ATOM Medical Corporation
Infa Warmers
Infa Warmer i Model 103TP Service Manual and Parts List Sept 2018
Service Manual and Parts List
132 Pages
Preview
Page 1
Equipment for neonatal and premature infants: Infant Warmer
Atom Infant Warmer Model 103TP
Service Manual & Parts List
TO THE OPERATOR AND THE PERSON IN CHARGE OF MAINTENANCE AND CARE OF THE UNIT: ●●This Manual describes various kinds of inspection needed to ensure proper operation of the Atom Infa Warmer i, including instructions for troubleshooting, those procedures to change certain settings which are not mentioned in the Operation Manual and important points to bear in mind in handling the unit. ●●Various kinds of inspection, including periodical inspection, are described in detail in this Manual. They should be carried out only by those who are fully familiar with the operation of the unit, having adequate technical knowledge and skills required in inspecting the unit in general. ●●If repairs seem to be required as a result of any inspection described in this Manual, either personnel with more advanced knowledge and skills should undertake the repair or you should contact your local Atom representative for repair service.
INTRODUCTION This Service Manual describes procedures for inspecting, repairing and changing the settings of the Infa Warmer i. It also describes precautions and troubleshooting in alarm conditions. The procedures described in this Service Manual should be carried out only by personnel trained in electricity and the operation of the unit. Neither Atom Medical Corporation nor its representative will be responsible for the quality and performance of the unit if the unit should not be handled as instructed or if unauthorized parts should be used in repairs. It should be noted that any responsibility arising from inspecting, repairing or changing the settings of the unit lies with the person who carried it out. Read this Service Manual carefully and familiarize yourself thoroughly with its contents before inspecting, repairing or changing the settings of the unit. Keep this Manual where it is readily accessible for reference when needed. For more detailed information on inspecting, repairing and changing the settings of the unit, contact your local Atom representative. Read the Operation Manual thoroughly before using the Service Manual.
SAFETY INFORMATION Please read the Manual carefully before operating the unit. Please follow the instructions when operating the unit.
Basic Instructions 1. Medical institutions are responsible for the maintenance, inspection and care of the unit. 2. When you choose to have the unit maintained and inspected or when the unit is found to be in need of repairs from the results of inspection, consult your local Atom representative and take one of the following measures. 1) Ask someone who has completed a training course specified by Atom and who has sufficient technical knowledge and skills to do the work. 2) Ask Atom to send its service engineer if a person who has completed a training course specified by Atom and who has sufficient technical knowledge and skills is not available. 3) Ask Atom for its approval of returning the unit for repair service. 3. Follow the instructions for safety. Read the operating precautions thoroughly before operating the unit. 4. Inspect the unit on a periodical basis. Periodical inspection is needed to use the unit in the optimum condition. 5. Ensure that the unit will not be used if it is found to be defective. Take proper measures to ensure that a defective unit will not be used by mistake before it is properly inspected, repaired and has its settings changed for normal operation. Such measures include indicating on the unit that it is under inspection/ repair and keeping it away from other devices that function correctly.
1
Definition of Warning Indication Three levels of warning indication are used throughout this Service Manual & Parts List and on the unit. They are defined as follows. A DANGER notice indicates an immediately hazardous situation which, if not avoided, DANGER: will result in death or serious injury, serious damage to property such as total loss of use of equipment, or fire. A WARNING notice indicates an indirectly (potentially) hazardous situation which, WARNING: if not avoided, will result in death or serious injury, serous damage to property such as total loss of use of equipment, or fire. A CAUTION notice indicates a hazardous situation which, if not avoided, can result in CAUTION: minor or moderate injury, partial damage to equipment, and loss of data stored in computers.
Definition of Symbols 1. Symbols to indicate danger, warning or caution Symbol
Title and indication General attention Indicates unspecified general danger, warning or caution. Caution: Hot surface Indicates that the surface can be dangerously hot under certain conditions.
2. Symbols to prohibit action Symbol
Title and indication General prohibition Indicates unspecified general prohibition. Prohibition of disassembly Indicates prohibition of disassembly of the unit where it may cause an electric shock or other hazards. Prohibition of use of fire Indicates prohibition of use of fire where an external use of fire may cause the unit to ignite under certain conditions. Prohibition of contact Indicates prohibition of contact.
3. Symbols to give instructions for action Symbol
Title and indication General instruction Indicates unspecified general action on the part of the user.
2
Connect a ground wire Instructs the user to connect the ground wire without fail where the unit is provided with a ground terminal. Remove the power plug from the power outlet Instructs the user to remove the power plug from the power outlet in the case of malfunction or when there is a threat of lightning. 4. Symbols of international standards (IEC) Symbol
Title and indication Equipment partially on Indicates that a part of the device is "ON." Equipment partially off Indicates that a part of the device is "OFF." Type BF applied part Indicates that the device is classified as Type BF in terms of the degree of protection against an electric shock. Bell silenced Indicates either a control switch to silence the bell permanently or temporarily or that the bell is silenced. Caution Indicates that the user needs to see a relevant accompanying document before operating the unit. Caution against electrostatic discharge Indicates that appropriate precautions must be taken against electrostatic discharge. Lighting Indicates a switch that controls lighting. Recycle mark (battery) Indicates that recycling is recommended. Manufacturer This symbol indicates the name and the address shown adjacent to the symbol is of the manufacturer. Date of manufacture Indicates the date when the unit was manufactured in the factory. See Operation Manual Follow Operation Manual.
5. Other symbols Symbol
Title and indication AC power indicator The indicator lamp comes on in green when the unit is connected to the AC power source. It comes on in red when no power is supplied to the unit.
3
Symbol
Title and indication
(Nellcor)
Interference Indicates that interference exists.
(Nellcor)
Pulse search Indicates that no pulse is detected. Fast response mode indicator Indicates that the response mode of the pulse oximeter is set to “Fast.” SatSeconds display Displays the selected SatSeconds alarm limit setting. Press Indicates that the lock function will be released when the key with this symbol is pressed. Setting Indicates that a setting is increased. Setting Indicates that a setting is decreased. Main screen display switch Indicates a switch to display the main screen. Trend screen display switch Indicates a switch to display the trend screen. Menu screen display switch Indicates a switch to display the menu screen. Pulse oximeter screen display switch Indicates a switch to display the pulse oximeter screen or the function related to the pulse rate. Weight screen display switch Indicates a switch to display the weight screen or the function related to weighing the infant. Battery Indicates whether the unit is being powered by a battery or not. Radiant warmer Indicates that the radiant warmer is in manual control. Setting Indicates set values or the setting procedure. Skin temperature Indicates the function related to the infant’s skin temperature or that is in servo control.
4
Symbol
Title and indication Temperature Indicates, with this symbol alone or with other symbols, a detected temperature or a set temperature alarm. Peripheral temperature Indicates the function related to the infant’s peripheral temperature (skin temperature 2). Load capacity Indicates the maximum load capacity. Prohibition of lubrication Indicates that lubrication is prohibited. Avoid getting caught Indicates that the user must avoid getting caught in the gap in the device. Locking of the baby guard Indicates that the user must check whether the baby guard is securely locked.
Precautions on Jamming
WARNING When installing and using the unit, follow the information indicated in "11-3. EMC Level and Classification." Electric surgical knives, portable and mobile communication equipment, and other devices which generate high-frequency noise can cause jamming to various kinds of electric equipment for medical use and thus result in malfunction. Since portable and mobile communication equipment, etc. are often used in medical facilities, some measures should be taken to prevent jamming due to such devices. Portable and mobile communication equipment and other devices which generate high frequency should not be used near the unit during its operation to prevent malfunction of the unit due to jamming. Using accessories and cables other than replacement parts sold by Atom may cause the emissions of the device to increase and immunity to decrease. Do not use the unit next to other devices. When it is necessary to use the unit next to other devices, observe and test that the unit operates normally with the required device configuration.
Responsibility for Care of Equipment
CAUTION It is the user (a hospital, a doctor's office, a clinic) that is responsible for the operation, maintenance and care of the electric equipment for medical use. The equipment should be used only by medical personnel.
5
Prohibition of Modification
WARNING Do not disassemble or modify the unit. Otherwise, a fire, an electric shock or injury may result.
Periodical Inspection
CAUTION Proper periodical inspection is needed to use the unit in the optimum condition.
In Case of Trouble
CAUTION If any abnormal condition or trouble should occur to the unit, indicate on the unit that it is out of order and contact your local Atom representative or service engineer immediately. See the end of this Operation Manual for whom to contact. If any abnormal condition or trouble should occur, do not use the unit until it has been repaired completely by a service engineer so as to prevent possible danger.
6
CONTENTS INTRODUCTION... 1 PLEASE READ WITHOUT FAIL [1]
[2]
Operating Precautions... 10 1-1.
DANGER... 10
1-2.
WARNING... 11
1-3.
CAUTION... 13
[7]
Disassembly and Replacement Procedures... 60 7-1.
Parts Identification... 15 2-1. 2-2.
[3]
DISASSEMBLY AND REPLACEMENT
Main Body... 15 Support Column... 16
Display Screens... 17 3-1.
3-2. 3-3. 3-4. 3-5. 3-6. 3-7.
General Description and Switching of the Screens... 17 3-1-1. General Description of the Screens... 17 3-1-2. Screen Transition Diagram... 18 List of Operations Available on Each Screen... 19 Main screen... 19 Pulse Oximeter Screen... 22 Weight Screen (Option)... 25 Menu Screen... 27 Trend Screen... 30
7-2.
MAINTENANCE [4]
Maintenance Inspection... 31 4-1. 4-2. 4-3. 4-4.
[5]
Cleaning and Disinfection... 37 5-1. 5-2.
[6]
Inspection before Use... 32 Quarterly Inspection... 34 Inspection Checklist... 35 Periodical Replacement Parts... 36 Main Body... 37 Others... 38
Troubleshooting... 40 6-1. 6-2. 6-3.
Alarms... 40 Troubleshooting... 43 Troubleshooting Flowchart... 44 6-3-1. Troubleshooting When a Control Failure Has Occurred... 44 6-3-2. Troubleshooting When the Unit Does Not Function Properly... 45
7-3.
Canopy... 60 7-1-1. Replacing the Upper Front Canopy Cover... 61 7-1-2. Replacing the Lower Front Canopy Cover... 61 7-1-3. Replacing the Heater Module... 62 7-1-4. Replacing the Heater... 63 7-1-5. Replacing the Microswitch... 64 7-1-6. Replacing the LED Lighting Unit... 64 7-1-7. Replacing the Lighting Lamp Cover... 65 Support Column... 66 7-2-1. Replacing the Rear Support Column Cover... 66 7-2-2. Replacing the Front Support Column Cover... 66 7-2-3. Replacing the Connection Cable for SpO2 board... 66 7-2-4. Replacing the SpO2 Board... 67 7-2-5. Replacing the Insulating Substrate... 67 7-2-6. Replacing the Detection Board... 68 Power Unit... 69 7-3-1. Replacing the Power Indicator Board... 70 7-3-2. Replacing the Speaker... 71 7-3-3. Replacing the Control Power Transformer...71 7-3-4. Replacing the Fuse for the Main Board... 72 7-3-5. Replacing the Buzzer... 72 7-3-6. Replacing the Battery for Power Failure Alarm... 73 7-3-7. Replacing the Main Board... 73 7-3-8. Replacing the Light Control Knob... 74 7-3-9. Replacing the Power Switch/ Light Control Switch... 74 7-3-10. Replacing the Pilot Lamp Board... 75 7-3-11. Replacing the Alarm Lamp Board... 75
How to Operate the Service Menu Screen... 96 8-1. 8-2.
8-3.
[9]
How to Enter the Service Menu... 96 Operating the Service Menu Screen... 97 8-2-1. Setting the Key Click Sound... 97 8-2-2. Setting the Language... 97 8-2-3. Setting the Baby Check Alarm... 98 8-2-4. Setting the SC Alarm Threshold... 99 8-2-5. Apgar Timer Settings... 99 8-2-6. Alarm History... 99 8-2-7. Adjusting the Weight Monitor with a 5 kg Weight... 100 8-2-8. Setting History... 101 Calibrating the Touch Panel... 101
Block Diagram... 102
[10] Parts List... 103 10-1. Fig.1 Overall View and Accessories... 104 10-2. Fig.2 Canopy... 106 10-3. Fig.3 Support Column... 108 10-4. Fig.4 Power Unit... 110 10-5. Fig.5 Mattress Platform... 112 10-6. Fig.6 Main Body and Base... 114
DISASSEMBLY AND REPLACEMENT
APPENDIX [11] Technical Information... 116 11-1. Principle of Operation... 116 11-1-1. Description of the Principle of Operation... 116 11-1-2. Illustration of the Principle of Operation... 117 11-2. Technical Data... 118 11-3. EMC Level and Classification... 122
INFORMATION
[12] Disposal... 126
APPENDIX
7-6.
[8]
MAINTENANCE
7-5.
INFORMATION
PLEASE READ WITHOUT FAIL
7-4.
7-3-12. Replacing the Button Battery... 76 7-3-13. Replacing the LCD Board... 76 7-3-14. Replacing the LCD Display... 76 7-3-15. Replacing the CCFL Inverter... 77 7-3-16. Replacing the Cable for Power Supply Part... 77 7-3-17. Replacing the Cable for Detection Board... 78 7-3-18. Replacing the Cable for Lighting Lamp/Tilting Knob... 78 Table Unit... 79 7-4-1. Replacing the Front Baby Guard/Side Baby Guard/ Rear Baby Guard... 79 7-4-2. Replacing the Pivot Shaft... 81 7-4-3. Replacing the Rotary Damper... 82 7-4-4. Replacing the Lock Lever... 83 7-4-5. Replacing the Mattress Platform... 84 7-4-6. Replacing the Corner Grip... 84 7-4-7. Replacing the Table Cover... 85 7-4-8. Replacing the Wire for Tilting Unit... 86 7-4-9. Replacing the Tilting Unit for Mattress Platform... 86 7-4-10. Replacing the Table... 87 Main Body... 89 7-5-1. Replacing the Cord Cleat... 91 7-5-2. Replacing the AC Inlet... 91 7-5-3. Replacing the Fuse for the Main Body... 91 7-5-4. Replacing the Fuse for the HiLow Drive Board... 92 7-5-5. Replacing the Hi-Low Drive Board... 92 7-5-6. Replacing the Drive Power Transformer... 93 Base... 93 7-6-1. Replacing the Caster... 93 7-6-2. Replacing the Relay Board for Hi-Low Pedal... 94 7-6-3. Replacing the Foot Switch... 94 7-6-4. Replacing the Hi-Low Stand Actuator... 95
Operating Precautions
[1] Operating Precautions Please follow the operating instructions described in this Service Manual & Parts List for the safe use of the unit. The unit should be operated only by those who have been trained and instructed properly in its operation under the supervision of qualified medical personnel familiar with the currently known risks and benefits associated with the use of a radiant warmer. The unit should be operated only for its intended use.
1-1.
DANGER
Death or serious injury, damage to equipment or a fire will result if the instructions given below are not followed. Monitor the infant’s skin temperature when operating the unit. The operator must monitor the infant’s skin temperature when operating the unit. Avoid unattended use of the unit. Do not leave the unit unattended with the baby guards folded down. If the baby guards are left folded down, the infant may fall out and get fatally injured. Never leave the unit unattended with the baby guards folded down. Do not tilt the mattress platform with the baby guards folded down. If the mattress platform is tilted with the baby guards folded down, the infant may fall out. When you pull up and lock the baby guards in position, be sure to check that no mattress sheet or linen is caught between any of the baby guards and the mattress platform. Make sure that the baby guards are locked securely. If linen or other object is caught between any of the baby guards and the mattress platform, remove it. Otherwise, the baby guards will not be locked completely. The baby guards may swing down and the infant may fall out of the unit and suffer serious or fatal injury. Stop using the unit immediately and seek repair if the locking mechanism of the baby guards or other related parts should be found faulty in any way. The infant may fall out. Never place a body warmer or any other possible ignition source in or near the unit. Use of oxygen will increase the risk of explosion or fire. Body warmers or other devices in which fire is used or which will generate a spark may cause an explosion or a fire if used near the unit. Do not use the unit in the presence of a flammable anesthetic gas. The unit may cause an explosion or a fire if used in the presence of such a gas. Do not use ether, alcohol or any other ignitable substance. Even a small amount of ether, alcohol or any other ignitable substance may cause a fire when mixed with the oxygen supplied to the unit. Do not hang any flammable materials on the canopy. It may cause a fire.
10
Operating Precautions
PLEASE READ WITHOUT FAIL
Ground the unit securely. Otherwise, a leakage current may cause an electric shock. In order to complete the ground connection, connect the power cord only to a properly grounded 3P power outlet including a ground terminal. Do not operate the unit if you have any doubt about its ground connection. Do not use a device generating high frequency near the unit. To prevent malfunction of the unit due to jamming, do not use electric surgical knives, portable and mobile communication equipment and other devices which generate high frequency near the unit during its operation. Analyze arterial gas levels repeatedly when a high oxygen environment is required. When the infant requires a high oxygen environment, it is extremely important and essential to periodically analyze arterial gas levels. Follow the doctor’s instructions in measuring the oxygen concentration because ignoring essential requirements may increase the risk of retinopathy of prematurity and other adverse effects. Do not give a shock to the unit or let it hit anything. The screws or fixed parts may become loose. If the unit is to be used beyond its expected life span, overall repairs including replacement of parts must be carried out.
1-2.
WARNING
Death or serious injury due to a fire or an electric shock will result if the instructions given below are not followed. Be sure to follow the doctor’s instructions in setting the infant’s skin temperature. Be sure to follow the doctor’s instructions in supplying oxygen. Use only oxygen for medical use. Be sure to bear in mind the following precautions during oxygen supply. • Do not place a body warmer, a flashlight, oils and fats, or flammable vaporizable matters near the unit. • Use pure cotton for the infant’s clothing, bed sheets, etc. Do not use any material that is easily charged with static electricity. • Use pure cotton or fire-proof materials for the clothing of doctors, nurses and ambulance staff who handle this unit. Bear in mind the following precautions while using oxygen supply equipment. • If oil, grease or a grease-like substance should get in contact with pressurized oxygen, a violent spontaneous ignition may occur. Do not let such substances stick to the oxygen pressure regulator, the oxygen cylinder valve, piping, connections and other parts of oxygen supply equipment. • On a high-pressure oxygen cylinder, use only a tested pressure reducing valve or pressure regulating valve indicated specifically for oxygen supply. Do not use such a valve for any gas other than air or oxygen. It is dangerous to use a valve to supply a gas other than air or oxygen and then to supply oxygen again.
11
Operating Precautions
Smoking is prohibited in the room where the unit is installed. Do not place any possible ignition sources in the room. Avoid damaging the power cord. A damaged power cord may cause a fire or an electric shock. • Do not get the power cord caught between the unit and the wall, a shelf or the floor. • Do not place the power cord near a heating apparatus or heat it. • Do not put anything heavy on the power cord. • Always grasp the power plug with your hand to remove the power cord from the power outlet. A damaged power cord should be replaced immediately with a new one. Use only the power cord supplied with the unit. Otherwise, a fire or an electric shock may result. Do not touch the heater with your hand when the heater is on. You may burn yourself. Do not touch the power plug with a wet hand. Touching the power plug with a wet hand may cause an electric shock. Do not disassemble or modify the unit. Disassembling or modifying the unit may cause a fire, an electric shock or injury. Do not install the unit where it will be exposed to excessive humidity, dust or steam. Installing the unit in such a place may cause a fire or an electric shock. Before cleaning and disinfecting the unit, be sure to turn the power switch off, remove the power plug, and allow the heater temperature to drop sufficiently. The power outlet should be located near the unit to prevent accidental contact with a trailing power cord. Use a separate power outlet for each unit. Do not put many loads on one power outlet. In order to complete the ground connection, connect the power cord only to a properly grounded 3P power outlet including a ground terminal. Do not operate the unit if you have any doubt about its ground connection. Ground peripheral electric equipment securely. Never connect the unit to a power outlet other than that specified. The unit should be serviced only by qualified personnel. Be sure to inspect the unit at the start of each day. Operating the unit without inspecting it before at the start of each day may let a defect pass unnoticed and cause a potentially unfavorable outcome.
12
Operating Precautions
PLEASE READ WITHOUT FAIL
Check for conformity to the applicable standard if the unit is going to be connected to another medical device for use as a system. Accessory equipment connected to the analogue and digital interfaces must comply with the relevant IEC standards (e.g. IEC 60950-1 for information technology equipment). Furthermore all configurations should comply with IEC 60601-1. But the items that are not specified as part of the system should not be connected. Anybody who connects additional equipment to the signal input or signal output configures a medical system, and therefore should make the medical system comply with the requirements of IEC 60601-1 on his or her own responsibility. If in doubt, consult your local Atom representative. During transport, set the incubator to the lowest level. If not set to the lowest level, injury or damages may result when the incubator loses balance while being moved and hits a surrounding object, or falls down. Use the unit on a flat and stable surface. Placing the unit on a surface tilted over 10 degrees or on an unstable location may cause it to fall down.
1-3.
CAUTION
Injury or damage to surrounding objects may result if the instructions given below are not followed. Be sure to clean and disinfect the unit before using it for the first time after purchase. The unit is shipped without being disinfected. During the standby mode, be sure to preheat the unit in order to keep the mattress surface temperature stable. Place the infant on the mattress only after the mattress surface temperature has stabilized. Do not twist or pull the cords by force. If any defect should be found, ask an expert for repair without attempting to repair it yourself. When feeding the cords and tubes into the unit, be very careful not to let them wind or tighten around the patient. Remove the power plug from the power outlet before moving the unit to another place or when the unit is not going to be used for a long time. Moving the unit to another place with the power plug connected to the power outlet will damage the power cord and may cause a fire or an electric shock. Remove the power plug from the power outlet before cleaning and disinfecting the unit. Cleaning and disinfecting the unit with the power plug connected to the power outlet may cause an electric shock. Install the unit out of reach of small children.
13
Operating Precautions
Do not operate the unit with the canopy covered with a cloth, etc. Operating the unit covered with a cloth or pressed tightly against the wall may cause a fire or an electric shock due to overheating. Do not install the unit in direct sunlight or near a heating apparatus. Do not expose the unit to extraordinarily high temperature or excessive humidity. Do not place anything heavy on the unit. Do not let the unit hit anything, fall, or drop inadvertently. Do not attach any peripheral device that is not specified by Atom to the unit. Check the operation of the peripheral devices. If a device transmitting or receiving weak signals is installed near the unit, it may be affected by the electromagnetic waves generated by the latter. Check the operation of the peripheral devices for any effect before using the unit in clinical settings. Stop using the unit immediately if any trouble is detected. Be sure to cover the mattress with a mattress sheet. The mattress is not breathable and may cause a bedsore. When taking radiographs with the supplied X-ray cassette tray, read the included documentation and instruction manuals for the radiation device, and make sure that necessary protection measures are taken.
14
Parts Identification
2-1.
PLEASE READ WITHOUT FAIL
[2] Parts Identification Main Body
UPS
No. ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨ ⑩*
Name Canopy Lighting lamp Mattress Baby guard Foot switch Caster Mattress platform tilting lever Mattress platform Tube introduction slit packing Resuscitator
No. ⑪ ⑫ ⑬ ⑭ ⑮ ⑯ ⑰ ⑱ ⑲ ⑳
230V~ 50/60Hz
Name Operation panel Alarm lamp Heater Connector for weight monitor I/O port (communication connector) Lock lever X-ray cassette tray Cassette tray (handle) Connector for the Power Pack i (UPS) AC inlet
* The figure shows the unit in which the Resuscitation Unit (P type, E type) or the Blender Unit (P type, E
type) is installed. For information on the Resuscitation Unit (P type, E type) or the Blender Unit (P type, E type), see the Service Manual of the Unit concerned. 15
Parts Identification
2-2.
Support Column
∗
No. ① ② ③ ④ ⑤* ⑥ ⑦ ⑧
Name Operation panel Light control switch Power indicator Power switch Resuscitator Skin temperature probe connecting port 1 Skin temperature probe connecting port 2 Connector for SpO2
* The figure shows the unit in which the Resuscitation Unit (P type, E type) or the Blender Unit (P type, E type)
is installed. For information on the Resuscitation Unit (P type, E type) or the Blender Unit (P type, E type), see the Service Manual of the Unit concerned.
16
Display Screens
3-1.
General Description and Switching of the Screens
3-1-1.
General Description of the Screens
PLEASE READ WITHOUT FAIL
[3] Display Screens
[Start screen] The start screen appears when the power is applied. It is automatically switched to the main screen.
[Main screen] If any other screen is being displayed, touch
,
and the main screen will be displayed. This screen displays the heater setting, the heater output, the skin temperature 1, the skin temperature 2, SpO2, the pulse rate, the timer, etc.
[Pulse oximeter screen] Touch on any other screen, and the pulse oximeter screen will be displayed. This screen highlights the timer, SpO2, and the pulse rate. Pulse waves are also displayed. In addition, the heater setting, the heater output, the skin temperature 1, the skin temperature 2, etc. are displayed.
[Weight screen] (Option) Touch on any other screen, and the weight screen will be displayed. This screen highlights weight functions such as the weighing procedure and the weight trend graph. In addition, the heater setting, the skin temperature 1, the skin temperature 2, SpO2, the pulse rate, the timer, etc. are displayed.
17
Display Screens
[Trend screen] Touch on any other screen, and the trend screen will be displayed. This screen displays trend graphs.
[Menu screen] on any other screen, and the menu screen Touch for advanced settings will be displayed. Use this screen to select advanced settings. In addition, the set temperature, the skin temperature 1, the skin temperature 2, SpO2, the pulse rate, etc. are displayed. When this screen is displayed, you can only select advanced settings of each menu item or switch the screen to another.
3-1-2.
Screen Transition Diagram Pulse oximeter screen
Weight screen
Main screen
Menu screen
Trend screen : [Main screen] display switch : [Pulse oximeter screen] display switch : [Weight screen] display switch : [Trend screen] display switch : [Menu screen] display switch 18
Display Screens
List of Operations Available on Each Screen
PLEASE READ WITHOUT FAIL
3-2.
Table 1. List of Setting Operations Available on Each Screen Setting (measuring) Operation
Trend screen
Main screen
Menu screen
Pulse oximeter screen ∗1
Weight screen ∗1
Temperature-related operation (manual/servo) Setting SpO2/pulse rate upper/lower alarm limits ∗1 Weighing the infant ∗1 Timer ∗2 Selecting other advanced settings : Setting (measuring) operation available : Setting operation not available : The function not provided ∗1: Option ∗2: Types selectable on the menu screen
3-3.
Main screen Heat area (See p.21)
Pulse area (See p.20) Timer area (See p.21)
Message area (See p.21) Clock and other indicators area (See p.20) Screen display switch area (See p.20)
19
Display Screens
[Pulse area] ∗ Touch this area to start the pulse oximeter-related setting operation. No.
Name Description Plethysmograph Indicates changes in the artebar rial flow. Displays a detected SpO2 value %SpO2 display digitally. Pulse rate Displays a detected pulse rate display digitally. Displays a pulse rate upper Pulse rate alarm alarm limit above and a pulse limits display rate lower alarm limit below digitally. Displays an SpO2 upper alarm SpO2 alarm limit above and an SpO2 lower limits display alarm limit below digitally. Interference Comes on when interference indicator is detected. (Nellcor only) Pulse search Comes on when no pulse is indicator detected. (Nellcor only)
No.
Name Main screen display switch Trend screen display switch Menu screen display switch Pulse oximeter screen display switch Weight screen display switch
No.
Name
[Screen display switch area] Description Touch this switch to have the main screen displayed. Touch this switch to have the trend screen displayed. Touch this switch to have the menu screen displayed. Touch this switch to have the pulse oximeter screen displayed. Touch this switch to have the weight screen displayed.
[Clock and other indicators area]
20
Description Touch this switch to silence an Alarm silence active audible alarm temporariswitch ly or reset an alarm condition. Displays a date in the month/ Date display day/year format. Displays a time in the Time display hour:minute format. Comes on only when the inBattery indicator ternal battery (Power Pack i : option) is used