perfusor_space_service_manual_ver_2-0-1.pdf
Page 1
0-12
Perfusor® Space
Service Manual
Version 2.0.1 English
0-
Page 2
0
This Service Manual is valid for:
Designation
Part No.
Infusion syringe pump Perfusor® Space . . . . . . . . . . . . 0871 3030
This Service Manual is available under
the following part number:
Designation
Part No.
Languages of this Manual
The Service Manual for this unit can be supplied in the following
languages:
Service Manual Perfusor® Space, English. . . . . . . . . . . 8713 9020
Designation
Part No.
Service Manual Perfusor® Space, German . . . . . . . . . . 8713 9010
Service Manual Perfusor® Space, English (US). . . . . . 8713 9020U
Service Manual Perfusor® Space, French . . . . . . . . . . . 8713 9030
The complete Service Manual contains
the following pages:
0-2
Page 0-1 to page 0-12
Page 1-1 to page 1-12
Page 2-1 to page 2-30
Page 3-1 to page 3-60
Page 4-1 to page 4-2
Page 5-1 to page 5-2
Page 6-1 to page 6-2
Page 7-1 to page 7-6
Page 8-1 to page 8-4
Page 9-1 to page 9-6
Page 10-1 to page 10-2
Page 11-1 to page 11-2
Perfusor® Space 2.0.1 gb
Page 3
0-
Table of Contents
0
Important Preliminary Remarks
Service Work
Technical Safety Checks
Current Versions
Revision Service
Quality Management
Checks and Repair
Notes on ESD
Spare Parts and Test Equipment
Setting Off
Page
Page
Page
Page
Page
Page
Page
Page
Page
0-5
0-5
0-5
0-5
0-6
0-6
0-6
0-7
0-7
Contact Persons
Technical Training
Entry for Technical Training
Ordering of Spare Parts and Test Equipment
Service Hotline
Return of Spare Parts and Test Equipment
Safety Officer
(§ 30 MPG)
Translation
Page
Page
Page
Page
Page
0 - 11
0 - 11
0 - 11
0 - 11
0 - 11
Page
Page
0 - 11
0 - 11
System Overview
Description
System Overview
Physical Construction
Function
Unit Software
Service Program
Technical Data
Options
Accessories
Page
Page
Page
Page
Page
Page
Page
Page
Page
1-1
1-1
1-2
1-3
1-6
1-7
1 - 12
1 - 12
1 - 12
Unit Diagnosis / Calibration
General
Alarms and Error Codes
CAN error states
The Most Important Error Modes
Device Check
Calibration
Procedural Instructions for Calibration
Trouble Shooting
Page
Page
Page
Page
Page
Page
Page
Page
2-1
2-3
2-8
2-9
2 - 10
2 - 15
2 - 15
2 - 29
Disassembly / Assembly
General
Battery Module
Housing Foot
Operating Unit
Upper Part of Housing
Release Button
Loudspeaker
Drive
Syringe Holder with Piston Brake
Processor PCB
Assembly / Installation
Pole Clamp SP adapter kit
Checks after Repair
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
3-1
3-9
3 - 16
3 - 17
3 - 22
3 - 24
3 - 24
3 - 25
3 - 32
3 - 39
3 - 40
3 - 54
3 - 58
Perfusor® Space 2.0.1 gb
0- 3
Page 4
0
Table of Contents
Servicing the Unit
Cleaning
Servicing the Battery
Page
Page
4-1
4-1
Technical Safety Check (TSC)
Perfusor® Space
Page
5-1
Technical Safety Check (TSC)
Power Supply SP
Page
6-1
Procedural Instructions on the TSC
Visual Inspection
Electrical Safety
according to IEC/EN 60601-1
or VDE 0750 and VDE 0751
Functional Inspection Perfusor® Space
Functional Inspection Power Supply SP
Page
7-1
Page
Page
Page
7-2
7-3
7-6
Test equipment
Special Tools
Page
Page
8-1
8-3
Page
9-1
Page
10 - 1
Page
11 - 1
Test Equipment and Special Tools
Spare Parts List
Revision Documentation
Index
0- 4
Description of Version
Perfusor® Space 2.0.1 gb
Page 5
0-
Service Work
Technical Safety Checks
Important Preliminary Remarks
0
The present manual is for your information only. The possession of
this manual does not authorize the performance of service work.
Service tasks may only be executed by persons, who
-
have received appropriate training on the system from
B. Braun
-
are included in the revision service
-
possess the necessary test equipment and mechanical aids,
and
-
fulfill the personal requirements (training and knowledge).
The user is obliged to perform or to have performed the Technical
Safety Checks on those medial products for which these checks
have been prescribed by the manufacturer and to carry them out
according to the indications of the manufacturer as well as the
generally approved technical standards while adhering to the periods stated (§ 6 MP BetreibV).
B. Braun also recommends training on the Technical Safety
Checks, or to perform at least the steps indicated in the current
version of the manual, as:
-
the TSC requires that the instructions in the manuals are observed
-
the manuals are a reference for measurements
-
depending on the unit type, the Service Program must be
called which may lead to a dangerous unit condition in case
of inappropriate operation. Furthermore, a special service
connector may be necessary.
Current Versions
This manual version corresponds to the state when the manual
was written. B Braun reserves the right to make technical modifications. The state of the revision is indicated by the index number
in the footer of every page.
Revision Service
The possession of this manual does not automatically mean inclusion in the revision service. You will be included in the revision
service after:
Perfusor® Space 2.0.1 gb
-
technical training by B. Braun Melsungen or
-
a written order placed with the sales department of B. Braun
(fee required).
0- 5
Page 6
0
Important Preliminary Remarks
Responsibility of the Manufacturer
Quality Management
The manufacturer, person who assembles, installs or imports the
device can only be held responsible for safety, reliability and performance if
-
mounting, enhancements, new settings, changes or repairs
are carried out by duly authorized persons,
-
the electrical installation in the corresponding room meets
the requirements of the VDE 0107, VDE 0100 part 710 or
IEC 60364-7-710 and the national standards,
-
the device is used in accordance with the instructions for use
and the Service Manual,
-
the Technical Safety Checks are performed at regular intervals,
-
a current manual which corresponds to the revision state is
used when carrying out maintenance, repair and service,
-
the service technician takes part in the revision service,
-
the technician has participated in a technical training course
for the specific B. Braun unit.
B. Braun is certified in accordance with DIN EN ISO 9001 and
ISO 13485. This certification also includes maintenance and service.
The unit has the CE label. The CE label confirms that the device
corresponds to the “Directive of the Council for Medical Products
93/42/EC” of June 14, 1993.
Checks and Repair
Training may only be performed by B. Braun. The possession of the
manual does not authorize the performance of repairs. The instructions on electrostatic sensitive components (ESD standards)
must be observed.
After repair a device check or diagnosis is to be carried out.
Notes on ESD
0- 6
Semiconductors can be destroyed by electrostatic discharge. Especially MOS components can be damaged by interference from
electrostatic fields, even without discharge via contact. This type
of damage is not immediately recognizable. Unit malfunctions
can even occur after a longer period of operation.
Perfusor® Space 2.0.1 gb
Page 7
Important Preliminary Remarks
0
Each workstation must be equipped according to the recommendations with the necessary static protective measures, if ESD
components or boards are handled.
Each workstation must be equipped with a conductive table surface. The conductive surface, the soldering iron or the soldering
stations must be grounded via protective resistors.
Chairs must be of antistatic design. The floor or floor mats should
be of electrically conductive material.
Fig.: 0 - 1
Spare Parts and Test Equipment
Personnel must wear conductive wristbands which are connected
to a central ground potential via protective resistors, e.g. the
ground contact of a wall outlet. Furthermore it is recommended
that personnel wear cotton clothing and electrically conductive
shoes to prevent electrostatic charge.
Only use original spare parts from the manufacturer. Do not
tamper with assembly groups which can only be exchanged completely. The spare parts required are listed in the repair descriptions.
Service personnel are responsible for the calibration of their test
equipment. Original test equipment can be calibrated at the
works of B. Braun. Further information is available upon request.
Setting Off
Additional notes and warnings are set off as follows:
Note
Is used for additional or special notes concerning information and
working steps.
CAUTION
Is used for working steps which may result in damage to the unit,
system or to a connected device.
! WARNING
IS USED FOR WORKING STEPS WHICH MAY RESULT IN PERSONAL
INJURY.
Perfusor® Space 2.0.1 gb
0- 7
Page 8
0
Important Preliminary Remarks
References to chapters are shown as follows
(see “Setting Off“ ➨ pg. 0 - 7)
References to figures and tables are shown as follows
Fig.: 2 - 3 or Table 2 - 1
References to item numbers in figures are shown as follows
(Fig.: 1 - 1 / Item 1)
In this case “Fig.: 1 - 1“ is the figure number and “Item 1“ the item
number within the figure.
When the Service Manual is stored as pdf-file, these references
are displayed green. Click with the mouse button on a reference
to jump to the corresponding source.
Menu commands are described as:
Menu File.
0- 8
Perfusor® Space 2.0.1 gb
Page 9
Important Preliminary Remarks
List of Abbreviations
Perfusor® Space 2.0.1 gb
0
Abbreviations which are not generally known, but are used in this
manual, are listed below.
CAN
Controller Area Network
CE
Communauté Européenne
(European Communities)
CS
Calibration Step
DIN
Deutsche Industrie Norm
(German Industrial Standard)
EN
European Standard
ESD
Electrostatic Discharge
FuP
Function Microprocessor
IEC
International Electrotechnical
Commission
ISO
International Standardization
Organization
ISP
Infusomat® Space
ISPS
Infusomat® Space, Silicon
ISPP
Infusomat® Space, PVC
KuP
Monitoring Microprocessor
LCD
Liquid Crystal Display
MOS
Short for the following
company name:
MOS Technology, Inc.
(Commodore Semiconductor
Group)
PCA
Patient-Controlled Analgesia
PSP
Perfusor® Space
SP
Space (System)
SPC
SpaceCover
SPCC
SpaceCover comfort
SPCS
SpaceCover standard
SPCO
SpaceCom
SPCT
SpaceControl
SPS
SpaceStation
TEMP
Temperature
TS
Troubleshooting Step
TSC
Technical Safety Checks
0- 9
Page 10
0
0 - 10
Important Preliminary Remarks
UTS
Unit Test Step
VDE
Verband der Elektrotechnik,
Elektronik und
Informationstechnik e.V.
(German electrical engineering
association)
Perfusor® Space 2.0.1 gb
Page 11
0 - 12
0-
Contact Persons
0
Technical Training
Via local representative.
Entry for Technical Training
Application for a technical training course must be made via the
responsible representative.
Ordering of Spare Parts and Test Equipment
Please contact your local B. Braun subsidary.
International Technicians (Intercompany)
Irene Marchel
Fax:
+49 (0) 5661 / 75 - 38 57
e-mail: irene.marchel@bbraun.com
Service Hotline
Karl Tippel, Martin Heusner
Phone:
+49 (0) 56 61 71 - 35 25
Fax:
+49 (0) 56 61 71 - 35 26
E-Mail: karl.tippel@bbraun.com
E-Mail: martin.heusner@bbraun.com
Return of Spare Parts and Test Equipment
B. Braun Melsungen AG
Schwarzenberger Weg 73-79
Wareneingang Werk C
34 212 Melsungen
Germany
Safety Officer
(§ 30 MPG)
Dr. Ludwig Schütz
e-mail: ludwig.schuetz@bbraun.com
Perfusor® Space 2.0.1 gb
0 - 11
Page 12
0
Contact Persons
For your notes:
0 - 12
Perfusor® Space 2.0.1 gb
Page 13
1 - 12
1-
Description
System Overview
1
The Perfusor® Space (PSP) is according to IEC/EN 60601 resp. IEC/
EN 60601-2-24 a transportable infusion syringe pump for administrating fluids in the nutritional therapy and infusion technique
as well as for home care applications.
The medical specialist must decide on suitability for application
on the basis of the warranted properties and the technical data.
System Overview
The Space system is a modular design of modern infusion technology for stationary, mobile or private use. The key modules and
their connection to the peripheral devices are shown in Fig.: 1 1.
All the pump types, Perfusor® Space, Infusomat® Space and Infusomat® Space P, as well as the other devices of the system are of
modular design. Up to three pumps can be connected together
mechanically using L rails on the bottom of the unit and grooves
on the top. They can then be fastened to a drip stand or appropriate rail using the pole clamp.
1
2
The SpaceControl module can be used to extend operation. One
single pump can be inserted onto this module. The pump is then
connected via connectors to the module.
3
4
5
Fig.: 1 - 1 Space system
Legend of fig. 1 - 1:
ItemDesignation
1
SpaceCover
2
Infusion pump Infusomat® Space
3
Infusion syringe pump Perfusor® Space
4
SpaceControl
5
SpaceStation
Perfusor® Space 2.0.1 gb
The SpaceStation module allows the set-up of a complete pump
system with up to 24 pumps. Up to four pumps can be installed in
every SpaceStation. The pumps are supplied with power via the
integrated power supply and the built-in connectors. The pumps
are connected to the optional SpaceCom via these connectors.
SpaceControl can also be integrated into the system.
Up to six SpaceStations can be set-up as a column with a total of
24 pumps. SpaceStation placed next to each other can be connected via special connection cables, if the maximum number of
24 pumps in maximum three columns is not exceeded.
SpaceCover Standard or SpaceCover Comfort forms the top of
each column. Alarms are signalled by a row of LEDs and a loudspeaker in the SpaceCover Comfort.
1- 1
Page 14
1
System Overview
Physical Construction
1
2
5
4
3
6
8
7
Fig.: 1 - 2 Perfusor® Space
Legend of fig. 1 - 2:
ItemDesignation
6
Connector “P2“ for SpaceStation module, external 12 V DC
and accessories
Syringe holder with piston brake
7
Connector “P3“, connection to SpaceControl module
4
Operating Unit
8
Battery compartment cover
5
Syringe area
1
Perfusor® Space
2
Drive head
3
1- 2
Perfusor® Space 2.0.1 gb
Page 15
System Overview
1
The Perfusor® Space housing mainly consists of the bottom part
and the upper part.
The battery module is inserted in the rear of the housing upper
part. The opening is covered by the battery compartment cover.
The operating unit is attached to the front of the bottom part with
two hinges. This operating unit covers the area for the syringes.
The complete drive assembly, consisting of lead screw and drive
head with driving tube is located directly behind the syringe area
in the bottom part of the housing. The housing bushing for the
driving tube is located in the side of the housing.
The syringe holder is mounted in the right side of the housing bottom part.
The processor PCB with the permanently connected external connectors “P2” and “P3” is located at the bottom of the housing bottom part.
Function
There are two power options for the Perfusor® Space:
-
via the inserted battery module
-
via an external 12 V DC power supply (e.g. SpaceStation,
SpaceControl, an external power supply or from an ambulance car) connected to connector “P2”
The voltage supplied is converted to the internal voltages required
through a voltage transforming and monitoring circuit on the
processor PCB.
An independent circuit in the battery module monitors the battery
cells and controls their charge condition.
The Perfusor® Space is connected to a SpaceControl by connector
“P3”.
The function processor controls all the functions of the Perfusor®
Space. Data is stored in a non-volatile memory which also controls the external data transfer.
The control microprocessor monitors all important responses of
the function processor to incoming information. If a response
does not correspond with that expected by the control microprocessor, an error message is generated and the device is switched to
a safe stop state.
The drive motor is monitored by a detector for speed and direction
of rotation. The extended end position of the drive head is detected by a switch on the processor PCB.
Perfusor® Space 2.0.1 gb
1- 3
Page 16
1
System Overview
The pressure in the infusion system is measured through a strain
gauge measuring in the drive head and monitored in the device
electronics. The data from the strain gauge is continuously compared with the limit values which are calculated dependent on the
selected syringe type and the pressure settings. When the limit
values are exceeded an alarm is automatically triggered and the
pressure in the infusion system is reduced. The maximum pressure
is additionally limited by a second, independent system. This maximum pressure limitation is performed using the motor current
control.
The syringe size detection is performed via the syringe holder. The
syringe holder is connected to a potentiometer. The syringe size is
determined from the resistance of the potentiometer.
The syringe is fixed with the syringe holder and the axial fastening
device. The syringe piston is fastened with two claws in the drive
head. When a syringe is inserted the syringe piston is held by the
piston brake, until the piston has been caught by the claws.
Keyboard and display as well as the syringe area are illuminated.
1- 4
Perfusor® Space 2.0.1 gb
Page 17
System Overview
1
Fig.: 1 - 3 Block diagram Perfusor® Space
Perfusor® Space 2.0.1 gb
1- 5
Page 18
1
System Overview
Unit Software
Approved Software Versions
Position
1 2 3 4 5 6 7 8 9 10
Digit
6 8 8 C 0 3 0 0 0 1
Revision level
Hardware
Software group
Device type: Perfusor® Space
Fig.: 1 - 4
688A030032
-
Basic software
688A030035
-
Improved functions
688A030040
-
Improved functions
-
Languages French and Swedish added
688B030002
-
Improved functions
688B030003
-
CAN bus functioning
688C030001
-
Dose calculation
-
Changed CAN log
688D030001
-
Drug list data base
-
Changed user language
688E030003
-
Improved functions
-
Piggyback
-
Soft limits
688F030006
-
PCA
-
Changed claw configuration
-
Optimized alarm handling
688G030002
-
Improved functions
688H030002
-
1- 6
TCI
Perfusor® Space 2.0.1 gb
Page 19
System Overview
1
Software Update of the Unit
The instructions for updating the software are supplied with the
software itself.
CAUTION
If the device is disconnected while the software is being updated
or the device or PC is switched off, a component of the software
may be seriously damaged so that repairs are no longer possible.
In such a case the software cannot be updated via the PC and the
device must be returned to B. Braun.
Service Program
Approved Version
Note
Please note that text and / or functions of the Service Program
may change depending on the software version. The following
screen illustrations are only examples and represent the state
when the manual was printed.
-
0.0.28
-
1.0.0
-
1.1.2
-
1.1.3
-
1.1.4
-
1.2.1
-
1.3.5
-
1.5.0
-
2.0.1
-
3.1.0
-
4.0.0
-
5.1.0
-
6.1.0
Starting the Service Program
Note
Installation and further operation of the Service Program is described in its separate instructions for use.
Perfusor® Space 2.0.1 gb
1- 7
Page 20
1
System Overview
1.
Start the “HiBaSeD.exe” program (History, Barcode, Service,
Drug list) on the PC. The Service Program is loaded and started and the initial window of the Service Program is displayed.
2.
Read the notes carefully.
3.
Mark the field “I accept all conditions” and then the field
“Yes” to confirm that you have read the notes.
Note
Click the field “English” to switch the language of the notes over
to English.
Fig.: 1 - 5
4.
Enter the password and confirm it by clicking the field “Start”.
Fig.: 1 - 6
The Service Program checks the PC interfaces for connected
devices of the Space system. Units that were found are displayed for a short moment on the screen.
Fig.: 1 - 7
1- 8
Perfusor® Space 2.0.1 gb
Page 21
System Overview
1
The work window of the Service Program appears on the
screen. All devices recognized are listed in the left column.
Fig.: 1 - 8
5.
Perfusor® Space 2.0.1 gb
Activate the desired device from the list on the left in the
work window with a double-click. The device data is then displayed below the device name.
1- 9
Page 22
1
System Overview
If the unit software version is not compatible with the Service
Program version, a window opens prompting the operator to
change the Service Program version. This window displays a
compatibility list of the Service Program- and unit software
versions.
If Service Program- and unit software versions are compatible, all the Service Program functions are activated.
Fig.: 1 - 9
Fig.: 1 - 10
1 - 10
Perfusor® Space 2.0.1 gb
Page 23
System Overview
1
Service Program Version
1.
Open the “HiBaSeD“ window via Help ➨ Info .... The current
version of the Service Program is shown in this window.
2.
Close the window by clicking “OK”.
Fig.: 1 - 11
Compatibility List
1.
Open the “Unit - Compatibility” window via Help ➨ Compatibility. This window displays the compatibility of the HiBaSeD-version and the unit software version.
2.
Close the window by clicking “OK”.
Fig.: 1 - 12
Quit the Service Program
Perfusor® Space 2.0.1 gb
1.
Exit the Service Program via Application ➨ Quit.
2.
Disconnect a power supply which might be connected from
the unit.
3.
Switch off the unit.
4.
Remove the battery module.
5.
The device can be restarted after appr. 10 seconds.
1 - 11
Page 24
1
System Overview
Technical Data
All technical data is indicated in the instructions for use.
Options
The functions of the individual options are detailed in the instructions for use.
Perfusor® Space
Designation
Part No.:
Power supply Euro . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3110A
Power supply UK. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3111A
Power supply USA / Japan . . . . . . . . . . . . . . . . . . . . . . 0871 3112A
Power supply Australia. . . . . . . . . . . . . . . . . . . . . . . . . 0871 3113A
Power supply South Africa. . . . . . . . . . . . . . . . . . . . . . 0871 3115A
Accessories
Designation
Part No.:
Charger SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3170
battery charging station
Connection cable staff call SP. . . . . . . . . . . . . . . . . . . . 0871 3232
Power supply cable 12 V . . . . . . . . . . . . . . . . . . . . . . . . 0871 3231
for ambulance cars
CombiLead SP 12 V . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3133
connection cable, pump - pump
InterfaceLead SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3234
interface cable RS232
InterfaceLead SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3230
interface cable CAN SP
SpaceClamp SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3130
The SpaceClamp is a holder attached on beds for one
or several Space system pumps.
Short stand SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3135
Space PCA kit (PCA button) . . . . . . . . . . . . . . . . . . . . . 0871 3554
Syringe Anti Removal Cap PSP . . . . . . . . . . . . . . . . . . . 0871 3556
1 - 12
Perfusor® Space 2.0.1 gb
Page 25
2 - 30
2-
Unit Diagnosis / Calibration
General
2
! WARNING
While testing the unit and trouble shooting the operator/service
technician must work with voltages up to 115 / 230 V AC.
THESE VOLTAGES MAY CAUSE INJURIES WHICH ARE DANGEROUS
TO LIFE AND LIMB.
-
The national and international safety regulations are to be
adhered to.
Before each disassembly and assembly of a unit subsystem check
the connectors, plug contacts and connections for corrosion and
tight fit. These fault types are not described again in the following
trouble shooting list.
The following equipment and gauges are necessary for testing the
unit and/or performing troubleshooting:
-
PC
-
Service connector SP
-
Service Program HiBaSeD
-
Interface cable
-
Syringe 2 ml / 3 ml
-
Syringe 10 ml
-
Syringe 30 ml
-
Diameter gauge 32.0 mm
-
Diameter gauge 23.4 mm
-
Diameter gauge 15.7 mm
-
Diameter gauge 9.0 mm
-
Length gauge PSP
-
Syringe gauge “#Lehre OPS 50“with push-button plate and
motor power test adapter for Perfusor® Space
There are pictures of the gauges in Chapter “Special Tools“ (➨
pg. 8 - 3).
CAUTION
Take special care when carrying out measurements on an open
and switched-on unit. Short circuits and wrong measuring methods can cause serious damage to or destroy the subsystems of the
device.
Perfusor® Space 2.0.1 gb