FMS 2000 Fluid Management System Training Guide
Training Guide
72 Pages

Preview
Page 1
Rapid Infuser
Overview
Setup
Infusing
Troubleshooting
Accessories
Rapid Infuser
Overview Overview
Rapid Infuser
Indications for Use For use in high blood loss surgical procedures,
Setup
Infusing
Troubleshooting
Accessories
trauma and anyy situation where rapid p replacement and warming of blood or replacement fluid is required.
Overview Overview
Rapid Infuser
Controlled Rapid Infusion Roller Pump
Setup
Infusing
– Eliminates need for pressure bag infusion – Pumps P ffrom 2.5 2 5 ml/min l/ i to t 750 ml/min l/ i – Continuous or bolus operation Constant Display of:
Troubleshooting
Accessories
– Flow Rate – Total volume infused – Fluid Temperature – In-line In line pressure
Overview Overview
Rapid Infuser
Cautions DO NOT mix calcium containing solutions with citrated blood
Setup
Infusing
Troubleshooting
Accessories
products DO NOT infuse i f platelets l t l t or cryoprecipitates i it t through th h the th Rapid Infuser
Overview Overview
Efficient Fluid Warming Electromagnetic
Setup
Infusing
Troubleshooting
Accessories
induction heatingg – 16 stainless steel rings – Heats to 37.5 degr. C – Heats to 39 degr. C betw. 10 & 50 ml/min – No heating < 10 ml/min
Rapid Infuser
Overview Overview
Air Management System Ensures patient
Setup
Infusing
Troubleshooting
safety by preventing infusion of air – Two air detectors – Air Ai ttrap – Infusion line clamp – Air vent (recirc.) (recirc ) line Automatic air purge
Accessories
every 500 ml infused
Rapid Infuser
Overview Overview
Rapid Infuser
Air Sensors “Fluid Out” air
Setup
Infusing
Troubleshooting
Accessories
sensor ensures system does not run dry “Air “Ai D Detection” i ” airi sensor for additional safety Keep air sensors clean & dry
“AIR DETECTION” Air Sensor “FLUID OUT” Air Sensor
Overview Overview
Pressure Monitoring Continuously displays
Setup
Infusing
Troubleshooting
Accessories
system pressure Stops infusion with pressure > 400 mmHg or sudden pressure spike Slows flow rate to maintain a safe line ppressure ((300 mmHg) g) when infusing through a small catheter
Rapid Infuser
Overview Overview
Rapid Infuser
Caution Catheter C h size i andd llengthh iinfluence fl flflow rates
Setup
Infusing
Troubleshooting
Accessories
Flow (ml/min) 750 600 400 200 0
M i Maximum achievable hi bl flows fl (at ( t 300 mmHg, H cannula l ≤ 2”)
22 g
20 g
Crystalloid
18 g
Packed Cells
16 g
14 g
12 g
Cannula Size
Overview Overview
Dual Patient Line Reduces in-line restriction
Setup
Infusing
Troubleshooting
Accessories
Increases flow rate to the
patient Be sure to: – check for patency of each line – maintain KVO
Rapid Infuser
Overview Overview
Rapid Infuser
Battery Operation Automatic as soon as AC is disconnected
Setup
Infusing
Troubleshooting
Accessories
Maximum infusion rate of 50 ml/min No heating, at least 30 minutes of operation Automatically A t ti ll recharges h when h AC iis presentt
Overview Overview
Rapid Infuser
IV Pole Clamp
Setup
Lift up on the handle of the Infusing
Troubleshooting
Accessories
pole clamp to open Take care to hold the carrying handle if ever unlocking the pole clamp
Shown locked (up to unlock)
Overview Overview
Rapid Infuser
Keep in Mind…
Setup
Screen is sensitive touch control for Infusing
Troubleshooting
Accessories
rapid response Flow rate in ml/min (not ml/hr) Displayed temperature is actual fluid temperature
Setup
Rapid Infuser
Overview
Setup
Infusing
Troubleshooting
Accessories
Setting up the Belmont Rapid Infuser
Setup Overview
Rapid Infuser
Fluid Path Heater H t Magnet
Setup
Pressure Sensor
Temperature Probe (warm) Air Detector ((“FLUID OUT”)
Infusing
Air Detector Troubleshooting
Accessories
(“AIR DETECTION”)
Rollerhead Pump
Valve Wand Temperature Probe (cold)
Setup Overview
Unpacking the Disposable Tubing Set Before beginning, ensure
Setup
Infusing
Troubleshooting
Accessories
Rapid Infuser
that the 2 dark blue luer fittings are tight 1- Remove the reservoir chamber 2- Pull out the interlock block from the tray 3- Lift out the disposable holding the reservoir chamber and the heat exchanger
Setup Overview
Setup
Infusing
Troubleshooting
Accessories
Rapid Infuser
Select Reservoir Standard St d dR Reservoiri Part of disposable set 3 bag spikes 120 mL 250 micron filter
LLarge R Reservoiri Separate accessory 5 bag spikes 3L 160 micron filter
Setup Overview
Loading the Tubing Set: Step #1 Load circular heat
Setup
Infusing
Troubleshooting
Accessories
g with red exchanger arrow pointing up and g the red tubingg aligned with the red line on the system y
Rapid Infuser
Setup Overview
Loading the Tubing Set: Step #2 Snap the reservoir chamber
Setup
Infusing
Troubleshooting
Accessories
into the reservoir support, or replace with 3L reservoir & place in large clamp Load L d interlock i l k block bl k iinto airi detector with the arrow pointing into the machine Thread pump tubing over pump head
Rapid Infuser
Setup Overview
Loading the Tubing Set: Step #3 Place pressure chamber
Setup
Infusing
Troubleshooting
Accessories
into pressure transducer cavity (be sure chamber is fully seated into the well) ell) Thread tubing into 2nd air detector Place tubing on both sides of valve
Rapid Infuser