Stimuplex® HNS 12
Regional Anesthesia
Nerve Stimulator for
Peripheral Regional Anesthesia
GB
HC-RA-LLDorder_00739_Grunewald 090358
Technical Service Manual
Page 2
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 2
Manufacturer:
Distributor:
STOCKERT GMBH
Bötzinger Straße 72
79111 Freiburg
Germany
B. Braun Melsungen AG
Hospital Care Division
34209 Melsungen
Germany
Phone: +49 (0)761-20716-0
Fax: +49 (0)761-20716-20
E-mail: Info@stockert.de
http://www.stockert.de
Phone: +49 (0)5661-71-0
Fax: +49 (0)5661-71-4845
E-mail: StimuplexHNS12@bbraun.com
http://www.bbraun.com
In the US and Canada distributed by:
B. Braun Medical Inc.
824 Twelfth Avenue
Bethlehem, PA 18018-3524
Phone: 1-800-854-6851
Fax: 1-610-758-9020
E-mail: inquiry@bbraunusa.com
http://www.bbraunusa.com/stimuplex
Caution:
The Stimuplex® HNS 12 may only be used in applications for which this product is intended.
Federal (US) law restricts this device to sale by or on the order of a physician.
Stimuplex® HNS 12 Technical Service Manual
Order number: 6043572 – Rev. C – valid for software version 1.003.x
Meliseptol® and Stimuplex® are registered trademarks of B. Braun Melsungen AG
Page 3
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 3
Contents
3
Contents
Page
1.
1.1
1.2
Technical data................................................................................................................................................5
Battery ..............................................................................................................................................................5
Cleaning and disinfecting the Stimuplex® HNS 12..............................................................................6
2.
2.1
2.2
Maintenance and technical safety checks (with battery installed) ........................................7
Visual inspection of Stimuplex® HNS 12 including accessories ......................................................7
Functional check of Stimuplex® HNS 12 ................................................................................................7
3.
3.1
3.2
3.3
3.4
3.5
Instructions for technical safety checks ............................................................................................9
Visual inspection of Stimuplex® HNS 12 and accessories ................................................................9
Functional check of Stimuplex® HNS 12 ................................................................................................9
Checking the operating controls and monitoring equipment ..........................................................9
Checking the accessories ..........................................................................................................................10
Checking the stimulation signal ..............................................................................................................11
4.
Replacing the electrode cables ............................................................................................................12
5.
Repairs ............................................................................................................................................................12
6.
Test Equipment Logbook Stimuplex® HNS 12 ..............................................................................14
Page 4
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 4
4
The Stimuplex® HNS 12 is a nerve stimulator used for locating peripheral nerves in regional anesthesia and pain therapy.
The Stimuplex® HNS 12 is a class lla medical device according to Council Directive 93/42/EEC.
The device should not be used until it has been subjected to an on-site function test and the persons
responsible for operating the device are instructed now to use the device with the aid of the user
manual.
Page 5
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 5
1. Technical data
5
Type:
Instrument type:
Battery:
Power consumption:
Stimulation current:
Stimulation voltage:
Stimulation frequency:
Stimulus duration:
Stimuplex® HNS 12
BF
9V (alkaline)
6 mA (8 mA max.)
Î = 5 mA (max.) (0–12 kΩ)
Û = 95 V (max.)
1 Hz / 2 Hz ±1% / SENSe (3 Hz)
0.05 ms – 0.10 ms – 0.30 ms – 0.50 ms – 1.00 ms ±1%
SENSe (0.10 ms - 0.10 ms - 0.15 ms to 1.0 ms)
Allowable load impedance:
0 kΩ–12 kΩ
Current measuring accuracy:
± 0.02 mA
Impedance measuring range:
1 kΩ – 90 kΩ for target stimulation current > 0.5 mA
Impedance measuring accuracy:
±10% / ±20% for target stimulation current > 1mA / <= 1mA
Sound pressure level:
51 dB / 54 dB / 63 dB for stimulation / warning / error
Weight:
250g
Operational environmental conditions: 0 – 50° C, max. 90 % relative humidity, no condensation
1.1 Battery
The battery charge status must be checked regularly (MENU info = state of charge indicated in
“percent” and “volt”). Whenever the battery symbol is flashing, the current application can be completed without problems, but the battery should be replaced afterwards.
Please do not replace the battery when the device is switched on. This may damage the LCD
display.
If the voltage is less than 6.00 V, an error message will be displayed and the device will switch off
automatically. Replace battery.
If the Stimuplex® HNS 12 nerve stimulator is not intended to be used for a long period of time, the
battery must be removed to prevent leakage.
The battery compartment is located in the base of the nerve stimulator. When installing the battery,
ensure correct polarity and that the battery contacts are not pressed too tightly together to ensure
solid contact.
Only use 9 V alkaline manganese batteries – e.g. VARTA 4922, DURACELL MN 1604.
These will provide you with an optimum operating time. Rechargeable batteries must not be used as
they may cause the device to malfunction.
Page 6
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 6
6
WARNING!
If the battery is leaking, Stimuplex® HNS 12 should no longer be operated for reasons of safety.
It must be returned to the manufacturer for correct cleaning and a safety check.
1.2 Cleaning and disinfecting Stimuplex® HNS 12
Use only soft, damp cloths to clean and disinfect the Stimuplex® HNS 12 and accessories, e.g. the
electrode cable. Water or soap are particularly suitable for this purpose. Ensure that no moisture
penetrates the Stimuplex® HNS 12.
Use only wipe disinfection, no spray disinfection! Avoid condensation!
White spirit or commercially available methanol-free disinfectant in an ethyl alcohol base can be used
for disinfection.
Meliseptol® or white spirit can be used for wipe disinfection.
Caution! The following substances must not be used: trichloroethylene, acetone, butanone, benzene
methyl ethyl ketone, benzene, methanol, cellulose thinner, and 2-propanol or any other organic solvents, acids and sodium hydroxide solution. Disinfectants containing iodine or dyes can discolor the
housing and should not be used.
Page 7
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 7
2. Maintenance and technical safety checks (with battery installed)
We recommend running the following simple checks before starting up operation for the first time
and then regularly (e.g. before daily use):
- Visual inspection of device and accessories
- Functional check of device
2.1 Visual inspection of Stimuplex® HNS 12 including accessories
The following components must be clean and undamaged and not show any contamination:
•
•
•
•
•
•
•
•
•
Housing
Labels
Operating controls (instant set keys)
Keypad
LCD display
Controller dial (digital control knob)
Connecting ports
Electrode cable (e.g. insulation or electrodes not broken)
Test resistor
If included:
• Stimuplex® Pen
• Stimuplex® Remote Control
• Stimuplex® Switch
Never use damaged equipment or accessories!
After visual inspection is successful, carry out the function test.
2.2 Functional check of Stimuplex® HNS 12
Switch on the nerve stimulator. Target stimulation current = 0.00 mA, LED is not flashing, and no
sound is audible. Battery symbol may not be flashing (battery voltage > 7 V). If SENSe is enabled in
the setup/options menu please use Hz quick access button to set frequenzy to 1 Hz or 2 Hz.
• Connect electrode cable to the front of Stimuplex® HNS 12 (to the middle 4-polar plug).
The plug connector configuration prevents wrong polarity connection. – Leave the circuit open.
7
Page 8
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 8
8
• Use the mA function key to set the current range to 5.00 mA – Appears in display.
• Use the ms function key to set the stimulus duration to 0.30 ms – Appears in display.
• Use the Hz function key to set stimulation frequency to 1 Hz – Appears in display.
• Using the control knob, slowly increase the target stimulation current from 0.00 mA to 5.00 mA.
Throughout the entire procedure, warning display must read “Patient current lower than your setting”
and patient current = “0.00 mA”. The LED flashes red in a 1-Hz rhythm and the warning signal is
audible in a 1-Hz rhythm.
• Close the circuit using the 10 kΩ test resistor supplied as an accessory. All warnings (display, LED,
sound) must disappear. The display reads 10 kΩ ± - tolerance. If the warnings do not disappear,
briefly connect the electrode cable to the contacts at the ends of the cable in order to rule out
that test resistor is defective.
• Using the digital control knob, slowly reduce the target stimulation current from 5.00 mA to 0.00
mA. Throughout the entire procedure, the warning display must not read “Patient current lower
than your setting”, and the LED must be flashing green in a 1-Hz rhythm (if applicable yellow for
target stimulation current < threshold current if option is activated).
In the event that the nerve stimulator does not behave as described, its use must be discontinued.
The nerve stimulator and the electrode cable must be technically inspected.
If you have changed any settings please reestablish them and then switch off the nerve stimulator
so that any configured settings (from the medical team) are available when the device is switched
back on.
Page 9
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 9
3. Instructions for technical safety checks
9
These tests may be performed by qualified staff only and in accordance with the following test
instructions (Section 6, para. 4 MPBetreibV). An equipment logbook of the technical safety checks
must be kept (see appendix) and filed for the next safety inspection (Section 6, para. 3 MPBetreibV).
The following measuring and testing equipment is required:
Oscilloscope (storage oscillograph) for checking the output signal
Settings:
- Time base = 20 µs/div (500 ms/div) for stimulation frequency measurement
- Resolution = 500 mV, 1 V, 2 V, 10 V/div (full screen)
- Mode = High resolution is recommended for maximum accuracy (HiRes) or
sample
- Trigger = falling slope, e.g. 30 %
Ohmmeter for checking test resistance
Test resistor 10 kΩ +/- 1%
9 V battery alkaline, battery voltage > 7 V (a fresh battery is recommended).
3.1 Visual inspection of Stimuplex® HNS 12 and accessories
See Section 2.1. – Enter results in equipment logbook.
3.2 Functional check of Stimuplex® HNS 12
See Section 2.2. – Enter results in equipment logbook.
3.3 Checking the operating controls and monitoring equipment
• Check instant set keys (keypad)
mA key: Press the mA key repeatedly. In the display, the value is highlighted by a border for approx.
3 seconds. Only a value highlighted by a border can be changed by repeatedly pressing the key.
Adjustable current range settings are 1.00 mA or 5.00 mA.
ms key (SENSe disabled only for 1 Hz and 2 Hz): Press the ms key repeatedly. In the display, the
value is highlighted by a border for approx. 3 seconds (see above). Adjustable settings are 0.10 ms,
0.30 ms, 1.00 ms stimulus duration (0.05 ms and 0.50 ms, if applicable).
Hz key: Press Hz key repeatedly. In the display, the value is highlighted by a border for approx.
3 seconds (see above).
Adjustable stimulation frequency settings are 1 Hz, 2 Hz or SENSe (if enabled).
Page 10
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 10
10
MENU and ON/OFF keys:
When the stimulation screen is displayed, press I arrow key ➝ Main Menu is displayed.
Use to select “Setup” menu.
Press key. “Setup” menu is displayed.
Press key. Main menu reappears in the display.
Press ON/OFF key briefly. The stimulation screen reappears.
• Check digital control knob
In the stimulation screen turn control knob clockwise; this increases the target stimulation current;
turn knob counter clockwise to reduce target stimulation current.
• LCD display: Readability check. If necessary, increase contrast by selecting Contrast menu.
Using the digital control knob, set target stimulation current = 5.00 mA.
• Check LED display with open circuit – LED lights up red.
• Check LED display with closed circuit – LED lights up green.
• Check Sound with open and closed circuit. If necessary, increase volume by selecting Volume
menu.
Enter results in equipment logbook.
3.4 Checking the accessories
• Measure 10 kΩ ± 1 % test resistance with the ohmmeter. Enter results in equipment logbook.
• Check electrode cable as described in Section 2.2. Enter results in equipment logbook.
If available:
• Connect Stimuplex® Remote Control to front of Stimuplex® HNS 12 (3-polar plug on left). The
plug connector configuration prevents any wrong polarity connection.
• Use the remote control buttons to increase target stimulation current and reset to 0.00. Enter
results in equipment logbook.
• Connect the Stimuplex® Switch to front of Stimuplex® HNS 12 (4-polar plug) and check the
device as described in the Stimuplex® Switch Instructions for use on page 4.
Page 11
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 11
11
• Connect Stimuplex® Pen to electrode cable. Increase target stimulation current. The warning
“Patient current lower than your setting” appears in the display. Then close the circuit with the
test resistor. The warnings (display, LED, sound) must disappear. Enter results in equipment logbook.
3.5 Checking stimulation signal
To prevent disturbances, perform all tests only when the 9-volt battery is installed.
Connect the electrode cable using the 10 kΩ +/- 1% test resistor – connect the oscillograph with
GND clip on the red clamp parallel to test resistor (see page 15 for tolerances).
Set stimulus duration = 0.10 ms – target stimulation current = 1.00 mA.
• Pulse form rectangular, negative.
• Measure stimulus duration – Enter results in equipment logbook.
Set stimulus duration = 0.10 ms – target stimulation current = 0.30 mA.
• Measure actual stimulation current – Enter results in equipment logbook.
Set stimulation frequency = 2 Hz – target stimulation current = 1.00 mA.
• Measure stimulation frequency – Enter results in equipment logbook.
Set stimulus duration = 0.10 ms – target stimulation current = 0.50 mA.
• Measure actual stimulation current – Enter results in equipment logbook.
Set stimulus duration = 0.10 ms – target stimulation current = 1.00 mA.
• Measure actual stimulation current – Enter results in equipment logbook.
Stimulus duration = 0.10 ms – current range = 5.00 mA. Using the control knob, set target stimulation current = 5.00 mA.
• Measure actual stimulation current – Enter results in equipment logbook.
• Check impedance display with 10 kΩ test resistor. Enter results in equipment logbook.
If faults are detected, return the device – including accessories – to the manufacturer
or distributor for repair.
Page 12
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 12
12
4. Replacing the electrode cable
A faulty electrode cable can be easily replaced on-site with a new one. The polarized connector prevents any wrong connection of electrode cable.
– B. BRAUN Code number for replacement electrode cable: 4892070
5. Repairs
Electrical and electronic equipment may only be repaired by the manufacturer or by an organization
expressly authorized by the manufacturer.
Unauthorized persons opening or attempting to repair Stimuplex® HNS 12 can create a dangerous
situation and will cause all warranty claims to be null and void.
Any order for repairs must be accompanied by a detailed description of the fault (see next page).
Please send repairs
Please send repairs from
from all EU countries to: all other countries to:
Stockert GmbH
B. Braun Melsungen AG
Bötzinger Straße 72
LO-WF-DE08P
79111 Freiburg
Attn.: Mr. Grützmacher
Germany
34209 Melsungen
Germany
For further request,
please contact:
B. Braun Melsungen AG
HC-SY-DE08 P Attn.: Mr. Krapf
Pfieffewiesen
34212 Melsungen
Germany
e-mail: Heiko.Krapf@bbraun.com
Telephon (0 56 61) 71-16 75
Telefax (0 56 61) 75-16 75
Page 13
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 13
13
ERROR REPORT
Product: STIMUPLEX® HNS 12
Connection Cable
Serial-No. ____________
Others: _______________________
Customer
(Clinic)
(Stamp)
Contact person: ___________________________________________
Phone no.: _________________________
Fax-no.: _________________________
Date when error appeared: _______________________
Exact description of the error:
In which situation did the error appear:
Device settings:
Accessories in use:
________________________________________
________________________________________
________________________________________
________________________________________
________________________________________
Patient hazards? No
Cat.-No.
_____________
_____________
_____________
_____________
_____________
LOT-No.
______________
______________
______________
______________
______________
Yes Description: _______________________________________
Error report recorded by: __________________________________ on: ____________________
Please return the error report together with the defect device!
Page 14
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 14
14
6. Test Equipment Logbook Stimuplex® HNS
Serial No.
________________
Test equipment for the test:
Measuring device
Ohmmeter
Oszilloscope
Type
Calibrated
Next calibration
Visual inspection
Operating elements
Housing
Labels
Keypad (instant set keys)
LC Display
Control knob
Plugs
Electrode cable
Test resistor
Stimuplex® Remote Control if available
Passed
Failed
Stimuplex® Pen if available
Passed
Failed
Passed
Failed
Functional check
Operational and monitoring elements check
Operational and monitoring elements
Keys
Control knob
LC Display
LED display green (yellow, if applicable), red
Sound, warning signal
Page 15
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
Seite 15
15
Accessories functional check
Accessories
Test resistor 10 kΩ +/- 1 %
Electrode cable
Stimuplex® Remote Control, if available
Stimuplex® Pen, if available
Passed
Failed
Stimulation signal check
Test
Pulseform
Stimulation duration
Stimulation frequency
Stimulation current
Stimulation current
Stimulation current
Stimulation current
Impedance
Stimuplex® HNS 12 Setting
0.10 ms, 1.00 mA
0.10 ms, 1.00 mA
2 Hz, 1.00 mA
0.10 ms, 0.30 mA, 10 kΩ +/- 1%
0.10 ms, 0.50 mA, 10 kΩ +/- 1%
0.10 ms, 1.00 mA, 10 kΩ +/- 1%
0.10 ms, 5.00 mA, 10 kΩ +/- 1%
0.10 ms, 5.00 mA, 10 kΩ +/- 1%
Tolerance
Square, negative
0.098 ms – 0.102 ms
1.98 Hz – 2.02 Hz
0.26 mA – 0.34 mA
0.46 mA – 0.54 mA
0.96 mA – 1.04 mA
4.96 mA – 5.04 mA
9 kΩ – 11 kΩ
Remarks:
Technical inspector:
Date:
Signature:
Passed
Failed
Page 16
6043572_Stimuplex_HNS12_TSM_16s_GB
13.05.2009
11:23 Uhr
B. Braun Melsungen AG
Carl-Braun-Straße 1
34212 Melsungen
Germany
Tel +49 (0) 56 61-71-0
Date of last revision April 2009
W. 01,2.01.08/6
6043572
0509
www.bbraun.com
Seite 16