Program Mode
Alarm Troubleshooting
1. Press OPTS
until PROGRAM MODE is
displayed
2. Press ON
.
Air In Line!
4. Enter the volume for Step 1 and press OK
.
6. Review the Total VTBD and press OK
Battery
Too Low!
• Plug pump into wall outlet.
Door Open!
• Close door.
Dwnstream
Occlusion!
• Open clamps.
.
.
7. Press START
• Remove air from line.
.
5. Repeat Steps 3 and 4 until all steps are
complete and press OK
Ramp Taper Mode
1. Press OPTS
until RAMP TAPER MODE is
displayed
.
2. Press ON
.
4. Enter the Ramp Up time in hours and minutes
and press OK
.
5. Enter the Taper Down time in hours and minutes
and press OK
• Check occlusion pressure
setting.
and minutes and press OK
.
7. Review the calculated Max Rate and
8. Press START
Pump
Stopped!
• Pump is stopped.
• Either start infusion or press
to extend hold time
3 minutes.
.
6. Enter the Total Time for the infusion in hours
press OK
• Check for kinked tubing.
• Check IV site.
3. Enter the total bag volume to be delivered and
press OK
Infusion Pump
• Check that tubing is properly
inserted in air-in-line detector.
.
3. Enter the rate for Step 1 and press OK
• Check for air in air-in-line
detector.
Vista™ basic
Silence alarms by pressing
.
Refer to Instructions for Use for complete
operating instructions.
.
Quick
Reference
.
B. Braun Medical Inc.
824 Twelfth Avenue
Bethlehem, PA 18018
1-800-227-2862
www.bbraunusa.com
Rx only
Vista™ and Working Together for Excellence in Safety
are a trademarks of B. Braun Medical Inc.
©2002, B. Braun Medical Inc.
950853-Rev B 06/02
Page 2
Important Keys
Piggyback Mode
Yes, No, OK, etc.
Number
Keys
: returns pump to last
• No
programmed infusion.
7
8
9
0
On/Off
(Power)
4
5
6
C
Alarm Silence
1
2
3
START
STOP
5. Is this a new therapy?
• Yes
: clears current infusion data.
Start/Stop
6. Prime set?
• Yes
: follow prompts to prime tubing.
• No
Alarm Light
Infusing Light
1. Press OPTS
until PIGGYBACK is displayed
.
2. Press ON
.
3. Press END
.
4. Press PIGY
5. Enter Rate/VTBD or VTBD/Time as in
Continuous Mode.
6. Press START
.
: begin programming infusion.
Continuous Mode
1. Enter the rate using the number keys.
Setting the Pump
1. Press
Changing the Rate
1. While infusing,
enter the new rate
using the number
keys.
(On/Off).
2. Press button on top of door to open.
3. Insert tubing into pump.
•
•
•
•
Push open tube pincher A
Place tubing in tube pincher A
Insert tubing along channel B
Push tubing into Air In Line Detector
2. Press OK
C
2. Press VTBD
.
3. Enter the bag volume using the number keys.
4. Press OK
.
NOTE:
If the OK key is not pressed within 10 seconds
of the change, the infusion will continue at the
original rate.
A
Changing the VTBD
B
1. Stop infusion.
2. Press VTBD
.
5. Press START
C
4. Close Door.
.
3. Press
.
4. Enter the new
volume to be
delivered.
5. Press OK
.
6. Press
950853-Rev 06/02
.
.