User Manual
45 Pages
Preview
Page 1
Remember It is important to adhere to the following specified conditions: •
Ambient temperature: +10°C to +35°C
•
Humidity: 20% to 80% RH (non-condensing)
•
Mobile phones, cordless phones and gas emitting appliances might interfere with the instrument and should therefore be kept away from the instrument. Interference could make it impossible to perform a measurement.
•
Exhaled breath contains water vapour which can condense inside the instrument. When excessively used in a short period, there is a risk for condensation of water inside NIOX VERO®. Normally a maximum of 10 exhalations/hour can be performed with NIOX VERO® during continuous use. However, it is possible to perform 20 exhalations in one hour if the instrument is paused for a minimum of 30 minutes prior to the next session of exhalations. Exhalations include failed and successful measurements.
•
Avoid spilling water or other fluids on the instrument or sensor.
•
Always use a closed case or bag (NIOX VERO® bag recommended) for transportation and storage of NIOX VERO®.
•
It is recommended, after inserting a new sensor, to wait for three hours with the instrument switched on before performing a measurement.
•
Operational life-time for NIOX VERO® Instrument: Maximum 5 years in use or 15 000 measurements or the expiration date stated on the device, whichever comes first. 5 years in use shall be defined as starting from the first NO measurement taken using the device.
•
Operational life-time for NIOX VERO® Sensor: Maximum 12 months after opening package and installed in NIOX VERO® or expiration date as stated on the sensor, whichever comes first.
WARNING! Use of substances containing alcohol close to the NIOX VERO® instrument may cause erroneous measurement results. DO NOT clean the instrument or handle with alcohol or any spray or wipe containing alcohol! Do not use substances containing alcohol on or close to the NIOX VERO® instrument. This includes any cleaning agents used to clean the facility, or other equipment in the area, as well as alcohol wipes or sprays used on patients. CAUTION!: Do not use NIOX VERO® in the proximity of areas where volatile substances such as organic fluids or disinfectants are being used. Special attention should be paid to aerosols and disinfection baths, either open vessels or ultrasonic baths.
Table of contents
1 Important information ...3 1.1 Before using NIOX VERO® Airway Inflammation Monitor ... 3 1.2 About this manual... 3 1.3 Compliance ... 3 1.4 Responsible manufacturer and contacts ... 3 1.5 Warnings ... 3 1.6 Intended use ... 4
6.2 Installation of NIOX® Panel ... 20 6.3 Connect to a PC via USB... 20 6.4 Connect to a PC via Bluetooth... 20 6.5 Setup... 21 6.6 Firmware update ... 22 6.7 Using NIOX Panel... 23 7 Troubleshooting ... 24
2 Product description ...5
7.1 Alert codes and actions... 24
2.1 NIOX VERO® accessories and parts... 5 2.2 Instrument ... 6
8 Preventive care ... 28
3 Installation and set up ...7 4 User interface ...10 4.1 Main and settings view ... 10 4.2 Main View... 10 4.3 Settings view ... 11 5 Using NIOX VERO® ...12 5.1 Start the instrument from power save mode ... 12 5.2 Register patient ID (optional)... 12 5.3 Measure FeNO... 12 5.4 Demonstration mode ... 15 5.5 Measure ambient NO ... 16 5.6 Change settings ... 17 5.7 Turn off the instrument ... 19 6 Using NIOX VERO® with NIOX® Panel ...19 6.1 Warnings ... 19
NIOX VERO® User Manual English 000191-14
8.1 General care ... 28 8.2 Change disposals ... 29 8.3 Operational life-time... 31 8.4 Disposal of instrument and accessories ... 31 8.5 Return shipments... 32 9 Safety information ... 32 9.1 Warnings... 32 9.2 Cautions... 32 9.3 Substances disturbing FeNO measurement ... 33 9.4 Electromagnetic immunity... 33 9.5 Electromagnetic emissions ... 33 9.6 Operating conditions ... 34 10 Reference information ... 35 10.1 Buttons and descriptions... 35 10.2 Symbols and descriptions ... 35 10.3 Symbol explanation... 36 11 Technical data ... 37
1
Table of contents
11.1 Dimensions and weight ... 37 11.2 Electrical data... 37 11.3 Noise level... 37 11.4 Exhaled NO - performance data... 37 11.5 Linearity... 37 11.6 Precision... 37 11.7 Accuracy... 37 11.8 Method comparison... 37 11.9 Inhalation parameters... 37 11.10 Exhalation parameters ... 38 11.11 Essential performance... 38 11.12 Memory capacity ... 38 11.13 Patient filter ... 38 11.14 Bluetooth ... 38 11.15 Rechargeable battery capacity... 38 11.16 Instructions for transport and storage... 39 12 NIOX VERO® parts and accessories ...40 12.1 Parts included in NIOX VERO® package (Article No. 12-1100) ... 40 12.2 Accessories ... 40 13 Vigilance ...41
2
NIOX VERO® User Manual English 000191-14
Chapter 1 Important information
1
Important information
1.1
Before using NIOX VERO® Airway Inflammation Monitor
NIOX VERO® may only be operated as directed in this manual by trained healthcare professionals. Trained status is achieved only after careful reading of this manual. Read the entire instructions for use and make certain that you fully understand the safety information.
It contains numbered step-by-step instructions with screens and illustrations. Choices within steps are displayed with bullet points.
1.3
Compliance
NIOX VERO® is CE-marked according to In Vitro Diagnostics Device Directive 98/79/EC. NIOX VERO® is RoHS compliant. NIOX Panel is CE marked according to In Vitro Diagnostics Directive 98/79/EC.
Symbol
Description
1.4
WARNING
Indicates a potentially hazardous situation that, if not avoided, can result in bodily harm or injury.
CAUTION
Indicates a potentially hazardous situation that, if not avoided, can damage a product or system, cause loss of data or harm to business.
Mailing address: Circassia AB P.O. Box 3006 SE-750 03 Uppsala, Sweden
Note
Alerts the reader to important information on the proper use of the product, user expectations, error situations, and actions related to these.
1.2
About this manual
NIOX VERO® User Manual - English 000191, version 14, May 2020, for instruments with software version 1D2x-xxxx and 1E2x-xxxx. For instruction on how to view the software version number installed in the instrument, see page 18.
Visiting Address: Hansellisgatan 13 SE-754 50 Uppsala www.circassia.com www.niox.com
1.5 •
Information in this document is subject to change. Amendments will be made by Circassia as they occur. The User Manual provides instructions on how to operate NIOX VERO®. •
000191-14 NIOX VERO® User Manual English
Responsible manufacturer and contacts
Warnings
Do not use substances containing alcohol on or close to the NIOX VERO® instrument. This includes any cleaning agents used to clean the facility, or other equipment in the area, as well as alcohol wipes or sprays used on patients. NIOX VERO® should only be operated by healthcare professionals.
3
Chapter 1 Important information •
•
• • • • • • • • •
• • •
4
Operate NIOX VERO® as stated in this manual. Circassia accepts no responsibility for damaged equipment or faulty results, if the equipment is not handled according to this manual. When selecting an accessory for your NIOX VERO® product keep in mind that an accessory not recommended by Circassia may result in loss of performance, damage to your NIOX VERO® product, fire, electric shock, injury or damage to other property. The product warranty does not cover product failure or damage resulting from use with non approved accessories. Circassia takes no responsibility for health and safety problems or other problems caused by the use of accessories not approved by Circassia. NIOX VERO® should not be used adjacent to or stacked with other electronic equipment. Only use the power supply provided. Pull the plug when disconnecting NIOX VERO® from the power outlet. Use only the breathing handle supplied by Circassia. No modification of NIOX VERO® instrument, handle or sensor is allowed. Do not drop the instrument or subject it to strong impact. Do not use a damaged NIOX VERO® instrument or damaged components. Keep the instrument and sensor out of water. Ensure that no liquid is spilled or dropped on the instrument or sensor. Do not heat or dispose of the instrument or sensor in fire. Refer to “Disposal of instrument and accessories” on page 31. NIOX VERO® and the NO scrubber in the breathing handle contains potassium permanganate. Used or expired instruments and breathing handles should be disposed of as hazardous waste in accordance with local waste disposal regulations. The breathing handle must not be used after expiration date. Patient filters should be used immediately after opening. NIOX VERO® Sensor contains chemicals that could be harmful if
• • • • •
•
swallowed. Be careful when opening the sensor can. The inside of the opening may have sharp edges. Do not touch or clean the white sensor membrane. After inserting a new sensor it is recommended to wait for three hours with the instrument switched on before performing a measurement. Make sure to use the correct measurement mode, otherwise incorrect FeNO results might be obtained. Patient filters are labeled for single use only. Always use a new patient filter for each patient. Reuse between patients could increase the risk of cross-contamination or cross-infection. The same filter can be reused in one patient for multiple attempts in the same session. Do not use NIOX VERO® in the proximity of areas where volatile substances such as organic fluids or disinfectants are being used. Special attention should be paid to aerosols and disinfection baths, either open vessels or ultrasonic baths. Do not use the instrument in the presence of flammable anesthetic, vapors or liquids.
1.6
Intended use
NIOX VERO measures Nitric Oxide in human breath (Fractional exhaled Nitric Oxide, FeNO) and Nasal Nitric Oxide (nNO) in the aspirated air from the nasal cavity. FeNO FeNO is increased in some airway inflammatory processes such as asthma and decreases in response to anti-inflammatory treatment. FeNO measurements with NIOX VERO are quantitative, non-invasive, simple and safe and should be used as part of regular assessment and monitoring of patients with these conditions. NIOX VERO is suitable for patients age 4 and above for FeNO measurements.
000191-14 NIOX VERO® User Manual English
Chapter 2 Product description As measurement requires patient cooperation, some children below the age of 7 may require additional coaching and encouragement. NIOX VERO should be used as directed in the NIOX VERO User Manual. •
CAUTION! NIOX VERO for FeNO measurement can be operated with two different exhalation times, 10 seconds and 6 seconds. The 10 second test is the preferred mode. For children who are not able to perform the 10 second test, the 6 second test is an alternative. The 6 second test should be used with caution in patients over the age of 10 years. It should not be used in adult patients. Incorrect use of the 6 second exhalation mode may result in falsely low FeNO values, which can lead to incorrect clinical decisions. nNO
2 2.1
Product description NIOX VERO® accessories and parts
Nasal Nitric Oxide has been shown to be decreased in patients with Primary Ciliary Dyskinesia (PCD), and measurement of nNO can assist in the identification of cases of PCD according to ERS guidelines1. Measurement of nNO with the NIOX VERO Nasal Measurement Mode is noninvasive, simple, safe and repeatable in patients age 5 and above when measured according to the NIOX VERO Nasal Measurement Mode User Manual. Suspected cases of PCD following screening with nNO should be confirmed according to published recommendations for PCD diagnosis and management. 1. Lucas JS, Barbato A, Collins SA, et al. European Respiratory Society guidelines for
the diagnosis of primary ciliary dyskinesia. Eur Respir J 2017; 49: 1601090.
For instructions on operation of the Nasal Nitric Oxide measurement mode, refer to the NIOX VERO Nasal User Manual.
000191-14 NIOX VERO® User Manual English
(A) Breathing handle and handle cap, (B) Sensor (supplied separately), (C) Instrument (including stand), (D) Rechargeable battery, (E) NIOX® Panel USB memory stick, (F) USB cable, (G) Power adapter and power cord, (H) Patient filter (supplied separately) Note: Only accessories and parts supplied by Circassia may be used.
5
Chapter 2 Product description
2.2
Instrument
(H) ON/OFF button, (I) Power adapter port, (J) USB port
(N) Breathing handle holder, (O) Breathing handle port
(K) Battery LED - lit when battery is charging, (L) Standby LED - blinking in Standby/Sleep mode, (M) Touch panel Display
6
000191-14 NIOX VERO® User Manual English
Chapter 3 Installation and set up
3
Installation and set up
Open the package with care. Prior to installation, check that the package contains all the parts (see page 5). A screwdriver is required for opening the compartment lid and installation of sensor and battery. Remove the plastic film from the display. 1.
WARNING! Open the sensor can with care. The inside of the opening may have sharp edges. 3. Open the sensor package.
Carefully place the instrument with the display facing down on a flat and clean surface, then unscrew and remove the compartment lid. There is a taper on the side of the lid for better grip when opening.
WARNING! Do not touch or clean the white sensor membrane. CAUTION! The sensor should only be stored in its original unopened package or installed in a NIOX VERO® instrument. 4. 2.
Insert the sensor and turn the swivel clockwise until locked.
Open the sensor can.
000191-14 NIOX VERO® User Manual English
7
Chapter 3 Installation and set up 5. Open the battery package. Note: Only use the correct rechargeable battery supplied by Circassia. (Type No BJ-G510039AA, Article No 12-1150)
Note:
Only attach the breathing handle supplied by Circassia. Article No 12-1010
Note:
Use care not to bend the handle tube.
6.
Insert the rechargeable battery and replace the lid. Tighten the screw by using a screwdriver.
Note:
The triangle should not be visible when assembled correctly.
7.
Take the breathing handle tube and push the end of the tube into the breathing handle port slowly until the triangle is no longer visible. The breathing handle and the patient filter are Applied parts Type B.
8
000191-14 NIOX VERO® User Manual English
Chapter 3 Installation and set up 8.
Attach the power adapter to the instrument and then to the power outlet. When installing the unit, either use a socket outlet with a readily accessible power switch, or connect the AC cord plug to an easily accessible socket outlet near the equipment. If a fault should occur during operation of the unit, use the power switch to cut the power supply, or remove the AC cord plug. Note: Only use the power adapter supplied by Circassia with the instrument. Article No 12-1120.
9.
Position the instrument with the stand folded out.
10. Start the instrument by sliding the ON/OFF button to ON and allow the instrument to start up and perform the internal check and measurement procedures.
000191-14 NIOX VERO® User Manual English
CAUTION! After inserting a new sensor it is recommended to wait for three hours with the instrument switched on before performing a measurement. 11. When the internal check is completed, the main menu appears.
12. Select the Settings button on the main menu. 13. Select Time and date.
9
Chapter 4 User interface This opens the time and date setting view.
4 4.1
User interface Main and settings view
This section describes the main view, settings view, menus and symbols. Buttons and symbols are further described on page 35.
4.2
Main View
14. Select between 12h US and 24h ISO time and date format. 15. Set time by pressing the button for hour. It changes color to blue. Change the value to the current hour by pressing the increase or decrease buttons. Repeat this procedure for minute, year, month and day. 16. Select OK to accept the changes and return to the main menu. The Undo button
closes the view without saving any changes.
17. Select the Settings button on the main menu. 18. Select the Breathing handle button. This opens the Breathing handle view. 19. Select the Reset Breathing handle button. The breathing handle information view opens to confirm insertion of the breathing handle. 20. Select the OK button to confirm insertion of a new breathing handle. This sets the remaining measurements to 1000 and expiry date one year from the current date. The Return button returns to Settings view without registering change.
10
(A) Status bar, (B) Instructive demonstration, (C) Patient ID, (D) Main menu, (E) Start measurement button
000191-14 NIOX VERO® User Manual English
Chapter 4 User interface
4.2.1
Main menu
4.3
Settings view
(a) Measurement mode 10s/6s (only shown when 6s is an option, for more information see “Enable/disable 6 second (6s) measurement mode” on page 19, (b) Demo, (c) Patient ID entry, (d) Settings
4.2.2
Status bar
(e) Battery status, (f) Bluetooth enabled (in this position a USB connection may be indicated instead ), (g) Breathing handle has expired or is about to expire - blinking symbol, (h) Instrument has expired or is about to expire blinking symbol, (i) Sound disabled, (j) Sensor status and number of remaining measurements, (k) Temperature outside of specification, (l) Humidity outside of specification, (m) Time
000191-14 NIOX VERO® User Manual English
(A) Modes configuration - see page 18, (B) Volume settings - see page 17, (C) Alert log - see page 18, (D) Instrument & Sensor info - see page 18, (E) Time and date settings - see page 17, (F) Measurement log - see page 17, (G) Ambient measurement - see page 16, (H) Breathing handle status and settings - see page 29, (I) Return to main view
11
Chapter 5 Using NIOX VERO®
5
Using NIOX VERO®
5.1
Start the instrument from power save mode
2. 3. 4.
If NIOX VERO® is in standby or sleep mode simply touch the display to activate it.
Enter up to 12 characters (alpha or numeric). Select the ABC-button to activate a keyboard with the alphabet. The 123-button changes view back to the numerical keyboard. Select OK button to confirm the registration. Use the Erase button to erase. Use the Undo button to undo a registration.
5.3
Measure FeNO
Verify proper preparations before performing a measurement with NIOX VERO®. A basic preventive inspection is recommended before each use (see page 29).
5.2
Register patient ID (optional)
Note:
If Patient ID is used, it must be entered before each measurement, even if it is the same patient. Local Regulations on Patient information privacy must be considered when using unique patient identifiers.
1.
12
Select the Register patient ID button from the main menu.
WARNING! Always use a new patient filter for each patient. Reuse between patients could increase the risk of cross-contamination or cross-infection. The same filter can be reused in one patient for multiple attempts in the same session.
5.3.1
Preparation for measurement
1.
Lift the breathing handle from the holder and remove the handle cap.
2.
Obtain a new patient filter. Attach the patient filter to the breathing handle. Make sure to twist the patient filter in place until it clicks into place.
000191-14 NIOX VERO® User Manual English
Chapter 5 Using NIOX VERO® Store the patient filters in the original box prior to use.
5.3.2 1.
Empty the lungs by breathing out thoroughly.
Note:
Do NOT use sharp objects to open the patient filter packaging. Do not touch the filter membrane.
2.
Close the lips around the mouthpiece on the patient filter so that no air leakage occurs.
Note:
Patient filters should be used immediately after opening.
Note:
There is a risk of leakage if the filter is not correctly attached to the breathing handle and this may result in incorrect measurement values.
Note:
Do not switch OFF the instrument during measurement procedure.
Note:
3.
Give the breathing handle to the patient and guide the patient to provide a breath sample as described in the next section.
000191-14 NIOX VERO® User Manual English
Measurement
3.
Inhale deeply through the patient filter to total lung capacity. During inhalation, the cloud on the display moves upwards. Note: The procedure is activated by inhaling air from the handle or by pressing the start measurement button.
13
Chapter 5 Using NIOX VERO® 4.
Exhale slowly through the filter while keeping the cloud within the limits as indicated on the display (the white lines).
6.
Exhale until the cloud has passed the flag.
7.
The instrument will analyse the sample and generate a result in approximately one minute. Note: Do not exhale or inhale through the patient filter during the analysis process. 5.
The instrument display and audio signals guide the user to the correct exhalation pressure.
A continuous sound indicates correct pressure with a frequency proportional to the pressure. An intermittent high frequency sound - too strong pressure An intermittent low frequency sound - too weak pressure Exhalation with: Pressure correct Pressure too strong Pressure too weak 8.
14
The result is then displayed: (A) Patient ID - if applicable, (B) FeNO value in ppb (parts per billion), (C) Measurement mode 10s/6s, (D) Measurement sequence number, (E) OK - returns to main view.
000191-14 NIOX VERO® User Manual English
Chapter 5 Using NIOX VERO®
5.3.3
Perform 6s NO measurement
CAUTION! NIOX VERO can be operated with two different exhalation times, 10 seconds and 6 seconds. The 10 second test is the preferred mode. For children who are not able to perform the 10 second test, the 6 second test is an alternative. • The 6 second test should be used with caution in patients over the age of 10 years. It should not be used in adult patients. Incorrect use of the 6s exhalation mode may result in falsely low FeNO values, which can lead to incorrect clinical decisions. 6 second measurement mode is not activated by default, refer to “Enable/disable 6 second (6s) measurement mode” on page 19.
1.
Select the Animation button on the main menu.
2.
Select which animation to use (Cloud, Balloon or Meter)
3.
Select the Demo button.
4.
Select the Forward button to move to the following sequence.
5.
The Undo button closes the demonstration and returns to animation select.
6.
Select OK button
7.
The Undo button returns to the main menu without saving changes.
a.
Inhalation through the breathing handle.
b.
Exhalation through the breathing handle with correct pressure.
•
1.
Change to 6s Measurement mode by selecting the 10s button (green man symbol) on the main menu.
2.
The button changes to 6s measurement mode (orange, small child symbol).
3.
The 6s measurement mode is illustrated with an orange start button.
4.
Perform measurement as instructed in the “5.3.2 Measurement” section.
5.
Wait for the result.
6.
The result screen displays the icon for 6s measurement.
Note:
The device will always return to the default 10s mode after a 6s measurement.
5.4
Demonstration mode
To help professionals in guiding patients, the instrument contains three animated demonstrations with visual and audio guides of the different stages of a measurement procedure.
000191-14 NIOX VERO® User Manual English
to confirm the changes
15
Chapter 5 Using NIOX VERO®
c.
Exhalation through the breathing handle with pressure too weak.
d.
Exhalation through the breathing handle with pressure too strong.
5.5
Measure ambient NO
Note:
An ambient measurement may be requested by technical support during troubleshooting.
Note:
An ambient measurement is counted as one measurement on NIOX VERO® Sensor and the instrument.
16
1.
Attach a patient filter to the breathing handle until it clicks into place.
2.
Select the Settings button on the main menu.
3.
Select Ambient Measurement button.
4.
Select the Start measurement button.
5.
The progress bar is visible until the measurement is finished and the result is displayed: Ambient measurement value (in ppb), measurement mode and measurement sequence number.
000191-14 NIOX VERO® User Manual English
Chapter 5 Using NIOX VERO®
5.6
Change settings
5.6.3
5.6.1
Change time and date
All measurement results are stored in the instrument and can be viewed at any time.
View measurement logs
1.
Select the Settings button on the main menu.
1.
Select the Settings button on the main menu.
2.
Select the Time and Date button. For more details refer to page 9.
2.
Select Patient Measurements log view button.
3.
The selected log will display the following:
5.6.2 1.
Change sound volume
Select the Settings button on the main menu.
2.
Select the Sound button.
3.
The settings for sound and volume opens.
4.
Select Decrease/Increase to adjust volume.
5.
The volume bar indicates the set volume.
6.
Select the OK button to save settings and return to the settings view.
(A) Patient ID - if defined, (B) FeNO value, (C) Measurement date and time, (D) Measurement mode 10s/6s, (E) Measurement sequence number, (F) Return to settings (G) Backward, (H) Forward
4.
Browse through the measurement logs using the Backward and Forward buttons.
5.
Select the Return button to return to settings.
The Undo button closes the view without saving changes. 7.
The status bar indicates mute status when the sound volume is set to zero.
000191-14 NIOX VERO® User Manual English
17
Chapter 5 Using NIOX VERO®
5.6.4
View alert logs
3.
This opens the Instrument information view displaying the following:
Alerts are stored in the instrument and can be viewed at any time. The alert codes are for Circassia Technical Support use. 1.
Select the Settings button on the main menu.
2.
Select Alert log button.
(A) Numbers of remaining measurements on the instrument, (B) Instrument serial number, (C) Software version number (D) Instrument expiration date, (E) Return to settings, (F) Numbers of remaining measurements on the sensor, (G) Sensor serial number, (H) Sensor expiration date, (I) Enter configuration code (only used on request from Circassia) (A) Alert code (for technical support purpose only), (B) Date and time of alert, (C) Scroll list (blue) (D) Return - returns to previous view, (E) Backward (F) Forward 3. Select the Return button to return to settings.
5.6.5
5.6.6
Turn QC functionality on or off
Note:
An optional External Quality Control (QC) procedure is available for NIOX VERO®. Refer to the NIOX VERO® External Quality Control user manual. Contact your local Circassia representative for further information.
View instrument information
Detailed information about the instrument and sensor can be viewed.
1.
Select the Settings button on the main menu.
1.
Select the Settings button on the main menu.
2.
Select the Instrument button.
2. 3.
Select the Modes configuration button. Select the check box to activate QC functionality.
18
000191-14 NIOX VERO® User Manual English
Chapter 6 Using NIOX VERO® with NIOX® Panel
5.6.7
Enable/disable 6 second (6s) measurement mode
1.
Select Settings in the main menu.
2.
Select Modes configuration.
3.
Check the 10s/6s icon to enable using the 6s mode. Uncheck to disable.
4.
Press OK.
5.6.8
Activate Nasal measurement mode
To unlock Nasal measurement mode in NIOX VERO a configuration code is required.
6
Your NIOX VERO® instrument can be used together with NIOX® Panel. NIOX® Panel is a PC application and visual aid allowing you to operate the instrument from your PC. The NIOX Panel software is an accessory to an In Vitro Diagnostic Device.
6.1 • •
Contact the local Circassia sales representative for further information. •
5.7 1.
Turn off the instrument To turn off the instrument, slide the ON/OFF button to OFF.
• • • Note:
Before transportation remove the used patient filter (if still attached) and attach the handle cap.
Note:
Always use a closed bag or case (NIOX VERO® bag recommended) for transportation and storage of the instrument.
000191-14 NIOX VERO® User Manual English
Using NIOX VERO® with NIOX® Panel
Warnings
NIOX® Panel shall only be operated by trained healthcare professionals. Operate NIOX® Panel as stated in this manual. Circassia accepts no responsibility for damaged equipment or faulty results, if the equipment is not handled according to this manual. When selecting an accessory for your NIOX® Panel product keep in mind that an accessory not recommended by Circassia may result in loss of performance, damage to your NIOX® Panel product, fire, electric shock, injury or damage to other property. The product warranty does not cover product failure or damage resulting from use with non approved accessories. Circassia takes no responsibility for health and safety problems or other problems caused by the use of accessories not approved by Circassia. If the equipment is used in a manner not specified by Circassia, the protection provided by the equipment may be impaired. Modification of NIOX® Panel application is not allowed. Do not use damaged components.
19