BELMONT MEDICAL TECHNOLOGY
FMS 2000 Fluid Management System Setup and Use Guide
Setup and Use Guide
44 Pages

Preview
Page 1
Belmont Instrument Corporation Features
Setup
Accessories
Specs
FMS 2000 Fluid Management System
Fluid Management Features
Control Fluid Delivery Avoid Hypothermia
Setup
Avoid Air Embolus Avoid Vessel Trauma
Accessories
Specs
Indications for Use Features
Setup
Accessories
Specs
For use in high blood loss surgical
procedures, trauma and any situation where rapid replacement and warming of blood or replacement fluid is required.
Overview Features
Controlled Rapid Infusion Efficient Fluid Warming
Setup
Active Air Detection and Handling Line Pressure Control
Accessories
Specs
Simple Setup and Operation
Controlled Rapid Infusion Features
Setup
Accessories
Specs
3 Touch Buttons To Control Infusion Rate
– UP and DOWN infuses from 2.5 to 1000 ml/min – BOLUS infuses fixed volumes – Separate key for 500 ml/min
Controlled Rapid Infusion Features
Roller Pump
– Eliminates need for pressure bag infusion Setup
Accessories
Specs
Vigilance by Constant Display of:
– Flow Rate – Total volume infused – Fluid Temperature – In-line pressure
Efficient Fluid Warming Features
Electromagnetic
induction heating Setup
Accessories
Specs
– 16 stainless steel rings
Efficient Fluid Warming Features
Warms fluid from 4oC to 37.5 oC in a single
pass
(4 oC Input Fluid)
40
35
Setup
Accessories
.
Temperature (oC)
30
25
20
Flu id temperature entering the heat exchanger Fluid t mperature exiting the heat exchanger (display)
15
Fluid temperature exiting the patient line
10
Specs
5
Prime
0 0 sec
Patient Line Prime
20 sec
Max Rate (1000ml/min)
40 sec
60 sec
Time o
80 sec
o Note: Initially 4 C fluid is warmed to near 10 C by the disposable and tubing at ambient temperat
30 minutes ure.
Air Detectors Features
Setup
Accessories
Specs
Fluid out air detector
– Ensures system does not run dry – No need to de-air bags – Ultrasonic air detector – Interlock design for easy loading
Air Detectors Features
Setup
Accessories
Specs
In-line Air Detector
– Ultrasonic air detector – 2nd air detector located in patient infusion line – Easy to load
High Pressure Alarm Features
Setup
Accessories
Specs
Pressure Sensor – Continuously displays system pressure – Stops infusion with pressure > 300mmHg or sudden pressure spike – Slows flow rate to maintain a safe line pressure when infusing through a small catheter
Simple Setup and Operation Features
No dedicated operator reduces manpower Setup and prime < 1 minute
Setup
Heats instantly Displays operational clues and diagrams
Accessories
Specs
Setup- IV pole Mounting Features
Setup
Accessories
Specs
Mount the system on an IV
pole with a maximum pole diameter of 1.25 inches (3.18 cm) and a minimum base diameter of 24 inches (61 cm) Do not mount the system higher than 30 inches (76 cm) off the ground to prevent tipping
Setup –IV Pole Mounting Features
Setup
Accessories
Specs
Lift up on the handle of the pole clamp to
open Position the pole clamp around the IV pole and push down on the handle to lock the clamp in place
Setup –IV Pole Mounting Features
Setup
Accessories
Specs
Install the Support
Assembly (clamp and washer) directly below the FMS 2000 Clamp the reservoir support onto the IV pole above the unit
Setup- Removing the disposable Features
Setup
Accessories
Specs
1. Remove the reservoir
chamber 2. Pull out the interlock block from the tray 3. Lift out the disposable holding the reservoir chamber and the heat exchanger
Setup- Loading the Disposable: Step #1 Features
Setup
Accessories
Specs
Load circular heat
exchanger with red arrow pointing up and the red tubing aligned with the red line on the system
Setup - Loading the Disposable: Step #2 Features
Setup
Accessories
Specs
Snap the reservoir
chamber into the reservoir support Load interlock block into air detector with the arrow pointing into the machine Thread pump tubing over pump head
Setup- Loading the Disposable: Step #3 Features
Setup
Accessories
Specs
Place pressure chamber
into pressure transducer cavity Thread tubing into 2nd air detector Place tubing on both sides of valve
Setup- Powering Up Features
Switch the circuit breaker ON AC POWER PRESENT appears at startup
Setup
Accessories
Specs