24-Hour Intra-Aortic Balloon Product Hotline: 800-447-IABP Worldwide: 617-389-8628
®
Quick Start-up Guide 1.
Power On
®
2.
Establish ECG Connection
3.
Connect FiberOptix™ sensor and cal key BEFORE INSERTION
4.
Verify auto-zero via presence of green light bulb and audible alert (if applicable)
5.
Insert IAB in usual fashion
24-Hour Intra-Aortic Balloon Product Hotline: 800-447-IABP Worldwide: 617-389-8628
®
Light Bulb Icon Legend Black light bulb with blue square n
FiberOptix™ IAB not corrected ®
Blue light bulb n
FiberOptix™ IAB not zero'd prior to insertion
Green light bulb n
FiberOptix™ IAB zero'd prior to insertion
White light bulb n
FiberOptix™ IAB cal value manually adjusted
Intra-Aortic Balloon Sizing Recommendations 30cc
40cc
50cc
Height
147-162 cm 4' 10''- 5' 4''
162-182 cm 5' 4'' - 6' 0''
> 182 cm > 6' 0''
BSA
< 1.8m2
> 1.8m2