Baylis Medical Company

Reusable RFP-100A Connector Cable Instructions for Use

23 Pages

Instructions for Use Reusable RFP-100A Connector Cable English... Français... Español... Português... Nederlands... Italiano... Deutsch... Čeština... Dansk... Suomo... Norsk... Svenska...  1 3 5 6 9 11 12 14 16 18 20 21  Baylis Medical Company Inc. 5959 Trans-Canada Highway Montreal, Quebec, Canada, H4T 1A1 Tel: (514) 488-9801/ (800) 850-9801 Fax: (514) 4887209 www.baylismedical.com EU Authorized Representative: Quality First International OÜ Laki 30, 12915 Tallinn Estonia Telephone: +372 610 41 96 Email: enquiries@qualityfirstint.ee  © Copyright Baylis Medical Company Inc., 2009-2022 Baylis Medical and the Baylis Medical logo are trademarks of Baylis Medical Technologies Inc. All other trademarks or registered trademarks are property of their respective owners.  English Carefully read all instructions prior to use. Observe all contraindications, warnings and precautions noted in these instructions. Failure to do so may result in patient complications. Caution: Federal (U.S.A.) law restricts this device to sale by or on the order of a physician. I. DEVICE DESCRIPTION The Reusable RFP-100A connector cable connects the Baylis Medical Company RFP-100A Radiofrequency Puncture Generator (RFP-100A Generator) to Baylis Medical approved radiofrequency puncture devices. This Cable enables radiofrequency (RF) power to be delivered from the Generator to the puncture device. Detailed information concerning the RFP-100A Generator is contained in a separate manual that accompanies the Generator (RFP-100A Generator Instructions for Use). In addition, detailed information concerning the RF puncture devices is contained in separate manuals that accompany these devices. The dimensions for the Reusable RFP-100A connector cable can be found on the device label and in section VII “Product Specifications.” The Reusable RFP-100A connector cable has a four-pin connector on one end that mates with the RFP100A Generator and a connector at the other end, which mates with the puncture device. II. INDICATIONS FOR USE The intended use of the Reusable RFP-100A connector cable is to connect the RFP-100A Generator to Baylis Medical approved puncture devices (RF puncture devices).  Page 1 of 23  III. CONTRAINDICATIONS The Reusable RFP-100A connector cable is not recommended for use with any other RF generator or any other device. IV. WARNINGS • The Reusable RFP-100A connector cable is a reusable device. Only use the validated cleaning and sterilization methods as described in section XI “Cleaning and Sterilization Instructions” to clean and sterilize the Connector Cable. No other cleaning and sterilization methods have been tested. Failure to properly clean and sterilize the device can cause patient injury and/or the communication of infectious disease(s) from one patient to another. • The Reusable RFP-100A connector cable must only be used with the RFP100A Generator and RF puncture devices. Attempts to use it with other RF Generators and devices can result in electrocution of the patient and/or operator. • Laboratory staff and patients can undergo significant x-ray exposure during radiofrequency puncture procedures due to the continuous usage of fluoroscopic imaging. This exposure can result in acute radiation injury as well as increased risk for somatic and genetic effects. Therefore, adequate measures must be taken to minimize this exposure. V. PRECAUTIONS • Do not attempt to use the Reusable RFP-100A connector cable or ancillary equipment before thoroughly reading the accompanying Instructions for Use. • Puncture procedures should be performed only by physicians thoroughly trained in the techniques of radiofrequency powered puncture in a fully equipped catheterization laboratory. • The sterile packaging should be visually inspected prior to use to detect any compromise. Ensure that the packaging has not been damaged. Do not use the equipment if the packaging has been compromised. • Visually inspect the cable to ensure there is no cracking or damage to the insulating material. Do not use the cable if there is any damage. • The Reusable RFP-100A connector cable is intended for use with RF puncture devices only. • Never disconnect the Reusable RFP-100A connector cable from the RFP100A Generator while the Generator is delivering RF power. • Never disconnect the Reusable RFP-100A connector cable from the RFP100A Generator by pulling on the cable. Failure to disconnect the cable properly may result in damage to the cable. • Do not twist the Reusable RFP-100A connector cable while inserting or removing it from the Isolated Patient Connector on the Generator. Twisting the cable may result in damage to the pin connectors. • Do not bend the cable. Excessive bending or kinking of the cable may damage the integrity of the cable and may cause patient injury. Care must be taken when handling the cable. • Take precautions to limit the effects that the electromagnetic interference (EMI) produced by the Generator may have on the performance of other equipment. Check the compatibility and safety of combinations of other physiological monitoring and electrical apparatus to be used on the patient in addition to the Generator. • Adequate filtering must be used to allow continuous monitoring of the surface electrocardiogram (ECG) during radiofrequency power applications. • During power delivery, the patient should not be allowed to come in contact with ground metal surfaces. • In order to prevent the risk of ignition make sure that flammable material is not present in the room during RF power application. Baylis Medical Company relies on the physician to determine, assess and communicate to each individual patient all foreseeable risks of the Baylis Medical Radiofrequency Puncture System. VI. ADVERSE EVENTS Adverse events associated with the use of this device are similar to those indicated for the Baylis Medical Radiofrequency Puncture System. VII. PRODUCT SPECIFICATIONS Model Number Strain Relief Colour Overall Useable Length Generator Connector Device Connector  RFX-BAY-TS Black at device end, blue at generator end 10 feet (3m) 4-pin (Plug) 4-pin (receptacle)  VIII. INSPECTION PRIOR TO USE Perform the following checks before the patient is presented for the procedure. These tests will allow you to verify that the equipment you will use is in proper working order. Do these tests in a sterile environment. Do not use defective equipment. KEY ITEMS  QUESTION?  Sterility  Is the connector cable sterile?  Visual Check  Have you done a visual check on the entire system?  WARNINGS AND EXPLANATIONS The Reusable RFP-100A connector cable is supplied sterile for its initial use. Inspect the packaging to ensure the package has not been damaged and sterility has not been compromised. Prior to each subsequent use it must be cleaned and sterilized.  Ensure connectors and the cable have no visible damage, such as discoloration, cracks, label fading, cable splice, or kinks. Do not use damaged equipment.  IX. EQUIPMENT REQUIRED Puncture procedures should be performed in a specialized clinical setting which may be equipped with a fluoroscopy unit, radiographic table, physiologic recorder, emergency equipment and instrumentation for gaining vascular access. X. DIRECTIONS FOR USE Once the RF puncture device is properly positioned at the puncture site, and the Generator is properly set up (following the instructions in the RFP-100A Generator  DMR RFX 3.3 V-12 15-Sep-2022
File Type: PDF
File Size: 836 KB
File Name: Baylis Medical Company - Boston Scientific - DMR RFX 3.3 V-12 - Reusable RFP-100A Connector Cable Instructions for Use - 2022-09.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions