AESCULAP Acculan 4 – Pre-cleaning and care information
63 Pages
Preview
Page 1
INHALT Vorreinigung3 Pflege18
CONTENT Pre-cleaning33 Care48
SAGITTALSÄGE UND SAGITTALSÄGEAUFSATZ
PFLE G E
VORREINIGUNG
30
STICHSÄGE UND STÜTZFUSS
PFLE G E
VORREINIGUNG
31
DETAILS FOR PRE-CLEANING
Phase
Step
T [°C / °F]
t [min]
Conc. [%]
Water quality
Chemistry / Note
I
Rinsing
RT (cold)
–
–
D-W
Until visually clean
II
Brushes
RT (cold)
–
–
D-W
Until visually clean
D-W: Drinking water RT: Room temperature
Phase I ❙ Mobilize non-rigid components during cleaning. ❙ Thoroughly clean the product under running water.
Phase II ❙ Mobilize non-rigid components during cleaning. ❙ Brush cannulation with cleaning brush TA011944 and difficult to access surfaces with a suitable plastic cleaning brush for at least 1 min. ❙ After manual preparation, check visible surfaces for residue and repeat the pre-cleaning process as needed.
This document is also valid for the Acculan 3Ti generation.
PRE- C L E AN IN G
CARE
33
SMALL DRILL (GA344)
1. Turn rotating sleeve of attachment coupling while brushing under running water
2. Move trigger while brushing under running water
3. Clean cannulation under running water with brush TA011944
PRE- C L E AN IN G
CARE
34
DRILL AND REAMING ATTACHMENTS WITH QUICK COUPLING FOR GA344 1. Move unlocking sleeve while brushing under running water
2. Clean cannulation under running water with brush TA011944
PRE- C L E AN IN G
CARE
35
JACOBS CHUCKS FOR GA344 1. CLOSE Jacobs chuck completely and brush under running water
2. OPEN Jacobs chuck completely and brush again under running water
3. Clean cannulation under running water with brush TA011944
PRE- C L E AN IN G
CARE
36
K-WIRE ATTACHMENT FOR GA344 1. Move adjustment sleeve and brush under running water
2. Turn adjustment sleeve on smallest diameter setting and brush under running water 3. Turn adjustment sleeve on largest diameter setting and brush cannulation with brush TA011944
PRE- C L E AN IN G
CARE
37
SAW ATTACHMENTS FOR GA344 1. Brush saw blade coupling and gap between saw blade coupling and saw head under running water
2. Set the saw blade coupling between two locking positions and brush under running water
3. After pre-cleaning set the saw blade coupling between two locking positions
1. Brush saw blade coupling under running water
PRE- C L E AN IN G
CARE
38
SCREW ATTACHMENTS FOR GA344 1. Move unlocking sleeve while brushing under running water
2. Clean cannulation under running water with brush TA011944
PRE- C L E AN IN G
CARE
39
DRILL AND REAMER
1. Turn rotating sleeve of attachment coupling while brushing under running water 2. Move trigger while brushing under running water
3. Clean cannulation under running water with brush TA011944
PRE- C L E AN IN G
CARE
40
DRILL- AND REAMER ATTACHMENTS WITH QUICK COUPLING 1. Move unlocking sleeve while brushing under running water
2. Clean cannulation under running water with brush TA011944
PRE- C L E AN IN G
CARE
41
JACOBS CHUCKS 1. CLOSE Jacobs chuck completely and brush under running water
2. OPEN Jacobs chuck completely and brush again under running water
PRE- C L E AN IN G
3. Clean cannulation under running water with brush TA011944
CARE
42
K-WIRE CHUCK 1. Move adjustment sleeve and brush under running water
2. Turn adjustment sleeve on smallest diameter setting and brush under running water
3. Turn adjustment sleeve on largest diameter setting and brush cannulation with brush TA011944
PRE- C L E AN IN G
CARE
43
SAGITTAL SAW AND SAGITTAL SAW ATTACHMENT 1. Brush saw blade coupling and gap between saw blade coupling and saw head under running water 2. Pull out saw head and brush gap under running water
4. Move trigger while brushing under running water
3. After pre-cleaning set the saw head between two locking positions
PRE- C L E AN IN G
CARE
44
RECIPROCATING SAW 1. Clean saw blade coupling with cleaning template GB440215 (five times) 2. Press prestle to its end position and brush under running water
PRE- C L E AN IN G
3. Move trigger while brushing under running water
CARE
45
PROTECTION GUIDE 1. Pull back locking sleeve and brush under running water
PRE- C L E AN IN G
2. After pre-cleaning pull back locking sleeve and keep it open by inserting cleaning template GB440215 into the gap
CARE
46
DERMATOME 1. Disassemble flap rod for cleaning
PRE- C L E AN IN G
2. Brush area around pusher dog under running water
CARE
47
DETAILS AUTOMATIC CLEANING AND DISINFECTION Phase
Step
T [°C / °F]
t [min]
Water quality
Chemistry / Note
I
Pre-rinse
< 25 / 77
3
D-W
–
II
Cleaning
55 / 131
10
FD-W
❙ Concentrate, alcaline: ❙ pH – 13 ❙ < 5 % anionic surfactant ❙ 0.5 % working solution ❙ pH – 11*
III
Intermediate rinse
> 10 / 50
1
FD-W
–
IV
Thermal disinfection
90 / 194
5
FD-W
–
V
Drying
–
–
–
mind. 10 min at max. 120 °C
D-W: FD-W: *Recommended:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) B. Braun Helimatic Cleaner alkaline
❙ Insert the product in its proper position in the ECCOS fixation ❙ Follow the product specific instruction for use ❙ Connect the interior rinse device with the rinse connector of the cleaning / disinfection device / rinse cart ❙ After automatic cleaning / disinfection, check visible surfaces for residues and repeat the cleaning / disinfection process as needed
CARE
PRE-CLEANING
48
DETAILS CARE : Spray drill and reamer, small drill and their attachments prior to each sterilization with oil spray for 2 sec. : In addition marked moving parts have to be lubricated occasional by spraying with the oil spray and moved afterwards.
ATTENTION ❙ Only use GB600 Sterilit Power Systems oil spray and the correpsonding adapters (GA600880 for Acculan 4 small drill, GA600840 for Acculan 4 Drill and Reamer)! ❙ Products may need final drying after automated cleaning. If compressed air is used for final dry off, ensure that air is sprayed from a distance of 30 cm in an oscillating motion across surfaces and that air stream is not aimed directly onto bearing seats.
CARE
PRE-CLEANING
49