BBraun
Hi-Line XXS Handpiece with shafts and reamers
56 Pages
Preview
Page 1
3.3
Aesculap® Hi-Line XXS Handpiece with shafts and reamers
Operating principle
The handpiece takes up the rotational speed of the motor and transfers the rotation to the coupled tool at a transmission ratio of 1:1. The handpiece can be coupled to the motor in 15 different positions. The handpiece shafts can be coupled to the handpiece and are therefore exchangeable. They can be fixed in eight different positions. Both straight and curved handpiece shafts are available.
Legend
4.
1 Handpiece 2 Threaded ring 3 Motor coupling 4 Tool unlocking lever 5 Shaft locking lever 6 Shaft connection 7 Handpiece shaft 8 Shaft coupling 9 Tool 10 Tool coupling 11 Reamer head 12 Spray nozzle 13 Rinsing tube 14 Tube connection
Non-compliance with the following instructions will preclude all responsibility and liability in this respect on the part of Aesculap. ► When setting up and operating the product, adhere to – national regulations for installation and operation, – national regulations on fire and explosion protection.
Preparation
4.1
First use Risk of injury and/or product malfunction due to incorrect operation of the medical system! ► Adhere to the instructions for use of any medical device. WARNING Risk of injury and material damage due to inadvertent actuation of the motor! ► Make certain the motor cannot be activated inadvertently during tool change. ► Change tools only with the motor secured by the safety stop.
Symbols on product and packages WARNING
Caution, general warning symbol Caution, see documentation supplied with the product
Risk of burns and other injuries due to bent or incorrectly coupled tools! ► Make certain the tool is coupled correctly. ► Only use perfectly straight tools.
Rotate “right” (clockwise) to unlock the handpiece, shaft and stem
Rotate “left” (counterclockwise) to lock the handpiece and stem
1.
5.
Working with the handpiece and the handpiece shafts
5.1
Connecting a handpiece to a motor
► Turn threaded ring 2 of handpiece 1 in the direction of the arrow in symbol as far as it will go. ► Install handpiece 1 on the handpiece coupling of the motor in such a way that the pin on the handpiece engages
Applicable to
in one of the notches of the handpiece coupling.
► For item-specific instructions for use and information on material compatibility, see also the Aesculap Extranet
at www.extranet.bbraun.com
2.
WARNING
► Turn threaded ring 2 in the direction of the arrow in symbol as far as it will go. ► Make certain that the threaded ring is fastened to its limit stop, because otherwise it may come loose due to
vibrations during operation. The handpiece is coupled now.
Safe handling
CAUTION Federal law restricts this device to sale by, or on order of a physician! ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Prior to use, check that the product is in good working order. ► Observe “Notes on Electromagnetic Compatibility (EMC)“, see TA022130. ► To prevent damage caused by improper setup or operation, and in order not to compromise warranty and manufacturer liability: – Use the product only according to these instructions for use. – Follow the safety and maintenance instructions. – Only combine Aesculap products with each other. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Keep the instructions for use accessible for the user. ► Always adhere to applicable standards.
5.2
Disconnecting a handpiece from a motor
► Turn threaded ring 2 in the direction of the arrow in symbol ► Detach handpiece 1 from the motor.
5.3
as far as it will go.
Connecting a handpiece shaft to the handpiece
► Turn shaft locking lever 5 in the direction of the arrow in symbol as far as it will go. ► Insert handpiece shaft 7 with shaft coupling 8 into shaft connection 6 of the handpiece. If necessary align bent
handpiece shafts. ► Turn shaft locking lever 5 in the direction of the arrow in symbol
5.4
as far as it will go.
Disconnecting a handpiece shaft from the handpiece
► Disconnect the tool, if necessary. ► Turn shaft locking lever 5 in the direction of the arrow in symbol ► Remove handpiece shaft 7.
as far as it will go.
Single-use product
WARNING
Infection hazard for patients and/or users and impairment of product functionality due to reuse. Risk of injury, illness or death due to contamination and/or impaired functionality of the product! ► Do not reprocess the product.
5.5
Coupling a tool
Note The tool is difficult to insert into/remove from the bent handpiece shafts due to the bend. ► To prevent cutting through the glove, be careful when touching the tool 9 where it is sharp, and if applicable use
gauze. The products GE802SU–GE818SU and GE832SU–GE848SU are sterilized by radiation and are supplied in sterile packaging. The product must not be reused. ► Do not use products from open or damaged sterile packaging. ► Prior to each use, inspect the product for loose, bent, broken, cracked or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Do not re-sterilize the product. ► Do not use the product after its use-by date.
jammed or are difficult to insert. ► Insert tool 9 into handpiece 1 as far as it will go, until the correct position is found.
The tool is locked automatically. ► If the pressure point cannot be passed with the tool: Swivel tool unlocking lever 4 in the direction of the arrow
in the symbol
Product description
3.1
Components required for operation
against the spring force and push the tool into handpiece 1 as far as it will go.
► Release tool unlocking lever 4.
The tool must remain fixed even when pulled.
5.6
3.
Uncoupling a tool
► Carefully hold tool 9 by the cutting edge. ► Swivel tool unlocking lever 4 in the direction of the arrow in symbol
Art. no.
Designation
GA740R, GD675, ...
Motor, motor system
GB790R
Handpiece
GB791R, GB793R
Handpiece shaft, straight
GB792R, GB794R
Handpiece shaft, curved
GE802SU–GE818SU GE832SU–GE848SU
Tools
3.2
► Insert tool 9 into the tool coupling on handpiece shaft 7 until the pressure point. Do not use tools that get
against the spring force and remove
the tool.
5.7
Connecting/Disconnecting the spray nozzle
Note A spray nozzle can be attached to the handpieces for the purpose of cooling the tools and, especially, to avoid heat necroses. Note If the connection of the spray nozzle at handpiece shaft is unstable, vibrations can cause the tip of the spray nozzle to touch the rotating tool, leading to metal chipping. Assignment of spray nozzles Handpieces (art. no.)
Rings
Spray nozzles (art. no.)
GB791R, GB792R
II
GB796SU
GB793R, GB794R
III
GB797SU
Intended use
The Hi-Line XXS handpiece GB790R with the shafts GB791R to GB794R are used in micro-, neuro-, spinal, ENT and maxillary surgery. The tools (e.g., GE802SU–GE818SU and GE832SU–GE848SU) are used to prepare the bones. The spray nozzles GB796SU and GB797SU are used to cool the tool's cutting edges.
Connecting the spray nozzle ► From the front, push spray nozzle 12 onto handpiece shaft 7. ► Push irrigation tube 13 over tube connection 14. To do this, hold spray nozzle12 at tube connection 14.
► If necessary, align the tip of spray nozzle towards the tool. ► Set the irrigation flow at the control unit, as described in the instructions for use of HiLAN irrigation pump GA530
6.
Validated reprocessing procedure
6.1
General safety instructions
or microspeed uni control unit GD670. Disconnecting the spray nozzle ► Pull off irrigation tube 13 from tube connection 14. ► Remove spray nozzle 12 from shaft 7.
5.8
Function checks
► Always perform a trial run prior to using the system. ► Make certain the handpieces and handpiece shaft are in proper condition. ► Make certain the handpiece is coupled correctly. ► Verify that the handpiece shaft is coupled correctly. ► Make certain the tool is coupled correctly. ► Check for damage to, and irregular running noises, excessive vibrations and excessive heat-up of the handpiece
and the handpiece shaft. ► Check that the handpiece/handpiece shaft is only warm to touch after a short period of use, no hotter. ► Check all cutting edges for any mechanical damage.
5.9
Safe operation Risk of burns to patient and user due to hot motor/hot tool! ► Do not put the motor/hot tool down on the patient. ► Use a cloth to protect against burns when changing the motor or tool. WARNING
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at www.extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
6.2
Single-use products
Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product. WARNING
WARNING Risk of injury when using the product beyond the field of view! ► Apply the product only under visual control. WARNING Risk of injury from bent or snapped tools! ► Only use perfectly straight tools. ► Only apply mild pressure when using a tool.
Infection hazard for patients and/or users and impairment of product functionality due to reuse. Risk of injury, illness or death due to contamination and/or impaired functionality of the product! ► Do not reprocess the product!
Art. no.
Designation
GE802SU–GE818SU GE832SU–GE848SU
Tools
GB796SU, GB797SU
Spray nozzles
6.3
Reusable products
WARNING Eye injury caused by particles coming loose from the tool! ► Always wear protective goggles when using the motor system.
Art. no.
Designation
GB791R, GB793R
Handpiece shaft, straight
GB792R, GB794R
Handpiece shaft, curved
GB790R
Handpiece
WARNING Risk of burns and necroses from hot cutting edges of the tool! ► Tools must be continuously cooled during operation. ► Only use sharp tools. WARNING Coagulation of patient tissue if the handpiece or handpiece shaft is hot! ► Let the handpiece and handpiece shaft cool down. ► Do not use a handpiece or handpiece shaft whose bearings are damaged. WARNING Risk of injury due to small parts that could enter the operation site! ► Do not perform any change of reamer over the operating field. WARNING
WARNING
Risk of injury if the tool is uncoupled by activating the tool unlocking lever during operation! ► Never actuate the tool unlocking lever on the handpiece while the system is operating.
WARNING
Risk of injury or loss of function if handpiece shaft is disconnected during operation! ► Never actuate the shaft locking lever on the handpiece while the system is operating.
Risk of infection and contamination due to aerosol formation! ► Take appropriate protective measures, e.g. watertight protective clothing, face mask, protective goggles and a suction extraction system.
► Carry out sterile processing of the product prior to first use.
6.4
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
6.5
Dismantling prior to carrying out the reprocessing procedure
► Disassemble the product immediately after use, as described in the respective instructions for use.
6.6
Preparations at the place of use
► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
WARNING
6.7 Risk of injury and damage to the tool/system! The rotating tool may catch cover drapes (textile or other). ► Never allow the tool to touch cover drapes (textile or other) during operation. WARNING
Preparation before cleaning
► Carry out non-fixating/NaCl-free pre-cleaning immediately after use.
6.8
Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure Damage to micro instruments caused by incorrect handling! ► Protect products against overstraining. ► Do not bend the handpiece shaft. CAUTION CAUTION Note The handpiece shaft has a limited service life. When used as intended, a minimum of 30 uses are guaranteed. The end of the service life will manifest itself through a relatively quick increase in the temperature of the shaft. ► Do not run the handpiece without a tool coupled for any length of time. ► For fully exploiting the advantages of this power system, operate it at maximum motor speed whenever possible. ► Remove any surface contamination with a moistened, disposable wipe/paper tissue immediately after use.
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – are approved for plastic material and high-grade steel. ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 60 °C.
► Do not clean handpieces/handpiece shafts by ultrasonic treatment and do not immerse them in any liquids. To
avoid the risk of corrosion/malfunctioning, allow any fluid that has entered the product to drain out immediately.
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
Validated cleaning and disinfection procedure
infection procedure. Validated procedure
Special features
Reference
Manual cleaning and wipe disinfection
■ Drying phase: Use a lint-free cloth or compressed air for medical
Chapter Manual cleaning/disinfecting and subsection:
purposes
■ Ensure that the product is positioned in such a way that water will not enter the product e.g. through coupling interfaces. (Immediately remove any fluid that entered the product inadvertently.)
■ Handpiece Manual cleaning with immersion disinfection
for phase I
■ Shafts
■ The inner lumen must not be brushed
■ Rinse the shaft with cleaning disinfectant solution a minimum of 5 times; to do so, attach the appropriate silicone tube (e.g. ID=4 mm) onto the shaft coupling and syringe
■ Chapter Manual cleaning and wipe disinfecting Chapter Manual cleaning/disinfecting and subchapter:
■ Chapter Manual cleaning with immersion disinfection
for phase II
■ Rinse the shaft with water (with syringe and silicone tube) ■ Blow dry the inside of the shaft with an empty syringe or compressed air for phase III for phase IV
■ Rinse the shaft with water for phase V
■ Blow dry the inside of the shaft with an empty syringe or compressed air
■ Insert the handpiece in its proper position in the Eccos holder ■ Mechanical cleaning of the handpiece is only permitted with the Eccos holder GB605R
■ Mechanical cleaning of the handpiece shaft is only permitted
Chapter Mechanical cleaning/disinfecting and subsection:
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
with the wash tray GB696R. For this, observe instructions for use TA013855.
6.9
Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully.
■ Fill the shaft with disinfectant solution using a syringe
Mechanical alkaline cleaning and thermal disinfecting
Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe.
Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
6.10 Mechanical cleaning/disinfecting Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Manual cleaning/disinfecting
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
Mechanical alkaline cleaning and thermal disinfecting
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
I
Prerinse
<25/77
II
Cleaning
55/131
t [min]
Water quality
Chemical/Note
3
D–W
-
10
FD-W
■ Concentrate, alkaline:
Manual cleaning and wipe disinfecting Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Cleaning
RT (cold)
-
-
D–W
-
II
Drying
RT
-
-
-
-
III
Wipe disinfection
-
>1
-
-
Meliseptol HBV wipes 50 % Propan-1-ol
IV
Final rinse
RT (cold)
0.5
-
FD-W
-
V
Drying
RT
-
-
-
-
– pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11*
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) Room temperature
RT:
Phase I ► Clean the product under running faucet water, using a suitable cleaning brush until all visible residues have been removed from the surfaces. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. Phase II ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
6.11 Inspection, maintenance and checks Risk of malfunction of the handpiece or handpiece shaft during surgical use due to inadequate maintenance! ► Lubricate the handpieces and handpiece shafts after every cleaning process.
Phase III ► Wipe all surfaces of the product with a single-use disinfectant wipe. Phase IV ► After the specified exposure time (at least 1 min), rinse the disinfected surfaces under running FD water. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
Manual cleaning with immersion disinfection Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
CAUTION ► Allow the product to cool down to room temperature. ► Inspect the product after each cleaning and disinfecting cycle to be sure it is: clean, functional, and undamaged. ► Spray through the Hi-Line XXS handpiece and handpiece shafts with Aesculap STERILIT oil spray GB600 with
adapter GB600820 until clear oil emerges from the product. ► Check the product for any damage, abnormal running noise, overheating or excessive vibration. ► Set aside the product if it is damaged.
6.12 Packaging ► Follow the instructions for use for the applied packaging and storage systems (e.g. instructions for use TA009721
for Aesculap Eccos storage system).
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
► Insert the product in its proper position in the Eccos holder, or put it on a tray in such a way that the product is
protected against damage. ► Pack trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging will prevent a recontamination of the product.
6.13 Steam sterilization Note The product may only be sterilized when dismantled. ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets). ► Validated sterilization process
V D–W: FD–W:
Drying
RT
-
-
-
-
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed
– Disassemble the instrument – Steam sterilization through fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization through fractionated vacuum process at 134 °C/2 bar; holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
10.
6.14 Sterilization for the US market ■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization. ■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6, Aesculap recommends the following parameters: Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters* Sterilization method
Temp.
Prevacuum
Time
270 °F/275 °F
4 min
Designation
GB791R
Hi-Line XXS Handpiece shaft, straight II
GB793R
Hi-Line XXS Handpiece shaft, straight III
GB792R
Hi-Line XXS Handpiece shaft, curved II
20 min
GB794R
Hi-Line XXS Handpiece shaft, curved III
GE802SU–GE818SU GE832SU–GE848SU
Hi-Line XXS Tools (single-use)
GB600
Aesculap maintenance oil
GB600820
STERILIT Power Systems adapter
GB459R
Eccos storage mounting set
6.15 Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area. ► Store sterile packed single-use products dust-protected in a dry, dark and temperature-controlled room.
Maintenance
To ensure reliable operation of the handpieces, Aesculap recommends having the product serviced at least once a year. The shafts are not serviceable. For technical service, please contact your national B. Braun/Aesculap agency, see Technical Service.
8.
Art. no.
Minimum drying time
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc.
7.
Accessories/Spare parts
GB605R
Eccos storage for handpiece
GB606R
Eccos storage for handpiece shafts
GB696R
Shaft rinsing device
GB796SU
Spray nozzle II
GB797SU
Spray nozzle III
Troubleshooting list 11.
Malfunction
Detection
Cause
Remedy
Handpiece cannot be coupled to the motor
–
Threaded ring not in ”unlocked” position
Turn the threaded ring in the direction of the arrow in symbol
Handpiece shaft cannot be coupled to handpiece
Symbol
not visible
-
Designation
Class
GB790R
Hi-Line XXS Handpiece
II a
Put shaft locking lever in open position
GB791R
Hi-Line XXS Handpiece shaft, straight II
II a
Shaft coupling damaged or deformed
Use a new handpiece shaft
GB792R
Hi-Line XXS Handpiece shaft, curved II
II a
GB793R
Hi-Line XXS Handpiece shaft, straight III
II a
GB794R
Hi-Line XXS Handpiece shaft, curved III
II a
Automatic locking mechanism defective
Couple the tool with the tool unlocking lever actuated
Tool cannot be uncoupled
Tool cannot be uncoupled even with the tool unlocking lever fully actuated
Tool coupling is worn
Have product repaired by the manufacturer
Handpiece shaft cannot be coupled even with shaft unlocking lever fully open
Tool coupling is worn
A tool is coupled to it
Handpiece shaft cannot be dismantled when a tool is coupled to it
Uncouple the tool
Handpiece blocked
Motor runs with handpiece decoupled
Have product repaired by the manufacturer
Motor locked
Motor safety setting is in ”0” position
Turn the safety setting on the motor to the ”I” position
During normal operation loud running noise
Gear or bearing worn
Have handpiece repaired by manufacturer
Bearings are worn
Use a new handpiece shaft
Gear or bearing worn
Have handpiece repaired by manufacturer
Bearings are worn
Use a new handpiece shaft
Loud running noise
Art. no.
Shaft locking lever not in open position
Pressure point cannot be passed
Handpiece not running
Classification acc. to Directive 93/42/EEC
until it stops
Tool cannot be coupled to handpiece
Handpiece shaft cannot be coupled
Technical data
Have product repaired by the manufacturer
Performance Data Art. no.
Weight
Dimensions (mm)
GB790R
72 g
65 mm x 17 mm x 22 mm
GB791R
10 g
6 mm x 109 mm
GB792R
10 g
6 mm x 109 mm
GB793R
12 g
6 mm x 139 mm
GB794R
12 g
6 mm x 139 mm
Ambient conditions Operation
Storage and transport
Temperature
Relative humidity Handpiece/shaft heats up too much
9.
Significant heat-up after short period of operation
Technical Service
Atmospheric pressure
Risk of injury and/or malfunction! ► Do not modify the product. WARNING ► For service and repairs, please contact your national B. Braun/Aesculapp agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95 -1601 Fax: +49 (7461) 14 -939 E-Mail: [email protected] Or in the US: Aesculap Inc. Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 USA Aesculap Repair Hotline Phone: +1 (800) 214 -3392 Fax: +1 (314) 895 -4420 Other service addresses can be obtained from the address indicated above.
12.
Disposal
Note The user institution is obliged to process the product before its disposal, see Validated reprocessing procedure. ► Detailed information concerning the disposal of the product is available through your national B. Braun/Aesculap
agency, see Technical Service.
13.
Distributor in the US/Contact in Canada for product information and complaints
Aesculap Inc. 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 013837
10/13
V6