BBraun
Modular monopolar instrument system
41 Pages
Preview
Page 1
Aesculap®
A
Aesculap Surgical Instruments
Instructions for use/Technical description Modular monopolar instrument system Instructions for use/Technical description Modular monopolar instrument system Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapusa.com and clicking the "Products" menu. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-800-282-9000. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung Modulares monopolares Instrumentensystem Mode d’emploi/Description technique Système modulaire d’instruments monopolaires Instrucciones de manejo/Descripción técnica Sistema de instrumentos modular monopolar Istruzioni per l’uso/Descrizione tecnica Sistema di strumenti monopolari modulari Instruções de utilização/Descrição técnica Sistema de instrumentos monopolares modulares Gebruiksaanwijzing/Technische beschrijving Modulair monopolair instrumentensysteem Bruksanvisning/Teknisk beskrivning Modulärt monopolärt instrumentsystem Инструкция по примению/Техническое описание Модульная монополярная система инструментов Návod k použití/Technický popis Systém modulárních monopolárních nástrojů Instrukcja użytkowania/Opis techniczny Modułowy, jednobiegunowy system instrumentów Návod na použivanie/Technický opis Modulárny monopolárny nástrojový systém Kullanım Kılavuzu/Teknik açiklama Modüler monopolar ekipman sistemi 사용 설명서 / 기술 설명 재사용가능발조절식전기수술기용전극 (RU Monopolar, 형명개별기재 )
A1
6
A2 B
6
B1
B2 6
7
8
6
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com
1
7
10
8
9 3
2
Aesculap – a B. Braun company TA-Nr. 011945
07/16
V6
Änd.-Nr. 54983
4 5 - DIR 93/42/EEC
I
II
Handle with lock
Aesculap® Modular monopolar instrument system Legend 1 Handle (complete) 2 HF pin 3 Star wheel 4 Actuating lever 5 Switch (lubrication point) 6 Jaw insert (lubrication point) 7 Inner tube 8 Outer tube 9 Leaf springs 10 Nib
Symbols on product and packages Caution, general warning symbol Caution, see documentation supplied with the product
► Press actuating lever 4 to release the ratchet mechanism. ► To activate the ratchet mechanism, release actuating lever 4.
Switch 5 can be used for permanently deactivating the ratchet mechanism. ► To deactivate the ratchet mechanism on the handle: position switch 5 as shown in I. The switch clicks into the front position. ► To activate the ratchet mechanism on the handle: position switch 5 as shown in II. The switch clicks into the back position.
Disassembling Note For disassembling the product, also see additional series of illustrations A. ► Deactivate the ratchet mechanism on the handle with lock: position switch 5 as shown in I. ► Turn star wheel 3 proximally until it clicks into place. A red colored ring becomes visible. ► Detach the shaft from the handle: Hold jaw insert 6 at its tip and extract it from handle 1, together with inner
tube 7 and outer tube 8. Do not hold the moveable part of the handle. The moveable part of the handle will move up. ► Disassemble the shaft: Remove outer tube 8 and inner tube 7 from jaw insert 6.
Assembling Note For assembling the product, also see additional series of illustrations B. Star wheel 3 must be arrested in such a way that the red colored ring is visible. ► Deactivate the ratchet mechanism on the handle with lock: position switch 5 as shown in I. ► Assemble the shaft:
Intended use All professional disciplines that use endoscopy: ■ Cutting, preparation, and grasping of tissues ■ Biopsies ■ Suturing
Safe handling and preparation CAUTION Federal law restricts this device to sale by, or on order of a physician! ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts. ► To avoid damage to the working end: Carefully insert the product through the working channel (e.g. trocar). Risk of injury from ignition or explosion of flammable gases! Sparks may occur when using the HF device as directed. ► Observe the safety guidelines in the instructions for use of the HF device. WARNING
WARNING
Thermal injuries to patients/users due to insufficient insulation of leads in active accessories! ► Adjust the HF device to an appropriate setting to ensure that the peak output voltage does match or not exceed the accessory voltage rating specified for the product.
The manufacturer has tested the product and verified that its insulation can withstand 300 reprocessing cycles. In clinical practice, the service life will depend on the individual intraoperative usage and the hospital's specific reprocessing conditions. ► Adjust the HF power output (and argon flow rate) to the intervention to be carried out. Take into account clinical experience or reference values. ► Select the lowest possible HF power output. ► Keep the product’s contact surfaces clean during surgery. Remove encrusted tissue residues or body fluids with a moistened swab. The product is furnished with the following connection on the plug side: pin, 4 mm, sprung. Refer to our brochures to find a compatible cable. The accessory voltage rating of the product is 2 500 Vp. The accessory voltage rating must exceed or match the peak output voltage with which the product is operated in combination with a suitable HF device at an appropriate operating mode/setting (see IEC/DIN EN 60601-2-2). To avoid HF burns: ► Always keep the working end of the product in the user’s field of vision whenever the HF power is activated. ► Prior to activating the HF device, check that the working end of the product is not touching any electrically conductive accessories. ► Prior to each use, visually inspect the product for: damage or surface changes to the insulation. ► Never place the product on or next to the patient. ► When using accessories for endoscopy or laparoscopy, deactivate the automatic switch-on mode of the HF device. ► Follow the instructions for use of the HF device.
– Push inner tube 7 with the four flexible tongues 9 onto the jaw insert 6 in the direction of the working end as far as it will go, so that the two nibs 10 are guided into the flexible tongues 9. – Slide outer tube 8 over inner tube 7 onto jaw insert 6 as far as it will go. ► Hold the assembled shaft at jaw insert 6 with one hand. ► With the other hand, hold handle 1 behind star wheel 3 at the fixed handle part. The moveable part of the handle must still be freely moveable. ► Push the shaft into the handle. The moveable part of the handle will move down. When the stop is reached, star wheel 3 automatically clicks into place at the distal end. The red colored ring is not visible anymore. ► Check to make sure that the instrument is working properly by opening and closing the jaws.
Validated reprocessing procedure General safety instructions Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see also the Aesculap Extranet at https://extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
General information Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
Safe operation Disassembling the product before carrying out the reprocessing procedure Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product. WARNING Risk of injury when using the product beyond the field of view! ► Apply the product only under visual control. WARNING ► Open and close the distal end of the jaws: open and close the moveable part of the handle.
► Disassemble the product immediately after use, as described in the respective instructions for use. ► Open up instruments with hinges.
Preparations at the place of use
Manual cleaning/disinfection
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
Preparation before cleaning
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning /disinfection process if necessary.
Manual cleaning with ultrasound and immersion disinfection
► Dismantle the product prior to cleaning, see Disassembling.
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
Cleaning/disinfection Product-specific safety notes on the reprocessing procedure
CAUTION
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – are approved for use, for example, on aluminum, plastic materials, and high-grade steel, – do not attack softeners (e.g. in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 93 °C.
Immersion treatment in a 3 % H2O2 solution for approx. 5 minutes is a particularly effective and gentle method to dissolve encrustations from HF instruments. Subsequently, the debris can be removed by hand, with a medium-hard brush and or in an ultrasonic bath. This is followed by the conventional reprocessing steps. ► Use suitable cleaning/disinfecting agents if the product is put away in a wet condition. To prevent foam formation and reduced effectiveness of the process chemicals: Prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water. ► Carry out ultrasound cleaning: – as an effective mechanical supplement to manual cleaning/disinfection. – as a pre-cleaning procedure for products with encrusted residues, in preparation for mechanical cleaning/disinfection. – as an integrated mechanical support measure for mechanical cleaning/disinfection. – for additional cleaning of products with residues left after mechanical cleaning/disinfection.
Validated cleaning and disinfection procedure Validated procedure
Specific requirements
Reference
Manual cleaning with ultrasound and immersion disinfection
■ Cleaning brush, e.g., PM995R
Chapter Manual cleaning/disinfection and sub-chapter:
■ Shaft and handle
■ 20 ml disposable syringe ■ Drying phase: Use a lint-free
or GK469R
cloth or medical compressed air Mechanical alkaline cleaning and thermal disinfection
■ Handle only
■ Place the instrument in a tray that is suitable for cleaning (avoiding rinsing blind spots).
■ Connect components with lumens and channels directly to the rinsing port of the injector carriage.
■ Chapter Manual cleaning with ultrasound and immersion disinfection Chapter Mechanical cleaning/disinfecting and sub-chapter:
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
cleaning.
■ Place instruments in the tray with their hinges open.
■ Shaft only
■ Cleaning brush, e.g., PM995R or GK469R
■ 20 ml disposable syringe ■ Place the instrument in a tray that is suitable for cleaning (avoiding rinsing blind spots).
■ Connect components with lumens and channels directly to the rinsing port of the injector carriage.
■ Keep working ends open for cleaning.
■ Place instruments in the tray with their hinges open.
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully.
■ Keep working ends open for
Manual pre-cleaning with brush and subsequent mechanical alkaline cleaning and thermal disinfection
D–W: FD–W:
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and sub-chapter:
Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
■ Chapter Manual pre-cleaning with a brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
Mechanical cleaning/disinfecting Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Mechanical alkaline cleaning and thermal disinfecting Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical/Note
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Mechanical cleaning/disinfection with manual pre-cleaning
Steam sterilization
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
Note The product can be sterilized either in disassembled or in assembled condition.
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Note To avoid breakage due to stress crack corrosion, sterilize the instruments with the lock fully open or locked no further than on the first ratchet tooth. ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
Manual pre-cleaning with a brush
opening any valves and faucets).
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfectant cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed
► Validated sterilization process
– Steam sterilization using fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
Sterilization for the US market ■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization. ■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6 , Aesculap recommends the following parameters:
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe.
Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters* Sterilization method
Temp.
Time
Minimum drying time
Prevacuum
270 °F/275 °F
4 min
20 min
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc.
Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Storage
Mechanical alkaline cleaning and thermal disinfecting
Technical Service
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
Risk of injury and/or malfunction! ► Do not modify the product. WARNING
– pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Inspection, maintenance and checks
CAUTION
► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Damage (metal seizure/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) at the marked lubrication points, using maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598).
► Allow the product to cool down to room temperature. ► After each complete cleaning, disinfecting and drying cycle, check that the instrument is dry, clean, operational,
and free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components). ► Dry the product if it is wet or damp. ► Repeat cleaning and disinfection of products that still show impurities or contamination. ► Check that the product functions correctly. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Technical Service. ► Assemble dismountable products, see Assembling. ► Check for compatibility with associated products.
Packaging ► Appropriately protect products with fine working tips. ► Store products with ratchet locks fully opened or locked no further than in the first notch. ► Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected. ► Pack trays appropriately for the intended sterilization process (e.g. in sterile Aesculap containers). ► Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
► For service and repairs, please contact your national B. Braun/Aesculap agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1602 Fax: +49 (7461) 16-5621 E-Mail: [email protected] Or in the US: Aesculap Inc. Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 Aesculap Repair Hotline Phone: +1 (800) 214-3392 Fax: +1 (314) 895-4420 Other service addresses can be obtained from the address indicated above.
Accessories/Spare parts For accessories and spare parts, see brochure C766 or visit http://www.endoscopy-catalog.com
Disposal ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging!
Distributor in the US/Contact in Canada for product information and complaints Aesculap Inc. 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 011945
07/16
V6
Änd.-Nr. 54983