■ Aesculap is not responsible for any complications arising from wrong indication, wrong choice of implant, incorrect combination of implant components and operating technique, the limitations of the treatment method, or inadequate asepsis.
®
Aesculap MACS TL® MACS TL® polyaxial screw XL Intended use The MACS TL® polyaxial screw XL is used in combination with the other MACS TL® implants, the MACS TL® system and the MACS II instrument set with pre-assembled retaining screw SX804T for the ventral monosegmental and multisegmental stabilization of the lumbar and thoracic spine. Note Use pre-assembled retaining screw SX804T only with the MACS II instrument set, see instructions for use TA013373.
■ The instructions for use for individual Aesculap implant components must be followed. ■ The implant components were tested and approved in combination with Aesculap components. If other combinations are used, the responsibility for such action lies with the operating surgeon.
■ Do not, under any circumstances, combine implant components from different manufacturers. ■ Do not, under any circumstances, use damaged or surgically removed components. ■ Implants that have been used before must not be reused. ■ Delayed healing can cause implant breakage due to material fatigue. ■ The implant components applied, along with their article numbers, the name of the implant, as well as the batch number and serial number (if available) must be documented in all patient records.
■ Postoperatively, individual patient information, as well as mobility and muscle training, is of particular importance.
■ Damage to load-bearing bone structures can result in loosening of the components, bone or implant fractures
The MACS TL® system consists of: ■ Anchoring screws for the vertebral bodies ■ Tensioning elements ■ Fixation nuts and clamping screws ■ Connecting plates and, if required, connecting rods of the Aesculap Spine System Evolution, diameter 5.2 mm, as connection elements
or other acute complications. In order to detect any such problems as soon as possible, the condition of the implant must be followed up at regular intervals postoperatively with the aid of appropriate examination procedures. ■ The surgeon in charge of the therapy shall make any decision with regard to the removal of implant components that have been used, taking into account the potential risks of another operation and the difficulties involved in the removal of an implant.
Materials
Sterility
The materials used in the implant are listed on the packaging: ■ ISOTAN®F titanium forged alloy Ti6Al4V acc. to ISO 5832-3 MACS TL® and ISOTAN® are registered trademarks of Aesculap AG, 78532 Tuttlingen / Germany.
System configurations The MACS TL® system is available as a polyaxial twin-screw system with plate or 2 rods, see Fig.
■ The implant components are supplied in an unsterile condition. ■ The implant components are packaged individually. ► Store the implant components in their original packaging and only remove them from their original and protec-
tive packaging immediately prior to application. ► Use the implant system storage devices for processing, sterilization and sterile setup. ► Ensure that the implant components in their implant system storage devices do not come into contact with each
other or with instruments. ► Ensure that the implant components are not damaged in any way.
Prior to initial sterilization and subsequent resterilization, the implant components must be cleaned using the following validated reprocessing procedure:
Indications Surgically installed implants serve to support normal healing processes. They are not intended for use either as replacements for natural body parts or to bear loads over the long term if healing does not occur. Use for: ■ Fractures ■ Spinal tumor ■ Degenerative disc disorder (spondylolisthesis, spondylolysis, spinal stenosis) ■ Post-trauma instability
Contraindications Do not use in the presence of: ■ Fever ■ Acute or chronic vertebral infections of a local or systemic nature ■ Pregnancy ■ Severe osteoporosis or osteopenia ■ Medical or surgical conditions that could negatively affect the outcome of the implantation ■ Severely damaged bone structures that could prevent stable implantation of implant components ■ Bone tumors in the region of implant fixation ■ Wound healing disorders ■ Inadequate patient compliance ■ Foreign body sensitivity to the implant materials ■ Cases not listed under indications Note In case of osteopenia, the surgeon must weigh up the risks that could jeopardise the therapeutic success of the MACS TL® system. If it is to be expected that the standard screw shall not provide sufficient stability, the polyaxial screw XL can be used.
Side effects and interactions The application or improper use of this system entails the following risks: ■ Implant failure resulting from excessive load – Warping or bending – Loosening – Breakage ■ Inadequate fixation ■ Failed or delayed fusion ■ Infection ■ Fractured vertebral body or bodies ■ Injuries to – Nerve roots – Spinal cord – Blood vessels – Organs
Safety notes
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation. Note For up-to-date information about reprocessing and material compatibility, see also the Aesculap Extranet at https://extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
Validated reprocessing procedure Mechanical alkaline cleaning and thermal disinfection ► Process the implant in its system storage device.
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical/Note
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfection
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline For implant components that are to be resterilized:
CAUTION Federal law restricts this device to sale by, or on order of a physician! This system is not licensed to be screwed onto, or fastened to, the posterior elements (pedicles) of the cervical, thoracic or lumbar spine. WARNING The implant has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration or image artifact in the MR environment. Scanning a patient who has this implant may result in patient injury. WARNING
■ It is the operating surgeon's responsibility to ensure that the surgical procedure is performed properly. ■ General risk factors associated with surgical procedures are not described in in the present instructions for use. ■ The operating surgeon must have a thorough command of both the hands-on and conceptual aspects of the established operating techniques.
■ The operating surgeon must be fully conversant with bone anatomy, including the pathways of nerves, blood vessels, muscles, and tendons. ■ It is the operating surgeon's responsibility to ensure the correct combination of implant components and their implantation.
WARNING
Intraoperative contamination with blood, secretions, and other fluids may render the affected component unsuitable for resterilization! ► Handle the implants with new gloves only. ► Keep the implant system storage devices covered or closed. ► Process implant system storage devices separately from instrument trays. ► Clean implants must not be processed together with contaminated implants. ► Process the implant components individually and separately if no implant system storage devices are available, ensuring that the implant components are not damaged in the process. ► Mechanically clean and disinfect the implant components. ► Do not reuse surgically contaminated implants! Direct or indirect contamination may render implants unsuitable for resterilization! ► Do not reprocess implants that have been directly or indirectly contaminated with blood.
Inspection, maintenance and checks
Disposal
► Allow the product to cool down to room temperature. ► Inspect the product after each cleaning and disinfecting cycle to be sure it is: clean, functional, and undamaged. ► Immediately set aside damaged or inoperative products.
► Adhere to national regulations when disposing of or recycling the product, its components and its packaging!
Packaging ► Place the product in its holder or on a suitable tray. ► Pack trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Make sure that the packaging will prevent a recontamination of the product during storage.
Sterilization – Steam sterilization through fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Application The MACS TL® system is described in detail in the O.R. manual. Detailed user instructions can be found in the product documentation. The O.R. manual can be ordered from the manufacturers at any time. The operating surgeon shall devise an operation plan that specifies and accurately documents the following: ■ Selection of the implant components and their dimensions ■ Positioning of the implant components in the bone ■ Location of intraoperative landmarks The following conditions must be fulfilled prior to application: ■ All requisite implant components are ready to hand. ■ Operating conditions are highly aseptic. ■ All requisite implantation instruments must be available and in working order, including specialized Aesculap implantation systems. ■ The operating surgeon and operating room team are thoroughly familiar with the operating technique and with the available range of implants and instruments; information materials on these subjects must be complete and ready to hand. ■ The operating surgeon is fully conversant with the rules governing medical practice, the current state of scientific knowledge, and the contents of relevant scientific articles by medical authors. ■ Information has been obtained from the manufacturer in the event of an ambiguous preoperative situation and if implants are present in the area to be treated Caution: For a secure implantation of the anchoring screws, careful measurement of the vertebral body's diameter at the planned screw position and direction is crucial. A screw length smaller than the measured diameter has to be chosen. The surgical procedure and following information has been explained to the patient, and the patient's consent has been documented: ■ In the case of delayed or incomplete fusion, the implants can break and loosen due to high loads. ■ The life-span of the implant depends on the patient’s body weight. ■ The implant components must not be overloaded by extreme strains, hard physical labor or sports. ■ Corrective surgery may be necessitated by implant loosening, fracture or loss of correction. ■ Smokers present an increased risk of bone fusion failure. ■ The patient must undergo medical check-ups of the implant components at regular intervals. The implantation of the MACS TL® system requires the following application steps: ► Select the suitable MACS TL® system version and anchoring screws on the basis of the indication, the preoperative planning and the bone situation found during the operation. ► Determine the optimum implant position of the MACS TL® system: – Preposition the implant with Kirschner wires and – Prepare the bone correctly, using cannulated instruments. ► Apply the implantation instruments in the correct way. ► To avoid internal stress on, and weakening of, the implant: avoid scoring or scratching of the implant components. ► Do not bend or warp any metal implants with the exception of the rods of the Aesculap Spine System Evolution. ► Do not bend or re-bend the rods excessively. ► Use only the bending instruments from the MACS TL® instrument set to bend the rods. ► Mount the centering sleeve on the tensioner element, using the special mounting instrument. ► Insert the implant components using the special insertion instruments. ► Do not tighten or screw in the polyaxial screw XL completely, so that the tensioner element can still be moved. ► Prior to unlocking the insertion instrument, remove the Kirschner wire through the cannulated insertion element, using the appropriate removal instrument. ► Remove the insertion instrument. ► Ensure that the centering sleeves remain mounted, temporarily, on the tensioner elements. ► Distract and insert placeholder, if necessary. Risk of injury caused by screws loosening or tearing out during reposition, distraction or compression, especially in cases of poor bone quality! ► Apply force carefully and well-dosed. ► Maintain permanent visual control (direct or through imaging techniques).
► Insert stabilizing plate or rods.
Implant stabilizing plate and locknut in such a way that the inscribed side (top side) is visible for the user. Injury can result from nuts with insufficient clamping stability. ► Set nuts in place correctly. ► Using a torque wrench, tighten the nuts to a torque of 15 Nm. WARNING ► Remove centering devices. ► Tighten the polyaxial screws XL evenly. ► Insert the stabilizing screw through the appropriate aiming sleeve. ► Comply absolutely with the safety instructions of the Aesculap cement application system. ► Insert clamping screw.
WARNING
Aesculap Implant Systems LLC 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 011153
► Validated sterilization process
WARNING
Distributor in the US/Contact in Canada for product information and complaints
Risk of injury caused by insufficiently tightened fixing screws of the polyaxial elements! ► Set clamping screws in place correctly. ► Using a torque wrench, tighten the fixing screws to a torque of 10 Nm.
Further information on Aesculap implant systems is always available from B. Braun/Aesculap or the appropriate B. Braun/Aesculap office.