BBraun
Aesculap endo-lenses
27 Pages
Preview
Page 1
Aesculap®
3
4
5
Aesculap Endoscopic Technology
1 Instructions for use/Technical description Aesculap endo-lenses Gebrauchsanweisung/Technische Beschreibung Aesculap Endoobjektive Mode d’emploi/Description technique Endo-objectifs Aesculap Instrucciones de manejo/Descripción técnica Endoobjetivos Aesculap Istruzioni per l’uso/Descrizione tecnica Endobiettivi Aesculap Instruções de utilização/Descrição técnica Objectiva endoscópica Aesculap Gebruiksaanwijzing/Technische beschrijving Aesculap endo-objectieven Bruksanvisning/Teknisk beskrivning Aesculap endoobjektiv Инструкция по примению/Техническое описание Эндоскопические объективы Aesculap Návod k použití/Technický popis Endoobjektivy Aesculap Instrukcja użytkowania/Opis techniczny Obiektywy endoskopowe Aesculap Návod na použivanie/Technický opis Endoobjektívy Aesculap
2
1
A 3
2
Kullanım Kılavuzu/Teknik açiklama Aesculap endoskop objektifi
1
B Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap – a B. Braun company TA-Nr. 010039
06/14
- DIR 93/42/EEC Technical alterations reserved
V6
Änd.-Nr. 50175
4
1
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
Aesculap® Aesculap endo-lenses
Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation.
Legend
Note If there is no final sterilization, then a virucidal disinfectant must be used.
A B 1 2 3 4 5
Zoom endo-lens PV122S Normal lens PV123S Glass surfaces Retaining claws Push buttons Focusing ring Zoom ring
Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at www.extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
General information Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
Symbols on product and packages Symbol
Explanation Caution, general warning symbol Caution, see documentation supplied with the product
Follow the instructions for use
Date of manufacture
Preparations at the place of use
Applicable to ► For item-specific instructions for use and information on material compatibility, see also the Aesculap Extranet
at www.extranet.bbraun.com
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
Preparation before cleaning
Intended use Aesculap endo-lenses are used to project the recorded operating field onto the CCD chip of the camera. The Aesculap endo-lens is suitable for all endoscopes with an eyepiece funnel according to DIN 58105 and Aesculap 1/2“ cameras with a standardized V-mount or C-mount thread. Lens
Connecting thread
Focal length
Zoom endo-lens PV122S
V-mount-/C-mount thread
25-1.97 in
Normal lens PV123S
V-mount-/C-mount thread
1.18 in
Damage to the product due to improper handling when reprocessing! ► Disconnect the camera head and endo-lens from each other prior to reprocessing. CAUTION
Cleaning/disinfection Product-specific safety notes on the reprocessing procedure
Safe handling and preparation CAUTION Federal law restricts this device to sale by, or on order of a physician! ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts.
Safe operation Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product.
Damage to, or destruction of the product caused by mechanical cleaning/disinfection! ► Only clean and disinfect the product manually. CAUTION Risk of damage to the product due to inappropriate cleaning/disinfecting agents! ► Only use cleaning/disinfecting agents approved for surface cleaning. Follow the manufacturer’s instructions for the respective cleaning/disinfecting agent. CAUTION
Validated cleaning and disinfection procedure Validated procedure
Specific requirements
Reference
Manual cleaning and wipe disinfection
■ When cleaning products with
Section Chapter Manual cleaning/ disinfection and subsection:
WARNING
movable hinges, ensure that these are in an open position and, if applicable, move the joint while cleaning.
■ Chapter Manual cleaning and wipe disinfecting
■ Drying phase: Use a lint-free cloth or medical compressed air
■ Ensure that the product is posi-
Risk of injury when using the product beyond the field of view! ► Apply the product only under visual control.
tioned in such a way that water will not enter the product e.g. through coupling interfaces. (Immediately remove any fluid that entered the product inadvertently.)
WARNING ► Press both push buttons 3 and spread the retaining claws 2 apart. ► Insert the endoscope. ► Release the push buttons 3. ► Turn the endoscope to check that it is connected correctly. ► Adjust the image sharpness to the relevant depth range with the focusing ring 4. ► Adjust the size of the image to the optimum size with the zoom ring 5 on the zoom endo-lens.
Validated reprocessing procedure General safety instructions Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
Manual cleaning/disinfection ► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning /disinfection process if necessary.
Manual cleaning and wipe disinfecting Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Cleaning
RT (cold)
-
-
D–W
-
II
Drying
RT
-
-
-
-
III
Wipe disinfection
-
>1
-
-
Meliseptol HBV wipes 50 % Propan-1-ol
IV
Final rinse
RT (cold)
0.5
-
FD-W
-
V
Drying
RT
-
-
-
-
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) Room temperature
RT:
Phase I ► Clean the product under running faucet water, using a suitable cleaning brush until all visible residues have been removed from the surfaces. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. Phase II ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure. Phase III ► Wipe all surfaces of the product with a single-use disinfectant wipe. Phase IV ► After the specified exposure time (at least 1 min), rinse the disinfected surfaces under running FD water. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
Inspection, maintenance and checks
CAUTION
Damage (metal seizure/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598).
► Allow the product to cool down to room temperature. ► After each complete cleaning, disinfecting and drying cycle, check that the product is dry, clean, operational, and
free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components). ► Dry the product if it is wet or damp. ► Repeat cleaning and disinfection of products that still show impurities or contamination. ► Check that the product functions correctly. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical Service. ► Check for compatibility with associated products.
Packaging ► Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected. ► Pack trays appropriately for the intended sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
Steam sterilization ► Use sterile covers JG904 from Aesculap. Follow the instructions for the use of the sterile covers.
Sterilization for the US market ■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization. ■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6, Aesculap recommends the following parameters: Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters* Sterilization method
Temp.
Time
Minimum drying time
Prevacuum
270 °F/275 °F
4 min
20 min
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc.
Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Technical Service Risk of injury and/or malfunction! ► Do not modify the product. WARNING ► For service and repairs, please contact your national B. Braun/Aesculap agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1602 Fax: +49 (7461) 16-5621 E-Mail: [email protected] Or in the US: Aesculap Inc. Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 Aesculap Repair Hotline Phone: +1 (800) 214-3392 Fax: +1 (314) 895-4420 Other service addresses can be obtained from the address indicated above.
Distributor in the US/Contact in Canada for product information and complaints Aesculap Inc. 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 010039
06/14
V6
Änd.-Nr. 50175