rabee_periodical_maintenance_schedule____spare_part.pdf
Page 1
ATOM Rabee Incubator
Maintenance Schedule and
Spare Parts List
For further information please contact our Service Department on (03) 9872 0222 or email service@parkerhealth.com.au
Page 2
Cleaning of ATOM Incubators
Chemicals Suitable
Tuffie Detergent Wipes (901DW100)
Osvan (Benzalkonium chloride aqueous solution)
Hyamine (Benzalkonium chloride aqueous solution)
Hibitane (Chlorhexidine aqueous solution)
Vinegar/Toothpaste/Eraser
(for removal of sticky residue?)
Chemicals Unsuitable
ZOFF
Glygent 20
Citrus Based Sticky Substance Remover
Eucalyptus Oil
Toluene
Benzine
Hollister
*NOTE: Use of these chemicals may cause
permanent damage to the incubator
Page 3
Maintenance Inspection
3-1.
Inspection before Use
Check the following points every time you are going to use the unit.
CAUTION
If you should notice any strange smell, strange noise, overheating or strange vibration in the inspection, stop using the unit immediately and contact your local Atom representative.
3-1-1.
Visual Inspection
MAINTENANCE
Item to check
Description
The main body and the hood should be neither broken nor deformed. (Otherwise, the infant and/or the user may get injured due to breakage, etc.)
Each access port cover should be attached securely to its corresponding
Access port cover
access port packing. It should be neither broken nor deformed. (Otherwise,
the air tight seal may be compromised.)
Snap-open access port
Each packing should be attached correctly to the access port. It should be
packing
intact. (Otherwise, the air tight seal may be compromised.)
Each packing should be attached correctly to the hood. It should be intact.
Tube introduction slit packing
(Otherwise, the air tight seal may be compromised.)
It should be neither broken nor deformed. (Otherwise, unsatisfactory control
Sensor module
may result from faulty detection by the sensors.)
Each knob should be held in place securely and should open and close the
Admittance panel operating
admittance panel reliably. (Otherwise, the infant may fall out of the baby
knob
compartment due to incomplete locking.)
With an admittance panel closed, hold the admittance panel operating knobs
Admittance panel operating
and pull the admittance panel toward you. It should not open. (Otherwise, the
lever
infant may fall out of the baby compartment due to incomplete locking.)
It should be neither broken nor deformed. (Otherwise, the infant may fall
Baby guard
out of the baby compartment due to incomplete attachment.)
Each access port should be held in place securely and should open and close
reliably. With the snap-open access port closed, the red line should not be
Snap-open access port
visible at the locking part. (Otherwise, the infant may fall out of the baby
compartment due to incomplete locking.)
The mattress platform should tilt smoothly when the tilting knob is operated.
Tilting unit
(Otherwise, the mattress platform may not tilt.)
It should turn the power on and off reliably. (Otherwise, it may not funcPower switch
tion.)
It should be fixed securely with the fixing device. (Otherwise, the display
Display
may become unstable or fall when a switch is operated.)
Each caster should rotate smoothly. (Otherwise, the unit cannot be relocated
Caster
easily.)
It should be clean. (Otherwise, air circulation may not be controlled propFilter
erly.)
An appropriate temperature should be displayed when the skin temperature
Skin temperature probe
probe is connected and the tip of the probe is held in your hand.
Skin temperature probe
The area around the connecting port should be neither broken nor unconnecting port
clean.
AC inlet
The AC inlet should be clean without any adherent medical fluid.
Appearance
23
Page 4
Maintenance Inspection
Item to check
Description
The plug should not be deformed.
The cord should not be damaged.
Height adjustment mechanism It should operate smoothly without making any strange noise.
Before connecting the power cord, check the following:
When the power switch is turned on, the power failure alarm indicator should
flash and an audible alarm should sound.
Activation of alarms
After connecting the power cord, check the following:
When the power switch is turned on, the alarm lamp should come on and
an audible alarm should sound.
It should be kept where it is readily accessible. (Otherwise, the unit may be
Operation Manual
operated incorrectly due to lack of accessible information.)
Power cord
3-1-2.
Functional Inspection
• Inspecting the basic functions of the incubator
(Preparation 1)
(1)
Connect the yellow skin temperature probe to the
skin temperature probe connecting port on the rear
of the main body.
Skin temperature probe
Skin temperature probe
connecting port
(2)
24
Open the humidity chamber cover down to the position shown in the figure on the right. Draw out the
humidity chamber until it stops halfway. Lift up the
humidity chamber slightly and take it out.
Allow the humidity chamber and the water in it to
cool down sufficiently before removing the humidity chamber. Otherwise, you may get scalded. Hold
the humidity chamber with both hands because the
humidity chamber filled with water is heavy.
Tube introduction
slit packing
Humidity chamber cover
Page 5
Maintenance Inspection
(Checking the displays and the audible alarm)
(3)
Turn on the power switch to activate the system.
Power switch
MAINTENANCE
Check:
①D
o all the indicators excluding the power failure
alarm indicator come on and does the audible
alarm sound immediately after the power switch
is turned on?
② I s a numerical value displayed on the incubator
air temperature display?
③ I s a numerical value displayed on the skin temperature display?
④ I s a numerical value displayed on the relative
humidity display?
⑤ I s the “humidity chamber off alarm” indicator
is on?
⑥A
re the alarm indicators other than the “humidity
chamber off alarm” indicator off?
✽✽ Digital indication includes “Lo·Hi” indication.
(Checking the humidity chamber)
(4)
Push the humidity chamber in which the cartridge
tank is placed gently into the main body until the
marking shown in the figure on the right is visible
and close the humidity chamber cover.
Check:
① Does the “humidity chamber off alarm” indicator go out
Humidity chamber cover
(Checking the temperature setting operation)
(5)
Set the temperature to 23.0°C in the manual control
mode.
Check:
①D
oes the numerical value on the set temperature
display change when either setting switch (
•
) is touched?
(Checking the relative humidity setting operation)
(6)
Set the relative humidity to 95%Rh.
Check:
① Does the numerical value on the set relative humidity display change when either setting switch
(
•
) is touched?
25
Page 6
Maintenance Inspection
(Checking the skin temperature probe alarm)
(7)
Set the temperature to 37.5°C in the servo control
mode.
(8)
Remove the skin temperature probe connected to
the skin temperature probe connecting port on the
rear of the main body.
Check:
①D
oes the skin temperature probe alarm occur?
Skin temperature probe
Skin temperature probe
connecting port
Tube introduction
slit packing
(Checking the system failure alarm)
(9)
Remove the sensor module connector connected to
the rear of the main body.
Check:
①D
oes the system failure alarm occur?
(Checking the power failure alarm)
(10) Remove the power cord.
Check:
① Does the power failure alarm indicator come
on?
② Does the audible alarm sound?
(Inspection completed)
(11) Turn off the power switch.
CAUTION
●● If any function is found to be unsatisfactory, refer to the “Troubleshooting” section and perform the
entire functional inspection once again.
26
Page 7
Maintenance Inspection
3-1-3.
Performance Inspection
Equipment required:
An oxygen flowmeter and an oxygen monitor are required to check the oxygen supply.
(Preparation)
Connect the skin temperature probe to the skin temperature probe connecting port on the rear of the
main body and place it 10cm above the center of the mattress surface.
(2)
Supply water in the humidity chamber and position it in the main body.
(3)
Turn on the power switch to activate the system.
(4)
Set the humidity setting to OFF.
MAINTENANCE
(1)
(Skin temperature control)
(5)
Set the temperature to 36.0°C in the servo control mode.
Check:
①D
oes the reading on the skin temperature display stabilize at 36.0±0.5°C?
(Incubator air temperature control)
(6)
Set the temperature to 36.0°C in the manual control mode.
Check:
① Does the reading on the incubator air temperature display stabilize at 36.0±1°C?
(Oxygen supply)
(7)
Supply oxygen through the oxygen supply port at 10L/min using an oxygen flowmeter.
Check:
① Measure the oxygen concentration with an oxygen monitor. Is the reading at least 65% O2?
✽✽ Stop the oxygen supply when checking is completed.
(Humidity control)
(8)
Set the relative humidity to 90%Rh.
Check:
① Is the reading on the relative humidity display 90±3%Rh
(Inspection completed)
(9)
Turn off the power switch.
CAUTION
●● If any performance is found to be unsatisfactory, refer to the “Troubleshooting” section, and perform
the entire performance inspection once again.
27
Page 8
Maintenance Inspection
3-2.
Quarterly Inspection
Check the following points every three months.
Item to check
Incubator air temperature control
Skin temperature
control
Humidity control
Fan
Humidifier
Power failure alarm
28
Procedure
Description
Set the temperature to 36.0°C in manual
control.
Place the skin temperature probe at about
10cm above the center of the mattress
surface in the incubator and set the temperature to 36.0°C in servo control.
Set RH to 90%Rh with the incubator air
temperature set to 32.0°C.
The displayed incubator air temperature
should be stable at 36.0±1°C.
The displayed skin temperature should be
stable at 36.0±0.5°C.
The displayed RH should be stable at
90±3%Rh.
It should be neither broken nor deCheck the exterior by visual inspection.
formed.
① Install the cartridge tank without any ① The no water alarm indicator should
water in it.
come on.
② Open the humidity chamber cover.
② The humidity chamber off alarm indicator should come on.
③ Install the humidity chamber with the ③ The humidity chamber off alarm indicaboiler cap off.
tor should come on.
The power failure alarm indicator should
Turn the power on, and then remove the
come on and an audible alarm should
power plug from the power outlet.
sound.
Page 9
Maintenance Inspection
3-3.
Inspection Checklist
WARNING
If any trouble is detected in any inspection, indicate on the unit that it is out of order and seek repair
without fail.
Contact your local Atom representative for further information on repairs.
)
Date of Inspection (
No.
.
.
)
Inspector (
Item to check
)
MAINTENANCE
Serial No. (
Judgment
Inspection before use (every time)
1
Are the main body and the hood free of breakage and deformation?
YES / NO
2
Are the access port covers free of breakage and deformation and attached securely to the hood?
YES / NO
3
Are the access port packings free of breakage and attached securely to the access ports?
YES / NO
4
Are the tube introduction slit packings free of breakage and attached securely to the hood?
YES / NO
5
Is the sensor module free of breakage and deformation and fixed securely to the hood?
YES / NO
6
Are the admittance panel operating knobs held in place securely and do they operate reliably? YES / NO
7
Are the admittance panels locked securely?
YES / NO
8
Are the baby guards free of breakage and deformation?
YES / NO
9
Are the access ports held in place securely and do they open and close reliably?
YES / NO
10 Does the tilting unit operate smoothly?
YES / NO
11 Does the power switch turn the power on and off reliably?
YES / NO
12 Is the fixing device of the display free of breakage and is the display fixed securely?
YES / NO
13 Does each caster rotate smoothly and do the caster locks function reliably?
YES / NO
14 Is the filter clean without any blackish dirt?
YES / NO
15
Is the correct skin temperature displayed when you hold the tip of the skin temperature probe
YES / NO
in your hand?
16 Is the area around the skin temperature probe connecting port free of breakage and dirt?
YES / NO
17 Is the AC inlet clean without any adherent medical fluid?
YES / NO
18 Is the plug of the power cord free of deformation and is the power cord free of damage?
YES / NO
19 Does the height adjustment mechanism operate smoothly without making any strange noise? YES / NO
20 Does the power failure alarm function properly and does the alarm lamp come on properly?
YES / NO
21 Is the Operation Manual readily accessible?
YES / NO
Quarterly inspection
1
Is the displayed incubator air temperature stable at 36.0±1°C in manual control?
YES / NO
2
Is the displayed skin temperature stable at 36.0±0.5°C in servo control?
YES / NO
3
Is the displayed RH stable at 90±3%Rh?
YES / NO
4
Does the oxygen concentration rise to 65%O2 or higher at 10L/min?
YES / NO
5
Is the fan free of breakage?
YES / NO
6
Do the “no water alarm” indicator and the “humidity chamber off alarm” indicator come on?
YES / NO
7
Does an audible alarm sound and does the power failure alarm indicator flash when the power
YES / NO
plug is removed from the power outlet?
29
Page 10
Maintenance Inspection
3-4.
Periodical Replacement Parts and How to Replace Them
Periodical replacement parts are those which gradually deteriorate and wear down with use. They need periodical replacement to keep the accuracy and performance of the unit at a proper level. Timing of replacement
varies with the frequency and conditions of use. Consult your local Atom representative for replacement.
Part name
30
Period of use
Reason for replacement
Filter
3 months
Dirtiness due to dust or defective air circulation control
due to clogging
Tube introduction slit
packing
1~2 years
Decreased air tightness due to breakage or deformation
Admittance panel packing
1~2 years
Decreased air tightness due to breakage or deformation
Iris access port frame
packing
1~2 years
Decreased air tightness due to breakage or deformation
Snap-open access por t
packing
1~2 years
Decreased air tightness due to breakage or deformation
Mattress
1~2 years
Decreased elasticity due to breakage or deformation
Fan motor
3 years
Decreased air circulation
Fan
3 years
Decreased air circulation
Snap-open access port
3 years
Incomplete locking
Humidity chamber
3 years
Insufficient humidification due to breakage or deformation
Battery for power failure
alarm
4 years
Defective power failure alarm function due to battery
life
Page 11
ATOM Rabee Incubator ‐ Periodical Replacement Parts
Part Number
Part Name
Period of Use
Parts List Item No:
60269
Filter (pack 5x)
3 months
92326
Tube introduction slit
assembly packing
1‐2 years
21
98682
Admittance panel packing
1‐2 years
7
92249
Iris access port packing
1‐2 years
20
98747
Snap‐open access port
packing
1‐2 years
13
60323
Mattress
1‐2 years
98572
Fan Motor
3 years
44
98573
Fan
3 years
43
60897
Snap‐open access port
3 years
98526
Humidity Chamber
3 years
50
92791
Rechargeable battery
(for power failure alarm)
4 years
65
Page 12
Parts Identification
PLEASE READ WITHOUT FAIL
[2] Parts Identification
2-1.
Main Body
<Fixed-height stand type unit>
<Hi-Low stand type unit>
①
①
②
③
②
③
④
⑫
⑪
⑩
④
⑫
⑪
⑩
⑦
⑤
⑥
⑨
⑦
⑤
⑥
⑨
⑧
⑧
⑬
No.
① Display
Name
No.
⑧ Caster
Name
②
Iris access port
⑨
Humidity chamber cover
③
Sensor module
⑩
Mattress
④
Admittance panel operating lever
⑪
Tube introduction slit packing
⑤
Snap-open access port
⑫
Admittance panel operating knob
⑥
Admittance panel
⑬
Foot switch
⑦
Mattress platform tilting knob
13
Page 13
Parts Identification
◾◾ Rear
⑭
⑮
⑯
⑰
⑱
⑳
⑲
No.
Name
⑭ Skin temperature probe connecting port
14
⑮
Connector for sensor module
⑯
Power switch
⑰
Filter cover
⑱
Oxygen supply port
⑲
Connector for the display
⑳
AC inlet
Page 14
Parts Identification
PLEASE READ WITHOUT FAIL
◾◾ Inside
No.
Name
Mattress platform
No.
Heater
Name
X-ray cassette tray
Fan
Mattress platform tray
Baby guard
Middle board
Tube introduction slit packing
Fan cover
15
Page 15
Rabee Incu i
No. Code No.
Drawing
Description
Unit of
quantity
Remarks
1
98700
2
98731
(w/Snap-open Access Port, Inner Panel,
Operating Knob, Packing)
set
3
98701
Admittance Panel
pce
4
98702
Inner Panel
pce
5
98703
Admittance Panel Hinge (Side)
pce
6
98744
Front Hood Cover
set
7
98682
Front Admittance Panel Packing
pce
Common to : Incu i
8
98723
set
Common to : Incu i
9
98722
set
Common to : Incu i
10
98713
Admittance Panel Lock Holder (Right)
pce
Common to :
Dual Incu i, Incu i
11
98712
Admittance Panel Lock Holder (Left)
pce
Common to :
Dual Incu i, Incu i
12
98681
Admittance Panel Operating Lever
pce
Common to :
Dual Incu i, Incu i
13
98747
Snap-Open Access Port Packing
pce
Common to :
Dual Incu i, Incu i
Slope Hood
pce
Admittance Panel Assy
Admittance Panel Operating Knob Assy
(Right)
Admittance Panel Operating Knob Assy
(Left)
Common to :
Dual Incu i, Incu i,
V-2200, V-2100G,
Neo-Servo
Common to :
Dual Incu i, Incu i,
V-2200, V-2100G,
Neo-Servo
14
92887
Inner Wall Lever (Right)
pce
15
92888
Inner Wall Lever (Left)
pce
16
98599
Inner Wall Bearing
pce
Common to :
Dual Incu i, Incu i
pce
Common to :
Dual Incu i, Incu i,
V-2200, V-2100G,
Neo-Servo
17
92892
Inner Wall Shaft Bearing
for the Type with Stopper
-1-
Page 16
Rabee Incu i
No. Code No.
Drawing
Description
Unit of
quantity
Remarks
18
98706
Admittance Panel Rotary Shaft
pce
Common to :
Dual Incu i, Incu i
19
98587
Sensor Module Holder
pce
Common to :
Dual Incu i, Incu i
pce
Common to :
Incu i, V-2200,
V-2100G, Neo-Servo,
V-808,
pce
Common to :
Dual Incu i, Incu i, V-2200,
V-2100G, Neo-Servo,
V-808, V-707,
Infa Warmer i, V-505
pce
Common to :
Dual Incu i, Incu i
20
92249
Iris Access Port Frame Packing
21
92326
22
98595
23
98522
Sensor Module
pce
Common to :
Dual Incu i, Incu i
24
98554
Baby Guard (Front/Rear)
pce
Common to :
Dual Incu i, Incu i
25
98555
Baby Guard (Side)
pce
Common to :
Dual Incu i, Incu i
26
98556
set
Common to :
Dual Incu i, Incu i
27
98557
Mattress Platform Tray (w/Latch)
pce
Common to :
Dual Incu i, Incu i
28
98559
Middle Board (w/ Fan Cover)
set
Common to :
Dual Incu i, Incu i
29
98565
Middle Board
pce
Common to :
Dual Incu i, Incu i
30
98566
Fan Cover
pce
Common to :
Dual Incu i, Incu i
31
98558
Latch
pce
Common to :
Dual Incu i, Incu i
32
98561
Level Gauge
pce
Common to :
Dual Incu i, Incu i
33
98574
Tilting Unit
set
Common to :
Dual Incu i, Incu i
34
98575
Tilting Knob (Right)
pce
Common to :
Dual Incu i, Incu i
Tube Introduction Slit Packing
Tilting Arm Fixed Ring
for Tilting Arm
Mattress Platform
(w/X-ray Cassette Tray)
-2-
Page 17
Rabee Incu i
No. Code No.
Drawing
Description
Unit of
quantity
Remarks
35
98576
Tilting Knob (Left)
pce
Common to :
Dual Incu i, Incu i
36
98677
Control Box Cover
pce
Common to : Incu i
37
98699
38
98577
Power Switch
pce
Common to :
Dual Incu i, Incu i
39
98582
Position Detection Board
pce
Common to :
Dual Incu i, Incu i
40
98569
Heater for Main Unit (AC120V)
set
Common to :
Dual Incu i, Incu i
41
98570
Heater for Main Unit (AC230V)
set
Common to :
Dual Incu i, Incu i
42
98571
High Temperature Sensor
set
Common to :
Dual Incu i, Incu i
43
98573
Fan
pce
Common to :
Dual Incu i, Incu i
44
98572
Fan Motor
pce
Common to :
Dual Incu i, Incu i
45
98579
Filter Joint
pce
Common to :
Dual Incu i, Incu i
46
98580
Steam Cap
pce
Common to :
Dual Incu i, Incu i
47
98651
Steam Gate
pce
Common to :
Dual Incu i, Incu i
48
98581
Humidity Chamber Cover Packing
pce
Common to :
Dual Incu i, Incu i
49
98525
set
Common to :
Dual Incu i, Incu i
50
98526
set
Common to :
Dual Incu i, Incu i
51
98549
pce
Common to :
Dual Incu i, Incu i
Detection Board
for Temperature, Skin Temperature, Humidity
Humidity Chamber
(w/Cartrige Tank, Boiler Cap) (AC120V)
Humidity Chamber
(w/Cartrige Tank, Boiler Cap) (AC230V)
Cartridge Tank (w/Cap)
-3-
pce
Page 18
Rabee Incu i
No. Code No.
Drawing
Description
Unit of
quantity
Remarks
52
98548
Tank Cap
pce
Common to :
Dual Incu i, Incu i
53
98547
Boiler Cap
pce
Common to :
Dual Incu i, Incu i
54
98553
Humidity Chamber Handle
pce
Common to :
Dual Incu i, Incu i
55
98611
Humidity Chamber Cover
pce
Common to :
Dual Incu i, Incu i
56
98610
Humidity Chamber Cradle
set
Common to :
Dual Incu i, Incu i
57
98612
Humidity Chamber Cradle Packing
pce
Common to :
Dual Incu i, Incu i
58
98613
Humidity Chamber Connector
set
Common to :
Dual Incu i, Incu i
59
98616
Control Power Transformer (AC100-120V)
set
Common to :
Dual Incu i, Incu i
60
98617
Control Power Transformer (AC230)
set
Common to :
Dual Incu i, Incu i
61
98708
Main Board
pce
62
98725
Heater Control Board (w/Fuse) (AC120V)
pce
63
98726
Heater Control Board (w/Fuse) (AC230V)
pce
64
98626
Noise Filter
pce
Common to :
Dual Incu i, Incu i
65
92791
Battery for Power Failure Alarm
pce
Common to :
Dual Incu i, Incu i,
V-2200,V-2100G,
Neo-Servo, V-707
Infa Warmer i
66
98625
Buzzer
pce
Common to :
Dual Incu i, Incu i
67
98689
AC Inlet
pce
Common to : Incu i
68
98648
Filter Box
pce
Common to :
Dual Incu i, Incu i
-4-
Page 19
Rabee Incu i
No. Code No.
Drawing
Description
Unit of
quantity
Remarks
69
98649
Filter Cover
pce
Common to :
Dual Incu i, Incu i
70
98650
Filter Packing
pce
Common to :
Dual Incu i, Incu i
71
95684
Cord Cleat
pce
Common to:
Dual Incu i, Incu i,
Infa Warmer i
72
98737
Hi-Low Stand Actuator
pce
73
95687
Foot Switch
set
Common to:
Dual Incu i, Incu i,
Infa Warmer i
74
95686
Relay Board for Foot Switch
pce
Common to:
Dual Incu i, Incu i,
Infa Warmer i
75
98741
Hi-Low Control Box (AC120V)
pce
For Hi-Low Stand
76
98742
Hi-Low Control Box (AC230V)
pce
For Hi-Low Stand
77
99023
Breaker 3A(106-P106)
pce
78
98738
Power Indicator (AC100-120V)
pce
79
98739
Power Indicator (AC230V)
pce
80
98717
Stand Base Cap
pce
81
98693
Display
set
82
98694
Front Panel (w/Display Seal)
pce
For Display
83
98697
Rear Panel
pce
For Display
84
98695
LED Display Board
pce
For Display
85
98696
LED Control Board
pce
For Display
-5-
Page 20
Rabee Incu i
No. Code No.
Drawing
Unit of
quantity
Description
Remarks
86
98539
Speaker (w/Cable)
pce
For Display
Common to :
Dual Incu i, Incu i
87
98698
Display Base (w/Suction Disk)
pce
For Display
88
98546
Display Cable
pce
For Display
Common to :
Dual Incu i, Incu i
89
98637
Fuse F0.63A (Fast-blow)
pack
Common to :
Dual Incu i, Incu i
90
98636
Fuse F1.25A (Fast-blow)
pack
Common to :
Dual Incu i, Incu i
91
98634
Fuse F2A (Fast-blow)
(2pcs/pack)
pack
Common to :
Dual Incu i, Incu i
92
98633
Fuse F4A (Fast-blow)
(2pcs/pack)
pack
Common to :
Dual Incu i, Incu i
93
98641
Fuse T0.63A (Slow-blow)
(2pcs/pack)
pack
Common to :
Dual Incu i, Incu i
94
98643
Fuse T0.315A (Slow-blow)
(2pcs/pack)
pack
Common to :
Dual Incu i, Incu i
95
95653
pack
Common to :
Infa Warmer i
96
99154
Fuse T5A (Slow-blow)
97
98715
Caster (w/Stopper)
pce
98
98716
Caster
pce
99
99003
Power Cord (AC100-120V)
pce
Common to :
V-2100G, Neo-Servo,
V-808, V-707
100
99016
Power Cord (AC120V) UL
pce
Common to :
Incu i, Infa Warmer i
101
99017
Power Cord (AC230V)
pce
Common to :
Dual Incu i, Incu I,
Infa Warmer i
pce
Common to :
Incu i, V-2200,
V-2100G, Neo-Servo,
V-808, V-707
102
92257
(2pcs/pack)
(2pcs/pack)
Fuse T2.5A (Slow-blow)
(2pcs/pack)
(2pcs/pack)
Dust Cover
-6-
pack