Aesculap®
Aesculap Endoscopic Technology
Instructions for use/Technical description AdTec mini instrument system Instructions for use/Technical description AdTec mini instrument system Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapusa.com and clicking the "Products" menu. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-800-282-9000. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung AdTec mini-Instrumentensystem Mode d’emploi/Description technique AdTec Système d’instruments mini Instrucciones de manejo/Descripción técnica AdTec Sistema de instrumentos mini Istruzioni per l’uso/Descrizione tecnica AdTec Sistema di strumenti mini Instruções de utilização/Descrição técnica AdTec Sistema de instrumentos mini Gebruiksaanwijzing/Technische beschrijving AdTec mini-instrumentensysteem Bruksanvisning/Teknisk beskrivning AdTec mini-instrumentsystem Инструкция по примению/Техническое описание AdTec Система инструментов mini Návod k použití/Technický popis AdTec Systém nástrojů mini Instrukcja użytkowania/Opis techniczny AdTec System instrumentów mini Návod na použivanie/Technický opis AdTec mini nástrojový systém Kullanım Kılavuzu/Teknik açiklama AdTec mini-Alet Sistemi 사용 설명서 / 기술 설명 재사용가능발조절식전기수술기용전극 (RU Monopolar, 형명개별기재 )
A A1
A2 B
B1
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap – a B. Braun company TA-Nr. 013248
2017-03
V6
B2
Änd.-Nr. 54275
7
8
9
2
- DIR 93/42/EEC Technical alterations reserved
1
4 3
2
5 6
I
II
Handle with lock
Aesculap® AdTec mini instrument system Legend 1 2 3 4 5 6 7 8 9
Handle (complete) Movable part of the handle HF pin Star wheel Actuating lever Switch (lubrication point) Jaw insert (lubrication point) Pressure bar Outer tube
► Press actuating lever 5 to release the ratchet mechanism. ► To activate the ratchet mechanism, release actuating lever 5.
Switch 6 can be used for permanently deactivating the ratchet mechanism. ► To deactivate the ratchet mechanism on handle 1: – Position switch 6 as shown in I. The switch 6 clicks into the front position. ► To activate the ratchet mechanism on handle 1: – Position switch 6 as shown in II. The switch 6 clicks into the rear position.
Disassembling Note For disassembling the product, also see additional series of illustrations A. ► Deactivate the ratchet mechanism on handle 1 with lock:
– Position switch 6 as shown in I. ► Detach the shaft from handle 1:
Symbols on product and packages Caution, general warning symbol Caution, see documentation supplied with the product
– Push star wheel 4 back, hold it there, and remove the shaft. – Extract jaw insert 7, together with outer tube 9, from handle 1. ► Disassemble the shaft: – Remove outer tube 9 from jaw insert 7. – Rotate pressure bar 8 counterclockwise by 90° and extract it from jaw insert 7.
Assembling Note For assembling the product, also see additional series of illustrations B.
Intended use All professional disciplines that use endoscopy: ■ Cutting, preparation, and grasping of tissues ■ Biopsies ■ Suturing
Safe handling and preparation ► Ensure that the product and its accessories are operated and used only by persons with the requisite training,
knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial
sterilization. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts. ► To avoid damage to the working end: Carefully insert the product through the working channel (e.g. trocar).
Risk of injury from ignition or explosion of flammable gases! Sparks may occur when using the HF device as directed. ► Observe the safety guidelines in the instructions for use of the HF device. WARNING
WARNING
Thermal injuries to patients/users due to insufficient insulation of leads in active accessories! ► Adjust the HF device to an appropriate setting to ensure that the peak output voltage does match or not exceed the accessory voltage rating specified for the product. ► Use the instrument only with an insulated outer tube.
The manufacturer has tested the product and verified that its insulation can withstand 300 reprocessing cycles. In clinical practice, the service life will depend on the individual intraoperative usage and the hospital's specific reprocessing conditions. ► Adjust the HF power output to the intended surgical intervention. Take into account clinical experience or reference values. ► Select the lowest possible HF power output. ► Keep the product’s contact surfaces clean during surgery. Remove encrusted tissue residues or body fluids with a moistened swab. The plug end of the product is fitted with the following connector: pin, 4 mm, sprung. Refer to our brochures to find a compatible cable. The accessory voltage rating of the product is 2 kVp. The accessory voltage rating must exceed or match the peak output voltage with which the product is operated in combination with a suitable HF device at an appropriate operating mode/setting (see IEC/DIN EN 60601-2-2). To avoid HF burns: ► Always keep the working end of the product in the user’s field of vision whenever the HF power is activated. ► Prior to activating the HF device, check that the working end of the product is not touching any electrically conductive accessories. ► Prior to each use, visually inspect the product for: damage or surface changes to the insulation. ► Never place the product on or next to the patient. ► When using accessories for endoscopy or laparoscopy, deactivate the automatic switch-on mode of the HF device. ► Follow the instructions for use of the HF device.
Safe operation Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product. WARNING Risk of injury when using the product beyond the field of view! ► Apply the product only under visual control. WARNING ► To open and close the distal end of the jaws:
– Open and close the moveable part of the handle 2.
► Deactivate the ratchet mechanism on handle 1 with lock:
– Position switch 6 as shown in I. ► Assemble the shaft:
– Slide pressure bar 8 into jaw insert 7 and rotate it clockwise by 90°. – Slide outer tube 9 onto jaw insert 7 as far as it will go. ► Hold the assembled shaft at jaw insert 7 with one hand. ► With the other hand, hold handle 1 at star wheel 4. The moveable part of the handle 2 must remain freely moveable. ► Push the shaft into handle 1, making certain that the notch on pressure bar 8 is aligned with the mark on star wheel 4. The moveable part of the handle 2 moves upward. The shaft engages automatically when it reaches the stop. ► Test the instrument for proper functioning by opening and closing the jaws.
Validated reprocessing procedure General safety instructions Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at www.extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
General information Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
Disassembling the product before carrying out the reprocessing procedure ► Disassemble the product immediately after use, as described in the respective instructions for use. ► Open up instruments with hinges.
Preparations at the place of use
Manual cleaning/disinfection
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
Preparation before cleaning
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning /disinfection process if necessary.
Manual cleaning with immersion disinfection
► Dismantle the product prior to cleaning, see Disassembling.
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
Cleaning/disinfection Product-specific safety notes on the reprocessing procedure
CAUTION
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – are approved for the material in question (e.g., aluminum, plastics, highgrade steel), ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 93 °C.
Immersion treatment in a 3 % H2O2 solution for approx. 5 minutes is a particularly effective and gentle method to dissolve encrustations from HF instruments. Subsequently, the debris can be removed by hand, with a medium-hard brush and or in an ultrasonic bath. This is followed by the conventional reprocessing steps. ► Use suitable cleaning/disinfecting agents if the product is put away in a wet condition. To prevent foam formation and reduced effectiveness of the process chemicals: Prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water. ► Carry out ultrasound cleaning: – as an effective mechanical supplement to manual cleaning/disinfection. – as a pre-cleaning procedure for products with encrusted residues, in preparation for mechanical cleaning/disinfection. – as an integrated mechanical support measure for mechanical cleaning/disinfection. – for additional cleaning of products with residues left after mechanical cleaning/disinfection.
Validated cleaning and disinfection procedure Validated procedure
Specific requirements
Reference
Manual cleaning with immersion disinfection
■ Cleaning brush, e.g., GK469200 ■ 20 ml disposable syringe ■ When cleaning products with
Chapter Manual cleaning/disinfection and sub-chapter:
■ Handle
movable hinges, ensure that these are in an open position and, if applicable, move the joint while cleaning.
■ Chapter Manual cleaning with immersion disinfection
■ Drying phase: Use a lint-free cloth or medical compressed air Manual cleaning with ultrasound and immersion disinfection
■ Shafts
■ Cleaning brush, e.g., GK469200 ■ 20 ml disposable syringe ■ When cleaning instruments with movable hinges, ensure that these are in an open position and, if applicable, move the joint while cleaning.
Chapter Manual cleaning/disinfection and sub-chapter:
■ Chapter Manual cleaning with ultrasound and immersion disinfection
■ Drying phase: Use a lint-free cloth or medical compressed air Mechanical alkaline cleaning and thermal disinfection
■ Handle
■ Place the instrument in a tray that is suitable for cleaning (avoiding rinsing blind spots).
■ Connect components with lumens and channels directly to the rinsing port of the injector carriage.
Chapter Mechanical cleaning/disinfecting and sub-chapter:
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
■ Keep working ends open for
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
Manual cleaning with ultrasound and immersion disinfection Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
cleaning.
■ Place instruments in the tray with their hinges open. Manual pre-cleaning with brush and subsequent mechanical alkaline cleaning and thermal disinfection
■ Shafts
■ Cleaning brush, e.g., GK469200 ■ 20 ml disposable syringe ■ Place the instrument in a tray that is suitable for cleaning (avoiding rinsing blind spots).
■ Connect components with lumens and channels directly to the rinsing port of the injector carriage.
■ Keep working ends open for cleaning.
■ Place instruments in the tray with their hinges open.
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and sub-chapter:
■ Chapter Manual pre-cleaning with a brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully.
Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened.
Mechanical alkaline cleaning and thermal disinfecting
Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully.
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution
Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
III
Intermediate rinse
>10/50
1
FD-W
-
Mechanical cleaning/disinfecting
IV
Thermal disinfecting
90/194
5
FD-W
-
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
V
Drying
-
-
-
According to the program for cleaning and disinfection device
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Mechanical alkaline cleaning and thermal disinfecting
– pH = 11*
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical/Note
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Inspection, maintenance and checks
– pH = 13 – <5 % anionic surfactant
Damage (metal seizure/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598).
CAUTION
■ 0.5 % working solution – pH = 11*
► Allow the product to cool down to room temperature. ► After each complete cleaning, disinfecting and drying cycle, check that the instrument is dry, clean, operational,
III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Mechanical cleaning/disinfection with manual pre-cleaning Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
and free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components). ► Dry the product if it is wet or damp. ► Repeat cleaning and disinfection of products that still show impurities or contamination. ► Check that the product functions correctly. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Technical Service. ► Assemble dismountable products, see Assembling. ► Check for compatibility with associated products.
Packaging ► Appropriately protect products with fine working tips. ► Store products with ratchet locks fully opened or locked no further than in the first notch. ► Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected. ► Pack trays appropriately for the intended sterilization process (e.g. in sterile Aesculap containers). ► Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
Steam sterilization
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Note The product can be sterilized either in disassembled or in assembled condition.
Manual pre-cleaning with a brush
Note To avoid breakage due to stress crack corrosion, sterilize the instruments with the lock fully open or locked no further than on the first ratchet tooth.
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfectant cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
opening any valves and faucets).
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by ► Validated sterilization process
– Steam sterilization using fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Technical Service Risk of injury and/or malfunction! ► Do not modify the product. WARNING ► For service and repairs, please contact your national B. Braun/Aesculap agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1602 Fax: +49 (7461) 16-5621 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
Accessories/Spare parts For accessories and spare parts, see brochure C 28011 or visit http://www.endoscopy-catalog.com
Disposal ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging!