Guide and Parts Catalogue
16 Pages
Preview
Page 1
CRITICAL CARE
AquaVENT® Heated Breathing Circuits
Why Humidification? Normal Respiration During normal respiration, inhaled gases are warmed and moistened as they pass from the upper respiratory tract to the lower airways of the lungs.
22oC, 7mg.L
37oC, 44mg.L
When inhaled gases reach the tracheo-bronchial bifurcation they have been heated to 37° Celsius and are fully saturated with water vapour at 44mg/H2O.L of gas - commonly referred to as saturation vapour density.
Mucociliary Transport System The epithelia of the mucosal lining of the upper and lower airways is designed to provide filtered gases inhaled at optimal humidity levels and received at the alveoli at or close to saturation vapour density. The epithelia also provides defence mechanisms by trapping particles and microorganisms in mucous and transporting these particles by way of cilia vibration to the oropharynx where they can be ingested or expectorated. Interactions between cilia and mucous, which establish the mucociliary transport system, can be disrupted by various factors including inadequate humidification of inhaled gases.(1) Mucous
Periciliary Fluid Layer
Cilia
Ciliated Cell
Goblet Cell
Ciliated Cell
Ciliated Cell
Submucosal glands
01
2
Why Humidification?
Why Humidification in the Clinical Setting? Humidity management in patients with bypassed airways is essential to maintain normal lung function and help protect against hospital acquired infections. Active humidification is a gold standard for longer term patients in critical care. Oxygen and medical air delivered invasively as dry gases, over sustained periods of time, will damage the epithelia of the respiratory tract and prevent adequate functioning of the mucosal tissue in contact with the dry gases. Inadequate humidification, in patients with bypassed airways, can disrupt secretion mobility across the mucocilliary transport system, leading to thickening of
1. 2. 3.
secretions, infective secretions, atelectasis or systemic infection(2) and has been reported to increase work of breathing and cause dehydration.(3) Maintaining secretion transport across the mucosal epithelium leads to better patient outcomes(1). This can be achieved by delivering optimal humidification to the mechanically ventilated patient with bypassed airways. In respect of non-intubated patients, provision of non-invasive respiratory therapies without adequate humidification can cause patients to experience discomfort and irritation of the upper airway mucosa, causing lack of patient co-operation and poor therapy outcomes.
Nakagawa et al. (2012) Humidification in the intensive care unit, Part 8, Humidification and Mucus Transport in Critical Patients: Clinical and Therapeutic Implications pp. 217-225 Tim Op’t Holt. Humidification During Mechanical Ventilation: Current Trends and Controversies. [online] Available at: <http://www.clinicalfoundations. org/assets/foundations1.pdf> [Accessed 23 November 2012]. Ryan et al. (2002) Energy balance in the intubated human airway is an indicator of optimal gas conditioning. Critical Care Med. Vol.30 No.2 pp. 355-361
3
02
AquaVENT® Technology for Invasive Ventilation 6
4 5
Heater Humidifier • For use with heated breathing circuits. • Invasive or non-invasive mode selection. • Automatic temperature selection. • Low and high temperature alarm. • Over-temperature protection. • Real-time temperature tracking display allows heated plate chamber and airway temperature to be viewed. • Digital display. • Straightforward operation and set up.
Chamber Humidity Output
05
4
AquaVENT® Technology 8 1
Humidity Tracker (Typical Values)
7
3
2
Heated Wire • • • •
Unique figure-of-eight heated wire design. Electrical heating elements minimize formation of water condensate. Heating elements energize the tubing, creating an insulator to heat loss. Wire design and configuration within the circuit ensures that inspiratory gases heat to 40°C before cooling to 37°C at >90% humidity at the tracheo-bronchial bifurcation and gas temperature rises along the expiratory limb to minimize formation of water condensate in the tubing or ventilator exhaust path.
Chamber • •
• • •
Internal chamber design ensures ideal conditions to allow optimal humidification of gases from the ventilator. Compared to other commonly used chambers the AquaVENT® humidification chamber produces a level of humidity closer to 44mg/H2O.L over a range of tidal volumes. Refills automatically. Pinch clamp on feed line allows circuit leak testing with chamber in situ. Compatible with many common heater humidifiers.
5
06
AquaVENT® Technology
for Non-Invasive Ventilation
High flow humidified oxygen therapy •
Provision of high flow humidified oxygen therapy via nasal cannulae reduces CO2 rebreathing, decreases work of breathing, improves oxygenation through alveolar recruitment and causes less discomfort to the patient than low flow oxygen therapy. (1)
•
Armstrong Medical’s circuits for oxygen therapy are compatible with a variety of air oxygen blenders and nasal cannulae patient interfaces.
(1) Ricard JD. High Flow Nasal Oxygen in Acute Respiratory Failure. Minerva Anestesiologica 2012;78:836-41
Bi-Level & CPAP •
• • • •
Failure to humidify gas, even during short term non-invasive ventilation, results in patient discomfort and may affect the success of therapy. Humidification improves patient comfort and therefore tolerance and enhances airway function and secretion removal (2). An extensive range of heated circuits and mask interfaces suitable for the provision of Bi-level ventilatory support and High Flow CPAP are available. Circuits can be customised to include masks, headgear and PEEP Valves. Customisable circuit kits improve the efficiency of therapy implementation. As an alternative flow source to MaxVenturi®, MaxBlend® provides flows up to 120L.min and is available for High Flow CPAP applications.
(2) Branson R, Gentile M. Is humidification always necessary during noninvasive ventilation in the hospital? Respir Care 2012; 55(2):209-216.
05
6
AquaVENT® Technology
MaxVenturi® MaxVenturi® has been specially designed to interface with high flow humidification systems. This unit provides users with the ability to mix ambient air and oxygen without the need for piped medical grade air, making it the perfect solution for high flow therapy in almost any clinical environment.
Key Features • All-in-one oxygen/air mixer, MaxO2+ analyser and flow meter. • Flow rates of up to 60L.min. • Long battery life (approximately 5,000 hours). • Independent adjustable O2% and flow. • Built-in pole clamp.
7
06
AquaVENT® Protected by BioCote® Antimicrobial Silver Technology AquaVENT® heated breathing circuits are supplied with BioCote®, an antimicrobial silver additive to limit the numbers of microbes on the surface of the breathing circuit, protecting it from microbial colonisation. BioCote® technology in AquaVENT® heated breathing circuits uses inorganic silver ion particles dispersed homogenously throughout the polymer used to form components of the breathing circuit. BioCote® technology is based on silver, a safe, natural and sustainable antimicrobial.
3
2
1
Disclaimer:
www.biocote.com 13 11
8
This product does not protect users or others against disease causing bacteria, germs, viruses or other harmful organisms. This technology is not a substitute for good hygiene and/or cleaning practises.
BioCote®
How does BioCote® work? 1. Silver Ions Cause Conformational Changes in Microbial Proteins Silver ions combine with microbial proteins located in the cell wall and cytoplasm, which interferes with their normal functioning. These proteins are vital for the healthy functioning of the cell. This loss of function results in an overall inhibition of microbial growth.
2. Silver Ions Interfere with DNA Replication Silver ions are known to interfere in the way microbes copy their genetic material or DNA. Microbes, like most living cells, have to make copies of their DNA during reproduction. Silver ions stop the microbes replicating by blocking the copying of their genetic material.
3. Silver Ions Promote Formation of Reactive Oxygen Species Silver ions are known to promote the formation of harmful chemicals called reactive oxygen species (ROS) inside microbial cells. ROS are harmful to living cells because they cause significant damage to cell structures. Damage caused by ROS is a major contributor to functional decline, characteristic of aging, that results in further inhibition of microbial growth.
BioCote® antimicrobial protection is effective against a broad spectrum of micro-organisms.
9
11
10
xp
ra to r
pi
ns
®E
®I
AMVC1775/115 AMVC1775/110 AMVC1775/107 AMVC1775/106 AMVC1775/108 AMVC1774/106 AMVC1774/101
(1) Oxygen Analyser T-piece & Safety Valve
n
de
ol
eH
Tu b
sio
Ex te n
t-F ix
ira to r
r Lim Co nv b Tu er sio bi ng n Co Cl i nn ps Ci ec rc ui to tH rs (1 ) an ge r
CP AP
CP AP
fo r
Co m
He
yL im b at ed yL Hu Ex im m pi id b ra ific t o a r tio Hu yL n m Lim imb id ific b at W io at n er Ch Tr am ap be Fil te r r
n
No
EN T
EN T
aV
Aq u
aV
Aq u
Therapy Solutions
Invasive Ventilation
AMVC1775/115
4 4 4 4 4 4 4
1 1 1 1 1 1 1
AMVC1792/004 AMVC1792/113
io
at
ific
r
Ch am be
n
yL im b
Lim b
n
ira to r
sp
In
io
at
ific
T®
ur eL in Bi e -le ve lM as CO k 2 E xh ala tio Tu bi n Va ng lve Cl ip s
Pr es s
Fil te r
id
m
Hu
id
m
Hu
aV EN
Aq u
Therapy Solutions
Non - Invasive Ventilation
AMVC1792/004
3 3
1112
AMCP1409/021 AMCP1409/018 AMCP1409/012 AMCP1409/003 AMCPUK01213
11
12
2
e
b
Lim
Va lve
n
b
yL im
be r
m
Ch a
b
Lim
H 2 O Ox PE EP yg en Va lve Se Tn so pi ec rP e& or t Flo CP AP w By He Lim ad ge Tu b ar bi ng & S Cl ize ip r s
PE EP
te ns io
O
Ex
H
cm
20
m
5c
CP AP
n
io
at
re Lin
Pr es su
Fil te r
n
ira to r
sp
In
at io
ific
NT ®
ific
id
m
Hu
id
m
Hu
aV E
Aq u
Therapy Solutions
Non - Invasive Ventilation
AMCP1409/021
3 0 3 3 3
AMHO1509/001 AMHO1509/004 AMHO1509/005 AMHO1509/006
m id
m id
n
Lim
ira to r
sp
In
at io
ific
NT ®
yL im
b ific b at Ad io ju n sta Ch bl am e be TV e pi r nt ec ur e& iJ et Flo Tr ac w he By ot Lim om Tu b y bi M ng as k Cl ip s
Hu
Hu
aV E
Aq u
Therapy Solutions
Non - Invasive Ventilation
AMHO1509/001
3
3 3
3
1312
AquaVENT® Family Non Invasive Ventilation Face Masks Comfortable, well fitting, non invasive patient interfaces are key to patient compliance and therapy outcomes. Armstrong Medical’s range of masks and headgear are designed to increase compliance rates with anatomicaly shaped masks and improved sizing options.
Comfort-Fix Comfort-Fix tube holders provide a secure method of holding endotracheal and tracheostomy tubes in position. The tube holders use an adhesive-backed tab which affixes easily to the tubes and fits comfortably to the patient. The Velcro® locking strip securely holds the tube in place without requiring additional fixing methods. The Velcro® fasteners can easily adjust to fit most neck sizes, simply by lifting and placing neckband tabs.
Catheter Mounts Armstrong Medical offer an extensive range of catheter mounts for anaesthesia and critical care applications. Catheter mounts are available in limblite corrugated tubing, smooth bore tubing and extendible tubing. A variety of patient elbows are available for gas monitoring, airway suctioning and fibre optic bronchoscopy. Low deadspace catheter mounts are available in corrugated tubing and smooth bore tubing. Catheter mounts can be further customised by attaching associated products including filters, face masks and Spirale® DDS.
14
AquaVENT® Family
Spirale® DDS Spirale® DDS is a collapsible volumising spacer device designed to deliver aerosolised micro-drug particles from an MDI canister, specifically intended for the delivery of pulmonary medication in ventilated patients. Spirale® provides optimal drug delivery while eliminating environmental drug pollution and maintaining a closed breathing circuit. Spirale® is designed to remain in the patient circuit for up to 7 days and can be used with invasive or non invasive breathing circuits.
HME Filters This extensive range of filters offers the user high quality devices for delivery of passive humidification. High levels of filtration and humidification efficiency are present throughout the range. The range offers the following features: • • • •
HME filters with humidification efficiency up to 33mg/H2O.L. Different filter options including ports, angles and elbows to ensure the right choice of device. Conical fittings are in accordance with ISO standards to ensure good connection stability. Initial set up is simple, quick and secure.
Combi-Flex® Combi-Flex® 3-in-1 is a passive humidification alternative to AquaVENT® systems for the management of ventilated patients. A lightweight, single limb coaxial system which generates a thermally efficient micro-climate and provides optimal humidification, drug delivery and filtration. Combi-Flex® 3-in-1 incorporates a HEPA filter HME and is also available with Spirale® DDS which reduces the need to break the patient circuit for drug nebulisation.
15
Armstrong Medical Wattstown Business Park Newbridge Road Coleraine BT52 1BS Northern Ireland T +44 (0) 28 7035 6029 F +44 (0) 28 7035 6875 E [email protected] W www.armstrongmedical.net
Creating Support for Life
Creating Support for Life
All Armstrong Medical products are manufactured to quality systems under ISO 13485 and EC Directive 93/42/EEC. Distributed by:
AMVCAQUA_V3
Armstrong Medical manufacture a complete range of disposable respiratory products for anaesthesia and critical care applications. For supply of these products or any product within the Armstrong Medical range, please contact your local representative.