BBraun
Pneumatic clip applier (with insertable titanium clip cassette and CO2 cartridge) Instructions for Use
93 Pages
Preview
Page 1
Aesculap®
Aesculap Endoscopic Technology
en USA
de fr es it pt nl da sv fi lv lt ru cs pl sk hu sl hr ro bg tr el
Instructions for use Pneumatic clip applier (with insertable titanium clip cassette and CO2 cartridge) Note for U.S. users This Instructions for Use is NOT intended for United States users. Please discard. The Instructions for Use for United States users can be obtained by visiting our website at www.aesculapusa.com. If you wish to obtain a paper copy of the Instructions for Use, you may request one by contacting your local Aesculap representative or Aesculap's customer service at 1-800-282-9000. A paper copy will be provided to you upon request at no additional cost. Gebrauchsanweisung Pneumatischer Clip-Applikator (mit einsetzbarem Titanclip-Magazin und CO2-Patrone) Mode d’emploi Applicateur pneumatique de clips (avec barrette de clips en titane à insérer et cartouche de CO2) Instrucciones de manejo Aplicador de clips neumático (con cargador de clips de titanio intercambiable y cartucho de CO2) Istruzioni per l’uso Applicatore per clip pneumatico (con caricatore per clip in titanio e cartuccia CO2 inseribile) Instruções de utilização Aplicador pneumático de clipes (com carregador amovível de clipes de titânio e cartucho de CO2) Gebruiksaanwijzing Pneumatische clip-applicator (met inzetbaar titaniumclip-magazijn en CO2-patroon) Brugsanvisning Pneumatisk agrafapplikator (med agraf-magasin af titan til indsætning og CO2-patron) Bruksanvisning Pneumatisk clipsapplikator (med instickbart titanclipsmagasin och CO2-patron) Käyttöohje Paineilmatoiminen hakaskiinnitin (paikalleen asetettavalla titaanihakasmakasiinilla ja CO2-patruunalla) Lietošanas instrukcijas Pneimatiskais skavu aplikators (ar ievietojamu titāna skavu magazīnu un CO2 patronu) Naudojimo instrukcija Pneumatinis spaustukų aplikatorius (su įstatoma titano spaustukų dėtuve ir CO2 kasete) Инструкция по примению Пневматический клипаппликатор (с картриджем титановых клипсов и баллоном CO 2) Návod k použití Pneumatický aplikátor svorek (s nasazovacím zásobníkem s titanovými svorkami a patronou CO2) Instrukcja użytkowania Pneumatyczny aplikator klipsów (z wkładanym magazynkiem klipsów tytanowych i wkładem CO2) Návod na použitie Pneumatický aplikátor svoriek (s vkladateľným zásobníkom titánových svoriek) a bombičkou CO2 Használati útmutató Pneumatikus kapocsrakó (behelyezhető titán kapocs tárral és CO2 patronnal) Navodila za uporabo Pnevmatski klip aplikator (z vlagalnikom za titanove klipe in vložkom s CO2) Upute za uporabu Pneumatski aplikator za kopče (sa magazinom za kopče od titana i CO2 uloškom za umetanje) Manual de utilizare Aplicator de agrafă pneumatic (cu magazie de agrafe din titan și cu cartușe de CO2) Упътване за употреба Пневматичен клипс апликатор (с използване на магазин за титанови клипсове и патрон за CO2) Kullanım Kılavuzu Pnömatik klips aplikatörü (kullanılabilir titanyum klips magazinli ve CO2 kartuşlu) Οδηγίες χρήσης Πνευματικός εφαρμοστής κλιπ (με αφαιρούμενη κασετίνα κλιπ τιτανίου και φυσίγγιο CO2)
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com
6 1
6 2 3
5 4
A
12
11
10
9
8
10
B
10
10 13
14
C
D
4
5
E
F 9 15
G
H
I
J
5
Aesculap® – a B. Braun brand TA012509
2020-09
V6
7
Change No. 63464
16
12
K
M
L
en ®
Aesculap Pneumatic clip applier (with insertable titanium clip cassette and CO2 cartridge)
2.2
Areas of use and limitations of use
2.2.1
Intended use
The pneumatic clip applier (with insertable titanium clip cassette and CO2 cartridge) is used for the ligation of vessels and hollow organs and for marking anatomical structures. The product is used in combination with Aesculap titanium clip cassettes.
2.2.2
1 Star wheel 2 Screw handle 3 Cartridge adapter 4 Flap (catridge compartment) 5 Application lever (for closing the clip) 6 Retaining ring 7 Pushing rod 8 Inner tube 9 Shaft 10 Lubrication points 11 Space for titanium clip cassette 12 Jaws protector 13 Cartridge 14 Stop (cartridge adapter) 15 Cassette feeder 16 Stop (jaw protection)
Note Ligating clips are intended for use on vessels and tissue structures (e.g. the biliary tract) for which the surgeon believes ligating clips are the best tools. The size, type and material of the clip are selected by the surgeon, based on his experience, estimation and requirement.
2.2.3
2.2.4
Relative contraindications
The following conditions, individual or combined, may result in delayed healing or compromised surgical success: ■ Medical or surgical conditions that could prevent the success of the operation, e.g. connective tissue disorders (e.g. Marfan syndrome) or infections. In the case of relative contraindications, the user decides individually on the use of the product.
About this document
Risks, adverse effects and interactions
The following possible undesired effects or interactions currently know to the manufacturer are referenced within the framework of the legal information reporting requirement. These are procedural, not product-specific and include infection, migration/dislocation of the clip with the possibility of (re)opening of a closed cavity organ/vessel (such as bleeding), necrosis of the clipped structure, injury to the treated or neighboring tissue structures and sensitivity to the metal components of the clip.
Note General risk factors associated with surgical procedures are not described in these instructions for use.
1.1
Absolute contraindications
Do not use: ■ on the central nervous system ■ in case of foreign-body sensitivity for titanium ■ in contraceptive procedures such as tubal litigation, vasectomy
2.3
1.
Indications
■ Ligation of vessels and hollow organs ■ Marking of anatomic structures for radiological detection
Legend
Scope
These instructions for use apply for the following products:
2.4
Safety information
Art. no.
Designation
2.4.1
Clinical user
PL520R
CHALLENGER TI-P HANDLE
PL522R
SHAFT COMPL.D:5MM L:310MM
PL536R
SHAFT COMPL.D:10MM L:370MM
PL538R
SHAFT COMPL.D:10MM L:260MM
PL603R
MULTIF.CLIP APPL.TI-P 5/205MM F.SM-CLIPS
PL604R
MULTIF.CLIP APPL.TI-P 5/310MM F.SM-CLIPS
PL606R
MULTIF.CLIP APPL.TI-P 10/370MM F.ML-CLIP
PL608R
MULTIF.CLIP APPL.TI-P 10/260MM F.ML-CLIP
PL609R
SHAFT TI-P ALL.5/205MM F.HANDLE PL520R
General safety information To prevent damage caused by improper setup or operation, and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge and experience. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user. Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located.
► For article specific instructions for use and material compatibility and lifetime information, see B. Braun eIFU at
eifu.bbraun.com
1.2
Safety messages
Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.4.2
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows: WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result.
Product-specific safety information
Note The applicability of ligating clips (in the respectively available clip size) depends on the size of the structure and is at the discretion of the user. Use is also effected in the context of the vessel condition and/or whether clip application is generally indicated. The position of the clip must be checked by the user after application. ► To avoid damage to the jaws and the titanium clip cassette: carefully insert the pneumatic clip applier through
the trocar.
CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
► In order to prevent unintentional releasing of the shaft from the handle by retracting the star wheel: push the
2.
Clinical use
► Keep the jaws protector installed when the product is not in use. ► Only use the product with visual control.
2.1
Available sizes
retaining ring 6 onto the handle to the star wheel from above and rotate by 180°.
2.4.3
Designation
Art. no./Dimensions
Pneumatic clip applier (with insertable titanium clip cassette and cartridge), complete
PL606R
PL608R
PL604R
PL603R
Shaft diameter
10 mm
10 mm
5 mm
5 mm
Length
370 mm
260 mm
310 mm
205 mm
2.5
Handle only
PL520R
PL520R
PL520R
PL520R
Shaft only (including interior tube and pushing rod)
PL536R
PL538R
PL522R
PL609R
Pushing rod only
PL536801
PL538801
PL522800
PL601800
Inner tube only
PL536802
PL538802
PL522801
PL601801
Jaws protector
PL536810
PL536810
PL522810
PL522810
Retaining ring
PL520301
PL520301
PL520301
PL520301
Titanium clip cassette, size: small-medium including cartridge (12 cassettes, each with 12 clips and 12 cartridges)
-
-
PL574T
PL574T
Titanium clip cassette, size: medium-large including cartridge (12 cassettes, each with 8 clips and 12 cartridges)
PL579T
PL579T
-
-
Challenger Ti-P CO2-cartridges alone (3 cartridges)
PL575SU
Sterility
The product is delivered in an unsterile condition. ► Check the new product for damage after removing its transport packaging and clean prior to its initial sterilization.
Application
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Always carry out a function test prior to each use of the product. WARNING Risk of injury to the patient and/or damage to the instrument if incompatible trocars are used! ► When using the instrument with a trocar: – Check the compatibility of the trocar and instrument (with titanium clip cassette). – Guide instrument into or out of the trocar only with opened jaws. WARNING Risk of dislocation of clips by cutting tissue too close to clip! ► When cutting clipped structures, leave an area of tissue of at least the width of double the clip between the clip and cutting site. ► Ensure that the structure is not under tension when cutting. ► Insert the clip applier carefully, in a straight line and centrally through the trocar.
2.5.1
Setting up the pneumatic clip applier
Demounting the jaws protector ► Compress the jaw part protection 12 and carefully remove it in the direction of the shaft, see Fig. K. PL575SU
PL575SU
PL575SU
Assembling the pneumatic clip applier ► For instruction on how to assemble the pneumatic clip applier, see Chapter 3.12.
Inserting a cartridge Note The flap of the cartridge compartment must be closed in one swift, continuous movement; otherwise, at the moment of broaching, the cartridge pressure will force open the flap of the cartridge compartment and the cartridge will be emptied! ► Open flap of the cartridge compartment 4: pull lateral surfaces of the gray flap of the cartridge compartment 4
backward. ► Insert the cartridge 13 with the round side to the stop 14 into the cartridge adapter 3, see Fig. D.
The flat side of the cartridge 13 is visible. ► Close the flap of the cartridge compartment 4, see Fig. E, making sure that application lever 5 is not caught or
pushed back. The cartridge 13 is broached automatically. Note The pushing rod 7 may move forward and back in the shaft 9 when the cartridge is broached! Function check Note During function checks, pushing rod 7 moves only if the application lever is pressed far enough. ► Push application lever 5 of the pneumatic clip applier through to the limit stop and release it again, see Fig. H.
Application lever 5 moves back to its rest position. Any forward and backward movement of the pushing rod 7 must take place in the shaft 9 and be visible on the space for the titanium clip cassette 11. Inserting the titanium clip cassette ► Insert the titanium clip cassette with cassette feeder 15 in shaft 9, making sure that the feeder is not damaged in the process, see Fig. F. ► Push down the titanium clip cassette until the lugs at its sides engage in the pneumatic clip applier, see Fig. G. The pneumatic clip applier is ready for use now.
2.5.2
Operating the pneumatic clip applier
CAUTION Malfunction of the titanium clip cassette due to use of the same cartridge for more than one cassette! ► Replace the cartridge as well when changing the titanium clip cassette.
3.3
Reusable products
Influences of the reprocessing which lead to damage to the product are not known. Evidence of biocompatibility and processibility for the product was produced by the manufacturer after 50 processing cycles. A careful visual and functional inspection before the next use is the best option to determine if a product is no longer functional, see Chapter 3.11. The recommended annual servicing reduces the risk of failure and extends the service life of the product; see Chapter 4.1.
3.4
Preparations at the place of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
3.5
Preparing for cleaning
► Make sure the cartridge of the pneumatic clip applier has been removed, see Chapter 3.6.2. ► Fixation screws which permanently connect the various components must not be disassembled. ► Disassemble the product prior to cleaning, see Chapter 3.6. ► Open up instruments with hinges.
3.6
Disassembly
3.6.1
Removing the titanium clip cassette
► Close the jaws with the application lever 5 and withdraw the titanium clip cassette, see Fig. I.
3.6.2
Removing the cartridge
WARNING Alarm caused by residual gas hissing out when opening the flap of the cartridge compartment! ► Inform O.R. personnel of impending cartridge change. ► Open flap of the cartridge compartment 4: pull lateral surfaces of the gray flap of the cartridge compartment 4
backward. ► Remove cartridge 13 from cartridge adapter 3 against the direction of the arrow, see Fig. D.
3.6.3
Disassembling the pneumatic clip applier
Note The last clip in each titanium clip cassette is color-marked in order to indicate that the titanium clip cassette will be empty after the present application!
► With retaining ring 6 mounted: rotate the retaining ring 6 until the opening of the ring points downward in the
Note Do not close the jaws of the instrument when inserting/approaching the structure to be clipped.
► Pull off retaining ring 6 towards rear of handle, see Fig. A.
► Load clip: Push application lever 5 to the stop and release it again.
Application lever 5 moves back to its rest position. The next clip is automatically fed into the jaws after each complete push and release movement of application lever 5. Note In case of any malfunction at this stage, clip loading must be inspected and repeated extracorporeally. ► Close clip: Push application lever 5 of the pneumatic clip applier through to the limit stop, see Fig. H. ► Close the clip only in the clip guide in the jaws of the pneumatic clip applier. Make sure the clip is applied under
visual control. ► Check the fit and operability of the clip. ► If necessary apply another clip. ► Dispose of titanium clip magazine and cartridges after the operation.
2.5.3
Changing the titanium clip cassette and the cartridge
► Remove the empty titanium clip cassette, see Chapter 3.6.1. ► Remove empty cartridge 13, see Chapter 3.6.2. ► Insert new cartridge 13, see Chapter 2.5.1. ► Carry out function check, see Chapter 2.5.1. ► Insert new titanium clip cassette, see Chapter 2.5.1.
3. 3.1
Validated reprocessing procedure General safety information
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
direction of the application lever 5. ► Hold the shaft 9, pull the rotary ring 1 to the stop in the direction of the handle 2 and remove the shaft 9,
see Fig. J.
3.6.4
Disassembling the shaft
► Withdraw inner tube 8 from shaft 9. ► Pull the pushing rod 7 from the inner tube 8.
3.7
Cleaning/Disinfection
3.7.1
Product-specific safety information on the reprocessing method
Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Following the manufacturer's instructions, use cleaning and disinfecting agents, – that are approved for peek plastics and stainless steel, – that do not attack softeners (e.g., in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C. Contamination or obstruction of the valve mechanism due to impact of cleaning agents! ► Do not apply ultrasonic immersion for cleaning the handle. ► Compress jaws protector 12 and carefully guide it over the jaws to the stop 16 of the outer tube, see Fig. L. ► If the product can be fixated on storage devices safely and appropriately for cleaning: carry out mechanical
cleaning and disinfecting. ► Only dispose of the product dry, as otherwise contamination may penetrate into the shafts.
3.7.2
Validated cleaning and disinfection procedure
Validated procedure
Specific requirements
Reference
Manual cleaning with ultrasound and immersion disinfection
■ Suitable cleaning brush: e.g. PM995R or
Chapter Manual cleaning/disinfection and subsection:
■ Shaft
■ Drying phase: Use a lint-free cloth or medical
3.2
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, link to "AKI-Brochures", "Red brochure".
■ Chapter Manual cleaning with ultrasound and immersion disinfection
compressed air Mechanical alkaline cleaning and thermal disinfection
■ Screw handle
Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For up-to-date information about reprocessing and material compatibility, see B. Braun eIFU at eifu.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
■ Single use syringe 20 ml ■ Place the jaw part protection 12 onto the product.
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this.
GK469R
Manual pre-cleaning with ultrasound and brush, and subsequent mechanical alkaline cleaning and thermal disinfection
■ Shaft
■ Place the product on a tray that is suitable for cleaning (avoid rinsing blind spots).
■ Place the product on the tray with flap of cartridge compartment 4 open, see Fig. A.
■ Suitable cleaning brush: e.g. PM995R or GK469R
■ Single use syringe 20 ml ■ Place the jaw part protection 12 onto the product.
■ Place the product on a tray that is suitable for cleaning (avoid rinsing blind spots).
■ Connect components with lumens and channels directly to the rinsing port of the injector carriage.
Chapter Mechanical cleaning/disinfection and subsection:
■ Chapter Mechanical alkaline cleaning and thermal disinfection Chapter Mechanical cleaning/disinfection with manual pre-cleaning and subsection:
■ Chapter Manual precleaning with ultrasound and brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
3.8
Manual cleaning/disinfection
3.10 Mechanical cleaning/disinfection with manual pre-cleaning
► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning/disinfection process if necessary.
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
3.8.1
Manual cleaning with ultrasound and immersion disinfection
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
3.10.1 Manual pre-cleaning with ultrasound and brush
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
II
Intermediate rinse
RT (cold)
1
-
D–W
-
III
Disinfection
RT (cold)
5
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed fresh
V
Drying
RT
-
-
-
-
► Note the information on appropriate cleaning brushes and disposable syringes, see Chapter 3.7.2.
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed fresh ► Note the information on appropriate cleaning brushes and disposable syringes, see Chapter 3.7.2.
Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Chapter 3.7.2.
3.9
Mechanical cleaning/disinfection
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
3.9.1
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical/Note
Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
3.10.2 Mechanical alkaline cleaning and thermal disinfecting Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH ~ 13 – <5 % anionic surfactant
■ working solution 0.5% – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
In accordance with the program for the cleaning and disinfecting machine
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
3.11 Inspection ► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp.
3.11.1 Visual inspection ► Ensure that all soiling has been removed. In particular, pay attention to mating surfaces, hinges, shafts, recessed
areas, drill grooves and the sides of the teeth on rasps. ► If the product is dirty: repeat the cleaning and disinfection process. ► Check the product for damage, e.g. insulation or corroded, loose, bent, broken, cracked, worn or severely
I
Pre-rinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
scratched and fractured components.
– pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
► Check the product for missing or faded labels. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see
Chapter 4.2.
3.11.2 Functional test CAUTION Damage to sealing components caused by maintenance spray-lubrication of the handle! ► Do not spray-lubricate the handle. ► Only spray marked areas in the shaft, see lubrication locations 10 in Fig. B.
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alcaline
CAUTION Damage (metal cold welding/friction corrosion) to the product caused by insufficient lubrication! ► Lubricate the labeled lubricating points 10 before the functional test using the maintaining oil suitable for the sterilization procedure used (e.g. for steam sterilization STERILIT® I oil spray JG600 or STERILIT® I drop-feed lubricator JG598).
► Check visible surfaces for residues after mechanical cleaning/disinfecting.
► Assemble disassembled products, see Chapter 3.12. ► Check that the product functions correctly.
– Close jaws from the tip to the rear. – The tips must contact when the jaws are closed. – the closed jaws may not intersect. ► Check that all moving parts are working property (e.g. hinges, locks/latches, sliding parts etc.). ► Make sure the cartridge of the pneumatic clip applier has been removed, see Chapter 3.6.2. ► Check for compatibility with associated products. ► Immediately put aside inoperative products and send them to Aesculap Technical Service, see Chapter 4.2.
3.12 Assembly CAUTION Malfunctions caused by incorrect assembly sequence! ► Keep to the correct order of assembly.
3.12.1 Assembling the shaft ► Push the pushing rod 7 into the inner tube 8. ► Push the inner tube 8 into the shaft 9 using the pushing rod 7.
3.12.2 Assembling the pneumatic clip applier ► Pull the rotary ring 1 to the stop in the direction of the handle 2. ► Hold the rotary ring 1 on the stop and push in the assembled shaft 9. Press the shaft 9 against the handle 2 and
turn it until the marking lines are in line with the shaft 9 and star wheel 1 and the star wheel 1 snaps forward, see Fig. C. ► Install retaining ring 6 on handle 2 between star wheel 1 and application lever 5 and twist the ring by 180°.
3.13 Packaging ► Place the product in its holder or on a suitable tray. Make certain the jaws are protected. ► Package trays appropriately for the sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against contamination of the product during storage.
3.14 Steam sterilization CAUTION Loss of sterility of/damage to the handle caused by cartridge left in the pneumatic clip applier! ► Make sure that no cartridge is left in the pneumatic clip applier under any circumstances. ► If a cartridge remains after sterilization: check handle for functionality and sterilize again. Note The product can be sterilized either in disassembled or in assembled condition. ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g., by
opening any valves and faucets). ► Validated sterilization process
– Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► If several devices are sterilized at the same time in the same steam sterilizer: Ensure that the maximum permitted load according to the manufacturers’ specifications is not exceeded.
3.15 Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
4.
Maintenance and Service
4.1
Maintenance
To ensure reliable operation, Aesculap recommends maintaining the product as indicated on the maintenance label, see Fig. M. ► For corresponding services, contact your national B. Braun/Aesculap agency. see Chapter 4.2
4.2
Technical service
CAUTION Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. ► Do not modify the product. ► For service and repairs, please contact your national B. Braun/Aesculap agency. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected] Other service addresses can be obtained from the address indicated above.
4.3
Accessories/Replacement Parts
Accessories and spare parts are listed in the Endoscopy online catalog at www.endoscopy-catalog.com or in the Aesculap brochure C62311.
5.
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging.
6.
Symbols on product and packaging Maintenance labeling on the handle and shaft with a note regarding the next maintenance deadline, see Chapter 4.1
YYYY-MM TA012509
2020-09
V6
Change No. 63464