Maintenance Inspection
[9] Maintenance Inspection
In order to use the unit safely for a longer period, perform the maintenance inspections described below.
CAUTION
Medical institutions are responsible for performing the maintenance inspections.
They are allowed to entrust the maintenance inspections of the unit to an appropriate external
contractor.
Clean and disinfect the unit and its accessories before maintenance inspections, repairs, or disposal.
Inspection before use
Check the basic functional operation of each part of the unit every time you are going to use the unit.
Quarterly inspection
Check the operation of each function of the unit every three months.
Periodical inspection
Periodical inspection needs to be performed annually. Contact your local Atom representative for periodical
inspection.
Parts requiring periodical replacement
Some parts need to be replaced periodically depending on their period of use.
48
Page 3
Maintenance Inspection
9-1.
Inspection before Use
CAUTION
If you should notice any strange smell, strange noise, overheating, strange vibration, missing parts,
or defective function in the inspection, stop using the unit immediately and contact your local Atom
representative.
Check the following points every time you are going to use the unit.
If you should notice any defect, contact your local Atom representative.
(1)
Check the following points before connecting the power cord.
Item to check
Description
No part of the main body should be broken.
Power cord
The plug should not be deformed. The cord should not be damaged.
Power connection
The power connections should be clean with no medical fluid stains.
Caster
The casters should rotate smoothly. They should be locked reliably.
Baby guard
The baby guards should be attached securely to the mattress platform. No
hinge should be broken.
Mattress platform tilting
mechanism
The mattress platform should be locked reliably in a tilted position.
Probe connecting port
The area around the probe connecting ports should be clean and should
not be broken.
Control knob
The control knobs should be attached securely and should operate reliably.
Reflector
The reflector should be clean and should not be damaged.
Canopy tilting mechanism
The canopy should be held in place securely when tilted.
Canopy fixing screw
The screw should be attached securely where it should be.
Power failure alarm
The power failure alarm should activate when the power switch is turned on.
Tube introduction slit assembly
Caution/Warning sticker
Operation Manual
MAINTENANCE
Appearance
The assemblies should be attached correctly to the baby guards and they
should not be broken.
Each sticker should be affixed securely. (Otherwise, important safety information will not be made known appropriately.)
It should be kept where it is readily accessible.
49
Page 4
Maintenance Inspection
(2)
Check the following points after connecting the power cord.
Item to check
Mains on indicator
The mains on indicator should come on.
Height adjustment mechanism
The mechanism should operate smoothly without making a strange sound.
Control panel
Skin temperature probe
SpO2 probe
Various switches
Lighting lamp
50
Description
All the indicators on the control panel, except for the power failure alarm
indicator and the battery operation indicator, should come on and an audible alarm should sound when the power switch is turned on.
An appropriate temperature should be displayed when the skin temperature probe is connected and the tip of the probe is held with your hand.
SpO2 and pulse rate values should be displayed when the SpO2 sensor is
attached.
Each switch should function properly when operated according to the
Operation Manual.
Intensity of the lighting lamp should increase gradually as the light control
switch is turned clockwise. The lighting lamp should go out when the light
control switch is turned fully counterclockwise.
Page 5
Maintenance Inspection
9-2.
Quarterly Inspection
Check the following points every three months.
If any trouble is detected, indicate on the unit that it is out of order and contact your local Atom representative
immediately.
Item to check
Procedure
Description
Servo control
Set the skin temperature to 36.0°C in the
servo control mode, and warm and cool
the skin temperature probe.
Set temperature
alarm❋
Set the skin temperature to 36.0°C in the
servo control mode, and warm and cool
the skin temperature probe.
Set heat alarm
Set the heater output to 100% in the
manual control mode, and warm the skin
temperature probe.
Baby check alarm
Set the heater output to 100% in the
manual control mode.
Warm the skin temperature probe and
check that the number of bars on the
heater output indicator decreases when
the displayed skin temperature is 36.0°C
or higher and increases when it is below
36.0°C.
Check that the set temperature alarm indicator comes on and an audible alarm
sounds when the displayed skin temperature deviates from 36.0°C앐1.0°C or
36.0°C앐0.5°C.
Check that the set temperature alarm indicator flashes and an audible alarm
sounds when the displayed skin temperature becomes 40°C or higher.
Check that the baby check alarm indicator comes on and an audible alarm sounds
in 15 minutes.
❋ Either the set temperature 앐1.0°C or the set temperature 앐0.5°C is selectable for the temperature deviation limit at which the set temperature alarm will occur. Check that the set temperature alarm occurs when
a detected skin temperature deviates from the selected deviation limit.
Periodical Replacement Parts
Periodical replacement parts are those which gradually deteriorate and wear down with use. They need periodical replacement to keep the accuracy and performance of the unit at a proper level. Timing of replacement
varies with the frequency and conditions of use. Consult your local Atom representative for replacement.
Part name
Period of use
Tube introduction slit
assembly
1~2 years
Breakage, deformation
Mattress
1~2 years
Breakage, deformation
Table tilting unit
3~5 years
Less effective stopper
Rechargeable battery
4 years
Reason for replacement
MAINTENANCE
9-3.
Life
51
Page 6
Maintenance Inspection
9-4.
Inspection Checklist
WARNING
If any trouble is detected in any inspection, indicate on the unit that it is out of order and seek repair without fail.
Contact your local Atom representative for further information on repairs.
Serial No. (
No.
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
1
2
3
4
5
6
7
1
2
3
4
52
)
Date of Inspection (
.
.
)
Inspector (
)
Judgment
Item to check
Inspection before use (1) (before connecting the power cord)
YES/NO
Is the main body free of breakage and other defects?
YES/NO
Is the power cord free of damage and other defects?
YES/NO
Are the power connections clean with no medical fluid stains?
YES/NO
Do the casters rotate smoothly? Are they locked reliably?
Are the baby guards attached to the mattress platform securely? Are the hinges of the baby
YES/NO
guards free of damage?
YES/NO
Is the mattress platform locked reliably in a tilted position?
YES/NO
Is the area around the skin temperature probe connecting ports intact and clean?
YES/NO
Are the control knobs attached securely? Do they operate reliably?
YES/NO
Is the reflector intact and clean?
YES/NO
Is the canopy held in place securely when tilted?
YES/NO
Is the canopy fixing screw attached securely where it should be?
Does the power failure alarm activate when the power switch is turned on? (The power failure
YES/NO
alarm indicator should come on and an audible alarm should sound.)
YES/NO
Are the tube introduction slit assemblies free of breakage and attached reliably?
YES/NO
Are all the caution/warning stickers affixed securely?
YES/NO
Is the Operation Manual kept where it is readily accessible?
Inspection before use (2) (after connecting the power cord)
Does the mains on indicator come on when the power cord is connected to the power outlet? YES/NO
Does the height adjustment mechanism operate smoothly without making a strange sound? YES/NO
Do all the indicators on the control panel, except for the power failure alarm indicator and the
battery operation indicator, come on and does an audible alarm sound when the power switch YES/NO
is turned on?
YES/NO
Is an appropriate temperature displayed when the skin temperature probe is connected?
YES/NO
Are SpO2 and pulse rate values displayed when the SpO2 sensor is attached?
YES/NO
Does each switch function properly when operated according to the Operation Manual?
Does intensity of the lighting lamp increase gradually as the light control switch is turned
clockwise? Does the lighting lamp go out when the light control switch is turned fully counter- YES/NO
clockwise?
Quarterly inspection (each time)
Does the heater output decrease when the skin temperature probe is warmed? Does the numYES/NO
ber of bars on the heater output indicator change when the skin temperature probe is cooled?
Does the set temperature alarm indicator come on and does an audible alarm sound when the
skin temperature detected by the skin temperature probe deviates from the set temperature YES/NO
앐1.0°C or the set temperature 앐0.5°C?
Does the set temperature alarm indicator flash and does an audible alarm sound when the
skin temperature detected by the skin temperature probe is 40°C or higher?
Does the baby check alarm indicator come on and does an audible alarm sound in 15 minutes?
YES/NO
YES/NO
Page 7
ATOM Infa‐Warmer i
Periodical Replacement Parts
Part Number
Part Name
Period of Use
Parts List Item No:
92326
Tube induction slit
assembly
1‐2 years
55
60896
Mattress
1‐2 years
95676
Table tilting unit
3‐5 years
54
92791
Rechargeable battery
4 years
32
Page 8
■Infa Warmar i
3
1
10
40
11
9
5 6
39
42
7 8
12
4
41
29
33
34
2
35
43
27
28
30 31
67 32 40
13 14
26 25
36 37 38
15 16
20
21 22
41
24
17 18 19
23
-1-
42
Page 9
47
50
45
47
52
55
48
52
51
46
52
49
49
48
47
47
44
50
47
52
54
45
56 69 68
53
50
47
57 58
62
72
73
59 60
61
63
64
70
71
65
66
-2-
Page 10
Infa Warmer i
No. Code No.
Drawing
Description
Unit of
quantity
1
95621
Upper Front Canopy Cover
pce
2
95622
Lower Front Canopy Cover
pce
3
95623
Upper Rear Canopy Cover
pce
4
95624
Lower Rear Canopy Cover
pce
5
95626
Heater Module (120V)
set
6
95627
Heater Module (230V)
set
7
95629
Heater (120V)
pce
8
95630
Heater (230V)
pce
9
90342
Micro Switch
pce
10
95631
LED Lighting Unit
pce
11
95632
Lighting Lamp Cover
pce
12
94304
Canopy Fixing Screw
pce
13
95635
14
95636
15
95633
16
95634
17
95637
Front Support Column Cover
(w/o Add-on Unit)
Front Support Column Cover
(w/Add-on Unit)
Rear Support Column Cover
(w/o Add-on Unit)
Rear Support Column Cover
(w/Add-on Unit)
Panel Cover (w/o SpO 2 Unit)
-6-
pce
pce
pce
pce
pce
Remarks
Commont to:
V-505
Page 11
Infa Warmer i
No. Code No.
Drawing
Description
Unit of
quantity
Remarks
18
95638
Panel Cover (w/SpO 2 unit for Masimo)
pce
19
95639
Panel Cover (w/SpO 2 unit for Nellcor)
pce
20
95642
Insulating Substrate for SpO 2 unit
pce
Common to:
Dual Incu i, Incu i
21
95640
SpO2 Board (Masimo)
pce
Common to:
Dual Incu i, Incu i
22
95641
SpO2 Board (Nellcor)
pce
Common to:
Dual Incu i, Incu i
23
95643
Skin Temperature Probe Board
pce
24
95644
Relay Board
pce
25
95645
Light Control Knob
pce
26
95646
Power Switch
pce
27
95647
Light Control Switch
pce
28
95648
Speaker
pce
29
95649
Triac Board
pce
30
95651
Power Transformer (120V)
pce
31
95652
Power Transformer (230V)
pce
32
92791
Battery for Power Failure Alarm
pce
33
95654
Main Board
pce
34
95656
LED Board
pce
-7-
Common to:
Dual Incu i, Incu i,
Rabee Incu i, V-2200,
V-2100G, Neo-Servo,
V-707
Page 12
Infa Warmer i
No. Code No.
Drawing
Description
Unit of
quantity
35
95657
36
95658
37
95659
38
95660
39
95661
Cable for Power Supply Part
pce
40
95662
Cable for Lighting Lamp/Tilting Knob
pce
41
95663
Cable for Relay Board
pce
42
95664
Cable for Skin Temperature Probe Board
pce
43
95665
Connector Cap for I/O Port
pce
44
95688
Front Baby Guard
pce
45
95689
Side Baby Guard
pce
46
95668
Rear Baby Guard
pce
47
95669
Pivot Shaft
pce
48
95670
Right Cushioning Material
pce
49
95671
Left Cushioning Material
pce
50
94310
Rotary Damper
pce
51
95673
Mattress Platform
pce
Upper Operation Panel Sheet
Lower Operation Panel Sheet
(w/o SpO 2 Unit)
Lower Operation Panel Sheet
for Masimo SpO 2 Unit
Lower Operation Panel Sheet
for Nellcor SpO 2 Unit
-8-
Remarks
pce
pce
pce
pce
Common to
left and right
Common to:
V-505
Page 13
Infa Warmer i
No. Code No.
Drawing
Unit of
quantity
Description
52
95674
Corner Cover
pce
53
95675
Wire for Tilting Unit
pce
54
95676
Tilting Unit for Mattress Platform
pce
Remarks
Common to:
Dual Incu i, Incu i,
Rabee Incu i, V-2200,
V-2100G, Neo-Servo,
V-808, V-707, V-505
55
92326
Tube Introduction Slit Packing
pce
56
95677
Hi-Low Drive Board for Hi-Low Stand
pce
57
95681
Power Transformer (120V)
pce
58
95682
Power Transformer (230V)
pce
59
99030
Breaker 10A(106-10P)
pce
Common to:
V-2100G, V-808,
V-505
60
99025
Breaker 5A(106-10P)
pce
Common to:
V-808, V-505
61
95683
AC Inlet
pce
Common to:
V-505
62
95684
Cord Cleat
pce
Common to:
Dual Incu i, Incu i,
Rabee Incu i
63
95685
Connector Cap for Power Pack i
pce
64
98216
Hi-Low Stand Actuator
pce
Common to:
Dual Incu i, Incu i
65
95687
Foot Switch
set
Common to:
Dual Incu i, Incu i,
Rabee Incu i
66
95686
Relay Board for Foot Switch
pce
Common to:
Dual Incu i, Incu i,
Rabee Incu i
67
95653
Fuse T2.5A (Slow-blow)
(2pcs/pack)
pack
For Main Board
Common to:
Rabee Incu i
68
95678
Fuse T3.15A(Slow-blow)
(2pcs/pack)
pack
For Hi-Low Stand
Control Board (Primary)
Common to:
Dual Incu i, Incu i
-9-
Page 14
Infa Warmer i
No. Code No.
Drawing
Description
Unit of
quantity
Remarks
pack
For Hi-Low Stand Control
Board (Secondary)
Common to:
Incu i
69
95679
Fuse T8A(Slow-blow)
70
99207
Caster (w/Stopper)
pce
Common to:
Dual Incu i, Incu i
71
98652
Conductive Caster (w/Stopper)
pce
Common to:
Dual Incu i, Incu i
72
99208
Caster
pce
Common to:
Dual Incu i, Incu i
73
98653
Conductive Caster
pce
Common to:
Dual Incu i, Incu i
74
99014
Power Cord (3P/1.25mm 2 )
pce
Common to:
Incu i,V-505
75
94201
Dust Cover
pce
Common to:
V-505
76
95709
Suction Pressure Meter (Anti-Clock Wise)
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E
77
95730
Suction Pressure Meter (Clock Wise)
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E,
V-505
pce
Common to:
Resuscitation Unit P,
Blender Unit P
pce
Common to:
Resuscitation Unit P,
Blender Unit P
pce
Common to:
Resuscitation Unit E,
Blender Unit E
pce
Common to:
Resuscitation Unit E,
Blender Unit E
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E
pce
Common to:
Resuscitation Unit P,
Blender Unit P
pce
Common to:
Resuscitation Unit E,
Blender Unit E
78
94804
79
95710
80
95731
81
95718
82
95732
Suction Controller
83
95711
Knob for Suction Pressure
84
95712
85
95719
Meter Cover
Suction Controller
for P type (Anti-Clock Wise)
Suction Controller
for P type (Clock Wise)
Suction Controller
for E type (Anti-Clock Wise)
for E type (Clock Wise)
Suction Switching Unit
On-Off type for P type
Suction Switching Unit
On-Off type for E type
- 10 -
(2pcs/pack)
Page 15
Infa Warmer i
No. Code No.
Drawing
Description
Unit of
quantity
Remarks
86
94805
Suction Lever
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E,
V-505
87
95720
Exhaust for E type
pce
Common to:
Resuscitation Unit E,
Blender Unit E
88
95721
Silencer
pce
Common to:
Resuscitation Unit E,
Blender Unit E
89
95713
Blender
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E
90
95714
Knob for Oxygen Concentration
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E
91
95716
Flow Meter
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E
92
95715
Knob for Flow Meter
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E
93
95717
Pressure Valve for P MAX
pce
Common to:
Resuscitation Unit P & E
94
95726
Protection Valve for Pressure Meter
pce
Common to:
Blender Unit P & E
95
95543
Pressure Valve for PIP
pce
Common to:
Resuscitation Unit P & E
96
95545
In-circuit Pressure Meter
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E
97
95723
Connection Port for Patient Circuit
pce
Common to:
Resuscitation Unit P & E
98
95550
pce
Common to:
Resuscitation Unit P & E
99
95727
Protection Cap
pce
Common to:
Blender Unit P & E
100
95551
Outlet for Oxygen/Air
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E
101
95722
Connection Port for Pressure Meter
pce
Common to:
Blender Unit P & E
102
95549
Chamber
pce
Common to:
Resuscitation Unit P & E
Protection Cap
for Protection Valve for Pressure Meter
- 11 -
Page 16
Infa Warmer i
No. Code No.
Drawing
Description
Unit of
quantity
Remarks
For Suction Canister
Common to:
V-505, Yox S401A,
Yox S401D
For Suction Canister
Common to:
V-505, Yox S401A,
Yox S401D
103
94316
Rubber Seal
pce
104
94317
Connector for Suction Hose
pce
105
95701
Canister Head
pce
For Suction Canister
Common to:
Yox S401D
106
95702
Joint
pce
For Suction Canister
Common to:
Yox S401D
107
95703
Seal A (w/O-ring)
pce
For Suction Canister
Common to:
Yox S401D
108
95704
Canister (w/Holder & packing)
pce
For Suction Canister
Common to:
Yox S401D
109
95706
Canister Packing
pce
For Suction Canister
Common to:
Yox S401D
110
94202
Oxygen Humidifier Holder
pce
For Humidifier
Common to:
V-505
111
94323
pce
For Humidifier
Common to:
V-505, Safety Flow
112
95707
O-ring
pack
For Humidifier
Common to:
V-505, Safety Flow
113
95708
Bottle Packing for Humidifier Bottle
pce
For Humidifier
Common to:
V-505, Safety Flow
114
94325
Bubble Holder
pce
For Humidifier
Common to:
V-505, Safety Flow
115
94326
Bubble Element
pce
For Humidifier
Common to:
V-505, Safety Flow
116
94324
Humidifier Bottle
pce
For Humidifier
Common to:
V-505, Safety Flow
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E,
V-505
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E,
V-505
pce
Common to:
Resuscitation Unit P,
Blender Unit P,
V-505
(w/Bubble Element)
(5pcs/pack)
Piping Connecting Hose
117
95083
118
95082
119
Cap for Humidifier
for Air(Yellow)
(w/o Jack) 2m
Piping Connecting Hose
94331
for Oxygen(Green)
(w/o Jack) 2m
Piping Connecting Hose
for Suction(Black) Φ6-12 2m
- 12 -
Page 17
Infa Warmer i
No. Code No.
120
94332
121
94333
122
94334
123
94335
Drawing
Description
Outlet Connecting Hose
for Air/Oxygen (Green) Φ6-12 60cm
Outlet Connecting Hose
for Suction (Black) Φ6-12 60cm
Oxygen Supply Hose (Green)
Φ6-12 1.2m
Suction Hose 2m
124
125
126
127
128
129
130
131
132
133
134
135
136
- 13 -
Unit of
quantity
Remarks
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E,
V-505
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E,
V-505
pce
Common to:
Resuscitation Unit P & E,
Blender Unit P & E,
V-505
pce
Common to:
V-505, Yox S401A,
Yox S401D