AESCULAP® Acculan
en USA
Instructions for use/Technical description Battery GA666, GA676 Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapusaifus.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-800-282-9000. A paper copy will be provided to you upon request at no additional cost.
et
Kasutusjuhend/Tehniline kirjeldus Aku GA666, GA676
lv
Lietošanas instrukcijas/tehniskais apraksts Akumulators GA666, GA676
lt
Naudojimo instrukcija/techninis aprašas Akumuliatorius GA666, GA676
ru
Инструкция по примению/Техническое описание Аккумулятор GA666, GA676
cs
de
Gebrauchsanweisung/Technische Beschreibung Akku GA666, GA676
Návod k použití/Technický popis Akumulátor GA666, GA676
pl
fr
Mode d’emploi/Description technique Batterie GA666, GA676
Instrukcja użytkowania/Opis techniczny Akumulator GA666, GA676
sk
es
Instrucciones de manejo/Descripción técnica Batería GA666, GA676
Návod na použitie/Technický opis Akumulátor GA666, GA676
hu
it
Istruzioni per l’uso/Descrizione tecnica Accumulatore GA666, GA676
Használati útmutató/Műszaki leírás Akkumulátor GA666, GA676
sl
pt
Instruções de utilização/Descrição técnica Bateria GA666, GA676
Navodila za uporabo/Tehnični opis Baterija GA666, GA676
hr
nl
Gebruiksaanwijzing/Technische beschrijving Accu GA666, GA676
Upute za uporabu/Tehnički opis Baterija GA666, GA676
ro
da
Brugsanvisning/Teknisk beskrivelse Batteri GA666, GA676
Manual de utilizare/Descriere tehnică Acumulator GA666, GA676
bg
nb
Bruksanvisning/Teknisk beskrivelse Oppladbart batteri GA666, GA676
Упътване за употреба/Техническо описание Акумулаторна батерия GA666, GA676
tr
sv
Bruksanvisning/Teknisk beskrivning Batteri GA666, GA676
Kullanım Kılavuzu/Teknik açiklama Akü GA666, GA676
el
fi
Käyttöohje/Tekninen kuvaus Akku GA666, GA676
Οδηγίες χρήσης/Τεχνική περιγραφή Μπαταρία GA666, GA676
en
AESCULAP® Acculan Battery GA666, GA676 Symbols on product and packages Caution Observe important safety information such as warnings and precautions in the instructions for use. Two-dimensional machine readable code The code contains a unique serial number which can be used for electronic tracking of the individual instrument. The serial number is based on the global standard sGTIN (GS1). Manufacturer
Date of manufacture
Follow the instructions for use
Labeling of electrical and electronic devices pursuant to directive 2002/96/EG (WEEE)
Manufacturer's batch name Manufacturer serial number Manufacturer order number Temperature limits during transport and storage
Air humidity limits during transport and storage
Atmospheric pressure limits during transport and storage According to US federal law, this product may only be sold by a physician or by the order of a physician CE mark according to Regulation (EU) 2017/745 Medical device
Keep dry
0
Contents 1. 1.1 1.2 2. 2.1 2.2 2.3 2.4 3. 4. 4.1 4.2 4.3 5. 6. 6.1 6.2 6.3 7. 7.1 7.2 7.3 7.4 7.5 7.6 7.7 8. 9. 10. 11. 12. 13. 13.1 13.2 13.3 14.
About this document Scope Warnings General information Use/intended use: Main functions and design characteristics Indications Contraindications Safe handling Product description Scope of Supply Components required to operate the device Operating principle Preparation Working with the device System set-up Functional test Safe operation Processing process General safety information General notes Reusable products Preparing for cleaning Cleaning/Disinfection Wipe disinfection of electrical equipment Inspection, maintenance and checks Maintenance Troubleshooting Storage and transport Accessories/Replacement Parts Technical Service Technical Data Classification in accordance with Regulation (EU) 2017/745 Performance data, information and standards Ambient conditions Disposal
1 1 1 1 1 1 1 1 1 1 1 2 2 2 2 2 2 2 2 2 3 3 3 3 3 3 4 4 4 4 4 5 5 5 5 5
en 1.
About this document
3.
1.1
Scope
WARNING Risk of injury and damage to property due to improper handling of the product! ► Comply with instructions for use for the Acculan 4 machines and the Acculan 3Ti machines. ► Follow the instructions for use of all products used.
These instructions for use apply to the products listed below: Art. no.
Designation
GA666
Acculan battery
GA676
Acculan battery
Safe handling
Note For the instructions for use and further information on B. Braun / AESCULAP products please refer to the B. Braun eIFU website at eifu.bbraun.com.
WARNING Fire and explosion hazard due to battery poles shorted by fluids or metal components! ► Do not short-circuit, open, crush or incinerate battery.
1.2
CAUTION Damage to or destruction of the batteries due to reprocessing! ► Do not sterilize batteries.
Warnings
Warnings indicate risks to patients, users, and/or the product that could develop while using the product. Warnings are marked as follows: DANGER Indicates a potential danger. If it is not prevented, death or severe injuries may result. WARNING Indicates a potential danger. Minor or moderate injuries may result if this is not prevented. CAUTION Indicates a potential damage to property. The product may become damaged if it is not prevented.
2.
General information
2.1
Use/intended use:
Task/Function The rechargeable batteries supply power to the Acculan units and contain the circuitry required to activate and control the electric motors of the Acculan units in accordance with operator commands. Application Environment The product fulfills the requirements of Type BF in accordance with IEC/DIN EN 60601-1. For use in operating rooms, outside of the explosive hazard zone (e.g. areas with highly pure oxygen or anesthetic gases).
2.2
Main functions and design characteristics
Art. no.
DC voltage
Capacity
GA666
9.6 V
1.05 Ah
GA676
9.6 V
1.95 Ah
2.3
prior to first use to ensure it is in good working order; fully charge using charger GA320 or GA677 and verify functionality. ► To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: – Use the product according to these instructions for use. – Follow the safety and maintenance instructions. – Only combine Aesculap products with each other. – Follow the application advisories acc. to standard, see Extracts from relevant standards. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Keep the instructions for use accessible for the user. ► Always adhere to applicable standards. ► Do not use the product if it is damaged or defective. ► Set aside the product if it is damaged. ► Charge product fully after 3 months in storage. ► Keep product dry. ► Observe the instructions for use for the corresponding Acculan machines. Note The user must report all serious incidents occurring in connection with the product to the manufacturer and to the competent authority of the country in which the user is registered.
4.
Product description
4.1
Scope of Supply
Art. no.
Designation
GA666
Acculan NiMH battery short
Indications
The indications can be found in the instructions for use for the corresponding Acculan machines.
2.4
► Inspect the new product after removing its transport packaging and
Contraindications
– or – GA676
Acculan NiMH battery
TA011867
Instructions for use, Acculan battery GA666, GA676
The contraindications can be found in the instructions for use for the corresponding Acculan machines.
1
en 4.2
Components required to operate the device
■ Acculan 4 charger GA320 or Acculan 3Ti charger GA677 ■ Acculan 4 or Acculan 3Ti machine: Art. no.
Designation
GA330
Acculan 4 drill and reamer
GA331
Acculan 4 oscillating saw
GA334
Acculan 4 reciprocating saw
GA336
Acculan 4 oscillating saw straight
GA337
Acculan 4 oscillating sternum saw
GA340
Acculan 4 Dermatome
GA341
Acculan 4 Dermatome 0.1 mm
GA344
Acculan 4 small drill
GA668
Acculan 3Ti oscillating saw straight
GA670
Acculan 3Ti Dermatome
GA671
Acculan 3Ti small drill
GA672
Acculan 3Ti drill and reamer
GA673
Acculan 3Ti oscillating saw
GA674
Acculan 3Ti reciprocating saw
GT666
Acculan 4 TPLO saw
GT668
Acculan 3Ti TPLO saw
4.3
Preparation
► Prior to use, inspect the product and its accessories for any visible
damage. ► Use the products and their accessories only if they are in perfect tech-
nical condition.
6.
Working with the device
6.1
System set-up
► See instructions for use for the respective Acculan machine and the
Acculan 4 charger GA320 or Acculan 3Ti charger GA677.
6.2
Functional test
See the instructions for use for the respective Acculan machine.
2
Safe operation
WARNING Risk of injury and material damage due to inappropriate use of batteries! ► Charge battery only in the Acculan 4 charger GA320 or Acculan 3Ti charger GA677. ► Insert battery correctly (Acculan machine and charger). WARNING Risk of injury and material damage due to damaged batteries! ► Check batteries for any damage before charging. ► Do not use or recharge damaged batteries. CAUTION Total battery failure can occur if the batteries are left in the charger while it is disconnected from the power supply! ► Always keep the charger connected to the power supply. Indicator light for readiness for service must be lit. ► Always remove batteries from the charger when it is disconnected from the power supply. CAUTION Loss of capacity/performance due to prolonged storage of discharged batteries! ► Leave the batteries in the functional charger after they have been successfully charged (trickle charging). ► For proper loading and removal of the battery from the Acculan
machine, see the instructions for use for the respective Acculan machine.
Operating principle
The Acculan NiMH battery supplies the Acculan machine with voltage. The battery is charged in the Acculan 4 charger GA320 or Acculan 3Ti charger GA677. The charging process of a completely discharged battery can take up to 120 min. The speed and rotational direction are set depending on the trigger setting of the machines: ■ for saws and Dermatome: speed clockwise ■ for drills: speed clockwise, counter-clockwise, oscillation and pilgrim step
5.
6.3
7.
Processing process
7.1
General safety information
Note Observe national statutory regulations, national and international standards and directives, and any proprietary hygiene instructions for processing. Note The applicable national regulations concerning the reprocessing of devices must be observed for patients having Creutzfeldt-Jakob disease (CJD), suspected of having CJD, or possible variants of CJD. Note It should be noted that successful processing of this medical device can only be guaranteed following prior validation of the processing method. The operator/processor shall be responsible. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note See also B. Braun eIFU at eifu.bbraun.com for current information on processing and compatibility of material
en 7.2
General notes
7.5
Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surface and could cause corrosion.
7.3
Device-specific safety information on the reprocessing method DANGER Risk of electric shock and fire hazard! ► Do not use flammable or explosive cleaning or disinfecting solutions. ► Ensure that no fluids will penetrate the product. CAUTION Damage to, or destruction of the product caused by mechanical cleaning/disinfection! ► Only clean and disinfect the product manually. ► Do not sterilize the product under any circumstances.
Reusable products
CAUTION Risk of damage to the product due to inappropriate cleaning/disinfecting agents! ► Only use cleaning/disinfecting agents approved for surface cleaning. Follow the manufacturer’s instructions for the respective cleaning/disinfecting agent. ► Do not introduce any fluid onto the electrical contacts of the battery.
The service life of the product is limited by damage, normal wear, type and duration of the application, handling, storage and transportation of the product. Careful visual and functional testing prior to next use is the best way to identify a malfunctioning product.
7.4
Cleaning/Disinfection
Preparing for cleaning
► Remove the battery from the Acculan machine.
7.6
Wipe disinfection of electrical equipment
Phase
Step
T [°C/°F]
t [min] Conc. [%]
Water quality
Chemistry
I
Cleaning
RT
0.5
-
-
17 % 1-Propanol, 0.23 % Didecyldimethylammonium chloride*
II
Wipe disinfection
RT
≥3
-
-
17 % 1-Propanol, 0.23 % Didecyldimethylammonium chloride*
RT: Room temperature *Recommended: Meliseptol® Wipes sensitive (B. Braun) Phase I ► Remove any visible residues with a disposable disinfectant wipe. Phase II ► Wipe all surfaces of the optically clean product with a fresh, dispos-
able disinfectant wipe. ► Observe contact time (at least 3min).
7.7
Inspection, maintenance and checks
► Inspect the product after each cleaning and disinfecting cycle to be
sure it is clean, functional and undamaged. ► Set aside the product if it is damaged.
3
en 8.
Maintenance
The display "Battery replacement recommended" on the Acculan 3Ti charger GA320 (see TA014535) indicates when the battery should be replaced due to reduced capacity. No other maintenance measures must be performed on the battery. For corresponding services, contact your national B. Braun/Aesculap agency, see Technical Service.
9.
Troubleshooting
► Replace product: have defective products repaired by Aesculap Technical Service, see Technical Service.
Fault
Detection
Cause
Remedy
Battery not working
No signal tone when inserting the battery into the machine
Battery defective
Replace product.
Battery empties quickly
Symbol "Battery replacement recom- Battery cells at end of life mended" illuminated on the charger at the end of the charging process
Have battery cells replaced by Aesculap Technical Service.
10. Storage and transport
12. Technical Service
WARNING Risk of injury from burns and damage to product in case of short-circuit! ► When storing the product, ensure that the product cannot shortcircuit with other products or metal objects.
DANGER Danger to life of patients and users in case of malfunctions and/or failure of protective measures! ► Do not perform any servicing or maintenance work under any circumstances while the product is being used on a patient.
► In order to prevent damage during transport, ship product in the orig-
inal packaging and in an outer carton. ► If the original packaging is no longer available, ensure the following when packaging the product: – The battery contacts are protected against short-circuit. – The packaging completely encloses the product. – The product is secured against movement within the packaging. ► Completely charge the product upon receipt and before extended periods of storage.
11. Accessories/Replacement Parts Art. no.
Designation
TA015514
Instructions for use, battery GA666, GA676 (A4)
TA011867
Instructions for use, battery GA666, GA676 (flyer)
4
CAUTION Modifications of medical devices may result in a loss of potential guarantee/warranty claims and forfeiture of applicable licenses. ► Do not modify the product. ► Contact national B. Braun/Aesculap representative for service and repair. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95 -1601 Fax: +49 (7461) 14 -939 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
en 13. Technical Data
14. Disposal
13.1 Classification in accordance with Regulation (EU) 2017/745
WARNING Risk of infection from contaminated products! ► Observe national regulations when disposing of or recycling the product, its components, and their packagings.
Art. no.
Designation
Class
GA666
Acculan battery
I
GA676
Acculan battery
I
13.2 Performance data, information and standards GA666
GA676
Cell type
NiMH
NiMH
DC voltage
9.6 V
9.6 V
Capacity
1.05 Ah
1.95 Ah
Weight
approx. 0.304 kg
approx. 0.425 kg
Dimensions (L x W x H)
approx. 120 mm x 43 mm x 50 mm
approx. 140 mm x 43 mm x 50 mm
EMC
IEC/DIN 60601-1-2 IEC/DIN 60601-1-2
Compliance with standards
IEC/DIN 60601-1
Note The operator must process the product before disposal, see Processing process. The recycling pass can be downloaded from the Extranet as a PDF document under the respective article number. (The recycling pass includes disassembling instructions for the product, as well as information for proper disposal of components harmful to the environment.) Products carrying this symbol are subject to separate collection of electrical and electronic devices. Within the European Union, disposal is taken care of by the manufacturer as a free-of-charge service. ► For questions regarding the disposal of the product, please contact
your local B. Braun/Aesculap representative, see Technical Service.
IEC/DIN 60601-1
13.3 Ambient conditions Operation
Transport and storage
Temperature
10 °C to 40 °C
-10 °C to 40 °C Maximum recommended period of storage is 3 months; keep out of direct sunlight
Relative humidity
30 % to 75 %
10 % to 90 %
Atmospheric pressure
700 hPa to 1 060 hPa
500 hPa to 1 060 hPa
Note The atmospheric pressure of 700 hPa correlates with a maximum operating altitude of 3 000 m.
5