BBraun
macro-Line drill handpiece GB102R
40 Pages
Preview
Page 1
Aesculap®
Aesculap Power Systems
Instructions for use/Technical description macro-Line drill handpiece GB102R Gebrauchsanweisung/Technische Beschreibung macro-Line Bohrhandstück GB102R Mode d’emploi/Description technique Pièce à main de forage macro-Line GB102R Instrucciones de manejo/Descripción técnica Pieza de mano macro-Line GB102R Istruzioni per l’uso/Descrizione tecnica Manipolo perforatore macro-Line GB102R Instruções de utilização/Descrição técnica Cabo porta-brocas macro-Line GB102R Gebruiksaanwijzing/Technische beschrijving macro-Line boorhandstuk GB102R Bruksanvisning/Teknisk beskrivning macro-Line borrhandstycke GB102R Инструкция по примению/Техническое описание Наконечник дрели macro-Line GB102R Návod k použití/Technický popis Vrtací násadec macro-Line GB102R Instrukcja użytkowania/Opis techniczny Uchwyt wiertła macro-Line GB102R Návod na použivanie/Technický opis Vŕtacia rukoväť macro-Line GB102R Kullanım Kılavuzu/Teknik açiklama macro-Line delme el parçası GB102R
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap – a B. Braun company TA-Nr. 005333
01/14
- DIR 93/42/EEC Technical alterations reserved
V6
Änd.-Nr. 48673
11
12
1 2
3
6 7
4
GB103R
OPEN
5
8 CLOSE
10
GB105R
9
13 15
1
16
GB106R 17
GA031 14 GC523200
2
3 GB102
4. Aesculap® macro-Line drill handpiece GB102R Legend 1 Threaded stud 2 Wrenching flats 3 Clamping lever 4 Coupling (for motor unit) 5 Centering lug 6 Face plate (motor unit) 7 Nib 8 Lubrication point (for clamping lever) 9 Fixing screw 10 Toothing 11 Sleeve 12 Ring 13 Tightening key 14 Flat wrench 15 Sleeve 16 Wrenching flats 17 HUDSON chuck adapter
Symbols on product and packages
Preparation
Non-compliance with the following instructions will preclude all responsibility and liability in this respect on the part of Aesculap. ► Before using the macro-Line drill handpiece and its accessories, visually check them to be sure they are not damaged. ► Use the macro-Line drill handpiece and its accessories only if they are in perfect condition.
5.
Working with the macro-Line drill handpiece GB102R
5.1
System set-up Damage to the handpiece due to incorrect handling or operation! ► Couple the handpiece only while the motor is inactive. CAUTION
Note Products marked as unsterile must be sterilized prior to use, according to the respective instructions, see Steam sterilization.
Connecting the accessories Combinations of accessories that are not mentioned in the present instructions for use may only be employed if they are specifically intended for the respective application, and if they do not compromise the performance and safety characteristics of the products. ► Please address your B. Braun/Aesculap partner or Aesculap Technical Service with any inquiries in this respect; for a contact address, see Technical Service.
Connecting the macro-Line drill handpiece to the motor unit
Caution, general warning symbol Caution, see documentation supplied with the product
Damage to the drill handpiece due to incorrect handling or operation! ► Couple the drill handpiece only while the motor is inactive. CAUTION
1.
Applicable to
► For item-specific instructions for use and information on material compatibility, see also the Aesculap Extranet
at www.extranet.bbraun.com
2.
Safe handling
► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial
sterilization. ► Prior to use, check that the product is in good working order. ► To prevent damage caused by improper setup or operation, and in order not to compromise warranty and manufacturer liability: – Use the product only according to these instructions for use. – Follow the safety information and maintenance instructions. – Only combine Aesculap products with each other. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Keep the instructions for use accessible for the user.
3.
Product description
3.1
Scope of supply
Art. no.
Designation
GB102R
macro-Line drill handpiece
GC523200
Flat wrench
3.2
Components required for operation
► Release clamping lever 3 towards the OPEN position. ► Push the handpiece onto centering lug 5 until the handpiece touches face plate 6 and nib 7 engages in the
recess on the handpiece. ► Tighten clamping lever 3 towards the CLOSE position.
Moving the clamping lever to be parallel with the motor unit ► Tighten clamping lever 3 hand-tight. ► Loosen fixing screw 9 with hexagon wrench TA004801 so that the toothing 10 disengages. ► Move clamping lever 3 to the intended position and engage toothing 10. ► Tighten fixing screw 9 with hexagon wrench TA004801.
Uncoupling the macro-Line drill handpiece from the motor unit ► Release clamping lever 3 towards the OPEN position. ► Remove the handpiece.
Mounting chuck adapter GB103R/GB105R/GB106R The chuck adapters GB103R/GB105R can be used to clamp any smooth or trihedral tool shafts. GB103R ► Turn the chuck adapter onto threaded stud 1, hold ring 12, and turn sleeve 11. ► To fasten the chuck adapter: Apply flat wrench 14 at wrenching flats 2 and tighten clockwise. GB105R ► Turn the chuck adapter onto threaded stud 1. ► Apply and hold flat wrench 14 at wrenching flats 2. ► To fasten the chuck adapter: Turn the chuck adapter clockwise with flat wrench 14. GB106R ► Turn the HUDSON chuck adapter 17 onto threaded stud 1. ► To fasten the HUDSON chuck adapter 17: Apply and hold flat wrench 14 at wrenching flats 2. ► Apply flat wrench 14 at wrenching flats 16 and tighten clockwise.
One of the following components is necessary for operation:
Dismounting chuck adapter GB103R/GB105R/GB106R
Art. no.
Designation
GA156
Flexible shaft (1.6 m) with Elan EC and foot control
GB103R ► Apply and hold flat wrench 14 at wrenching flats 2. ► Open the chuck adapter completely and continue turning it until it becomes detached from threaded stud 1. ► Remove the chuck adapter from threaded stud 1.
GA172
Flexible shaft (2.1 m) with Elan EC and foot control
GA200
AIRLAN pneumatic motor
The following attachments can be coupled to the handpiece:
GB105R ► Apply and hold flat wrench 14 at wrenching flats 2. ► To loosen the chuck adapter: Turn the chuck adapter counterclockwise with tightening key 13. ► Remove the chuck adapter from threaded stud 1. GB106R ► To loosen the HUDSON chuck adapter 17: Apply and hold flat wrench 14 at wrenching flats 2. ► Apply flat wrench 14 at wrenching flats 16 and turn counterclockwise. ► Remove the HUDSON chuck adapter 17 from threaded stud 1.
Art. no.
Designation
GB103R
Keyless chuck adapter with 0.8 mm to 6.35 mm tightening range
5.2
GB105R
Chuck adapter with key with 0 mm to 6.35 mm tightening range
Note The function checks must be carried out prior to every surgical application of the product!
GB106R
HUDSON chuck adapter with 0 mm to 6.35 mm tightening range
Test run ► Check that the macro-Line drill handpiece is seated properly. ► Briefly run the macro-Line drill handpiece at maximum speed. Check for proper and reliable functioning.
3.3
Intended use
The macro-Line drill handpiece is used for drilling and reaming in osteosynthesis procedures and for trepanning in neurosurgery.
3.4
Operating principle
The macro-Line drill handpiece reduces the drive speed by the ratio 10:1 (e. g, motor speed 15 000 rpm, drill speed 1 500 rpm).
Function checks
5.3
Safe operation Risk of burns to patient and user due to hot motor/hot tool! ► Put down the motor/hot tool beyond reach of the patient. ► Use a cloth to protect against burns when changing the motor or tool.
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – be approved for plastic material and high-grade steel, – do not attack softeners (e.g. in silicone). ► Observe specifications regarding concentration, temperature and exposure time.
CAUTION
WARNING
WARNING
Risk of injury and material damage due to inappropriate use of handpieces or motors! ► Always follow the safety advice and information given in the instructions for use of the handpieces and motors. ► Do not exceed the specified maximum motor speed. Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product.
► Dry the product for at least 10 minutes at a maximum of 120 °C.
Note The indicated drying temperature is a guide temperature only. It must be checked taking into account the specific conditions (e.g. load) and adjusted if applicable. ► Do not clean motors/handpieces by ultrasonic treatment and do not immerse them in any liquids. To avoid the
risk of corrosion/malfunctioning, allow any fluid that has entered the product to drain out immediately.
Validated cleaning and disinfection procedure
WARNING Risk of injury when using the product beyond the field of view! ► Apply the product only under visual control. WARNING
Validated procedure
Special features
Reference
Manual cleaning and wipe disinfection
■ When cleaning instruments with movable
Chapter Manual cleaning/disinfecting and subsection:
hinges, ensure that these are in an open position and, if applicable, move the joint while cleaning.
■ Chapter Manual cleaning and wipe disinfecting
■ Drying phase: Use a lint-free cloth or com-
GB106R
pressed air for medical purposes
► To couple and uncouple the HUDSON chuck adapter 17: Pull sleeve 15 forward.
6.
Validated reprocessing procedure
6.1
General safety instructions
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing.
■ Ensure that the product is positioned in such a way that water will not enter the product e.g. through coupling interfaces. (Immediately remove any fluid that entered the product inadvertently.)
6.7
Manual cleaning/disinfecting
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
Manual cleaning and wipe disinfecting
Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at www.extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
6.2
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Cleaning
RT (cold)
-
-
D–W
-
II
Drying
RT
-
-
-
-
III
Wipe disinfection
-
>1
-
-
Meliseptol HBV wipes 50 % Propan-1-ol
IV
Final rinse
RT (cold)
0.5
-
FD-W
-
V
Drying
RT
-
-
-
-
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
6.3
infecting solution.
Dismantling prior to carrying out the reprocessing procedure
D–W: FD–W: RT:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) Room temperature
Phase I ► Clean the product under running faucet water, using a suitable cleaning brush until all visible residues have been removed from the surfaces. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. Phase II ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure. Phase III ► Wipe all surfaces of the product with a single-use disinfectant wipe. Phase IV ► After the specified exposure time (at least 1 min), rinse the disinfected surfaces under running FD water. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
► Disassemble the product immediately after use, as described in the respective instructions for use.
6.8 6.4
Preparations at the place of use
► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
6.5
Preparation before cleaning
► Carry out non-fixating/NaCl-free pre-cleaning immediately after use.
6.6
Inspection, maintenance and checks
► Allow the product to cool down to room temperature. ► Inspect the product after each cleaning and disinfecting cycle to be sure it is: clean, functional, and undamaged. ► After every cleaning and disinfecting cycle, lubricate the product at threaded stud 1, the lubrication point (for
clamping lever) 8, and the sliding points of the chuck adapter with one drop of STERILIT Power Systems oil from drip lubricator GA059. ► Check the product for any damage, abnormal running noise, overheating or excessive vibration. ► Inspect tools for broken, damaged or blunt edges. ► Set aside the product if it is damaged.
Cleaning/disinfection 6.9
Product-specific safety instructions for the reprocessing procedure Damage to, or destruction of the product caused by mechanical cleaning/disinfection! ► Only clean and disinfect the product manually. CAUTION Damage to the product due to inappropriate cleaning/disinfecting agents! ► Only use cleaning/disinfecting agents approved for surface cleaning. Follow the manufacturer’s instructions for the respective cleaning/disinfecting agent. CAUTION
Packaging
► Always observe the instructions for use of packaging and storage devices. ► Place the product on a tray in such a way that it is protected against damage. ► Pack trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging will prevent a recontamination of the product.
6.10 Steam sterilization
12.
► Prior to sterilization, remove any tools and adapters and uncouple the handpiece from the motor unit. ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
Note The user institution is obliged to process the product before its disposal, see Validated reprocessing procedure.
opening any valves and faucets).
► Detailed information concerning the disposal of the product is available through your national B. Braun/Aesculap
► Validated sterilization process
– Disassemble the instrument – Steam sterilization through fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
6.11 Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
7.
Maintenance
To ensure reliable operation, the product must be maintained at least once a year. For technical service, please contact your national B. Braun/Aesculap agency, see Technical Service.
8.
Troubleshooting list
Malfunction
Detection
Cause
Remedy
Intense/loud running noise
Motor unit can be pulled out from the handpiece
Motor unit not coupled correctly
Couple the motor unit correctly, see Connecting the macro-Line drill handpiece to the motor unit
Handpiece hot
Gearing worn
Have repaired by manufacturer
Lack of power at handpiece
Intense/loud running noise
Gearing worn
Have repaired by manufacturer
Inadequate power at handpiece
Intense/loud running noise
Blunt tool
Replace the tool
-
Motor unit not running correctly
See instructions for use of the motor unit
Coupling parts broken
Coupling defective
Have repaired by manufacturer
Motor unit running, but not powering the handpiece
Defective gearing
Have repaired by manufacturer
Tool not moving
9.
Technical Service Risk of injury and/or malfunction! ► Do not modify the product. WARNING
► For service and repairs, please contact your national B. Braun/Aesculapp agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95 -1601 Fax: +49 (7461) 14 -939 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
10.
Accessories/Spare parts
Art. no.
Designation
GA031R
Tightening key
GA059
STERILIT Power Systems drip lubricator
GB032207
Fixing screw
GB103R
Keyless universal chuck (0.8–6.35 mm)
GB105R
Universal chuck, incl. GA031R (0–6.35 mm)
GB106R
HUDSON chuck adapter
GC523200
Flat wrench
TA004801
Hexagon wrench
11.
Technical data
Classification acc. to Directive 93/42/EEC Art. no.
Designation
Class
GB102R
macro-Line drill handpiece GB102R
IIa
Designation
Disposal
Value
Size
∅ 40 mm
Length
125 mm
Length
approx. 320 g
Maximum motor unit speed
20 000 rpm
Gear ratio
10:1
agency, see Technical Service. TA-Nr. 005333
01/14
V6
Änd.-Nr. 48673