Bien-Air Medical Technology
Contra-Angles and Handpieces for Surgery Maintenance Guide Nov 2016
Maintenance Guide
3 Pages
Preview
Page 1
MAINTENANCE
OF CONTRA-ANGLES AND HANDPIECES FOR SURGERY AND IMPLANTOLOGY. Refer to the user manual for the product for the complete cleaning and sterilisation maintenance procedures. This medical device is supplied in "non sterile" condition. Please ensure that this procedure is followed before the device is used.
1
PREPARATION
A
B
A. REMOVING THE INSTRUMENT • Remove the rotary instrument from the chuck mechanism • Disconnect the contra-angle from the motor B. PRE-DISINFECTION • External disinfection Do not immerse the instrument in a disinfectant bath.
2 A
CLEANING B
A. CLEAN UNDER RUNNING WATER AT <38°C B. REMOVING DIRT/DEPOSITS • Carefully remove all traces of dirt or deposits from the nozzles using the Bien-Air cleaning wire • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth Do not immerse in an ultrasonic cleaner
C
C
C . INTERNAL CLEANING WITH AQUACARE • Shake the Aquacare before use • Use the Aquacare attachment 1 * • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 2 seconds • Use the Aquacare attachment 3 * • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Aquacare button down for 1 second
E
E
D. INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET • Shake the Spraynet before use • Use the Spraynet attachment 1 * • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Spraynet button down for 1 second • Clean the exterior of the contra-angle with a cloth soaked in Spraynet
* All the references for the attachments are given on the last page of the document.
Version 11/16
E
3
DISINFECTION B
A
A. DISINFECTION USING SURFACE DISINFECTANT (follow the manufacturer's instructions for use)
Do not immerse the instrument in a disinfectant bath OR
A
B. THERMAL DISINFECTION (follow the manufacturer's instructions for use)
4
LUBRICATION
A
A
A. LUBRICATION WITH LUBRIFLUID • Shake the Lubrifluid before use • Use the Lubrifluid attachment 1 * • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the Lubrifluid button down for 1 second • Then, run an operating test for 30 seconds at a low speed to remove any surplus lubricant • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
5 A
STERILISATION B
A. PLACE THE CONTRA-ANGLE IN A STERILISATION BAG B. STERILISE IN ACCORDANCE WITH THE MANUFACTURER'S INSTRUCTIONS • Instruments bearing the symbol can be sterilised in an autoclave
Version 11/16
Only sterilise instruments which have been cleaned and lubricated
* All the references for the attachments are given on the last page of the document.
MAINTENANCE ACCESSORIES
SPRAYNET 500 ml cleaning spray Pack of 6
LUBRIFLUID 500 ml lubricant Pack of 6
AQUACARE 500 ml cleaning spray Pack of 6
REF. 1600036-006
REF. 1600064-006
REF. 1600617-006
1
1
1
2
2
2
3
3
4
LUBRIMED Pack of 6 cartridges
BIEN-AIR CLEANING WIRE Pack of 10
REF. 1600915-001
REF. 1600037-006
REF. 1000001-010 Version 11/16
LUBRICARE Maintenance system
Bien-Air Dental SA Länggasse 60 Case postale 2500 Bienne 6, Switzerland Tel. +41 ( 0 )32 344 64 64 Fax +41 ( 0 )32 344 64 91 [email protected] www.bienair.com