caused by incorrect packaging. Original packaging should be kept for the possible return of the unit. Only original packaging ensures optimal protection of the equipment during transport. If it is necessary to return the equipment during the warranty period, the manufacturer is not liable for damages caused by incorrect packaging.
3.5 Contact Phone:
+41 44 729 70 80 +41 44 729 70 86
E-mail:
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 29.09.2018
9 / 99
4. System overview
4 System overview
Acutronic fabian +nCPAP medical ventillator
10 / 99
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 24.04.2019
4. System overview
4.1 Manifold user side
1. Connectors on the Manifold (Photo and cross-section schematic) Component name 1
Inspiratory tube port
2
Connection for nCPAP system based on flow generators (single limb systems)
3
Proximal pressure measuring tube port
4
Expiration tube port
5
Exhalation valve coil
6
Electronic safety valve coil
7
Safety valve (function: pneumatic pressure relief)
8
Pressure sensor
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 29.09.2018
11 / 99
4. System overview 9
Pressure sensor (P insp)
4.2 Rear panel
2. Connectors on the rear panel (photo and schematic)
1
Mains connection with fuse holder
2
Terminal studs for potential equalization
3
Connection for external voltage supply 12 – 24V DC From HW- Rev. 1.0 use only VDC cable with Part No. 151401 (See spare parts list!)
4
Nurse call, max. Switching voltage 50 VDC
5
Flow sensor connector
6 7 8 12 / 99
Network connection for data management ( For connection to network with minimum 3 KV galvanic isolation) USB connection for data output (log file) and system update; Connection for Masimo SpO2 sensor module. RS232 interface, service, PDMS fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 24.04.2019
4. System overview 9
Connection for compressed air supply 2.0 – 6.5 bar / 40 l/min
10
Oxygen sensor
11
Connection for oxygen “O2” supply 2.0 – 6.5 bar / 40 l/min
For adapting to the DISS system you need: AIR adapter NIST-female to DISS-male / 90 degrees O2 adapter NIST-female to DISS-male / 90 degrees
Warning: Do not connect Ethernet, Nurse call, USB, R232, Flow Sensor ports to anything other than specified devices.
4.2.1 External alarm connector (Nurse call) Plug: Manufacturer: Binder Series 712, 3 pins PIN 2/1
Normal open (NO)
2/3
Normal Closed (NC)
4.2.2 Connector for 12 - 24 VDC The fabian evolution can likewise be operated with an external power supply. An external supply with a voltage of 12 – 24 volts must be connected to the back of the device in this case. Plug: Manufacturer: Binder Series 680, 4 pins PIN 3/4
12 - 24 V DC
1/2
GND
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 29.09.2018
13 / 99
4. System overview
4.3 Hardware components 4.3.1 Mainboard
3. Mainboard
14 / 99
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 24.04.2019
4. System overview Marks in pictures 3. and 4. 1 2
3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18
Component name NetDCU connector Lithium Battery cell of the Real Time Clock (Type CR2032) UDP connector Touch screen connector Patient gas supply connector Blender connector Fan connector Loud speaker connector Backlight connector Safety relays LCD display connector Flow sensor connector Nurse call connector O2 sensor connector Programmer pin strips Hardware version (HW-Rev) Power PCB connector Keyboard connector
Comment
If the fabian does not show the correct time, even after refreshment, please exchange the battery of the Real Time Clock.
Monitor, Alarm-PIC, Boot, Conductor e.g: MIM Fabian Evolution V22 20160330, Rev.: 2.2
4. Hardware version sign on the Mainboard
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 29.09.2018
15 / 99
4. System overview
4.3.2 NetDCU The NetDCU PCB can be found on the Mainboard, connected via pin strips. The NetDCU contains an exchangeable SD memory card to store the trends, log files and special font characters. Log and trend files can be exported via USB. The NetDCU uses embedded WinCE operation system, therefore the used USB device has to be WinCE compatible (for example: USB1.1). For more information about the downloading of the log files, trends and device information please see Chapter 4.6. The Device ID of the fabian and licenses are stored in the inner memory of the NetDCU, therefore at replacing the NetDCU module, the fabian will get a new device ID (and a new service-password) and the licenses has to be reinstalled as well (See Licensing submenu and License handling, Chapter 5.8.).
5. NetDCU
16 / 99
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 24.04.2019
4. System overview
4.3.3 Power module The device can be operated from an external power source, but features an internal battery (NiMH Powerpack) also. For detailed description on the topic please read the Battery operation Chapter in the IfU. In this document the power module expression means the sum of the following components: - Power PCB - Powerpack NiMH/ Inner battery - Power supply “XP” - Power inlet module The most important components of the Power PCB are shown on the picture below.
6. Power PCB Numbers in picture 6. 1 2
Comp. name on PCB
5
LED 5
“Power OFF” LED 5V OK 4.2) 12 V OK 4.2) Power board in normal condition (See Chapter 4.2) State of charge
6
F3 (SMD, 3 AT, 125 V” MELF 6125”
Fuse3: Safety OFF function
3 4
LED 1
Function
LED 2 LED 3 LED 4
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 29.09.2018
(See Chapter 4.2) (See Chapter (See Chapter
(See Chapter 4.2) (See Chapter 4.2) 17 / 99
4. System overview 6.1mm x 2.7mm) 7 8
F2 (SMD, 3 AT, 125 V” MELF 6125” 6.1mm x 2.7mm) AKKU
Fuse2: 24V connection (Power off alarm) (See Chapter 4.2) NiMH Powerpack (Internal battery) connector
9
ICP2 programmer connector
10
External power supply connector
11
24V wires
12
Mainboard connection e.g: MIM AKKU_EVOL V.2.1 04.06.2012
Hardware version (HW-Rev)
Exchange of the Powerpack (Internal battery) is part of the 4 year preventive maintenance.
7. Power supply “XP” Power inlet module and Power Supply “XP” join to the same point of the back panel as ground potential (Rear panel schematics above refers to it, as terminal studs for potential equalization).
18 / 99
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 24.04.2019
4. System overview
8. Power inlet module and Power supply “XP” connection with potential equalizator
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 29.09.2018
19 / 99
4. System overview
4.3.4 Manifold In case of the failure of the Manifold please exchange the complete Manifold block, and do not disassemble it. Except if the damage is occurred on the proximal connector, the expiration valve membrane, the ribbon cable or the tubes. The exchange of the expiration valve membrane is part of the yearly preventive maintenance.
9. Main components of the Manifold (Left: removed Manifod, Right: fabian without housing, Manifold is still in its original position) Numbers in pictures 9. and 10. 1 2
Component name
Comment
Connection with the mainboard Expiratory port
3 4 5 6 7 8
Manifold PCB Pressure sensor inspiratory (P insp) Pressure sensor proximal (P prox) Purge flow Inspiratory tube Electromagnetic and pneumatic valves
9
Calibration potentiometer
10
Calibration potentiometer
“Patient gas” With expiratory valve and membrane. See in picture 10. Tube G Tube Y Tube B Tube R See in picture 1. with number 5 and 6 Pinsp 0mbar (clockwise +, anticlockwise -) Pinsp 60mbar (clockwise +, anticlockwise -)
Detailed description about the front side of the Manifold please see above (Chapter 4.1). 20 / 99
fabian +nCPAP evolution | SM-Rev. 2.0 Ref: 122010.02EN / Date: 24.04.2019