Model AR- 6400 series Continuous Wave & DualWave Safety Test Quick Guide Rev C

Safety Test Quick Guide

2 Pages

Safety test Quick guide  This is the “clamp test” for the main pump tubing set-up on the Arthrex Continuous Wave & DualWave Arthroscopic Pump (AR-6450, AR-6475, AR-6480) I  II  III  IV  1. Set up the pump tubing. a) First remove the protection cap ( I – IV ). b) Insert the connector fitting of the Pump Tubing into the Tubing Sensor Coupler. This step must be completed first to ensure accurate pressure measurements. c) Release the locking lever of the roller housing door and open the door completely. Allow the door to rest against the stop. The roller mechanism is now exposed.  green dot  d) Place the green-collared section of the Pump Tubing into the Tubing IN Guide indicated by the green dot. e) Guide the tubing boot over the rollers and insert the output side of the tubing boot into the Tubing OUT Guide.  locking lever  AR-6475  NOTE: The Pump Tubing is connected properly when the green connector on the Main Pump Tubing is aligned with the green dot on the front panel of the console. f ) Close and lock the roller housing door. g) Connect the tubing spikes to an bag.  2. After adjusting the required pressure using the Pressure Set buttons, remove the cap from the patient end of the tubing. AR-6480  green dot  www.arthrex.com