NIOX® Panel PC application for NIOX MINO® User Manual
NIOX® Panel User Manual US 000200-02 (EPM-000194)
Table of contents
1 Important information ...2
7.1 Cautions ... 22
1.1 Before using NIOX® Panel... 2 1.2 About this manual ... 2 1.3 Responsible manufacturer and contacts ... 2 1.4 Warnings... 2 1.5 Intended use ... 3 1.6 Training requirements... 3
8 Buttons and descriptions ... 23
2 Installation and setup...3 2.1 Installation... 3 2.2 Setup... 5 3 Navigation ...8 3.1 Interface... 8 3.2 Main view ... 8 4 Using NIOX® Panel...10 4.1 Register patient ID (optional) ... 10 4.2 Measure FeNO... 10 4.3 Demonstration mode ... 12 4.4 Measure ambient NO... 13 4.5 Change settings ... 13 4.6 External Quality Control (QC) procedure ... 16 4.7 Turn off NIOX® Panel ... 20 5 Troubleshooting...20 5.1 Alert codes and actions ... 20 6 Preventive inspections ...22 7 Safety information ...22 1
Chapter 1 Important information
1
Important information
Information in this document is subject to change. Amendments will be made by Aerocrine AB as they occur.
NIOX® Panel is a visual aid for operating NIOX® MINO Airway Inflammation Monitor by allowing the user to operate the NIOX MINO from their PC.
• Aerocrine, NIOX and NIOX MINO are registered trademarks of Aerocrine AB.
1.1
1.3
Before using NIOX® Panel
NIOX® Panel PC application (NIOX® Panel) shall only be operated as directed in this manual by trained healthcare professionals. Trained status is achieved only after careful reading of this manual. Read the entire instructions for use and make certain that you fully understand the safety information. The NIOX MINO User Manual describes how to use the device. Symbol
Description
WARNING!
Indicates a potentially hazardous situation that, if not avoided, can result in bodily harm or injury.
CAUTION!
Indicates a potentially hazardous situation that, if not avoided, can damage a product or system, cause loss of data or harm to business.
Note
Alerts the reader to important information on the proper use of the product, user expectations, error situations, and actions related to these.
1.2
About this manual
NIOX® Panel User Manual - US 000200 (EPM-000194), version 02, May 2015, for application with version number 1.0.X.X and instruments with software version 23XX. To view version number for your instrument, refer to NIOX MINO User Manual. For instruction on how to update firmware see page 6.
2
Responsible manufacturer and contacts
Mailing address: Aerocrine AB P.O. Box 1024 SE-171 21 Solna, Sweden Visiting address: Råsundavägen 18 SE-169 67 Solna, Sweden www.aerocrine.com www.niox.com
1.4
Warnings
• NIOX® Panel shall only be operated by healthcare professionals. • Operate NIOX® Panel as stated in this manual. Aerocrine accepts no responsibility for damaged equipment or faulty results, if the equipment is not handled according to this manual. • When selecting an accessory for your NIOX® Panel product keep in mind that an accessory not recommended by Aerocrine may result in loss of performance, damage to your NIOX Panel product, fire, electric shock, injury or damage to other property. The product warranty does not cover product failure or damage resulting from use with non approved accessories. Aerocrine takes no responsibility for
NIOX® Panel User Manual US 000200-02
Chapter 2 Installation and setup
health and safety problems or other problems caused by the use of accessories not approved by Aerocrine. • Modification of NIOX® Panel application is not allowed. • Do not use damaged components. • Do not disconnect the instrument during firmware loading. • The patient filter is for single use only. • All warnings in the NIOX MINO User Manual are also applicable.
1.5
Intended use
NIOX® Panel is an optional software accessory allowing an additional display for the NIOX MINO® Airway Inflammation Monitor Refer to the NIOX MINO® User Manual for NIOX MINO® intended use.
1.6
Training requirements
NIOX® Panel shall only be used as directed in this manual by trained healthcare professionals. Trained status is achieved only after careful reading of this manual. The User Manual provides instructions on how to operate NIOX® Panel. It contains numbered step-by-step instructions with screens and illustrations. Choices within steps are displayed with bullet points.
2
Installation and setup
2.1
Installation
NIOX Panel System requirements • Windows® XP*(32 bit), Windows® Vista(32 bit), Windows® 7 or Windows® 8.0/8.1 Pro (RT versions excluded) • .NET Framework 4.0 • 1 GHz or faster processor • 256 MB RAM (512 MB RAM recommended) • 250 MB of video graphics RAM • 250 MB of available hard-disc space • 1024x768 screen resolution * Windows XP with Service Pack 3
Note:
If the computer does not fulfill the system requirements, see note on page 10.
The NIOX® Panel software is supplied on a USB storage device. 1. Insert the USB storage device in the computer’s USB port. 2. In the folder NIOX Panel 1.0.X.X select the file named setup.exe 3. If .NET Framework 4.0, VC++ 2010,XNA or SQL Server Compact is not installed, an installation wizard for each of the programs opens, one at the time. 4. Select Yes. 5. Select Accept license agreement. 3
Chapter 2 Installation and setup
6. Wait until the programs are installed. 7.
4
8. Select Accept license agreement.
The Installation wizard for NIOX® Panel opens. Click Next.
NIOX® Panel User Manual US 000200-02
Chapter 2 Installation and setup
9. Accept or select a different folder for the program. Then select to allow NIOX® Panel to be accessible for only yourself (logged in) or everyone who logs in on the same computer. Click Next.
2.2.1 Connect the instrument via USB 1. Turn on the PC and monitor.
10. Start the installation by clicking Next.
2. Plug the USB cable into the USB port on the instrument and connect it to the USB port on the PC.
The icon is shown on the display to indicate that a connection is established and NIOX MINO® is running remote controlled. Note:
It may take a few seconds for the connection to establish.
3. Select the Start or the Windows button normally found in the left lower hand corner of your monitor. 4. Select NIOX Panel from the program list.
2.2.2 Note: Note:
If the firmware version is 2005 or earlier contact Aerocrine Customer Support.
Note:
Always connect the instrument by USB cable for firmware update.
11. Wait until the installation is finished. 12. When the installation is complete, click Finish.
Firmware update If NIOX MINO firmware is a version < 23XX the instrument must be updated.
13. The application is now available on the start menu.
2.2
Setup
Note:
In order for the NIOX® Panel to be able to communicate with NIOX MINO®, a USB cable is needed. Only USB cables supplied by Aerocrine may be used. 5
Chapter 2 Installation and setup
1. A view for automatic update is opened.
3. When the firmware has been uploaded the instrument will disconnect and perform an internal update, after completion the instrument will reconnect automatically. 4. The NIOX Panel opens. Note:
When starting NIOX Panel for the first time the connectivity details dialog box opens
2. Press OK and wait for the upload to finish.
2.2.3 NIOX Panel connectivity module The connectivity module in NIOX Panel utilizes Microsoft’s secure cloud service, Microsoft Azure, to automatically transmit technical data from the device via the internet to Aerocrine. CAUTION! Do not disconnect the instrument during load of firmware.
6
Technical data such as time stamp, alert codes and number of remaining measurements on the device and sensor are received. This information will ensure better service and support.
NIOX® Panel User Manual US 000200-02
Chapter 2 Installation and setup
Aerocrine recommends completion of the details dialog box to allow receipt of technical data and provide better service and support functions to its customers. Mandatory fields are marked with a *. Complete these, leave the box checked, and click OK to continue.
If the connection to Microsoft Azure is lost or the user has chosen to not send technical data, the cloud icon is crossed over . To decide at a later stage to allow Aerocrine to collect data, press cancel and the dialog box will open again next time NIOX Panel is started, or click on the cloud icon in the status bar. To reject collecting of Aerocrine technical data (not recommended) uncheck the box in the bottom of the window and click OK. Note:
Only technical data and no patient data is collected by Aerocrine.
Changing contact details To edit contact details click on the cloud icon open the contact details dialog box.
When connection to Microsoft Azure is established a cloud icon shown in the status bar.
in the status bar to
is
7
Chapter 3 Navigation
3
Navigation
3.1
Interface
3.2.1
Main menu
This section describes the main view. Buttons and symbols are further described on page 23.
3.2
Main view
(a) Patient ID entry, (b) Demo, (c) Quality Control, (d) Settings
3.2.2
Status bar
(a) USB icon, (b) QC warning, (c) Scrubber symbol (about to expire or has expired-blinking), (d) Instr symbol (about to expire or has expiredblinking), (e) Sound mute, (f) Remaining measurements on Sensor, (g) Connected to Microsoft Azure, (h) Time
(A) Status bar (time, sensor info and alert), (B) Animation in three steps, (C) Patient ID (if entered), (D) Start measurement button, (E) Main menu
8
NIOX® Panel User Manual US 000200-02
Chapter 3 Navigation
3.2.3
Settings view
3.2.4
Measurement view
(a) Measurement sequence number, (b) FeNO value in ppb (parts per billion), (c) Measurement mode 10 seconds exhalation time, (d) QC alert (if not performed the current day), (e) OK - returns to main view
(A) User Settings, (B) Device info (see page 14), (C) Volume (see page 13), (D) Scrubber info (see page 15), (E) Ambient measurement (see page 13), (F) Patient log (see page 15), (G) QC log (see page 19), (H) QC info (see page 19), (I) Return to main view
9
Chapter 4 Using NIOX® Panel
4
Using NIOX® Panel
4. Instruct the patient to perform steps 5-8. 5. Empty lungs.
4.1
Register patient ID (optional)
6. Close the lips around the NIOX MINO® patient filter so that no air leakage occurs.
Note:
If Patient ID is used, it has to be entered before each measurement, even if it is the same patient. Local Regulations on Patient information privacy must be considered when using unique patient identifiers.
7. Inhale deeply through the filter. The screen shows an animation where the sun rises. Note:
There are three optional animations (Cloud, Balloon or Meter). For instructions on how to change animation see page 12.
1. Select the Patient ID button from the main menu. 2. Enter up to 12 characters (alpha or numeric).
4.2
Measure FeNO
Verify proper operating conditions before performing a measurement with NIOX MINO®. A basic preventive inspection is recommended before each use (see page 22). Make sure a successful daily QC is performed, see page 17. Refer to NIOX MINO User manual for more detailed directions.
4.2.1 Perform FeNO measurement WARNING! The patient filter is for single use only.
8. Exhale slowly through the filter while keeping the balloon within the limits as indicated on the screen (the white lines).
1. Attach a new patient filter. 2. Turn the PC-monitor facing the patient. 3. Select Start measurement button on the display. Note:
10
If the computer does not fulfill the system requirements it may still be possible to run NIOX Panel. To start a measurement select the Win XP animation. It is recommended to press the “start measurement” button to initiate the measurement.
NIOX® Panel User Manual US 000200-02
Chapter 4 Using NIOX® Panel
9. The visual incentive, instruments- audio signals and top light guide the user to the correct exhalation pressure.
The instrument will analyze the sample and generate a result in approximately one and a half minutes.
A continuous sound indicates correct pressure with a frequency proportional to the pressure. • An intermittent high frequency sound - too strong pressure • An intermittent low frequency sound - too weak pressure Exhalation with: Correct pressure
Pressure too strong
Pressure too weak
CAUTION! Do not disconnect the instrument from the PC during measurement and analyzing process.
10. Exhale until the balloon has reached the end of the screen.
(A) Patient ID - if applicable, (B) Measurement mode, (C) FeNO value in ppb (parts per billion), (D) Measurement sequence number, (E) OK - returns to main view 11. Remove the patient filter.
11
Chapter 4 Using NIOX® Panel
4.3
Demonstration mode
4. Select OK to confirm the changes.
To help professionals in guiding patients, NIOX Panel contains an animated demonstration with a visual guide of the different stages of a measurement procedure. 1. Select the Animation button on the main menu. A view to select demo animation appears.
The Close button returns to the main menu. a.
2. Select which animation mode you which to use. Then select DEMO.
Inhalation through the patient filter. The sun rises.
b. Exhalation through the patient filter with correct pressure.
3. Select the Forward button to move to the following sequence.
12
c.
Exhalation through the patient filter with pressure too weak.
NIOX® Panel User Manual US 000200-02
Chapter 4 Using NIOX® Panel
d. Exhalation through the patient filter with pressure too strong.
4.4
Measure ambient NO
Note:
An ambient measurement is counted as one measurement on NIOX MINO® Sensor and the instrument.
5. The analyzing view is visible until the measurement is finished and the result is displayed.
1. Attach a patient filter. 2. Select the Settings button on the main menu. 3. Select Ambient Measurement button. 4. Select the Start Ambient measurement button.
(A) Ambient measurement value (in ppb), (B) Measurement mode, (C) Measurement sequence number, (D) OK - returns to main view
4.5
Change settings
4.5.1 Change volume 1. Select the Settings button on the main menu. 2. Select the Volume button. 13
Chapter 4 Using NIOX® Panel
3. Select the decrease/increase to adjust volume. 4. Select the Return button to confirm changes and return to settings. Note:
The status bar indicates mute status when the sound volume is set to zero.
4.5.2 View alert logs Alerts are stored in the instrument, and can be viewed at any time. The alert codes are for use by Aerocrine Customer Support. 1. Select the Settings button on the main menu. 2. Select the Device info button. 3. Select Alert log button. Note:
It may take a few seconds to load all alert codes.
(A) Alert code, (B) Return button - returns to settings view, (C) Date of alert, (D) Scroll bar, (E) Download service data (only upon request by Aerocrine) 4. Select the Return button to return to settings.
4.5.3 View instrument information Detailed information about the instrument and Sensor can be viewed. 1. Select the Settings button on the main menu. 2. Select the Device info button.
14
NIOX® Panel User Manual US 000200-02
Chapter 4 Using NIOX® Panel
3. This opens the Instrument information view displaying the following:
(A) Numbers of remaining measurements on the instrument, (B) Instrument serial number, (C) Instrument software version, (D) Instrument expiration date, (E) Return to settings, (F) Numbers of remaining measurements on the Sensor, (G) Sensor serial number, (H) Sensor expiration date, (I) Alert log button
3. This opens the Record view.The log will display the following:
(A) Measurement sequence number, (B) Patient ID - if entered, (C) Measurement mode, (D) Return to settings, (E) FeNO value, (F) Measurement date and time, (G) QC Alert, (H) Forward, Backward
4.5.4 View patient measurement logs All measurement results are stored in the instrument and can be viewed at any time.
4. Browse through the measurement logs using the forward and backward buttons. Select the Return button to return to settings.
1. Select the Settings button on the main menu.
4.5.5 View scrubber information 1. Select the Settings button on the main menu.
2. Select Patient log button.
2. Select the Scrubber button.
15
Chapter 4 Using NIOX® Panel
The Scrubber information view opens.
4.6
External Quality Control (QC) procedure
The external Quality Control is one of the procedures that ensures the system is operating within its specifications. Note:
The Quality Control function in NIOX MINO must always be activated as a daily QC measurement is mandatory when the instrument is clinically used.
A QC icon in the status bar on the screen indicates that a daily external QC assessment is needed or that there are not any currently qualified QC individuals who have completed their initial 4 qualification attempts.
(A) Scrubber icon, (B) Number of measurements left on Scrubber, (C) Return button, (D) Expiration date on Scrubber, (E) Change scrubber button 3. If all measurements on the Scrubber have been used, select the Change Scrubber button. 4. The Change Scrubber view opens. 5. Change Scrubber as instructed and select OK to verify.
The external Quality Control consists of two parts. One positive control from a qualified staff member with a stable FeNO value providing a normal biological FeNO sample and a negative control consisting of a NO free gas sample, generated from ambient air. To perform QC measurements with NIOX Panel a connection to the NIOX database is needed. This makes it possible to create the QC users and store the QC measurements in the database. For instructions on creating QC users in the database see NIOX Patient User Manual. Performing the QC measurements with NIOX Panel gives the QC User (after qualification) the opportunity to qualify a new instrument with just one measurement or several instruments on the same day.
4.6.1 Selection and qualification of QC testers A minimum of one individual (two individuals are recommended) needs to qualify for this procedure. If possible, identify a third individual as a backup. Identify the staff members who will perform the Quality Control and meet the following criteria:
16
NIOX® Panel User Manual US 000200-02
Chapter 4 Using NIOX® Panel
• Over 18 years of age.
• Avoid strenuous exercise at least 1 hour before the measurement.
• No ongoing cold or known airway disease.
• Preferably do not perform a measurement in case of:
• Non-smoker. • Preferably no allergies (except seasonal, see below) or asthma.
• Ongoing cold. • Acute seasonal allergy. • Always attach a new patient filter for each new QC tester.
A QC tester will be qualified over the course of four days within a seven day period.
1. Select QC.
• Expected stable FeNO values between 5 and 40 ppb.
Note:
After a QC tester has been qualified, if the most recent QC measurement is older than 30 days, then the qualification is suspended and the QC tester needs to re-qualify according to the qualification procedure. Perform four QC measurements, one per day within seven days, according to the QC measurement section, in order to qualify a QC tester.
A mean value is calculated from the three qualifying measurements that must be between 5-40 ppb for the QC tester to be qualified. The QC measurement on the fourth day (daily QC) must be within ± 10 ppb from the mean value and the negative control approved. If this has been met the Quality Control has passed for that staff member and the instrument is ready for clinical use. The moving mean value is recalculated when the QC tester performs a QC measurement after seven days.
4.6.2 QC measurement This procedure applies for qualification and daily QC measurements. The instrument will prompt for a daily QC procedure by showing in the status bar. Always consider the following in order to obtain reliable results.
2. Select the correct QC User. 3. Empty lungs. 4. Close the lips around the NIOX MINO® patient filter so that no air leakage occurs. 5. Inhale deeply through the filter. The screen shows an animation where the sun rises.
Before any measurement: • Avoid nitrate rich food within 3 hrs before the measurement. 17
Chapter 4 Using NIOX® Panel
7. Remove the patient filter and immediately attach the QC plug.
8. Press 6. Exhale slowly through the filter until the balloon reaches the end of the screen. The processing view appears.
18
to proceed.
9. The process indicator is visible until the measurement has been analyzed. The QC control result is displayed.
NIOX® Panel User Manual US 000200-02
Chapter 4 Using NIOX® Panel
(A) QC user limits (mean value +/- 10 ppb), (B) Control result in ppb, (C) QC measurement sequence number, (D) QC measurement date and time, (E) QC user (database user) 10. Remove the QC plug. Note:
Press
4.6.3 View QC logs 1. Select the Settings button on the main menu. 2. Select the QC log button. The QC log window opens.
to return to main menu.
During the qualification days of a new QC tester the result is displayed as presented below.
3. Browse through the measurement logs using the backward and forward buttons.
4.6.4 View QC information After day 3 the QC qualification procedure is ended. A qualified tester is displayed in the QC info view (A) Measured NO value in ppb, (B) Status button (green = within specifications, white = not within specifications), (C) Date of measurement, (D) QC measurement sequence number, (E) QC user Press
1. Select the Settings button on the main menu. 2. Select QC info button.
to return to main menu.
Repeat the QC test if the positive and/or the negative control fails. If the QC failure persists, discontinue use of NIOX MINO® and contact Aerocrine Customer Support.
19