Instructions for use/Technical description Adapter RT081R and RT046P, Insert RT086P and RT055P Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapusa.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-800-282-9000. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung Adapter RT081R und RT046P, Einsatz RT086P und RT055P Mode d’emploi/Description technique Adaptateur RT081R et RT046P, insert RT086P et RT055P Instrucciones de manejo/Descripción técnica Adaptador RT081R y RT046P, pieza de interposición RT086P y RT055P Istruzioni per l’uso/Descrizione tecnica Adattatore RT081R e RT046P, inserto RT086P e RT055P Instruções de utilização/Descrição técnica Adaptador RT081R e RT046P, inserto RT086P e RT055P Gebruiksaanwijzing/Technische beschrijving Adapter RT081R en RT046P, Inzetstuk RT086P en RT055P Bruksanvisning/Teknisk beskrivning Adapter RT081R och RT046P, insats RT086P och RT055P Инструкция по примению/Техническое описание Переходник RT081R и RT046P, вставка RT086P и RT055P Návod k použití/Technický popis Adaptér RT081R a RT046P, vložka RT086P a RT055P Instrukcja użytkowania/Opis techniczny Adapter RT081R i RT046P, wkładka RT086P i RT055P Návod na použitie/Technický opis Adaptér RT081R a RT046P, vložka RT086P a RT055P Kullanım Kılavuzu/Teknik açiklama Adaptör RT081R ve RT046P, Geçme vidaRT086P ve RT055P
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap® – a B. Braun brand TA009907
2020-04
V6
Change No. 62390
en
2.4 ®
Aesculap Adapter RT081R and RT046P, Insert RT086P and RT055P Legend 1 Fixation screw 2 Housing 3 Plastic insert 4 Locking (turn clockwise) 5 Lock nut 6 Unlocking (turn counterclockwise) 7 Connector 8 Locking sleeve 9 Support arm 10 Arrow symbol 11 Marking (secure/locked) 12 “Locked” symbol
1.
About this document
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
Scope
These instructions for use apply for the following products: Art. no.
Designation
RT081R, RT046P
Adapter with Insert
RT086P, RT055P
Insert
► For article-specific instructions for use as well as information on material compatibility and lifetime see B. Braun
eIFU at eifu.bbraun.com
1.2
Safety messages
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows: WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result. CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
2.
Clinical use
2.1
Areas of use and limitations of use
2.1.1
Intended use
The adapters are used to securely attach optics or trocars to a support arm, using the appropriate plastic insert. Suitable support arms: ■ Unitrac support arm RT040R ■ M-TRAC support arm FF168R
2.1.2
Indications
Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications. For indications, see Intended use.
2.1.3
Contraindications
Assembly
CAUTION Damage to the trocar and/or optic shaft can occur if they fall out of the adapter! ► Use the adapter with compatible plastic inserts only. ► Turn the locking screw 1 counterclockwise. ► Insert the compatible plastic insert 3 in housing 2. ► Turn locking screw 1 clockwise until plastic insert 3 is fixated. ► Insert the trocar and lens shaft into the plastic insert 3. ► Turn the locking screw 1 counterclockwise until the trocar or optical shaft are firmly seated. ► Make certain that mark 11 and arrow symbol 10 are aligned with each other. ► Slide the connector 7 through locking sleeve 8 and into support arm 9. Connector 7 should click into place. ► Turn locking screw 8 clockwise to the positive stop.
Mark 11 and “Locked” symbol 12 are aligned with each other. ► Tighten locking nut 5 in the direction of arrow 4 (clockwise).
3.
Validated reprocessing procedure
3.1
General safety instructions
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
3.2
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to "AKI-Brochures", "Red brochure".
3.3
Reusable products
2.2
Safety information
Influences of the reprocessing which lead to damage to the product are not known. A careful visual and functional inspection before the next use is the best opportunity to recognize a product that is no longer functional, see Inspection.
2.2.1
Clinical user
3.4
No known contraindications.
Preparations at the place of use
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user.
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located.
► Release lock nut 5 of the adapter by turning it as indicated by arrow 6 (counterclockwise). ► Rotate locking sleeve 8 counterclockwise to the positive stop.
Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
Sterility
The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
2.3
Application
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product.
3.5
Preparing for cleaning
► Disassemble the product immediately after use, as described in the respective instructions for use. ► Disassemble the product prior to cleaning, see Disassembly.
3.6
Disassembly
3.6.1
Removing the adapter from the support arm
Mark 11 and arrow symbol 10 are aligned with each other. ► To avoid damage to the adapter: Hold the adapter while detaching it from the support arm 9. ► Push locking sleeve 8 towards the adapter.
3.6.2
Removing the trocar or the optics shaft and the plastic insert
► Turn the locking screw 1 counterclockwise. ► Take the trocar and lens shaft out of the plastic insert 3. ► Remove the plastic insert 3 from housing 2.
3.7
Cleaning/Disinfection
3.8.2
Manual cleaning with ultrasound and immersion disinfection
3.7.1
Product-specific safety information on the reprocessing method
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Following the manufacturer's instructions, use cleaning and disinfecting agents that are approved for e.g. aluminum, plastic materials and stainless steel. ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C.
3.7.2
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
Reference
Manual cleaning with immersion disinfection
■ Place the product on a tray that
Chapter Manual cleaning/disinfection and subsection:
is suitable for cleaning (avoid rinsing blind spots).
■ RT086P ■ RT055P
■ Chapter Manual cleaning with immersion disinfection
Manual cleaning with ultrasound and immersion disinfection
■ Cleaning brush: for example
■ RT081R ■ RT046P
■ Disposable syringe 20 ml ■ Place the product on a tray that
TA011944
Chapter Manual cleaning/disinfection and subsection:
■ Chapter Manual cleaning with ultrasound and immersion disinfection
is suitable for cleaning (avoid rinsing blind spots). Manual pre-cleaning with brush and subsequent mechanical alkaline cleaning and thermal disinfection
■ RT086P ■ RT055P
Manual pre-cleaning with ultrasound and brush, and subsequent mechanical alkaline cleaning and thermal disinfection
■ RT081R ■ RT046P
3.8
■ Cleaning brush: for example, TA011944
■ Disposable syringe 20 ml ■ Place the product on a tray that
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
■ Chapter Manual pre-cleaning
Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe.
with a brush
■ Chapter Mechanical alkaline
■ Cleaning brush: for example,
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subsection:
■ Disposable syringe 20 ml ■ Place the product on a tray that is suitable for cleaning (avoid rinsing blind spots).
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subsection:
is suitable for cleaning (avoid rinsing blind spots).
TA011944
D–W: FD–W:
cleaning and thermal disinfecting
■ Chapter Manual pre-cleaning with ultrasound and brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
Manual cleaning/disinfection
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
3.8.1
Manual cleaning with immersion disinfection
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
infection procedure.
Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.9
D–W: FD–W: RT: *Recommended:
Drinking water Fully desalinated water (demineralized, microbiological, at least of drinking water quality) Room temperature BBraun Stabimed fresh
Mechanical cleaning/disinfection with manual pre-cleaning
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
3.9.1
Manual pre-cleaning with a brush
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
3.13 Storage
3.9.2
Manual pre-cleaning with ultrasound and brush
Phase
Step
D [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
4.
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
3.9.3
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
I
Prerinse
<25/77
II
Cleaning
55/131
t [min]
► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Water quality
Chemical
3
D–W
-
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ working solution 0.5% – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD–W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alcaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.10 Inspection ► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
3.10.1 Visual inspection ► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps. ► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
scratched and fractured components. ► Check the product for missing or faded labels. ► Check the products with long, slim shapes (in particular rotating instruments) for deformities. ► Check the product for damage to the spiral element. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical service.
3.10.2 Functional test CAUTION Damage (metal cold welding/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598). ► Assemble disassembled products, see Assembly. ► Check that the product functions correctly. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical service. ► Turn locking screw 1 clockwise into housing 2.
3.11 Packaging ► Place the product in its holder or on a suitable tray. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.
3.12 Steam sterilization ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
opening any valves and faucets). ► Validated sterilization process
– Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
Technical service
CAUTION Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. Note The user institution is obliged to reprocess the product before its disposal, see Validated reprocessing procedure. TA009907