Cleaning Instructions
10 Pages
Preview
Page 1
ATOM V-808
Cleaning Instructions
PARKER H e a l t h c a r e w w w . p a r k e r h e a l t h . c o m . a u
Parker Healthcare Pty Ltd Head Office & International Department:
16 Redland Drive, Mitcham VIC 3132 Australia Ph: +61 3 9872 0222 Fax: +61 3 9873 5911 Email: [email protected]
2014 Rev. 1.2
IMPORTANT Please ensure all steps of the cleaning regime are followed after each use.
Strip the linen from the incubator and process as per hospital protocol .
Remove the maƩress and wipe over with a damp cloth of warm water & hospital grade detergent.
MaƩress
DO NOT fully emerse maƩress in cleaning soluƟon.
Incubator Head End
Stopper
PLATFORM Open the head end admiƩance panel located on the leŌ side of the hood, draw out the maƩress plaƞorm toward you unƟl the stopper is engaged. To fully remove the maƩress plaƞorm, press the stopper to disengage the lock and remove.
STEP 3
Mattress Platform
Stopper Position
STEP 2
MATTRESS
STEP 1
LINEN
PLEASE FOLLOW HOSPITAL PROTOCOL FOR LINEN PROCESSING.
Wipe the maƩress plaƞorm with a soŌ cloth moistened with hospital grade detergent.
PARKER H e a l t h c a r e Page 1
STEP 4
ACCESS PORT Remove the iris access port cover. Immerse and clean in a hospital grade detergent.
Iris Access Port
STEP 5
SNAP-OPEN ACCESS PORT PACKING Remove the rubber packing from the snapopen access ports. Immerse and clean in a hospital grade detergent.
ADMITTANCE PANEL PACKING
Snap-open Access Panel Packing
STEP 6
Remove the rubber packing from the admiƩance panels. Immerse and clean in a hospital grade detergent.
TUBE INLET Remove the packing from the tube inlets. (both sides of incubator) Immerse and clean in a hospital grade detergent.
PARKER H e a l t h c a r e Page 2
Tube inlet
AdmiƩance panel packing
Front AdmiƩance Panel
Fixing Screw PosiƟons
1. Open the front admiƩance panel. Loosen the fixing screws for each inside panel (i.e. front, rear, top and right) and remove the inner panels. 2. Clean both sides of each inside panel with a soŌ cloth moistened with a hospital grade detergent.
Light release buƩons
Gently press the light release buƩons on both sides of the light housing and liŌ in an upward direcƟon.
STEP 8
REMOVING THE LIGHT
STEP 7
Fixing Screws
INSIDE PANEL (FRONT, REAR, TOP, RIGHT)
PARKER H e a l t h c a r e Page 3
REMOVING THE HOOD
NOTE: THE FOLLOWING OPERATION WILL REQUIRE TWO (2) PEOPLE.
1. Unclip the front metal
1. Metal side clips (2 clips - L & R)
side clips (L & R) to release the hood from the base. 2. Pull the release screw at the back of the hood on both sides of the
STEP 9
incubator. NOTE: Removal of the hood will require two (2) people to complete. Using one person on each side of the hood, liŌ up evenly and parallel to the base unƟl the hood has been cleared from the incubator body. Place on a flat surface and wipe over with a soŌ cloth dampended with hospital grade detergent.
PARKER H e a l t h c a r e Page 4
2. Release screw (2 screws - L & R)
Middle Deck Hold both sides of the middle deck and liŌ. Fan Cover
Wipe over middle deck plaƞorm with a soŌ cloth moistened with hospital grade detergent
STEP 10
MIDDLE DECK PLATFORM
FAN COVER & FAN REMOVAL
AŌer removal of the fan cover, clean it with a soŌ cloth moistened with a hospital grade detergent. To remove the fan, turn the fixing screw clockwise. Once the fixing screw has been removed, use 2 hands to liŌ the fan.
Fixing Screw
STEP 11
Fan Cover
Hold both sides of the fan cover and liŌ.
Immerse and clean in a hospital grade detergent.
Fan
PARKER H e a l t h c a r e Page 5
CONDITIONING CHAMBER
STEP 12
The condiƟoning chamber will appear when all the components menƟoned previously have been removed. Clean thoroughly with a soŌ cloth moistened with hospital grade detergent. Ensure all parts are completely dry before undertaking re-assembly. Re-assemble all removed parts in the reverse order of dis-assembly.
CondiƟoning Chamber
Check all parts are aƩached correctly.
STEP 13
Once re-assembled and checked, please ensure the incubator is plugged in, switched ON and the incubator temperature is set to 33 degrees.
SKIN TEMPERATURE PROBE Wipe the enƟre probe (including thermostat) with a soŌ cloth moistened with hospital grade detergent between use. Thermostat
PARKER H e a l t h c a r e Page 6
This page is intentionally left blank.
Distributed by:
PARKER H e a l t h c a r e w w w . p a r k e r h e a l t h . c o m . a u
Parker Healthcare Pty Ltd Head Office & International Department:
16 Redland Drive, Mitcham VIC 3132 Australia Ph: +61 3 9872 0222 Fax: +61 3 9873 5911 Email: [email protected]