Medical Endoscopy  TJF Type Q180V Pre-cleaning your TJF-Q180V The TJF-Q180V has a number of features enabling easier reprocessing. This quick reference guide provides an overview of the main improvements to the pre-cleaning procedure.  At the light source: • The distal cap of the TJF-Q180V is fixed and is therefore not removed prior to pre-cleaning • The sealed forceps elevator wire of the TJF-Q180V means that the elevator wire channel does not require flushing and rinsing  During manual cleaning: • Use one of the recommended brushes to brush the front and rear side of the forceps elevator • The MAJ-1888 brush can be used for heavy soiling or delayed reprocessing situations and enables deeper access to the forceps elevator • The sealed forceps elevator wire of the TJF-Q180V means that the elevator wire channel does not require flushing and rinsing  Before automated reprocessing:  This sheet is for quick reference only. For detailed reprocessing instructions, please refer to the TJF-Q180V reprocessing manual.  Specifications, design and accessories are subject to change without any notice or obligation on the part of the manufacturer.  Postbox 10 49 08 20034 Hamburg, Germany www.olympus-europa.com  E0429476 · 2.000 · 07/10 · PR  • Set and lock the forceps elevator to 45° before placing the endoscope into an automated washer disinfector to enable cleaning and disinfection of both sides of the forceps elevator