BBraun
S4 Occipital Cervical Thoracic System
27 Pages
Preview
Page 1
Safety notes Aesculap® S4 Occipital Cervical Thoracic System Intended use The S4 Occipital Cervical Thoracic System implants are used for the dorsal monosegmental and multisegmental stabilization of the occipitocervical junction and of the cervical and upper thoracic spine. The system consists of: ■ Occiput plates ■ Rods ■ Mini polyaxial screws ■ Set screws ■ Hook ■ Cross connector ■ Other connectors Appropriate implant components from S4 SPINE System® (e.g. rods) can also be used. Special instruments must be used for implanting these components, as well as for the distraction, compression and reduction of the lumbar and thoracic spine.
Materials The materials used in the implant are listed on the packaging: ■ ISOTAN®F titanium forged alloy Ti6Al4V acc. to ISO5832-3 ■ ISOTAN®P pure titanium acc. to ISO 5832-2 The titanium implants are anodized with a colored oxide layer. Slight changes in coloration may occur, but do not affect the implant quality. ISOTAN® is a registered trademark of Aesculap AG, 78532 Tuttlingen / Germany.
CAUTION Federal law restricts this device to sale by, or on order of a physician! ■ It is the operating surgeon's responsibility to ensure that the surgical procedure is performed properly. ■ General risk factors associated with surgical procedures are not described in in the present instructions for use. ■ The operating surgeon must have a thorough command of both the hands-on and conceptual aspects of the established operating techniques. ■ The operating surgeon must be fully conversant with bone anatomy, including the pathways of nerves, blood vessels, muscles, and tendons. ■ The operating surgeon is responsible for combining and implanting the implant components, with or without bone cement. ■ It is the operating surgeon's responsibility to ensure the correct combination of implant components and their implantation. ■ Aesculap is not responsible for any complications arising from wrong indication, wrong choice of implant, incorrect combination of implant components and operating technique, the limitations of the treatment method, or inadequate asepsis. ■ The instructions for use for individual Aesculap implant components must be followed. ■ The implant components were tested and approved in combination with Aesculap components. If other combinations are used, the responsibility for such action lies with the operating surgeon. ■ Do not, under any circumstances, combine implant components from different manufacturers. ■ Do not, under any circumstances, use damaged or surgically removed components. ■ Implants that have been used before must not be reused. ■ Damage to load-bearing bone structures can result in loosening of the components, bone or implant fractures or other acute complications. ■ To ensure the earliest possible detection of such implant dysfunction, the prosthetic joint must be checked periodically, using appropriate techniques. The implant has not been evaluated for safety and compatibility in the MR environment. It has not been tested for heating, migration or image artifact in the MR environment. Scanning a patient who has this implant may result in patient injury. WARNING
Indications Use for: ■ Fractures ■ Degenerative instability ■ Post-trauma instability ■ Tumors
■ Delayed healing can cause implant breakage due to material fatigue. ■ The attending physician shall make any decision with regard to the removal of implant components that have been used.
■ Damage to the load-bearing structures of the implant can lead to loosening of components, dislocation, migration, and other severe complications.
■ The implant components applied, along with their article numbers, the name of the implant, as well as the batch number and serial number (if available) must be documented in all patient records.
Contraindications Do not use in the presence of: ■ Fever ■ Acute or chronic vertebral infections of a local or systemic nature ■ Pregnancy ■ Severe osteoporosis or osteopenia ■ Medical or surgical conditions that could negatively affect the outcome of the implantation ■ Mental illness ■ Dependency on pharmaceutical drugs, drug abuse, or alcoholism ■ Adiposity ■ Severely damaged bone structures that could prevent stable implantation of implant components ■ Neuromuscular disorders or illnesses ■ Bone tumors in the region of implant fixation ■ Wound healing disorders ■ Inadequate patient compliance ■ Foreign body sensitivity to the implant materials ■ Cases not listed under indications
Side effects and interactions The application or improper use of this system entails the following risks: ■ Implant failure resulting from excessive load – Warping or bending – Loosening – Breakage ■ Inadequate fixation ■ Failed or delayed fusion ■ Infection ■ Fractured vertebral body or bodies ■ Injuries to – Nerve roots – Spinal cord – Blood vessels – Organs
■ Postoperatively, individual patient information, as well as mobility and muscle training, is of particular importance.
Sterility ■ The implant components are supplied in an unsterile condition. ■ The implant components are packaged individually. ► Store the implant components in their original packaging and only remove them from their original and protec-
tive packaging immediately prior to application. ► Use the implant system storage unit for sterilization and sterile prep.
Prior to initial sterilization and subsequent resterilization, the implant components must be cleaned using the following validated reprocessing procedure: ► Validated sterilization process – Steam sterilization using fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
Validated reprocessing procedure Sterilization ► Validated sterilization process
– Steam sterilization through fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer: ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
Application
Disposal
The operating surgeon shall devise an operation plan that specifies and accurately documents the following: ■ Selection of the implant components and their dimensions ■ Positioning of the implant components in the bone ■ Location of intraoperative landmarks The following conditions must be fulfilled prior to application: ■ All requisite implant components are ready to hand. ■ Operating conditions are highly aseptic. ■ All requisite implantation instruments must be available and in working order, including specialized Aesculap implantation systems. ■ The operating surgeon and operating room team are thoroughly familiar with the operating technique and with the available range of implants and instruments; information materials on these subjects must be complete and ready to hand. ■ The operating surgeon is fully conversant with the rules governing medical practice, the current state of scientific knowledge, and the contents of relevant scientific articles by medical authors. ■ The manufacturer has been consulted if the preoperative situation was unclear and if implants were found in the area operated on. The surgical procedure and following information has been explained to the patient, and the patient's consent has been documented: ■ In the case of delayed or incomplete fusion, the implants can break and loosen due to high loads. ■ The life-span of the implant depends on the patient’s body weight. ■ The implant components must not be overloaded by extreme strains, hard physical labor or sports. ■ Corrective surgery may be necessitated by implant loosening, fracture or loss of correction. ■ Smokers present an increased risk of bone fusion failure. ■ The patient must undergo medical check-ups of the implant components at regular intervals. The following steps need to be carried out in order to implant the S4 Occipital Cervical Thoracic System implant: ► Select the appropriate S4 Occipital Cervical Thoracic System variant and anchoring implants according to indication, preoperative planning, and intraoperative bone situation. ► When selecting S 4 Occipital Cervical Thoracic System screw types, make certain to select the correct diameter, length, and canal orientation.
► Adhere to national regulations when disposing of or recycling the product, its components and its packaging!
Incorrect application can result in trauma to the spinal cord and nerve roots! ► Align and screw in the instruments, the pedicle screw and the polyaxial screw only under X-ray control or using a navigation system. WARNING ► To avoid internal stress on, and weakening of, the implant: avoid scoring or scratching of the implant compo-
nents. ► Do not bend metal implants, except for the rods and cross connectors of the S4 Occipital Cervical Thoracic Sys-
tem. ► Do not bend back, or bend excessively, the rods and cross connectors. ► Only use bending instruments from the S4 Occipital Cervical Thoracic System instrument set for bending the rods
and cross connectors. ► For positioning the clamping screw, always use the clamping screw insertion instrument intended for this pur-
pose. ► For tightening and loosening the clamping screw, always use the screwdriver and countering instrument provided
for this purpose. ► Always tighten the set screws with the torque wrench intended for this purpose.
WARNING
WARNING
Risk of injury due to improper mounting of the set screw! ► Set the set screw in place correctly. ► Make certain that the rods are correctly positioned on the floor of the groove. ► Fully tighten the set screw, using the torque wrench. Correction loss due to insufficient fixation of the polyaxial head! ► Never loosen the connection of the polyaxial head again, once it has been tightened. ► Tighten the set screw only after every necessary corrective measure has been performed. Risk of damage to the implant due to overtightening of the set screw! ► Always tighten the set screw to the exact torque specified for this purpose.
WARNING Screw head expansion caused by tightening of the clamping screw without using the countering instrument! ► Always use the countering instrument when tightening the set screw. WARNING
WARNING
Damage to the implant caused by applying the reduction instruments too high at the implant! ► Always apply reduction instruments (e.g. distraction and compression forceps) below the rod at the implant.
Expansion of the implant head caused by derotation without derotation sleeves! ► Always use derotation sleeves for rod derotation. WARNING ► When using a connector (hook, cross connector and other connectors) use special S4 Occipital Cervical Thoracic
System mounting instruments. Risk of loosening if occiput screw is incorrectly inserted! ► Ensure that the occiput screw is screwed in vertically into the plate. ► Ensure that the head of the occiput screw engages in the plate. WARNING ► Always tighten the locking screws of the cross connector, clamp, hook and occiput plate with the torque wrench
and countering instrument for cross connectors intended for this purpose. Further information on Aesculap implant systems is always available from B. Braun/Aesculap or the appropriate B. Braun/Aesculap office.
Distributor in the US/Contact in Canada for product information and complaints Aesculap Implant Systems LLC 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 011796
2017-12
V6
Änd.-Nr. 57595