Instruction Manual
22 Pages
Preview
Page 1
ふ
AD-6121A
1WMPD4001194B
This Manual and Marks All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Institute: Product Safety Signs and Labels). The meanings are as follows:
WARNING CAUTION
A potentially hazardous situation which, if not avoided, could result in death or serious injury. A potentially hazardous situation which, if not avoided, may result in minor or moderate injury.
This is a hazard alert mark. This manual is subject to change without notice, at any time, to improve the product. The product specifications and the contents of this manual are subject to change without any obligation on the part of the manufacturer to update past products. Windows is a registered trademark of Microsoft Corporation.
2007
All rights reserved.
CONTENTS 1.
PRECAUTIONS ...2 1.1. Safety Precautions ...2 1.2. Handling Precautions ...2 1.2.1. Installation...2 1.2.2. Precautions for use ...2
2.
COMPLIANCE WITH COUNCIL DIRECTIVES ...3
3.
FEATURES ...3
4.
PART NAMES AND ACCESSORIES ...4 4.1.
5.
Display Symbols and Switches ...5
BEFORE USE ...6 5.1. Preparation Procedure ...6 5.1.1. Using the AC adapter ...6 5.1.2. Using the batteries ...6
6.
USING THE SCALE ...7 6.1. 6.2.
7.
Basic Weighing ...7 Weighing with a Tare ...8
CALIBRATION...9 7.1. 7.2. 7.3.
8.
About Calibration...9 Calibration Using a Calibration Weight...9 Adjusting the Gravitational Acceleration...12
DATA COMMUNICATION...13 8.1. RS-232C Serial Interface ...13 8.1.1. Interface specifications ...13 8.1.2. Data format ...13 8.1.3. Output timing...13 8.2. Connecting to a Personal Computer ...14 8.2.1. Connection procedure...14 8.2.2. Communication example ...14
9.
MAINTENANCE ...15 9.1. 9.2. 9.3.
Maintaining the Performance Quality ...15 Cleaning the Scale ...15 Storing the Scale ...15
10.
BEFORE ASKING FOR REPAIR ...16
11.
SPECIFICATIONS ...17
11.1. 11.2. 11.3. 12.
Specifications...17 External Dimensions...17 Options ...18
APPENDIX ...19
12.1. 12.2.
The Value of Gravity at Various Locations ...19 World Map ...20 -1-
AD-6121A
1. PRECAUTIONS 1.1. Safety Precautions CAUTION Do not place the scale on a soft, wet or slippery floor. Place the scale on a solid and level floor. Do not place the scale in areas where it can be tripped over or kicked. Do not step on the scale if your feet are wet. Step on the center portion of the scale gently. Do not step on the display window. When carrying the scale, be sure to hold it by the handle, and do not swing around or bump the scale. Do not disassemble or modify the scale. Keep the scale free of moisture, dust and foreign materials. Only qualified personnel can repair the scale. Attempting repairs yourself may cause a fire or damage to the scale. Damage caused by attempting to do the repair yourself will void the warranty.
1.2. Handling Precautions 1.2.1.
Installation
Place the scale on a flat floor free from vibrations and drafts. The floor should be solid enough to support the weight of the scale and the person being weighed. Place the scale where it is not exposed to direct sunlight. Place the scale where it is not exposed to external noises and strong electromagnetic waves. Place the scale where there are no corrosive, flammable, or explosive substances. Keep the temperature and the humidity of the installation site within the specified operating temperature range/humidity.
1.2.2.
Precautions for use
The maximum weighing capacity of the scale is 150 kg. Do not place anything on the scale that is beyond the weighing capacity. Stand still on the scale during weighing operation. Keep the cables of the AC adapter and other options away from the scale top surface. If they are in contact to the scale surface, a weighing error may occur. Press the ON/OFF switch and the ZERO switch gently. Do not kick them. AD-6121A
-2-
2. COMPLIANCE WITH COUNCIL DIRECTIVES This device conforms to the following Council Directives. Council directive 89/336/EEC
EN61326
EMC directive
Council directive 73/23/EEC
EN61010-1
Low voltage directive
3. FEATURES AD-6121A is a verified compact, lightweight scale that can be carried. Three displaying modes are available: Personal
Holds the weight on the display for about 5 seconds after the person being weighed steps off the scale.
Screening
Holds the weight on the display while the person being weighed is on the scale.
Scale
Displays the weight without holding it. Suitable to the weighing with a tare.
The displaying direction can be changed. The direction for a person other than the person being weighed to read and record the weight.
The direction for the person being weighed to read the weight
The display is equipped with the backlight so that the weight can be read in dark areas. (Only when the AC adapter is used.) The buzzer indicates the weighing conditions. The buzzer can be switched off. Power is supplied by either batteries or the AC adapter. Automatic power shut-off function, when batteries are used, turns off the power automatically after two minutes of non-operation.
-3-
AD-6121A
4. PART NAMES AND ACCESSORIES Handle LCD Spirit level
Weighing surface ZERO switch ON/OFF switch
Leveling foot
Buzzer switch
Terminal
Battery compartment
Mode selection switch AC adapter input jack Display direction switch Leveling foot ON/OFF switch
ZERO switch
Accessories
AC adapter
Instruction manual
Note The shape of the adapter plug may differ from that of the illustration, depending on the country where the scale is used. Please confirm that the AC adapter type is correct for your local voltage and receptacle type. AD-6121A
-4-
4.1. Display Symbols and Switches Stable mark Unit Display hold mark
Tare mark
Unit Low battery mark
Switches
Display symbols Symbols Stable mark
Description This symbol appears when the displayed weight is stable.
Tare mark
This symbol appears when tare operation is in progress.
Low battery mark
This symbol appears when the batteries are drained. Replace the batteries with four new type AA batteries.
Display hold mark
This symbol appears when the scale holds the weight on the display.
Switches Switches
Description
ON/OFF switch
Turns the scale on and off. Sets the displayed weight to zero. When this switch is pressed after placing an object of 3 kg or more on the scale, the tare mark turns on.
ZERO switch Mode selection switch
Selects how the scale holds the displayed weight.
Personal
Holds the weight when it becomes stable and displays the value for about 5 seconds after the person being weighed steps off the scale. Used when the person being weighed reads the value. When changes of 3 kg or more occurs in the value, or 5 seconds after the person being weighed steps off the scale, the display hold is released.
Screening
Holds the weight when it becomes stable and displays the value as long as the person being weighed is on the scale. When the person steps off the scale, the display hold is released. Used when a person other than the person being weighed reads the value. When changes of 3 kg or more occurs in the value, the display hold is released.
Scale
Does not hold the value. Used for weighing an object other than body weight.
Display direction switch
Selects the display direction. Used when the person being weighed reads the weight. Used when a person other than the person being weighed reads the weight.
Buzzer switch
Selects whether to sound the buzzer or not.
On
Sounds the buzzer when the scale is turned on or the scale holds the weight.
Off
Does not sound the buzzer.
Note The settings of the Mode selection switch, Display direction switch and Buzzer switch become effective after the ON/OFF switch is turned off and then turned on again. -5-
AD-6121A
5. BEFORE USE 5.1. Preparation Procedure Spirit level
1 Use the accessory AC adapter or four type AA alkaline batteries. Good No good
2 Place the scale on a solid floor and level the scale using the leveling feet so that the bubble in the spirit level is centered.
5.1.1.
Leveling foot
Using the AC adapter
1 Insert the AC adapter plug into the AC adapter input jack, and the AC adapter into an electrical outlet.
AC adapter input jack
AC adapter
5.1.2.
Using the batteries CAUTION
When installing the batteries, match the + (positive) and – (negative) terminals to those indicated in the battery compartment. Use four type AA batteries. Do not mix old and new batteries. Remove the batteries if the scale is not to be used for a long period of time. The batteries may leak and cause a malfunction. If the battery liquid spills on the body, wash the body part under running water immediately. When the Low battery mark appears, replace the batteries with four new type AA batteries. Please note that the Low battery mark does not appear when the batteries are completely drained. Even when the batteries are installed, connecting the AC adapter will make the AC adapter effective. When the batteries are used, the backlight does not turn on.
1 Remove the battery compartment cover. 2 Install four new type AA batteries into the battery compartment, taking care that the polarities (+) and (-) are correct. 3 Replace the battery compartment cover. AD-6121A
-6-
6. USING THE SCALE 6.1. Basic Weighing 1 Make sure that nothing is placed on the scale. AC adapter
2 Press the ON/OFF switch to turn the scale on. All the display segments turn on and then the zero display appears.
Nothing on the scale
ON/OFF switch
When the AC adapter is used, the display backlight turns on.
ZERO switch
3 Press the ZERO switch. If the displayed value is not zero, press the ZERO switch.
4 Step on the scale gently and read the weight.
Zero display
Step on
The scale displays the weight. When the value is stable, the Stable mark
turns on. Stable mark
Weight
5 Step off the scale gently. Step off
6 To weigh again, repeat steps 3 to 5.
7 Press the ON/OFF switch to turn the scale off.
6.2.
ON/OFF switch
-7-
AD-6121A
6.2. Weighing with a Tare In weighing with a tare, the scale stores the tare weight, turns on the tare mark net weight. The tare range is 3 kg to 150 kg.
and displays the
Net weight = Gross weight – Tare weight (Example of tare items: Clothes, plate and container) For weighing an object other than body weight, the scale mode, that does not hold the weight, is convenient.
Procedure for weighing with a tare 1 Make sure that nothing is placed on the scale.
AC adapter
2 Press the ON/OFF switch to turn the scale on.
ON/OFF switch
Nothing on the scale
All the display segments turn on and then the zero display appears. If the displayed value is not zero, press the ZERO
Place the tare
switch.
3 Place the tare on the scale gently.
ZERO switch
Place a tare item of 3 kg or more on the scale. When the value is stable, the Stable mark
Zero display
turns on.
4 Press the ZERO switch. The tare mark
Step on
turns on and the display is set to zero.
5 Place the object to be weighed on the scale and read the weight.
Stable mark
The scale displays the net weight. When the weight is stable, the Stable mark
turns on.
Net value (Weight)
6 Remove the object weighed with a tare from the scale. Step off
7 Press the ON/OFF switch to turn the scale off. ON/OFF switch
Releasing “Weighing with a tare” (Canceling the tare mark
)
With nothing on the scale, press the ZERO switch. Press the ON/OFF switch to turn off the scale and press the ON/OFF switch to turn on the scale again.
AD-6121A
-8-
7. CALIBRATION 7.1. About Calibration Calibration is the function to adjust the scale so that it can weigh accurately. When the verification seal is broken, re- verification is required. CAL switch cover
CAL switch
Break the verification seal and locate the CAL switch cover on the scale bottom. Slide the CAL switch cover in the direction indicated by the arrow. The CAL switch is on the board inside. When pressing the CAL switch, do not use a sharp object such as a ballpoint pen. It may cause the scale to malfunction. Bottom of the scale
7.2. Calibration Using a Calibration Weight 1 While holding down the CAL and ZERO switches, press and hold the ON/OFF switch for about 2 seconds. All of the display segments turn on. Release the switches after about 2 seconds.
2 While all the display segments are on (about 5 seconds), press the CAL switch 3 times. The display indicates [- - - -]. If the CAL switch is not pressed 3 times within about 5 seconds after step 1, the power will be turned off automatically.
3 Press the ZERO switch. The display indicates [C-0].
4 Press the ZERO switch several times until [C-5] is displayed. Each time the ZERO switch is pressed, the number increases by 1; [C-0] [C-1] [C-2] ….
5 Press the ON/OFF switch. The display indicates [20], which means the temperature for calibration is 20°C. Here, adjust the value as necessary.
-9-
AD-6121A
6 Press the ZERO switch to adjust the temperature setting. When the Display direction switch is set to ↑, each time the ZERO switch is pressed, the number increases by 1. When the Display direction switch is set to ↓, each time the ZERO switch is pressed, the number decreases by 1.
Display direction switch
The display value increments
The display value decrements
7 Press the ON/OFF switch. The scale stores the new temperature setting for calibration and [C-6] is displayed.
8 Press the ON/OFF switch. The display indicates [9798], which means the gravitational acceleration setting for the calibration site is 9.798 m/s2. Here, adjust the value as necessary.
9 Press the ZERO switch to adjust the gravitational acceleration. When the Display direction switch is set to , each time the ZERO switch is pressed, the number increases by 1. When the Display direction switch is set to , each time the ZERO switch is pressed, the number decreases by 1.
Note
Display direction switch
The display value increments
The display value decrements
The gravity values for various locations are listed on page 19.
10 Press the ON/OFF switch. The scale stores the new gravitational acceleration value and [C-7] is displayed.
11 Press the ZERO switch several times until [C-10] is displayed. Each time the ZERO switch is pressed, the number increases by 1; [C-8] [C-9] [C-10] ….
12 Press the ON/OFF switch. The display indicates the AD count for the current output of the thermo-sensitive resistor. Wait for the Stable mark to turn on.
13 When the Stable mark is displayed, press the ON/OFF switch. The scale stores the AD count for the current output of the thermo-sensitive resistor and moves to the next item [C-11]. When the ZERO switch is pressed in place of the ON/OFF switch, the scale moves to [C-11] without storing the AD count. AD-6121A
- 10 -
Stable mark
14 With [C-11] displayed, press the ON/OFF switch. The display indicates [CAL0]. Stable mark
15 Make sure that nothing is placed on the scale and wait for the Stable mark to turn on. 16 Press the ON/OFF switch. The scale stores the zero point data and moves to the next item [C-12]. When the ZERO switch is pressed in place of the ON/OFF switch, the scale moves to [C-12] without storing the zero point data.
17 With [C-12] displayed, press the ON/OFF switch. The display indicates [CAL1]. Stable mark
18 Place a 150-kg (weighing capacity) weight on the scale and wait for the Stable mark to turn on. 19 Press the ON/OFF switch. The scale stores the span data and moves to the next item [END]. When the ZERO switch is pressed in place of the ON/OFF switch, the scale moves to [END] without storing the span data.
20 After the scale indicates [END], the power is turned off automatically. Remove the weight from the scale. Calibration has been completed.
Note A room with a temperature of around 25 ºC with little temperature change is needed. Perform calibration only after the temperature of a scale and the temperature of the room are the same. Place the scale on a level surface, and after pre-load do not move the scale until calibration is over.
- 11 -
AD-6121A
7.3. Adjusting the Gravitational Acceleration Adjust the gravitational acceleration to that of the installation site. When the scale is transported to a different area, the adjustment is required.
Note The gravity values for various locations are listed on page 19.
1 Set the Mode selection switch to [Personal]. 2 While holding down the CAL switch, press and hold the ON/OFF switch for about 2 seconds.
Mode selection switch
All of the display segments turn on. Release the switches after about 2 seconds.
3 While all the display segments are on (about 5 seconds), press the CAL switch 3 times. The display indicates [9798], which means the gravitational acceleration setting for the installation site is 9.798 m/s2. If the CAL switch is not pressed 3 times within about 5 seconds after step 2, the power will be turned off automatically.
4 Press the ZERO switch to adjust the gravitational acceleration.
Display direction switch
When the Display direction switch is set to , each time the ZERO switch is pressed, the number increases by 1. When the Display direction switch is set to , each time the ZERO switch is pressed, the number decreases by 1.
5 Press the ON/OFF switch.
The display value decrements
The scale stores the new gravitational acceleration value and the power is turned off automatically.
AD-6121A
The display value increments
- 12 -
8. DATA COMMUNICATION 8.1. RS-232C Serial Interface The AD-6121A scale has an RS-232C serial interface installed as standard. Using the optional RS-232C cable (AX-KO2238), the scale can send the data to external devices such as a personal computer.
8.1.1.
Interface specifications
Transmission system Transmission form Baud rate Data bits Parity Stop bit Code Terminator
8.1.2.
: EIA RS-232C : Asynchronous, half duplex : 2400 bps : 7 bits : 1 bit, Even, Odd : 1 bit : ASCII : CR LF (CR: 0Dh, LF: 0Ah)
Data format
Data
Unit
Terminator
Stable header: Added when the Stable mark
is turned on.
Unstable header: Added when the weight value is not stable. Overload header: Added when the weight value is beyond the weighing capacity. When overloading, [E] or [-E] appears on the display. Space (
20h)
Carriage return (CR: 0Dh) Line feed (LF: 0Ah)
8.1.3.
Output timing Mode selection switch
Output timing
Personal, Screening
Outputs the weight value once when the scale holds it on the display.
Scale
Outputs the weight value 4 or 5 times per second after zero display.
- 13 -
AD-6121A
8.2. Connecting to a Personal Computer 8.2.1.
Connection procedure
The below is an example of connecting the scale to a Windows-based personal computer.
1 Insert the optional RS-232C cable (AX-KO2238) into the LINE (External output terminal) jack Optional RS-232C cable AX-KO2238 located on the bottom of the scale. 2 Insert the D-Sub 9-pin connector of the RS-232C cable into the RS-232C port (COM1 port) located on the personal computer.
8.2.2.
External output terminal LINE
Personal computer
Communication example
The below is an example of communication with a Windows-based personal computer.
1 Make a connection between the scale and the personal computer as described above. 2 Select [Start], [Program], [Accessories] and [Hyper Terminal]. If Hyper Terminal is not installed, install it using [Add Remove Programs]. 3 Double-click the Hyper Terminal icon to start Hyper Terminal. Click [Cancel] when a guide to the modem installation appears. 4 A new window opens. Give it a proper name and click [OK]. 5 The settings window appears. Select [Direct to COM1] for connection method and click [OK]. 6 Set the port on the COM1 properties window and click [OK]. Baud rate Data bits Parity Stop bit Flow control
: 2400 bps : 7 bits : 1 bit, Even, Odd : 1 bit : Hardware
7 Press the ON/OFF switch to turn the scale on. 8 The data is output as described in “8.1.3 Output timing” and displayed on the personal computer. AD-6121A
- 14 -
9. MAINTENANCE 9.1. Maintaining the Performance Quality Check the scale regularly to make sure that it weighs correctly, and perform inspection or calibration as necessary. For details on inspection or calibration, contact the local A&D dealer.
9.2. Cleaning the Scale To clean the scale, use a soft cloth that is moistened with water and a mild detergent solution. After moistening, the cloth must be squeezed or rung out, so that it is not dripping wet. Do not use organic solvents such as thinner, benzine or methanol to clean the scale. Do not use a disinfectant that contains chlorine to clean the scale. Do not expose the scale to strong ultraviolet radiation or high temperature water vapor.
9.3. Storing the Scale When storing the scale: Make sure that nothing is placed on the scale. Make sure that the ON/OFF switch is not left pressed. Remove the batteries. Do not store the scale: Where it is exposed to high temperature or high humidity. Where salinity, sulfur content or corrosive gases exist. Where chemicals are stored.
- 15 -
AD-6121A
10. BEFORE ASKING FOR REPAIR Before asking for repair, refer to the table below. When
What to do
Nothing appears on the display, even
Check whether the AC adapter or batteries are installed correctly.
when the scale is turned on.
Check whether the batteries are drained.
[----] remains on the display and does
Check that nothing is placed on the scale.
not change.
Check that nothing is touching the scale.
The difference between an estimated
Check that the scale is leveled.
weight and the actual weight is large.
Check that nothing is touching the scale.
The switches do not function.
Disconnect the AC adapter and connect it again. Then, turn the
The display does not change even
power on.
when the switches are pressed.
Remove the batteries and insert them again. Then, turn the power on.
[E-1] , [E-2] or [E-4] appears on the display. The buzzer remains sounding. [ E ] appears.
Ask for repair. Check whether the ON/OFF switch is left pressed. This error appears when the object on the scale is beyond the weighing capacity. Remove the object from the scale. This error appears when the weight value is less than 3 kg.
[ -E ] appears
Check that nothing is placed on the scale and nothing is touching the scale. Then, turn the scale off and turn it on again.
AD-6121A
- 16 -
11. SPECIFICATIONS 11.1. Specifications Type
AD-6121A
Maximum weighing capacity
150 kg
Minimum display
100 g
Minimum weighing capacity
2 kg
Maximum tare
Maximum weighing capacity
Display
LCD with backlight (Backlight: Only when the AC adapter is used.) Displaying direction can be changed.
Operating temperature range/humidity
+5°C to +35°C, 85%RH or less, No condensation
Power supply
AC adapter or four type AA batteries (Not included)
Battery life
Approx. 1000 weighing operations (Varies depending on the type and ambient temperature.)
Buzzer
Buzzer switch to turn the buzzer on and off
External dimensions
314 (D) x 315 (W) x 57 (H) mm
Weight
Approx. 3.0 kg
11.2. External Dimensions
Unit: mm
- 17 -
AD-6121A
11.3. Options RS-232C cable Use
AX-KO2238 RS-232C cable D-Sub 9-pin
Length
AD-6121A
2m
- 18 -