BBraun
MACS TL thoracic screwdriver
27 Pages
Preview
Page 1
Validated reprocessing procedure Aesculap® MACS TL® MACS TL® thoracic screwdriver
General safety instructions
Legend
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
1 Screw sleeve 2 Retaining tongues 3 Screw the knurled nut
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
Explanation
Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation.
Caution, general warning symbol Caution, see documentation supplied with the product
Note If there is no final sterilization, then a virucidal disinfectant must be used.
Symbols on product and packages Symbol
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing.
Note For up-to-date information about reprocessing and material compatibility, see also the Aesculap Extranet at https://extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
Applicable to ► For item-specific instructions for use and information on material compatibility, see also the Aesculap Extranet
at https://extranet.bbraun.com
Intended use The MACS TL® thoracic screwdriver is used to implant the MACS TL® thoracic stabilization screw.
Safe handling and preparation CAUTION Federal law restricts this device to sale by, or on order of a physician! Applying the MACS TL® thoracic screwdriver requires precise knowledge about spine stabilization and the biomechanical situation at the spine. Operative use of the MACS TL® instrument set is described in detail in the relevant O.R. manual. ► Prior to first application, thoroughly read the O.R. manual for the MACS TL® thoracic screwdriver. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts. ► To avoid damage to the working end: Carefully insert the product through the working channel (e.g. trocar).
General information Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
Disassembling the product before carrying out the reprocessing procedure ► Disassemble the product immediately after use, as described in the respective instructions for use.
Preparations at the place of use
Safe operation Mispositioning of the screw can cause extremely severe complications for the patient! ► Always maintain radiographic control through all steps of the operation. DANGER
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
Preparation before cleaning ► Dismantle the product prior to cleaning, see Disassembling.
Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product.
Cleaning/disinfection Product-specific safety notes on the reprocessing procedure
WARNING To insert the screw: ► Insert the tip of the thoracic screwdriver MACS TL® into the hexagonal socket of the screw. ► Turn knurled nut3 clockwise to fixate the screw on the MACS TL® thoracic screwdriver. ► Screw the screw into the bone through the guide sleeve using the MACS TL® thoracic screwdriver. ► Turn knurled nut 3 counterclockwise to loosen the screw from the MACS TL® thoracic screwdriver. ► Pull the thoracic screwdriver MACS TL® out.
CAUTION
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – are approved for (e.g. aluminum, plastics, high-grade steel), – do not attack softeners (e.g. in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 60 °C.
Disassembling The MACS TL® thoracic screwdriver will not function properly if the retaining tongues are bent or kinked! ► Do not bend or crimp the retaining tongues. CAUTION
► Do not use oxidizing chemicals (e.g. H2O2), which could cause bleaching/layer loss of the product. ► Use suitable cleaning/disinfecting agents if the product is put away in a wet condition. To prevent foam forma-
tion and reduced effectiveness of the process chemicals: Prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water. ► Carry out ultrasound cleaning:
► Unscrew screw sleeve 1. ► Unscrew knurled nut 3 and push it, together with the retaining tongues 2, over the shaft of the MACS TL® tho-
racic screwdriver.
Assembling
– as an effective mechanical supplement to manual cleaning/disinfecting. – as a pre-cleaning procedure for products with encrusted residues, in preparation for mechanical cleaning/disinfecting. – as an integrated mechanical support measure for mechanical cleaning/disinfecting. – for additional cleaning of products with residues left after mechanical cleaning/disinfecting.
Validated cleaning and disinfection procedure The MACS TL® thoracic screwdriver will not function properly if the retaining tongues are bent or kinked! ► Do not bend or crimp the retaining tongues.
CAUTION ► Push knurling nut 3 with retaining tongues 2 on to the shaft of the MACS TL® thoracic screwdriver in such a way
that the retaining tongues 2 are guided in the grooves of the shaft. ► Turn knurled nut 3 onto the thread, but do not tighten it. ► Screw the screw sleeve 1 onto the tip of the MACS TL® thoracic screwdriver and tighten it down.
Validated procedure
Specific requirements
Reference
Manual pre-cleaning with brush and subsequent mechanical alkaline cleaning and thermal disinfection
■ Suitable cleaning brush ■ 20 ml disposable syringe ■ Place the product in a tray that
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and sub-chapter:
is suitable for cleaning (avoiding rinsing blind spots).
■ Keep working ends open for cleaning.
■ Chapter Manual pre-cleaning with a brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
Mechanical cleaning/disinfection with manual pre-cleaning
Steam sterilization
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
Note The product may only be sterilized when dismantled.
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Manual pre-cleaning with a brush Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfectant cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
Rinsing
RT (cold)
1
II
-
D–W
-
► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets). ► Validated sterilization process
– Disassemble the product – Steam sterilization through fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several products at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
Sterilization for the US market ■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization. ■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6 , Aesculap recommends the following parameters:
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed
Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters*
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Sterilization method
Temp.
Time
Minimum drying time
Prevacuum
270 °F/275 °F
4 min
20 min
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc. WARNING for the US market If this device is/was used in a patient with, or suspected of having Creutzfeldt-Jakob Disease (CJD), the device cannot be reused and must be destroyed due to the inability to reprocess or sterilize to eliminate the risk of crosscontamination.
Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Mechanical alkaline cleaning and thermal disinfecting Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
I
Prerinse
<25/77
II
Cleaning
55/131
t [min]
Water quality
Chemical
3
D–W
-
10
FD-W
■ Concentrate, alkaline:
Risk of injury and/or malfunction! ► Do not modify the product. WARNING
– pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Inspection, maintenance and checks
CAUTION
Technical Service
Damage (metal seizure/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598).
► Allow the product to cool down to room temperature. ► After each complete cleaning, disinfecting and drying cycle, check that the product is dry, clean, operational, and
free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components). ► Dry the product if it is wet or damp. ► Repeat cleaning and disinfection of products that still show impurities or contamination. ► Check that the product functions correctly. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical Service. ► Assemble dismountable products, see Assembling. ► Check for compatibility with associated products.
Packaging ► Appropriately protect products with fine working tips. ► Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected. ► Pack trays appropriately for the intended sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
► For service and repairs, please contact your national B. Braun/Aesculap agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1602 Fax: +49 (7461) 16-5621 E-Mail: [email protected] Or in the US: Aesculap Implant Systems LLC Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 Aesculap Repair Hotline Phone: +1 (800) 214-3392 Fax: +1 (314) 895-4420 Other service addresses can be obtained from the address indicated above.
Accessories/Spare parts Additional instruments are required for implanting the MACSTL implants. The complete instrument set is listed in the product documentation.
Distributor in the US/Contact in Canada for product information and complaints Aesculap Implant Systems LLC 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 011177
2017-12
V6
Änd.-Nr. 57993