Instructions for use/Technical description Flexible shaft for drill bits Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapimplantsystemsifus.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-866-229-3002. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung Flexible Welle für Bohreinsätze Mode d’emploi/Description technique Arbre flexible pour inserts de forage Instrucciones de manejo/Descripción técnica Árbol flexible para brocas Istruzioni per l’uso/Descrizione tecnica Albero flessibile per inserti perforatori Instruções de utilização/Descrição técnica Eixo flexível para brocas Gebruiksaanwijzing/Technische beschrijving Flexibele as voor boorinzetstukken Brugsanvisning/Teknisk beskrivelse Fleksibel aksel til boreindsatser Bruksanvisning/Teknisk beskrivning Flexibel axel för borrinsatser Käyttöohje/Tekninen kuvaus Taipuisa akseli poranterille Lietošanas instrukcijas/tehniskais apraksts Elastīga vārpsta urbju uzgaļiem Naudojimo instrukcija/techninis aprašas Lankstusis velenas grąžtų galvutėms Инструкция по примению/Техническое описание Адаптер для сверлильных насадок для дрелей Návod k použití/Technický popis Flexibilní hřídel pro vrtací nástavce Instrukcja użytkowania/Opis techniczny Elastyczny wałek do nasadek wiertarskich Návod na použitie/Technický opis Pružný hriadeľ pre vŕtacie nadstavce Használati útmutató/Műszaki leírás Rugalmas tengely fúrófejekhez Navodila za uporabo/Tehnični opis Prožna gred za svedre Upute za uporabu/Tehnički opis Fleksibilno vratilo za svrdla Manual de utilizare/Descriere tehnică Arbore flexibil pentru ajutaje de burghiu Упътване за употреба/Техническо описание Гъвкав вал за накрайници за свредла Kullanım Kılavuzu/Teknik açiklama Delme uçları için esnek mil Οδηγίες χρήσης/Τεχνική περιγραφή Εύκαμπτος άξονας για μύτες τρυπανιού
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.bbraun.com AESCULAP® – a B. Braun brand TA013807
2021-03
Change No. 64356
1 4
3
1
A
B
C
en ®
AESCULAP Flexible shaft for drill bits
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this.
About this document
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
Scope
These instructions for use apply for the flexible shaft for drill bits. ► For article specific instructions for use and material compatibility and lifetime information, see B. Braun eIFU at
eifu.bbraun.com
1.2
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows: WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result. CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
Clinical use
2.1
Areas of use and limitations of use
2.1.1
Intended use
Drill bits (for example, NT424R, NT426R, NT429R) are inserted in to the flexible shaft to be used for pre-drilling drill holes for bone screws.
2.1.2
Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
3.2
Safety messages
2.
General safety information
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
Sleeve Flexible shaft Connecting link Drill Pin
1.
Validated reprocessing procedure
3.1
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing.
Legend 1 2 3 4 5
3.
Indications
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems:
■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes.
■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling.
The indications correspond to those of the implant system used.
► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause
2.1.3
► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at
corrosion.
Absolute contraindications
The absolute contraindications correspond to those of the implant system used.
2.1.4
www.a-k-i.org, link to "AKI-Brochures", "Red brochure".
Relative contraindications
3.3
The relative contraindications correspond to those of the implant system used. In the presence of relative contraindications, the user decides individually regarding the use of the product.
2.2
Safety information
2.2.1
Clinical user
3.4
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user. Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located. Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
Sterility
The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
2.3
Reusable products
Influences of the reprocessing which lead to damage to the product are not known. A careful visual and functional inspection before the next use is the best option to determine if a product is no longer functional, see Inspection.
Preparations at the place of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
3.5
Preparing for cleaning
► Disassemble the product prior to cleaning, see Disassembly.
3.6
Disassembly
► Push sleeve 1 in the direction of the flexible shaft 2 and turn counterclockwise.
The sleeve can be removed. The drill can be removed.
3.7
Cleaning/Disinfection
3.7.1
Product-specific safety information on the reprocessing method
Danger to the patient! The product must not be manually reprocessed. ► Reprocess the product only in a mechanical washer/disinfector. ► Reprocess the product only by manually pre-cleaning followed by cleaning in a mechanical washer/disinfector. Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents approved for aluminum, plastics and high-grade steel according to manufacturer’s instructions. ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C. ► With PVD coated products, do not use oxidizing process chemicals (e.g. H2O2), as these can cause bleaching or
layer loss.
Application
► Use suitable cleaning/disinfecting agents if the product is disposed of in a wet condition. To prevent foaming and
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product.
3.7.2
2.3.1
Validated procedure
Specific requirements
Reference
Manual pre-cleaning with ultrasound and brush, and subsequent mechanical alkaline cleaning and thermal disinfection
■ Suitable cleaning brush ■ 20 ml disposable syringe ■ Place the product on a tray that
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subsection:
Inserting the drill and engaging the lock
► Press the sleeve 1 in the direction of the flexible shaft 2, turn it a counter-clockwise direction and click the pin 5
into the unlocked position , see Fig. B. The flexible shaft has been unlocked. ► Insert the hexagon connection of the drill completely into the flexible shaft. ► Move the sleeve 1 back in the opposite direction and bring the pin 5 into the locking position The drill is now locked.
, see Fig. C.
degradation of the efficacy of the process chemicals: prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water.
Validated cleaning and disinfection procedure
is suitable for cleaning (avoid rinsing blind spots).
■ Connect components with lumens and channels directly to the rinsing port of the injector carriage.
■ Chapter Manual pre-cleaning with ultrasound and brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
3.8
Mechanical cleaning/disinfection with manual pre-cleaning
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
3.8.1
Manual pre-cleaning with ultrasound and brush
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
3.12 Steam sterilization ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
opening any valves and faucets). ► Validated sterilization process
– Disassemble the product – Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
3.13 Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
4.
Technical service
CAUTION Modifications to medical technology equipment result in loss of approval and can result in forfeiture of guarantee/warranty claims. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency.
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging.
Machine type: single-chamber cleaning/disinfection device without ultrasound
Note The user institution is obliged to reprocess the product before its disposal, see Validated reprocessing procedure.
Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
6.
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
3.8.2
Mechanical alkaline cleaning and thermal disinfecting
Symbols on product and packaging Locked position
– pH ~ 13 – <5 % anionic surfactant
Unlocked position
■ working solution 0.5% – pH = 11* TA013807
III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
In accordance with the program for the cleaning and disinfecting machine
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.9
Inspection
► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
3.9.1
Visual inspection
► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps. ► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
scratched and fractured components. ► Check the product for missing or faded labels. ► Check the products with long, slim shapes (in particular rotating instruments) for deformities. ► Check the product for damage to the spiral element. ► Check the cutting edges for continuity, sharpness, nicks and other damage. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical service.
3.9.2
Functional test
► Assemble disassembled products, see Assembly. ► Check that the product functions correctly. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Check rotating products (e.g. reusable drills and cutters) for bends and deformities. To do this, roll the product
on an even surface. ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical service.
3.10 Assembly ► Slide sleeve 1 onto flexible shaft 2 and with connecting link 3 lock the sleeve 1 in place, see Fig. A.
The sleeve and flexible shaft are connected. The pin 5 is in the locked position
.
3.11 Packaging ► Place the product in its holder or on a suitable tray. Ensure that sharp edges are covered. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.