HORIZON Nxt®
Operation Manual
©1999 B.Braun Medical Inc.
This manual may not, in whole or in part, be copied, photocopied, reproduced, translated, or converted to any electronic or
machine-readable form without prior written consent of B.Braun Medical Inc.
950520 Rev F
Page 2
Table of Contents
Chapter 1 INTRODUCTION.....................................................................................1-1
Warranty ..............................................................................................................1-1
Intended Use .......................................................................................................1-2
Epidural Administration ............................................................................... 1-2
Blood Infusion .............................................................................................. 1-3
q(x)o ............................................................................................................. 1-3
Chapter 2 PRODUCT DESCRIPTION .....................................................................2-1
Operating Modes..................................................................................................2-1
Operating States ...................................................................................................2-1
Illustrations ..........................................................................................................2-2
Important Keys ....................................................................................................2-4
Daisy Chaining ....................................................................................................2-4
Battery Operation.................................................................................................2-4
Chapter 3 PRINCIPLES OF OPERATION...............................................................3-1
Cassette Removal.................................................................................................3-1
Standard Mode .....................................................................................................3-2
For Primary Fluids ....................................................................................... 3-2
For Piggyback Fluids ................................................................................... 3-2
Piggyback Callback .............................................................................................3-3
Changing the Rate................................................................................................3-3
During Infusion (titration) ........................................................................... 3-3
With Pump on Hold ..................................................................................... 3-4
Chapter 4 MENUS ....................................................................................................4-1
Main Menu ..........................................................................................................4-1
Alternate Menu ....................................................................................................4-1
Change Alarm Volume ................................................................................ 4-2
Clear Total Infused ...................................................................................... 4-2
Dose Mode .................................................................................................... 4-2
Occlusion Limit ........................................................................................... 4-3
Alternate Menu Items .........................................................................................4-4
Calculate BSA .............................................................................................. 4-4
Set LCD Contrast.......................................................................................... 4-4
Chapter 5 CUSTOM MENU OPTIONS ..................................................................5-1
Check Battery Level ...........................................................................................5-2
Hold Extender ......................................................................................................5-3
Micro Rate / Limit ...............................................................................................5-3
Monitor Docking..................................................................................................5-4
Occlusion Warning ..............................................................................................5-1
Piggyback Callback .............................................................................................5-5
PROFILE Mode ..................................................................................................5-5
PROGRAM Mode ..............................................................................................5-5
q (x)° Schedule.....................................................................................................5-5
Schedule Next Run .............................................................................................5-9
Select MAX Pressure ..........................................................................................5-10
Set PROFILE Parameters ...................................................................................5-11
Horizon Nxt®
Modular Infusion System
950520 Rev F
6/00
Page 3
Set Time and Volume .........................................................................................5-11
Site Trend.............................................................................................................5-12
Exit from Menu....................................................................................................5-12
Chapter 6 DOSE MODE ...........................................................................................6-1
Drug Menus .........................................................................................................6-1
Epidural Category ................................................................................................6-2
Concentration Screen ...........................................................................................6-2
Data Screen ..........................................................................................................6-2
Change SetUp Parameters ..................................................................................6-3
Titrate Rate .........................................................................................................6-4
Titrate Dose .........................................................................................................6-4
Exit DOSE Mode ................................................................................................6-4
Chapter 7 SPECIAL MODES....................................................................................7-1
Profile Mode ........................................................................................................7-1
Change Profile Parameters: ...........................................................................7-1
Data Entry .....................................................................................................7-1
Review/Change PROFILE Data ....................................................................7-2
Restart PROFILE ...........................................................................................7-2
Immediate Ramp Down ................................................................................7-2
Exit PROFILE Mode ....................................................................................7-3
Program Mode ....................................................................................................7-4
Data Entry .....................................................................................................7-4
Review/Change PROGRAM ........................................................................7-5
Restart PROGRAM .......................................................................................7-6
Skip PROGRAM Periods .............................................................................7-6
Exit PROGRAM Mode .................................................................................7-6
Chapter 8 MAINTENANCE ....................................................................................8-1
Cleaning ..............................................................................................................8-1
Alarm Procedures ................................................................................................8-2
Alarm Troubleshooting Guide .............................................................................8-3
Cautions ..............................................................................................................8-5
Service and Performance Information .................................................................8-7
Specifications ......................................................................................................8-8
Chapter 9 Table 1 DRUG LIST ................................................................................9-1
Index ....................................................................................................................Index-1
Horizon Nxt®
Modular Infusion System
950520 Rev F
6/00
Page 4
Chapter 1 INTRODUCTION
Warranty
B. Braun Medical Inc., hereby warrants that reasonable care has been used
in the manufacture of each Horizon Nxt® and that it shall be free from
defects in material and workmanship for a period of two years after the date
of shipment, when properly used and maintained. Any Horizon Nxt® that
is found not to meet these standards within this two-year period will be
repaired after examination by B. Braun Medical or, at B. Braun Medical’s
option, will be replaced without charge. If the defect has been caused by
misuse, unauthorized modification, or abnormal conditions or operation,
repairs will be billed at B. Braun Medical’s then current charges. In this
case, an estimate will be submitted before work is started, if requested. The
defective device should be returned promptly to B. Braun Medical, properly
packaged, postage prepaid. Loss or damage in return shipment to B. Braun
Medical will be at purchaser’s risk. The foregoing express warranty, as
conditioned and limited, is in lieu of and excludes all other warranties not
expressly set forth herein, whether expressed or implied by operation of law
or otherwise including, but not limited to, any implied warranties of
merchantability or fitness, since handling, storage, and cleaning of these
devices, as well as factors relating to other matters beyond B. Braun
Medical’s control, directly affect these devices and the results obtained from
their use. B. Braun Medical shall not be liable for any incidental or
consequential loss, damage, or expense, directly or indirectly arising from
the use of this device other than replacement of all or part of it. B. Braun
Medical neither assumes, nor authorizes any other person to assume for it,
another or additional liability or responsibility in connection with this
device.
This warranty will not apply to any Horizon Nxt® or component thereof
which has been: (1) repaired by anyone other than an authorized B. Braun
Medical trained technician or repair facility; (2) altered in any way that, in
B. Braun Medical’s judgement, affects its stability or reliability; (3)
subjected to misuse, negligent handling, or accident; (4) used in any manner
not in accordance with the instructions furnished by B. Braun Medical in the
Operation and/or Service Manual.
This warranty extends only to the first purchaser or lessee and does not
extend to, and may not be enforced by, any other person.
Horizon Nxt®
Modular Infusion System
950520 Rev F
1-1
6/00
Page 5
Intended Use
The Horizon Nxt® is intended for use with B. Braun Medical Horizon®
I.V. sets to regulate the flow of primary and secondary fluids when positive
pressure is required. The infusion system is capable of delivering fluid from
a negative head height (when the I.V. fluid container is lower than the
pump), and provides clinically accepted volumetric accuracy for all
standard I.V. fluids, including blood, lipids and TPN.
Positive pressure is frequently a necessity but clinical experience shows that
high pressure limits can increase the severity of an infiltration without
causing an alarm. Because there is a need to control pump pressure settings,
the Horizon Nxt has user-selectable occlusion pressure limit settings which
start at 75 mmHg and extend to 500 mmHg. At rates of 400 mL/hr and
higher, the pressure setting is automatically increased to 400 mmHg. For
epidural infusions which require higher pressures, the pump has an extended
occlusion pressure setting of 750 mmHg.
Because some infusions require multiple flow rates during the delivery
cycle, the Horizon Nxt is equipped with two automatic multiple delivery
rate modes. The Profile Delivery mode incorporates a standard ramped
delivery regimen, and the Programmable Delivery mode can be set up to
sequentially deliver up to nine user-selected volumes of fluid each at a userselected rate. Both modes are designed for ease of and can be adapted to
home-care settings.
The Horizon Nxt also incorporates a Dose/Rate mode which automatically
calculates the rate when dose information is entered, or the dose when rate
information is entered.
The pump is equipped with distinct audible and visual alarm signals to
indicate KVO, low battery, and actual alarm conditions.
Epidural Administration
The Horizon Nxt can be used for epidural administration of anesthetic and
analgesic drugs.
Caution: Whenever the pump is being used for epidural infusions, the
pump’s secondary infusion features should not be used.
For epidural administration of anesthetic drugs, use indwelling catheters
specifically indicated for short-term (96 hours or less) anesthetic epidural
drug delivery.
Horizon Nxt®
Modular Infusion System
950520 Rev F
1-2
6/00
Page 6
For epidural administration of analgesic drugs, use indwelling catheters
specifically indicated for either short-term or long-term analgesic epidural
drug delivery.
When the Horizon Nxt® is being used for epidural infusions, the source
container and administration set should be clearly differentiated from source
containers and administration sets used for other routes of infusion.
Caution: To prevent inadvertent infusion of drugs which are not indicated
for epidural use, only Horizon microbore administration sets,
without injection sites, that are specified for epidural
administration of drugs should be used.
Following an occlusion alarm, a small amount of fluid may be delivered.
Use of a microbore set assures this is less than 0.1 mL.
The administration of drugs is restricted to those anesthetic and analgesic
drugs labeled for continuous epidural administration.
Warning: Serious injury to the patient may result from epidural
administration of drugs other than those specifically labeled for
epidural use.
Blood Infusion
The Horizon Nxt can be used for infusion of blood and blood products. For
blood infusion, a Horizon Pump Blood Set should be used. Do not attempt
to piggyback blood. The normal saline roller clamp on the blood tubing set
must be clamped off. The high viscosity of blood will cause the saline to
infuse before the blood if both clamps are left open.
q(x)°
The Horizon Nxt can be used for the periodic infusion of antibiotics,
chemotherapy or other IV infusions according to a schedule set by the user.
Assure that there is a maintenance IV infusing in the intervals between
pump infusions.
Horizon Nxt®
Modular Infusion System
950520 Rev F
1-3
6/00
Page 7
Chapter 2 PRODUCT DESCRIPTION
The Horizon Nxt® has been developed in reponse to the special needs of all
caregivers in the medical setting.
Operating Modes
The Horizon Nxt pump has five operating modes:
Standard Mode - for simple IV fluid delivery with piggyback capability.
Dose Mode - used for administration of medication where concentrations
and doses are critical. The pump will automatically calculate either the dose
or rate.
Profile Mode - offers a tapered delivery mode for infusing TPN or other
fluids requiring ramp up and ramp down time.
Program Mode - offers maximum flexibility in customizing IV fluid
administration. Up to 9 separate periods of fluid volume can be infused
sequentially.
q(x)° Mode - can be used for the periodic infusion of antibiotics,
chemotherapy or other IV infusions according to a schedule set by the user.
Assure that there is a maintenance IV infusing in the intervals between
pump infusions.
Operating States
There are 4 operating states:
Hold - silences an alarm, stops an infusion, or is a quick escape from the
menu.
Alarm - stops an infusion to notify the user of an error.
KVO - 3 mL/hr for delivery rates of 3 to 999.9 mL/hr, or the continuation
of the delivery rate for 0.1 to 2.9 mL/hr settings.
Run - starts the infusion.
Horizon Nxt®
Modular Infusion System
950520 Rev F
2-1
6/00
Page 8
Illustrations:
Figure 1: Front View
1. Alarm Indicator
2. Run Indicator
3. Hold Indicator
4. Battery Use Indicator
5. AC Use Indicator
6. Power Key
7. Primary Indicator Key
8. Piggyback Indicator Key
9. Primary Run Key
10. Piggyback Run Key
11. Data Keys
7
6
5
4
3
12. Hold Key
13. Menu Key
14. Cursor Keys
15. Enter Key
16. Piggyback Rate Display
17. Piggyback Volume Display
18. Information Screen
19. Primary Rate Display
20. Primary Volume Display
21. Door Lever
2
1
9
10
21
19
19
20
20
16
16
17
17
8
Horizon Nxt®
Modular Infusion System
950520 Rev F
18
18
12
11
2- 2
13
14
15
6/00
Page 9
Figure 2: Rear View
1. Panel Lock Out Switch
2. Serial Communication Data Port
1
2
B. Braun Medical
Carrollton, TX 75006
Figure 3: Inside the Door
1
2
3
1. Cassette Alignment Pins
2. Air-In-Line Detector
3. Free Flow Protection Clip
Horizon Nxt®
Modular Infusion System
950520 Rev F
2-3
6/00
Page 10
Important Keys
An understanding of the function and location of these keys will assist in the
use of the pump.
From Figure 1:
• Power key (6) turns the pump on and off
• Data keys (11) allow input of data values
• Hold key (12) silences an alarm, stops an infusion, is a quick escape from
the menu, and extends the Hold state an additional 3 minutes every time it
is pushed if Hold Extender is active.
• Menu Key (13) lets the user see options using a menu screen
• Cursor keys (14) move the cursor bar on the Information Screen
• Enter key (15) validates or accepts the selection made
From Figure 2:
• Panel Lock Out Switch locks the keypanel to prevent tampering. The
switch can only be used while the pump is infusing.
Daisy Chaining
Daisy chaining allows up to 5 Horizon Nxt® or Horizon® pumps to be
plugged into each other with only one pump plugged into a wall outlet. On
the side of each Horizon Nxt pump is an electrical outlet. Only another
Horizon or Horizon Nxt pump should be plugged into that outlet.
Caution: To daisy chain more than 5 pumps, verify that the total leakage
current as a system does not exceed 100 uA.
Battery Operation
The Horizon Nxt has been equipped with an internal, rechargeable battery
which will power the pump automatically if the plug is disconnected during
use. Battery power is provided as a back-up system only, and the length of
the battery life is affected by the rate of delivery. Whenever possible, the
pump should remain plugged into an electrical outlet. The Battery Use
indicator will be lit when the pump is operating on its battery.
The pump must remain plugged into a hospital-grade electrical outlet for at
least 24 hours to fully recharge the battery from a totally discharged
condition.
Horizon Nxt®
Modular Infusion System
950520 Rev F
2- 4
6/00
Page 11
Chapter 3 PRINCIPLES OF OPERATION
Caution: Use only with B. Braun Medical Horizon® I.V. Sets
1) To attach the pump to the I.V. pole:
a)
b)
c)
d)
Press down on the pole clamp lever.
Position the pump on the pole (make sure the pole is up against
the slot of the pump).
Release the pole clamp.
Lock by rotating the pole clamp lever 1/4 turn, toward the rear
of the pump.
2) Plug the pump’s power cord into a hospital-grade electrical outlet, or the
AC Receptacle on the side of another Horizon or Horizon Nxt pump.
3) Press the Power key. If the Information Screen says:
"PROGRAM Operation Was Interrupted..." choose an option using the
CURSOR keys. Press the ENTER key.
"PROFILE Operation Was Interrupted..." choose an option using the
CURSOR keys. Press the ENTER key.
4) Open the door of the pump by pulling the Door Lever.
5) Prepare the IV fluids as described on the set package. Close the
downstream roller clamp.
6) Position the cassette over the Cassette Alignment Pins, with the metal
disc toward the pump. Thread the tubing through the Air-In-Line
Detector. Insert and press the Free Flow Protector Clip into the
receptacle.
Warning: Unrestricted fluid flow may occur if the Free Flow Protector
Clip is not properly installed in the receptacle.
7) Close the door of the pump.
8) Open all clamps on the pump set.
Note: Do not use a pump with visible damage or with bent, damaged or
missing components.
Cassette Removal
1) Press the HOLD key. Open the door of the pump.
2) Grasp the Free Flow Protector Clip and pull it from the Free Flow
Protector receptacle while grasping the tubing as it enters the left side of
the pump. Gently pull with both hands to remove the cassette from the
pump. Grasping the Free Flow Protector Clip and removing it from the
pump will automatically stop the flow of fluid in the tubing.
3) If gravity infusion is desired, close the roller clamp, open the Free Flow
Protector Clip and adjust the flow with the roller clamp.
Horizon Nxt®
Modular Infusion System
950520 Rev F
3-1
6/00
Page 12
Standard Mode
For Primary Fluids:
1)
2)
3)
4)
Press the PRIMARY INDICATOR key.
Press the DATA keys to enter the rate and volume.
Press the PRIMARY RUN key to start the infusion.
Press the PANEL LOCK OUT SWITCH to make the key panel tamper proof, if
desired.
The time left, total infused, and occlusion pressure limit are shown on the
Information Screen. Monitor the infusion according to hospital policy.
After the infusion is complete, the pump will alarm and go into the KVO
state.
For Piggyback Fluids:
The Piggyback option can only be enabled by Biomedical personnel. The
piggyback option is not available while operating the pump in the Dose
Mode.
Caution: Do not attempt to infuse both fluids simultaneously using this
method.
1) Prepare IV fluids according to the set package. Use primary sets with
check valves.
2) Connect the piggyback set to the upper injection site (the site above the
pump) on the primary set.
3) Lower the primary bag at least 8 inches.
Note: To minimize or prevent fluid flow from the primary container during
Horizon Nxt®
ModularInfusion System
950520 Rev F
3-2
6/00
Page 13
a piggyback infusion (sympathetic flow), it may be necessary to
lower the primary bag more than 8 inches or clamp off the primary
tubing. Sympathetic flow increases significantly when the
piggyback rate is greater than 125 ml/hr, and clamping the primary
tubing is recommended at rates greater than 125 ml/hr.
4) Press the PIGGYBACK INDICATOR key.
5) Enter the Piggyback rate and volume.
Note: The programmed Piggyback volume to be infused must be equal to
the amount of fluid in the Piggyback container to prevent incorrect
delivery. Automatic piggyback is not intended for the infusion of
fluids requiring flushing before and after administration because of
incompatibility.
6) Press the PIGGYBACK RUN key.
When the Piggyback volume is complete, the Piggyback Rate and Volume
Displays will blank. The Primary Rate and Volume Displays will light up.
The infusion will now be delivered at the primary rate. Make sure the
primary tubing is not clamped.
Note: If the secondary container is not empty, the remaining secondary
fluid will be delivered at the primary flow rate.
Piggyback Callback
This feature causes the pump to alarm and continue the infusion at a KVO
rate after the programmed piggyback volume is infused. Piggyback
Callback must be selected in the User menu.
Changing the Rate
During Infusion (titration):
1) Press the DATA keys to change the rate.
2) Press the ENTER key or the PRIMARY RUN key to validate data.
Note: If the ENTER key or RUN key is not pressed within 4 seconds of
entering a new rate, a series of error beeps will occur and the rate
change will be cancelled. The pump will then display the message
"Rate change aborted", and will continue to infuse at the old rate.
An intermittent alarm will continue to sound to notify the user that a
rate change has been attempted and cancelled. Pressing any key will
silence the alarm.
With Pump on Hold:
Horizon Nxt®
Modular Infusion System
950520 Rev F
3-3
6/00
Page 14
1) Press the HOLD key.
2) Press the DATA keys to change the rate.
3) Press the RUN key to start the infusion.
Horizon Nxt®
ModularInfusion System
950520 Rev F
3-4
6/00
Page 15
Chapter 4 MENUS
The Horizon Nxt® can be customized to fit your specific needs. The User
Menu and Alternate Menu contain options identified by each individual
hospital. The following options will always appear in one of the 2 menus.
Main Menu
Change Alarm Volume changes the volume of the audible alarm.
Clear Total Infused clears the total IV fluid infused shown on the
Information Screen.
Dose Mode calculates the rate or dose information necessary for use with
many critical drug infusions.
Occlusion Limit sets the occlusion pressure limit to 75, 100, 200, 300, 400
or 500 mmHg.
Alternate Menu lists options used less often by the user.
Exit from Menu returns to the Information Screen.
Alternate Menu
Calculate BSA calculates the body surface area (BSA).
Set LCD Contrast changes the readability of the Information Screen.
Biomed Options allows access to the Biomed Menu.
Exit from Menu returns to the Information Screen.
Horizon Nxt®
Modular Infusion system
950520 Rev F
4-1
6/00
Page 16
Change Alarm Volume:
The volume of the alarm can be changed to low, medium or high.
1) Press MENU key
2) Press CURSOR keys to select Change Alarm Volume
3) Press ENTER key
1) Press CURSOR keys to change the
intensity of the alarm
2) Press ENTER key
Change Alarm Volume
Low
Medium
High
Move Cursor & Press ENTER
The pump will return to the previous Hold or Run state.
Clear Total Infused
This feature clears the Total Volume displayed on the Information Screen.
Note: The volume can be cleared while the pump is running.
1) Press MENU key
2) Clear Total Infused is already selected
3) Press ENTER key.
1) Press ENTER key.
2) The total volume displayed
on the Information Screen
goes to zero.
The present Total Infused
of 125.02 ml will be
zeroed. Is this OK?
Yes No
Cursor
& Press ENTER
The pump will return to the previous Hold or Run state.
Dose Mode:
The Dose Mode is fully described in the Dose Mode Chapter.
Horizon Nxt®
Modular Infusion System
950520 Rev F
4 -2
6/00
Page 17
Occlusion Limit
This feature lets the user change the occlusion pressure limit to 75, 100, 200,
300, 400, or 500 mmHg. At rates of 400 mL/hr and higher, the pressure
setting is automatically increased to 400 mmHg. Occlusion pressure is a
threshold value.
For Example: If the occlusion pressure is set to 300 mmHg, the pump will
infuse at the lowest pressure possible but will alarm if the
pressure required to infuse the IV fluid rises above the 300
mmHg value.
A gravity infusion of D5W with the IV container 3 feet above the IV site is
equivalent to approximately 70 mmHg.
The occlusion warning feature is optional. The occlusion warning feature
gives an audible warning (3 beeps) and a visual indication in the LED
display to let the user know an occlusion is occuring downstream from the
pump. This warning occurs 5 times at 5-second intervals before alarming
“Occlusion.” This feature is designed to alert the clinician of downstream
resistance before an alarm occurs which stops the infusion.
Caution: The higher the pressure limit, the less sensitive the pump is to
changes in fluid resistance which may be caused by positional
IV’s or infiltrations. High pressure limits can increase the
severity of an infiltration without an alarm condition.
Note:
1)
2)
3)
4)
5)
To select a pressure of 750 mmHg, refer to Select MAX
Pressure.
Press MENU key
Press CURSOR keys to select Occlusion Limit
Press ENTER key
Press DATA keys to change the pressure limit
Press ENTER key
The pump will return to the previous Hold state.
Horizon Nxt®
Modular Infusion system
950520 Rev F
4-3
6/00
Page 18
Alternate Menu Items:
Calculate BSA:
Use this feature to calculate body surface area (BSA).
1) Press MENU key
2) Press CURSOR keys to select Calculate BSA
3) Press ENTER key
Note: Height can be measured in centimeters (cm) or inches (in). Weight
can be measured in grams (g), kilograms (kg) or pounds (lb). The
height and weight parameters are changed by pressing theCursor to
select the parameter and pressing the up or down Cursor key.
1) Press CURSOR keys to select “Ht” and
“Wt” values
2) Press the DATA keys to enter
the information
3) The BSA value is displayed
4) Press ENTER to exit
Calculate BSA
Ht = 100 cm Wt = 70 kg
BSA = 1.23 m2
Press ENTER when done
Set LCD Contrast:
This option allows the user to change the readability of the Information
Screen.
1) Press MENU key
2) Press CURSOR keys to select Set LCD Contrast
3) Press ENTER key
1) Press right and left CURSOR keys
to vary the contrast
2) Press ENTER key to validate
EXIT
Adjust cursor
until
this message is clear.
Then press ENTER key.
The pump will return to the previous Hold or Run state.
Horizon Nxt®
Modular Infusion System
950520 Rev F
4 -4
6/00
Page 19
Chapter 5 CUSTOM MENU OPTIONS
The following items can be added to either the User or Alternate
Menus by the Biomedical Department. The Custom Menu items are
listed in alphabetical order.
Check Battery Level graphically displays the state of the charge of
the battery.
Hold Extender lengthens the hold time to a maximum of 120
minutes. This option must be turned on by the Biomedical
Department.
Micro Rate / Limit limits the infusion rate. (Useful for Pediatric and
Neonatal settings).
Monitor Docking allows the Horizon Nxt® with MedScan to
communicate with select monitoring systems to visually display
current infusion parameters. This option must be turned on by
the Biomedical Department.
Occlusion Warning is designed to alert the clinician of downstream
resistance. An audible warning (three beeps) and a visual
indication will occur 5 times at 5-second intervals before
alarming "Occlusion." This option must be turned on by the
Biomedical Department.
Piggyback Callback continues the primary infusion at a KVO rate
and alarms once the piggyback is complete.
PROFILE Mode incorporates a ramped delivery regimen based on
the patient's specific needs.
PROGRAM Mode lets the user program a maximum of nine
selected volumes of fluid.
q(x)° Schedule allows the programming of periodic IV infusions.
Schedule Next Run delays the start of an infusion for up to 24 hours.
Select MAX Pressure sets the occlusion pressure limit to the
maximum value of 750 mmHg.
Set PROFILE Parameters alters the ramping process.
Set Time and Volume allows the delivery rate to be calculated based
on time. (for example: 1000 ml to be infused over 8 hours)
Site Trend graphically displays the resistance to downstream fluid
flow. A Resistance Factor can be enabled by the Biomedical
Technician which will display a numerical value both on the
Run/Hold screen and when Site Trend graph is displayed.
Each of the above options are described more fully in this chapter.
The options are listed in alphabetical order.
Horizon Nxt® with MedScan
Modular Infusion System
950520 Rev F
5-1
6/00
Page 20
Check Battery Level:
Three states of the battery are possible: discharging, charging or
fully charged. Prior to unplugging the pump for transport purposes,
this feature makes it easy to determine the state of the battery.
1)
2)
3)
4)
5)
Press MENU key
Press CURSOR key to select Alternate Menu
Press the MENU key
Press CURSOR key to select Check Battery Level
Press ENTER key
One of the three graphs below will be displayed:
Discharging:
The pump is unplugged or running
on battery.
Check Battery Level
DISCHARGING
EXIT
Charging:
The pump is not fully charged.
Check Battery Level
CHARGING
Fully Charged:
The pump is 95% or more charged.
Check Battery Level
FULLY CHARGED
EXIT
Press the ENTER key to return to the Hold or Run state.
Horizon Nxt® with MedScan
Modular Infusion System
950520 Rev F
5-2
6/00
Page 21
Hold Extender
The normal Hold Time is 3 minutes. Use this option to extend that
time to a maximum of 120 minutes.
1) Press HOLD key to add time in
3 minute increments, or
2) Press CURSOR keys to select the
hours or minutes and use DATA
keys to change the value
3) Press ENTER key.
Delay = 0 hr 23 min 40 sec
Press HOLD to add 3 min
Cursor
to change
Cancel
"Hold" is displayed on the LEDs. Press the RUN key to cancel the
pause and begin infusing. Press the ENTER key to cancel the
extended pause and return to the standard 3 minute hold. Press the
MENU key to cancel the pause and access the Menu system.
Micro Rate / Limit
This option is useful in environments where low rates are critical,
such as pediatrics or neonatal nurseries. The maximum rate limit
can be set as low as 5.0 ml/hr or as high as 99.9 ml/hr.
Note: If the pump is turned on with the Micro Rate Limit activated,
a Teddy Bear shows on the LCD screen for 4 seconds.
1)
2)
3)
4)
5)
6)
Horizon Nxt® with MedScan
Modular Infusion System
950520 Rev F
Press HOLD key
Press MENU key
Press CURSOR keys to select Alternate Menu
Press ENTER key
Press CURSOR keys to select "Micro Rate / Limit"
Press ENTER key
5-3
6/00
Page 22
1) Press right CURSOR key to select
On
2) Press down CURSOR keys to select
Rate Limit
3) Press DATA keys to set a rate limit
4) Press ENTER key
Set Micro Status:
Off On
Rate Limit = 59.9 ml/hr
Press ENTER when done
Note: If a rate exists or is entered that is higher than the Rate Limit
set, an audible signal occurs. The user will need to lower the
rate or exit this feature.
1) Press RUN key to start the
infusion.
MICRO
Time Left = 7 hr 34 min
Total Infused = 24.1 ml
Occlusion Limit = 300 mmHg
Monitor Docking
Monitor Docking allows the Horizon Nxt® with MedScan to
communicate with select monitoring systems to display current
infusion parameters. Parameters displayed may vary depending on
which monitoring system is being used. Refer to Monitor Docking
Guide for operations and troubleshooting.
Set-up
Clamp the Serial Data Communication IR (Infrared) Module over
the Serial Data Communication fluid resistant window on the back
of the Horizon Nxt pump whose information you wish displayed.
1) Grasp the Serial Data Communication IR Module with index
finger on top and thumb on grooved surface on the bottom.
2) Hook the top of the unit onto the pump, and with thumb press the
grooved surface to clamp the bottom of the unit on the pump.
Occlusion Warning
Occlusion Warning is designed to alert the clinician of
downstream resistance. An audible warning (three beeps) and
visual a indication will occur 5 times at 5-second intervals before
alarming "Occlusion." This option must be turned on by the
Biomedical Department.
Horizon Nxt® with MedScan
Modular Infusion System
950520 Rev F
5-4
6/00
Page 23
Piggyback Callback
Piggyback Callback, if activated, continues the primary infusion at
a KVO rate and causes the pump to alarm. This feature may be
accessed during the Run or Hold states.
1)
2)
3)
4)
5)
Press MENU key
Press the CURSOR key to select Alternate Menu
Press ENTER key
Press CURSOR keys to select “Piggyback Callback”
Press ENTER key
1) Press CURSOR keys to select On
2) Press ENTER key
Note: Piggyback Callback will
return to the Off position when the
pump is turned off.
When Piggyback Callback
is ON, KVO will alarm as
soon as the Piggyback is
done.
Set Callback:
Off
On
PROFILE Mode
The Profile Mode is described in Chapter 7 (Special Modes).
PROGRAM Mode
The Program Mode is described in Chapter 7 (Special Modes).
q(x)° Schedule
This option allows the programming of periodic antibiotics,
chemotherapy or other IV infusions. For example, antibiotics are
written to be infused "q 6° " (every 6 hours). It is important to have
a maintenance IV infusing. This option can only be enabled by
Biomedical personnel.
Note: A cassette must be loaded in the pump to program this
feature. The piggyback option is not available while using
this feature. The q(x)° option cannot be used while in DOSE,
PROFILE or PROGRAM Modes.
1)
2)
3)
4)
Horizon Nxt® with MedScan
Modular Infusion System
950520 Rev F
Press HOLD key
Press MENU key
Press CURSOR keys to select q(x)° Schedule
Press ENTER key
5-5
6/00
Page 24
1) Press right CURSOR keys to select
OK
2) Press ENTER key
This feature allows
for periods of no
infusion. Ensure IV
patency prior to start
Cancel
OK
To Enter Data:
1) Press CURSOR keys to select "q(x)
hrs and min"
2) Press the DATA keys to set the
time interval between infusion.
(For example: q 6 hrs 00 min)
Dose q(x) 01 hr 00 min
# Doses = 4
Dose Rate = 200.0 ml/hr
Dose Vol = 100 ml
Press ENTER when done
Note: The time interval can be 1 to 24 hours. There must be at least
15 minutes between each run.
1) Press down CURSOR key to select "# of Doses"
2) Press DATA keys to set the total number of doses to be infused.
Note: The total number of doses may not exceed 40.
1) Press down CURSOR key to select "Dose Rate"
2) Press the DATA keys to set the infusion rate in ml/hr.
1) Press CURSOR keys to select "Dose Vol"
2) Press the DATA keys to set the volume of each dose (for example:
4 doses, each 50 ml)
Note: The Dose Volume of each dose has to be greater than 1 ml
but not more than 2000 ml.
3) Press the ENTER key
To change the time:
1) Press up CURSOR key to select
"Present Time is"
2) Press DATA keys to set the correct
time
Present Time is
09:37a
Start dose at
10:35a
VALIDATE
Note: The time is displayed in military time. Use 00:00 to represent
Horizon Nxt® with MedScan
Modular Infusion System
950520 Rev F
5-6
6/00
Page 25
midnight and 12:00 to represent noon. Verify that the Time
is correct.
1)
2)
3)
4)
Press CURSOR to select "Start Dose at"
Press DATA keys to set the time the infusion is to begin
Press the CURSOR keys to select "Validate"
Press ENTER key
Note: "q(x) Schedule Overlap"
message appears when doses
entered are not at least 15 minutes
apart. When this occurs the q(x)°
feature must be exited to cancel the
present infusion regimen.
"Wait Is" shows the amount of time
until delivery starts
q(x) Schedule Overlap
Press HOLD to Silence Alarm
q(x) Start 10:35a
DATA
Wait Is 0 h 49 m
3 Doses left
Total Infused = 2.2 ml
Note: After pressing Enter, the pump goes into a waiting period.
"SCHd dLY" shows on the LED’s. No fluid is infusing
during this period.
1) Press ENTER key for the data screen
2) Follow the steps above to change the data, if necessary
3) Press ENTER key
At the q(x)° Start time, the pump will begin the infusion. During the
infusion "SCHd run" shows on the LED’s. The number of doses
remaining is shown on the LCD. After the infusion is complete, the
pump will automatically return to the waiting period. The pump will
continue to "infuse and wait" for the total number of doses entered.
q(x) Delivery
DATA
Time Left = 0 hr 29 min
Total Infused = 2.7 ml
Occlusion Limit = 300 mmHg
3 Doses left
To Exit:
Horizon Nxt® with MedScan
Modular Infusion System
950520 Rev F
5-7
6/00