Electronic Wheelchair and Multipurpose Scales M500660-Rev005-UM-e Operating Manual March 2021
Operating Manual
16 Pages
Preview
Page 1
Operating Manual Electronic Wheelchair and Multipurpose Scales
M500660 Series M500660
M500660-210326-Rev005-UM-e
2
M500660-210326-Rev005-UM-e
Content
Introduction ... 4 Safety ... 5 Assembly and set up ... 5 Power supply ... 6 Replacing the rechargeable battery ... 6 Display ... 7 Front View and side view of the display ... 7 Display Description ... 7 Key pad... 8 Weighing ... 8 Correct weighing ... 8 How to change the unit ... 8 Tare function ... 9 Weighing with preset tare values ... 9 HOLD function ... 9 BMI (Body Mass Index) ... 9 Evaluation of the Body Mass Index ... 10 Connecting additional devices ... 10 Print out a record ... 10 Error Messages ... 10 Cleaning and Maintenance ... 11 Cleaning ... 11 Safety information ... 11 Disposal... 11 Storage and Transport ... 11 Warranty – Liability ... 11 Technical specifications ... 12 Declaration of Conformity... 13
M500660-210326-Rev005-UM-e
3
Introduction
Dieses Symbol bedeutet: „Gebrauchsanweisung beachten“. Um präzise Ergebnisse zu erzielen, lesen Sie die Gebrauchsanweisung sorgfältig durch und folgen Sie den darin enthaltenen Instruktionen. Bewahren Sie die Gebrauchsanweisung für späteres Nachschlagen auf. Your ADE electronic wheelchair-scale is a quality, sturdy and durable product manufactured in accordance with national regulations in Germany. The scale is used to evaluate the weight of persons sitting in a wheelchair, able to walk on their own or by using any kind of walking aid or lying on an emergency litter. The scale could be used for adipose persons as well. ADE instruction manuals have been prepared to provide the user with all the necessary knowledge about the safe use of ADE scales and their related accessory equipment. For further questions about how to use the products or about this or other service manuals, feel free to contact your local ADE representative or visit our web page at www.ade-hamburg.de. In order to achieve the best possible results kindly read and follow the operation manual in all respects. This operation instruction applies to the following models of electronic scales of the M500660-series with the different adaptations (see table 1). In order to achieve the best possible results kindly read and follow the
Multipurpose and wheelchair scales 1 platform 800 x 800mm / M500660-C 1400 x 800 mm 1 ramp Configuration of the basic 1 eyelevel version M500660: 1 mains adapter 1 indicator (display) Reference number Description of supplementary equipment to the standard version M500660-01 1 foldable column M500660-02 1 set of side boundaries [table 1]
M500660
M500660-02
[picture 1]
[picture 2] M500660-01
4
M500660-C
M500660-210326-Rev005-UM-e
Equipment supplied
All wheelchair scales and multipurpose scales are being delivered in an already assembled condition with handrails, side boundaries or columns. The equipment supplied includes: 1) One indicator including the correspondent set of screws 2) One wall mounting device and one cable extension for model numbers: M500660 3) One mains adapter 4) Operation manual including declaration of conformity Safety The ADE-display unit corresponds to the latest state of the art technology and complies with standard regulations and directives. Kindly follow the safety instructions in this operating manual, the country-specific installation standards as well as the applicable safety conditions and accident prevention regulations in all aspects.
Operate the scale only with the original ADE mains adapter.
Always connect the mains adapter to the scale BEFORE connecting it to the electrical outlet
Disconnect from the electrical outlet BEFORE disconnecting the adapter from the scale.
In case the mains adapter became wet or show visible damages the scale can no longer be operated by using this mains adapter. In this case please contact your local dealer or service partner.
The voltage and power supply as mentioned on the type plate must match the mains voltage and power type at the place of use.
When using the scale with a mains unit ensure that the supply cable is routed in such a way as to exclude any type of potential trip hazard.
Remove the mains plug before cleaning the unit.
Do not operate the unit in hazardous or unstable conditions.
Set up the unit as far away as possible from equipment that generates electromagnetic or other interference. Portable and mobile HF-communication systems may influence electronic medical devices
All actions and operations described in this operation manual may only be carried out by professional personnel authorized by ADE or official appointed dealers.
Assembly and set up The ADE wheelchair scale is already been delivered in a fully assembled condition. Please insure that your scale is being placed free of any vibration and at a dry place. The indicator/ display of your ADE scale is able to be mounted on walls, on sockets, handrails or on stand-on devices for the table depending on the model you have selected. Please insure that your scale is being placed free of any vibration and at a dry place. Take the indicator and the cable extension out of the packaging and follow the instruction for assembly. Alignment of the scale 1) Place the scale on a plain, dry and level surface. 2) Undo the locknuts of the four feet of the scale and unscrew them until they got contact to the surface. 3) The air bubble in the spirit level must be right in the centre of the circle. 4) Please ensure that the scale has no other contact to the floor than by the four feet. 4) For the model with the Reference M500660-01 with a foldable column: a) Release the handle (pic 4a) at the bottom of the column or the handrail. b) Upright the column or the handrail. c) Finally re-tighten the handle (pic 4b) at the bottom of the column or the handrail again.
M500660-210326-Rev005-UM-e
5
(Pic 4a)
(4-b)
Please ensure that the foldable handrail and column of the scale is fixed and stable in order to avoid any unexpected moving of same. After transport or movement of the scale follow step 1 to 4 again.
Power supply Power is supplied either by the rechargeable battery in the display unit or by the mains power via the supplied power unit. The rechargeable battery pack is consisting of 4xAA NIMH rechargeable batteries. Use the AC adapter to load the rechargeable batteries. The imprinted voltage value at the mains adapter must match the local mains voltage. When the battery is being charged for the first time a sufficient charging period of approx. 24 hours is necessary to achieve a full battery load. The wall socket must be located near to the unit and be easily accessible. Please use only the mains adapter supplied by ADE. The mains adapter is charging the battery pack at the same time.
Replacing the rechargeable battery The battery compartment is located at the back of the display
1) Open the battery compartment. 2) Remove the battery cables carefully from the housing, connect the wires and insert the battery into the housing. 3) Close the battery compartment again. Only use original rechargeable batteries and charger units from ADE. The warranty does not cover any damage caused by using other types of batteries or charger units.
6
M500660-210326-Rev005-UM-e
Display
Front View and side view of the display
Display Description
Name
Description Status indicator for the current battery charge level. When this symbol appears is the scale in battery operation mode or will be charged. The scale is within the zero level.
The scale is at a standstill, meaning the weight is stable and no Standstill status indica- movement is on the platform. This is a mandatory precondition for functions such as resetting, printing and taring). tor (14) : In case the LCD is not displayed is the scale not in the standstill mode “Hold”
Displayed when the hold function is activated.
“Ib”
Displayed when the weighing unit is displayed in pounds.
“kg”
Displayed when the weight is displayed in kilograms.
M500660-210326-Rev005-UM-e
7
Key pad
Name
Description
HOLD
Activates the Hold-Function.
Output of data to serial communication port RS 232.
BMI
Enter, after the user steps on the scale, in order to proceed with the BMI mode.
UNIT
Switch button between primary kgs and secondary units Lbs.
TARE PRESET ZERO ON/OFF
Activates the tare function and the Preset Tare Function.
Switch the scale on/off and activates the zero function.
During key operation use this key to switch between the units.
During key operation use this key to switch between the units.
During key operation use this key to switch between the numbers downwards.
During key operation use this key to switch between the numbers upwards. Enter button to confirm the value.
Weighing
Correct weighing -
Press the start key ON/OFF to enter the normal operating mode of the scale with no load on the platform. An automatic function test will be perform and the zero setting of the scale. When `0.0´appears you are in the kg weighing mode. Additionally the display shows the symbol ` ´ to confirm stability of the zero setting.
-
The scale is now ready to be used.
-
Place the person to be weighed on the wheelchair and ensure that he is not moving.
-
The weight could be only determined when there is no contact of the feet to the floor. For this reason ensure that the person´s feet rest on the footrest.
-
After the weight is being displayed and the symbol ` achieved.
-
To turn off the scale press ON/OFF for a period of minimum 3 seconds. Using the scale on mains it may remain permanently switched on or can be shut off manually by pressing key.
´ appears the utmost accuracy has been
How to change the unit The scale offers the possibility to select between the weighing mode kg and lb. To change into another mode press the UNIT key. Please check weighing mode after switching on to get proper result.
8
M500660-210326-Rev005-UM-e
Tare function The tare function allows ignoring a fixed weight (e.g. clothes) that will be deducted after each weighing procedure. Proceed as follows: -
Press the start key with no load on the scale.
-
Place the additional weight of the object(s), which needs to be ignored, on the scale and store the result by pressing the TARE key.
-
The display is being reset to zero. The scale is now in NET weighing mode which is also indicated on the display.
-
Remove now the objects from the platform. The ignored weight is being displayed with a minus.
-
Place the person to be weighed and the additional weight (e.g. clothes) on the scale.
-
The scale determines the weight of the person without the additional weight. You may now use the scale for as many weighing procedures and the stored value will always be deducted as long as the scale is not turned off.
-
Enter the tare button again to get of the tare mode.
-
The tare value is likewise erased by the scale until pressing the tare button or being switched off.
Weighing with preset tare values To input a pre set tare value press down the tare button more than 3 seconds. NET will appear at the screen. Follow the same steps as before and place a load on the platform. The last stored tare value will appear at the display and ignored by the scale.
HOLD function The Hold function is being used if you like to hold the results at the display after the weight/load has been removed from the scale. Use this button to take care of the person to be weighed first and then note the weight. -
Press HOLD key while scale is under load.
-
“HOLD” is being displayed at the screen.
-
The weight remains saved in the display after unloading the scale.
-
For deactivation of the Hold function press again the HOLD key.
BMI (Body Mass Index) The Body Mass Index is the ratio between height and weight squared. The BMI is accepted world wide – also by WHO (World Health Organization) - helping to evaluate the state of nutrition and with that value also the state of health of a person. The result is a tolerance value. -
Start the scale by entering the ON/OFF Button. The scale is automatically being set to zero.
-
Help the person to be weighed onto the platform.
-
Take care that the feet are being rested on the footrest.
-
Wait until the weight is being displayed in a stable condition and
-
Press the BMI key shortly.
-
The scale starts automatically with the last input. Use the arrow buttons to enter the height of the person to be weighed. Confirm the height with the enter button .
-
The BMI is now being displayed at the indicator.
-
To correct the given height press the BMI button once again and the display returns to the height mode.
-
To get back into normal weighing mode and leave the BMI modus press the tare button.
M500660-210326-Rev005-UM-e
´ appears on the screen.
9
Evaluation of the Body Mass Index Compare the determined value which corresponds to those being used by the WHO. BMI below 18,5
The person is under the regular standard and does not got enough weight. An increase of weight is recommended. Please consult a professional.
BMI between 18,5 and 24,9
The person got a normal weight. If he feels well there is no need for a change.
BMI between 25 and 30
The person is slightly overweighed to pre-obese. A reduction of weight is indicated or, is necessary if the state of health is already affected (e.g. diabetes, high end blood pressure, heart diseases).
BMI over 30
Reduction of weight is imperative. To visit a doctor is highly recommended.
Connecting additional devices Connection of printer, PC or remote-display will be done with the RS-232 interface. For connection of additional devices kindly ask your local representative or authorized ADE dealer for further instructions.
Print out a record The scales must be at standstill. The standstill status symbol TER key to send the data to the serial interface.
must appear at the screen. Press the EN-
Note: Printing is not possible if the net weight is 0. The print function only produces one print-out. A reset is required to print out another copy.
Error Messages Error Codes
Description
Solution
Load cell wires to indicator are incorrectly connected; or shorted, or opened; or ADC/ load cell are being damaged
Make sure wires are in good order and condition and correctly connected. Replace load cell or ADC chip. Service is required.
Weight reading exceeds the zero limit after switching the scale on.
Make sure scale platform is empty.
Weight reading is below the zero limit after switching the scale on.
Install platform on scale. Otherwise the zero calibration/ adjustment have to be done by authorized personnel.
Weight reading exceeds load limit.
Reduce load on scale until weight value can be displayed.
Weight reading below load limit. Configuration or Calibration parameters are not correctly set. Capacity parameters are not correct.
Service by authorized personnel required. Contact your local dealer / supplier for calibration purpose. Contact your local dealer / supplier for a new set-up.
Calibration error. The incoming signal of loaded weight is too small, too big, unstable or un-linear.
Contact your local dealer / supplier for a new set-up.
Not turn on.
Power cord is not properly connected. Power outlet not supplying electricity. Batterie is discharged. Other failure.
Check power cord connections. Make sure power cord is plugged into the power outlet. Check power source. Replace batteries.
Cannot zero the display or will not zero when turned on.
Load on scale exceeds allowable limits. Load on scale is not stable. Load cell damage.
Remove load from the scale. Wait for load to become stable. Service required.
10
M500660-210326-Rev005-UM-e
No Battery symbol or display of Lo.bAt
Batteries are discharged.
Charge batteries.
Cleaning and Maintenance
Cleaning Before cleaning disconnect the scale from the socket. For cleaning just use a damp cloth or common disinfection fluids. Do not use any aggressive cleaning fluids, abrasive or acid cleaners or white spirits. During he process of cleaning take care that no water or cleaning fluids are penetrating the scale. Safety information Safe operation is no longer guaranteed
If the adapter shows visible signs of damage
If the adapter no longer functions
After extended storage in humid areas
In these cases, contact your service partner for your own safety! Never use other adapters and batteries than those provided with the device by the manufacturer. Otherwise there is a risk of interference with other electric medical devices! Disposal Electrical appliances and batteries should not be disposed of in domestic refuse. Always dispose of defective or unserviceable devices and batteries in accordance with the applicable laws and regulations! Please ensure batteries are empty prior to disposal, since otherwise there is a risk of short circuit!
Storage and Transport Please keep all original parts and pieces for an eventual sending back of the scale in order to avoid any possible damage during transport, which automatically leads to non warranty. Disconnect all cables before transport and avoid unnecessary damages.
Warranty – Liability ADE will repair or replace this product within a period of 2 years from date of delivery applying to defects occurred due to poor material or workmanship (presentation of invoice is necessary). All moveable parts – batteries, cables, mains units, rechargeable batteries etc. – are excluded. Warranty does not include natural wear, damages caused by incorrect or negligent handling, overuse or unsuitable or improper storage. The cost of transport will be borne by the customer should the equipment be located anywhere other then the customer’s premises. In the event of transport damage, claims under warranty can be accepted subject to using the complete original packaging and the equipment secured the way it was packed and secured originally. All the packaging should therefore be retained. Any warranty claim will be rejected if the equipment is opened by persons not authorized expressively by ADE to do so. Customers abroad are kindly requested to contact their local dealer for any warranty claim, customer service and spare parts.
M500660-210326-Rev005-UM-e
11
Technical specifications Power supply Mains Adapter:
AC Adapter: 6-9V DC, ≥500mA, central negative
Battery Requirement:
Rechargeable Lithium Ion Cylindrical (18650x2) Cell 3.7V 4000mAH
Serial Communications EDP Port Printer Port Both Ports
Full duplex RS-232 38400,19200,9600, 4800, 2400, 1200 bps Output-only RS-232 or active 20 mA current loop, 9600, 4800, 2400, 1200 bps 7 or 8 data bits; even, odd, or no parity
Environmental Operating Temperature Storage Temperature Humidity Altitude
–10 to +40°C (legal); –20 to +70°C 0–90% relative humidity 2000 m (6500 ft) maximum
Housing Dimensions Weight Material
12
18.6 cm x 9.5 cm x 10 cm 0.4 kg (without battery) ABS plastic, white
M500660-210326-Rev005-UM-e
Declaration of Conformity
Declaration of conformity by the manufacturer This product has been manufactured in accordance with the harmonized European standards. They conform to the regulations of the EC-directives listed below: EMC Directive 2014/30/EU Low Voltage Directive 2014/35/EU RoHS Directive 2011/65/EU in the currently valid versions. This declaration loses its validity if modifications are made to the scale without our approval. Hamburg, March 2021 ADE Germany GmbH Neuer Hoeltigbaum 15 22143 Hamburg / Germany
Manufacturer:
M500660-210326-Rev005-UM-e
ADE Germany GmbH Neuer Hoeltigbaum 15 22143 Hamburg / Germany ℡ +49 40 432 776-0 +49 40 432 776-10 [email protected] www.ade-germany.de
13
14
M500660-210326-Rev005-UM-e
M500660-210326-Rev005-UM-e
15
authorised dealer:
16
M500660-210326-Rev005-UM-e