Instructions for use/Technical description OCT instrument set Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapusa.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-800-282-9000. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung/Technische Beschreibung OCT-Instrumentarium Mode d’emploi/Description technique Instrumentation OCT Instrucciones de manejo/Descripción técnica Instrumental para TOC Istruzioni per l’uso/Descrizione tecnica Strumentario per trapianti osteo-condrali OCT (Osteo-Chondral Transplantation) Instruções de utilização/Descrição técnica Jogo de instrumentos OCT Gebruiksaanwijzing/Technische beschrijving OCT-instrumentarium Brugsanvisning/Teknisk beskrivelse OCT-instrumentsæt Bruksanvisning/Teknisk beskrivning OCT-instrumentuppsättning Käyttöohje/Tekninen kuvaus OCT-instrumenttisarja Lietošanas instrukcijas/tehniskais apraksts OHT instrumentu komplekts Naudojimo instrukcija/techninis aprašas OCT instrumentų rinkinys Инструкция по примению/Техническое описание Комплект инструментов OCT Návod k použití/Technický popis Nástroje OCT Instrukcja użytkowania/Opis techniczny Instrumentarium do przeszczepów chrzęstno-kostnych Návod na použitie/Technický opis Inštrumentárium OCT Használati útmutató/Műszaki leírás OCT műszerkészlet Navodila za uporabo/Tehnični opis Komplet instrumentov za OHP Upute za uporabu/Tehnički opis Komplet instrumenata za osteohondralnu transplantaciju Manual de utilizare/Descriere tehnică Set de instrumente pentru transplant osteocondral Упътване за употреба/Техническо описание Комплект инструменти OCT Kullanım Kılavuzu/Teknik açiklama OCT alet seti Οδηγίες χρήσης/Τεχνική περιγραφή Σετ οργάνων μεταμόσχευσης οστεοχόνδρινου ιστού
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap® – a B. Braun brand TA010387
2020-10
V6
Change No. 62945
en
2.3 ®
Aesculap OCT instrument set Legend 1 2 3 4 5 6 7
Handle Plug extractor Donor trephine DONOR 8 Plunger D 8 Recipient trephine RECIP 8 Plunger R 8 Cap
1.
About this document
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
Scope
These instructions for use apply for the OCT instrument set. ► For article-specific instructions for use as well as information on material compatibility and lifetime see B. Braun eIFU at eifu.bbraun.com
1.2
Safety messages
Application
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product. WARNING Risk of injury when using the product beyond the field of view! ► Apply the product only under visual control. ► When using the OCT instrument set, follow the instructions in the operating manual for osteochondral trans-
plantation. ► For the indications osteochondrosis dissecans or sclerotized bone: Use a pointed burr instead of recipient tre-
phine RECIP 5. ► To extract osteochondral plugs with the trephines, insert cap 7 into handle 1.
3.
Validated reprocessing procedure
3.1
General safety instructions
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows:
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result.
Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this.
CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
2.
Clinical use
2.1
Areas of use and limitations of use
2.1.1
Intended use
The OCT instrument set (OsteoChondral Transplantation) is used for the treatment of cartilage defects at the femoral condyles. The OCT instrument set is used to make osteochondral plugs and recipient drill holes. Note The trephines of the OCT instrument set are extremely precise and fine instruments that allow osteochondral plugs to be extracted and their recipient drill holes to be made accurately, in line with the requirements of osteochondral transplantation. Consequently, the trephines must be applied with special care. Therefore the general note in these instructions for use regarding damage to the trephines must be observed closely.
2.1.2
Indications
Note The manufacturer is not responsible for any use of the product against the specified indications and/or the described applications. For indications, see Intended use.
2.1.3
Contraindications
No known contraindications.
2.2
Safety information
2.2.1
Clinical user
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user. Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located. Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
Product
Product-specific safety information Risk of injury caused by incorrect operation of the product! ► Attend appropriate product training before using the product. ► For information about product training, please contact your national B. Braun/Aesculap agency. Trephine breakage caused by overstraining, especially in hard bones! ► Limit the service life of the trephines to a maximum of 20 plug extractions. ► Be careful to avoid deformations caused by excessive bending or torsion moments at the trephine tip when driving in the trephine and during the extraction of the osteochondral plug.
2.2.3
Sterility
The product is delivered in an unsterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
3.2
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to "AKI-Brochures", "Red brochure".
3.3
Reusable products
Influences of the reprocessing which lead to damage to the product are not known. A careful visual and functional inspection before the next use is the best opportunity to recognize a product that is no longer functional, see Inspection.
3.4
Preparations at the place of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
3.5
Preparing for cleaning
► Disassemble the product immediately after use, as described in the respective instructions for use. ► Disassemble the product prior to cleaning, see Disassembly.
3.6
Disassembly
3.6.1
Removing the trephine
► Unscrew plug extractor 2 out of handle 1. ► Take plunger 4 out of trephine 3 or plunger 6 out of trephine 5. ► Rotate the rapid connector on handle 1 as far as it will go in the direction of the arrow and pull trephine 3 or 5
off handle 1.
3.7
Cleaning/Disinfection
3.7.1
Product-specific safety information on the reprocessing method
Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Following the manufacturer's instructions, use cleaning and disinfecting agents – that are approved for (e.g. aluminum, plastics, high-grade steel), – that do not attack softeners (e.g., in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C.
3.7.2
3.9.1
Manual pre-cleaning with ultrasound and brush
Validated procedure
Validated cleaning and disinfection procedure Specific requirements
Reference
Phase
Step
t [min]
Conc. [%]
Water quality
Chemical
Manual cleaning with ultrasound and immersion disinfection
■ Suitable Cleaning brush ■ Keep working ends open for
D [°C/°F]
Chapter Manual cleaning/disinfection and subsection:
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
cleaning.
■ Chapter Manual cleaning with ultrasound and immersion disinfection
■ When cleaning instruments with movable hinges, ensure that these are in an open position and, if applicable, move the hinge while cleaning.
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh
■ Drying phase: Use a lint-free cloth or medical compressed air Manual pre-cleaning with ultrasound and brush, and subsequent mechanical alkaline cleaning and thermal disinfection
■ Suitable Cleaning brush ■ Place the product on a tray that is suitable for cleaning (avoid rinsing blind spots).
■ To flush the product: Use a
Chapter Mechanical cleaning/disinfecting with manual pre-cleaning and subsection:
■ Chapter Manual pre-cleaning with ultrasound and brush
■ Chapter Mechanical alkaline
flushing nozzle or flushing sleeve.
■ Keep working ends open for
cleaning and thermal disinfecting
cleaning.
■ Place the product on the tray with all product links and joints open.
3.8
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Manual cleaning/disinfection
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution.
3.9.2
Mechanical alkaline cleaning and thermal disinfecting
Machine type: single-chamber cleaning/disinfection device without ultrasound
► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
Phase
3.8.1
Manual cleaning with ultrasound and immersion disinfection
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.9
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
Mechanical cleaning/disinfecting with manual pre-cleaning
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ working solution 0.5% – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD–W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alcaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.10 Inspection ► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
3.10.1 Visual inspection ► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps. ► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
scratched and fractured components. ► Check the product for missing or faded labels. ► Check the products with long, slim shapes (in particular rotating instruments) for deformities. ► Check the product for damage to the spiral element. ► Check the cutting edges for continuity, sharpness, nicks and other damage. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Tech-
nical service.
3.10.2 Functional test CAUTION Damage (metal cold welding/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598). ► Assemble disassembled products, see Assembly. ► Check that the product functions correctly. ► Check that all moving parts are working properly (e.g. hinges, locks/latches, sliding parts etc.). ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Technical service.
3.11 Assembly 3.11.1 Mounting the trephine ► Rotate the rapid connector on handle 1 as far as it will go in the direction of the arrow and insert trephine 3 or 5. ► Push plunger 4 or 6, narrow end first, into trephine 3 or 5. ► Screw plug extractor 2 at the grip end of handle 1 into handle 1.
3.12 Packaging ► Appropriately protect products with fine working tips. ► Place the product in its holder or on a suitable tray. Ensure that sharp edges are covered. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.
3.13 Steam sterilization ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
opening any valves and faucets). ► Validated sterilization process
– Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
3.14 Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
4.
Technical service
CAUTION Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. WARNING Risk of injury due to sharp-edged and/or pointed products! ► When disposing of or recycling the product, ensure that the packaging prevents injury by the product. Note To avoid the release of toxic incineration products, the incineration of the product, its components and its packaging must be carried out in an appropriate incinerator (>1 000 °C). Note The user institution is obliged to reprocess the product before its disposal, see Validated reprocessing procedure. TA010387