Bioptics
faxitron Users Manual Rev 3.0 July 2012
Users Manual
117 Pages

Preview
Page 1
User’s Manual
Document #: 05-5280-04 Version: 3.0 18 July 2012 Page 1 of 117
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
© 2005-2012 Faxitron Bioptics, LLC
All rights reserved. No parts of this work may be reproduced in any form or by any means - graphic, electronic, or mechanical, including photocopying, recording, taping, or information storage and retrieval systems - without the written permission of the publisher. Products that are referred to in this document may be either trademarks and/or registered trademarks of the respective owners. The publisher and the author make no claim to these trademarks. While every precaution has been taken in the preparation of this document, the publisher and the author assume no responsibility for errors or omissions, or for damages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it. In no event shall the publisher and the author be liable for any loss of profit or any other commercial damage caused or alleged to have been caused directly or indirectly by this document.
Printed: JuLY 2012
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
2
Table of Contents ... 2 Forward ... 5 SECTION 1: 1.0.0 1.1.0
User’s Manual – BioVision ... 6 Introduction ... 7 Warning and Caution Symbols in this Manual... 7
SECTION 2: 2.0.0 2.1.0 2.2.0 2.3.0 2.3.1 2.3.2 2.3.3 2.3.4 2.4.0 2.5.0 2.6.0
Radiation Safety and X-Rays ... Introduction ... X-Rays ... Sources of Radiation ... Biological Effects of Radiation... Deterministic and Stochastic Effects ... Radiosensitivity ... Risks from Radiation Exposure ... Conclusions on Health Risks ... ICRP Dose Limits... Risk Management ... References...
8 9 9 9 11 11 12 13 13 13 14 15
SECTION 3: 3.0.0 3.0.1 3.0.2 3.0.3 3.0.4 3.0.5 3.1.0 3.2.0 3.3.0 3.4.0 3.4.1 3.4.2 3.4.3 3.4.4
About Your BioVision ... Overview ... Service and Troubleshooting ... Schedule of Maintenance ... Regulatory Requirements ... Physical Attributes ... Input Power ... Safety ... Shielding and Attenuation ... ALARA ... X-Ray Sub-System ... X-Ray Generating and Safety Components ... X-Ray Control System ... Emergency Stop and Interlocks ... System and Interlock Diagrams ...
16 17 17 17 17 18 18 19 20 20 21 21 26 35 37
SECTION 4: 4.0.0 4.1.0 4.2.0
BioVision Installation and Set-up ... BioVision Specimen DR System Precautions ... BioVision Specimen DR System Setup ... Moving the BioVision Specimen DR System...
39 40 41 42
SECTION 5: 5.0.0 5.1.0 5.1.1 5.1.2 5.2.0 5.2.1 5.2.2 5.2.3
Quick Start – Basic Operation of the BioVision DR System ... Quick Start Overview ... Power Up Sequence ... System Start Up ... Calibration ... Image Acquisition... Entering Patient Data Manually ... Selecting Patient Data from Database ... Selecting Patient Data from a Worklist ...
45 46 47 47 48 50 52 54 55
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
3
5.2.4 Acquiring Images ... 5.2.5 Starting Exposure ... 5.2.6 Image Storage ... 5.2.7 Advanced Software Operation... 5.3.0 System Shut Down ... SECTION 6: 6.0.0 6.1.0 6.2.0 6.3.0 6.4.0 6.4.1 6.4.2 6.4.3 6.4.4 6.4.5 6.4.6 6.4.7 6.4.8 6.4.9 6.4.10 6.4.11 6.4.12 6.4.13 6.4.14 6.5.0 6.5.1 6.5.2 6.5.3 6.5.4 6.5.5 6.5.6 6.5.7 6.5.8 6.5.9 6.5.10 6.5.11 6.5.12 6.6.0 6.7.0 6.8.0 6.9.0 6.10.0
57 58 60 60 60
Vision Software Operation ... 62 Overview of Vision Software ... Software Start-up ... Advanced Calibration ... AEC Set-up ... Advanced Imaging ... Patient Data Entry ... Original Contrast ... Adjust Contrast ... Invert ... Full Resolution... Enhance Image ... Open Database ... PACS Print ... PACS Store ... Back To Procedure ... Device Control Panel ... Exit ... View Display ... Tools... Image Tools ... Original Contrast ... Invert ... Magnify ... Edge Enhance ... ROI Contrast ... ROI Window/L ... Ruler ... ROI Statistics... ROI Histogram... ROI Profile ... Line Profile ... Built-In Window Level ... Database... DICOM Functionality ... User Annotations ... About Help ... About Vision Software ...
63 64 65 71 72 73 78 78 79 80 80 80 81 81 81 81 82 82 84 89 89 89 90 90 90 91 91 92 93 93 94 94 95 104 112 115 115
Index ... 116
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
4
Foreword This manual was created by Faxitron Bioptics, LLC. It is intended to guide the advanced user on how to setup, install and use the BioVision hardware and the Vision Software System, both developed by Faxitron Bioptics, LLC. The user must first setup the hardware, than initiate the Vision Software. Please review the required setup procedures in Section 4 of this manual before continuing.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
5
SECTION Introduction and Symbols
This manual was created by Faxitron Bioptics, LLC. It is intended to guide the user on how to setup, install and use the BioVision hardware and Vision Software, both developed by Faxitron Bioptics, LLC.
1
1.0.0 Introduction
The user must first setup the hardware, than initiate the Vision Software. At a minimum, please review the required setup and safety procedures in this manual before attempting to operate the system. Thank you from all of us at Faxitron Bioptics, LLC.
Additional Information For further technical assistance contact: Faxitron Bioptics, LLC 3440 E. Britannia Dr. Suite# 150 Tucson, AZ 85706 (520) 399-8180 (520) 399-8182 Fax www.faxitron.com
The following symbols are used in this manual and online help system:
Warns the reader of potential electrocution should the reader ignore the warning and deliberately use specialized tools to open a sealed electrical cabinet and handle its contents. Note: No tools are provided by Faxitron Bioptics, LLC.
1.1.0 Warning and Caution Symbols in this Manual
Warns the reader of a hazardous condition. This symbol is generally accompanied by a more specific symbol and appropriate instructions as to the nature of the hazard, the expected outcome of the hazard, and information on avoiding the hazard.
Cautions the reader to use care in lifting heavy equipment in order to avoid personal injury and damage to the equipment.
Warns the reader that ionized radiation is emitted into the labelled area when the x-ray beam is energized. The Faxitron BioVision has safety interlocks to prevent the labelled area from being accessed while the x-ray beam is energized.
Warns the reader to use appropriate protection (such as surgical gloves or an approved specimen container) to avoid exposure to potentially infections materials such as patient tissue samples.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
7
SECTION 2
Radiation Safety and X-Rays
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
2
SECTION Radiation Safety and XRays
Radiation is energy in the form of waves or particles. High frequency, high energy radiation, which has sufficient energy to displace an electron from its orbit around a nucleus, is referred to as ionizing radiation. X-rays, gamma rays, beta particles, alpha particles, and neutrons are all forms of Ionizing radiation and can be given off by radioactive material, stars, and highvoltage equipment.
2
2.0.0 Introduction
We cannot see, feel, taste, smell, or hear ionizing radiation, so monitoring equipment is required for us to detect it.
X-Rays X-rays are high frequency, high energy radiation having sufficient energy to displace an electron from its orbit around a nucleus, and are called “ionizing” radiation. X-rays are capable of traveling long distances through air and most other materials. The amount of energy carried by radiation is directly proportional to the frequency, and inversely proportional to the wavelength. X-rays have a relatively short wavelength and high frequency, thus posses a great deal of energy.
2.1.0 X-rays
X-rays can produce biological changes in tissue that can be beneficial when used in radiation therapy. However, x-rays can also be harmful to biological organisms because of their ability to damage chromosomes. Sources of X-Ray radiation in medical facilities include: X-ray Machines. X-Ray machines are used for treatment (radiation therapy) and diagnostic purposes. Diagnostic X-Ray machines are used to X-Ray various parts of the body, chest, leg, breasts, etc for diagnostic purposes. Today, in the US alone, diagnostic radiology accounts for two- thirds of our dose from man- made sources. X-Ray machines can produce high levels of ionizing radiation. Cabinet X-ray machines. Cabinet x-ray machines are enclosed, self- shielded, interlocked irradiation chambers. The machine can only operate when the chamber door is securely closed. The exposure rates at every location on the exterior meets the rate specified for uncontrolled areas.
Sources of Radiation We are all exposed to radiation every day. According to the NCRP, National Council on Radiation Protection, the average background dose in the United States is 360 mrem/year (3.6 mSv/yr). These exposures are mostly from natural sources of radiation, such as radon, cosmic radiation, and natural deposits in the earth. Even our bodies contain natural radioactivity!
2.2.0 Sources of Radiation
There are 2 sources of radiation: Natural and Man-Made. Natural Sources of Radiation include Cosmic rays, Terrestrial gamma rays, Radionuclides in the body (except radon), and Radon and its decay products. The worldwide average annual effective dose from natural sources is estimated to be 2.4 mSv (240mrem). There are factors that affect the effective doses from these natural sources of radiation: The cosmic ray dose rate depends on height above sea level and latitude, the terrestrial gamma-ray dose rate depends on local geology, and the dose from radon decay products depends on local geology and housing construction and use. See Table 1 for annual effective doses from these natural sources.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
9
SECTION Radiation Safety and XRays
2
2.2.0 Cont. Table 1 Annual effective dose to adults from natural sources Annual effective dose (mSv) Typical Elevated * Cosmic Rays 0.39 2.0 Terrestrial gamma rays 0.46 4.3 Radionuclides in the body (except radon) 0.23 0.6 Radon and its decay products 1.3 10 TOTAL (rounded) 2.4 *The elevated values are representative of large regions. Even higher values occur locally. Source of exposure
Medical X-Rays are a major source of man-made radiation. Medical X-rays utilize ionizing radiation for both the diagnosis and treatment of injuries and disease. UNSCEAR (United Nations Scientific Committee on the Effects of Atomic Radiation) estimates that individual exposure to medical radiation ranges from 0.4 to 1 mSv annually. Exposure from medical radiation exceed those from all other man-made sources, but amount to less than half the exposure to natural background radiation, The doses of ionizing radiation used in diagnostic X-rays are usually quite low, ≤50KeV.
The chart below shows the contributions from various sources of radiation.
Figure 2.2.1 Sources of Radiation Dose in the Unites States. From NCRP 160, Fig. 1.1. Percent contribution of various sources of exposure to the total collective effective dose (1,870,000 person-Sv) and the total effective dose per individual in the U.S. population.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
10
SECTION Radiation Safety and XRays
Exposure to ionizing radiation can have adverse health effects by causing atoms and molecules to change. When ionizing radiation passes through tissue, large amounts of energy are transferred to individual molecules. This energy transfer causes electrons to be dislodged from atoms, initiating a variety of chemical and physical effects. The most critical of these effects is damage to DNA molecules. The body has mechanisms in place that normally repair this damage, but extensive or certain types of damage may not be repaired. When the body is unable to repair the damage the cell may die or be mutated. Mutated cells have the potential to turn into cancers. The extent of damage is proportional to the energy imparted, or the dose of ionizing radiation received.
2
2.3.0 Biological Effects of Radiation
Health effects of exposure to x-ray radiation come in two general types, direct of indirect. Xrays are thought to create radicals in exposed cells of your body that may break or modify chemical bonds within critical biological molecules. As a result (1) cells may be injured or damaged, although many cells repair themselves, resulting in no residual damage, (2) cells may die, which millions of body cells do every day and are replaced in a normal biological process, (3) or cells may incorrectly repair themselves resulting in a biophysical change. Finally, X rays may pass through the body with no interaction. The damage to cells from ionizing radiation can result in: Production of free radicals Breakage of chemical bonds Production of new chemical bonds and cross-linkage between macromolecules Damage to molecules which regulate vital cell processes (e.g. DNA, RNA, proteins) Dose Rate, Total Dose Received, Energy of the radiation, Area of the body exposed, the individual’s sensitivity, and Cell sensitivity, are all factors that determine the biological effects of radiation exposure. There are two types of effects from exposure to ionizing radiation: deterministic effects and stochastic effects.
DETERMINISTIC AND STOCHASTIC EFFECTS Potential biological effects depend on how much and how fast a radiation dose is received. Radiation doses can be grouped into two categories, acute and chronic dose.
2.3.1 Deterministic and Stochastic Effects
Acute Dose and Deterministic Effects Acute radiation exposure is a high radiation dose occurring over a short period of time, normally seconds to days. The effects caused by acute radiation doses are called deterministic; the severity of the effect is determined by the amount of dose received. Deterministic effects usually have some threshold level - below which, the effect will probably not occur, but above which the effect is expected. Above the threshold, the severity of the effect increases as the dose increases. For example, a whole body dose of about 300 rems (3 Sv), 60 times the annual occupational dose limit, if received within a short time period (e.g., a few hours) will cause vomiting and diarrhea within a few hours; loss of hair, fever, and weight loss within a few weeks; and about a 50 percent chance of death without medical treatment. These effects would not occur if the dose 300 rems (3 Sv) were accumulated gradually over many years. There is also a difference between whole body and partial body exposures. Where an acute dose of 600rem (6 Sv) to the hand would cause skin reddening; recovery would occur over the following months and no long-term damage would be expected. However, an acute dose of that magnitude to the whole body could cause death within a short time without medical treatment.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
11
SECTION Radiation Safety and XRays
The utilization of appropriate radiation protection mechanisms and occupational exposure dose limits reduce the likelihood of these effects occurring.
2
2.3.1 Cont.
The occupational dose limits are far below the thresholds for deterministic effects. Chronic Dose and Stochastic Effects A chronic dose is a relatively small amount of radiation received over a long period of time (e.g. years). The body is better equipped to tolerate a chronic dose than an acute dose, since the body has time to repair damage because a smaller percentage of the cells need repair at any given time. The body also has time to replace dead or non-functioning cells with new, healthy cells. However, this does not mean there is no risk related to chronic radiation exposure. The term stochastic means ‘random’, with the implication that low levels of radiation exposure are not certain to produce an effect. These low levels, below the threshold for deterministic effects, may or may not result in detrimental effects to the body. However keep in mind that: 1) there is no threshold level of radiation exposure below which we can say with certainty that effects will NOT occur, and 2) Doubling the radiation dose, doubles the probability that effects will occur. Chronic exposure may produce only effects that can be observed some time following initial exposure. These may include genetic effects and other effects such as cancer, precancerous lesions, benign tumors, cataracts, skin changes, and congenital defects. Generally, the greatest concern is the development of some form of cancer. Immediate effects are not seen below doses of 25 rem (0.25 Sieverts). Latent effects may appear years after a dose is received. It has been found through studies of populations exposed to ionizing radiation (> 0.5 Sv) that the probability of a fatal cancer occurring from radiation exposure is approximately 5 percent per Sv.
RADIOSENSITIVITY Radiosensitivity is the probability of a cell, tissue, or organ suffering an effect per unit dose of radiation. Radiosensitivity is highest in cells which divide rapidly (highly mitotic) or are relatively non-specialized (undifferentiated). For this reason the basal epidermis, bone marrow, thymus, gonads, and lens cells are highly radiosensitive. Muscle, bones, and nervous system tissues have a relative low radiosensitivity. Also, radiosensitivity is greatest during the fetal stage and becomes progressively smaller through adolescence and adulthood.
2.3.2 Radiosensitivity
Basic Law of Radiobiology: Biological effects are directly proportional to the mitotic index and the mitotic future of the exposed cell, and inversely proportional to the degree of differentiation. Mitosis refers to the natural division of a cell nucleus during cell reproduction; differentiation refers to the cell’s degree of specialization to perform a specific function in an organism.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
12
SECTION Radiation Safety and XRays
RISKS FROM RADIATION EXPOSURE It is estimated that approximately 20% (1 in 5) of all deaths in the United States are due to some type of cancer. If every member of a population of 1 million were to receive 10 mrem (0.1 mSv) of radiation, it is possible that 5 additional deaths would be observed. Remember that out of this population of 1 million, about 200,000 will die of cancer, making these few additional deaths statistically impossible to detect. Additionally, the risk of cancer death is 0.08% per rem (10 mSv) for doses received rapidly (acute) and might be 2 times (0.04%, or 4 in 10,000) less than that for doses received over a long period of time (chronic).
2
2.3.3 Risks from Radiation Exposure
From currently available data, the NRC has adopted the risk value for an occupational dose of 1 rem (0.01 Sv) as representing a risk of 4 in 10,000 of developing a fatal cancer. The International Commission on Radiological Protection (Publication 103, 2007) indicates a value of 5.5 % per sievert for cancer and 0.2 % per sievert for heritable effects after exposure to radiation at low dose rate The risk associated with the diagnostic uses of ionizing radiation are normally limited to late stochastic effects, which are estimated to occur at a frequency of perhaps 0.01% for an average examination (deterministic skin damage may occur after fluoroscopy in extreme cases). At the individual level, these risks are almost always small compared to the benefit of diagnosis and treatment.
CONCLUSIONS ON HEALTH RISKS We assume that any radiation exposure, no matter how small, carries with it some risk. However, we know that on average these risks are comparable to or smaller than risks we encounter in other activities or occupations that we consider safe. Since we have extensive control over how much radiation exposure we receive on the job, we control and minimize this risk. The best approach is to keep our dose As Low As Reasonably Achievable, or ALARA – a term we will discuss in detail later. Minimizing the dose minimizes the risk.
2.3.4 Conclusions on Health Risks
ICRP DOSE LIMITS The international Commission on Radiological Protection (ICRP) has established radiation dose limits based on available data. These dose limits are maximum allowed values for whole body exposures. The dose unit of measure is called the Sievert (Sv) which accounts for various factors to determine a quantity called effective dose. The dose limits were last updated in 2007.
2.4.0 ICRP Dose Limits
For members of the public the dose limit is 1mSv/yr (0.001 Sieverts per year) For occupational workers (those who work in the nuclear industry the dose limit is 20 mSv/yr (0.02 Sieverts per year) The dose limits exclude the effects of background radiation.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
13
SECTION Radiation Safety and XRays
ICRP Recommended annual dose limits for individual organs or tissues Radiation weighted dose in Workers Public Lens of the eye 150 mSv 15 mSv Skin 1,2 500 mSv 50 mSv Hands and feet 500 mSv 1 The limitation on effective dose provides sufficient protection for the skin against stochastic effects. An additional limit is needed for localised exposures in order to prevent tissue reactions. 2 Averaged over 1 cm2 area of skin regardless of the area exposed.
Determining Factors The effects of x-ray exposure depend upon the duration of exposure, how fast the dose is delivered, Energy – How much energy was in the x-ray, the total dose – the magnitude of the dose, and whether the exposure is whole body or localized. Low energy (<50KeV) X-rays can cause damage only to skin or outer part of body while high energy X-rays can penetrate the body to the internal organs. A large acute dose delivered at once would have a greater effect than the same dose administered over time as incremental fractions.
2
2.4.0 Cont.
2.5.0 Risk Management
In medical environments risk of radiation exposure can be minimized by: Avoiding all unnecessary exposures to radiation Using Doses that are AS LOW AS REASONABLY ACHIEVABLE (ALARA) Following equipment manufacturers operating instructions Following equipment manufacturers preventive maintenance instructions Utilizing appropriate shielding
ALARA The ALARA concept is based on the assumption that any radiation dose, no matter how small, can have some adverse effect. Under ALARA, every reasonable means of lowering exposure is used. There are three general rules to reduce a person's exposure to any type of ionizing radiation. 1. Reduce the time you are exposed to the radiation source. Reducing the exposure time reduces the radiation dose. 2. Increase the distance between yourself and the radiation source. Increasing the distance from a source of radiation significantly reduces the radiation dose. Doubling the distance from a radiation source means one-fourth the dose rate. Tripling the distance gives one-ninth the rate. 3. Increase shielding between yourself and the radiation source. The use of appropriate shielding greatly reduces dose. The material used and thickness of the shied depends on the source of the radiation. Lead is a common shielding material. Consult your Radiation Safety Officer for the rules and guidelines of specific to your facility.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
14
SECTION Radiation Safety and XRays
REFERENCE SOURCES
2
2.6.0 References
Radiation Safety Training and Reference Manual, California Institute of Technology, Training and Reference Manual, March 1995 (Revised June, 1996) Radiation Safety Manual(Revised March 2010), Stanford University, Veterans Affairs Palo Alto, Health Care System, Environmental Health and Safety, Stanford University, Stanford California Radiation and X-Ray Training - Environmental Health and Safety - University of North Carolina at Chapel Hill USF (University of Southern Florida) Radiation Safety – Research X-Ray Safety Manual, Radiation Safety Office – 2003 University of Oklahoma Chemical Crystallography Lab, 11-APR-2011 European Nuclear Society – Glossary of Nuclear Terms http://www.euronuclear.org/info/encyclopedia.htm United Nations Scientific Committee on the Effects of Atomic Radiation, UNSCEAR 1993, Report to the General Assembly, With Scientific Annexes, UNITED NATIONS PUBLICATION, Sales No. E.94.IX.2, ISBN 92-1-142200-0 JEFFERSON LAB - Thomas Jefferson National Accelerator Facility (Jefferson Lab) Ionizing Radiation Effects and Their Risk to Humans, T.R. Goodman, MD Yale University School of Medicine, New Haven, CT IMAGE WISELY – Radiation Safety in Adult Medical Imaging http://www.imagewisely.org/Imaging-Professionals/Imaging-Physicians/Articles/IonizingRadiation-Effects-and-Their-Risk-to-Humans.aspx Seibert 2004: Seibert J A “X-ray imaging physics for nuclear medicine technologists.” Part1: Basic principles of x-ray production J Nucl Med Technol32 139-47 Shultis 2005: Shultis J K and Faw R E 2005 Radiation shielding technology Health Phys 88 297-322 Simpkin 1995: Simpkin D J 1995 Transmission Data for Shielding Diagnostic X-Ray Facilities Health Phys 68 704-709
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
15
SECTION 3
About Your BioVision System
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
3
SECTION About Your BioVision System
The BioVision Digital Specimen Radiography (DSR) System is a stand-alone cabinet digital X-ray imaging system intended to provide rapid verification that the correct tissue has been excised during percutaneous biopsy.
3
3.0.0 Overview
Performing the verification directly in the same biopsy procedure room enables cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls. This device is intended to be operated wherever the medical professionals deem appropriate, including a surgical suite or a room adjacent to a surgical suite. The BioVision Digital Specimen Radiography System employs the use of Vision, a full featured and powerful image acquisition and data manipulation software. Vision software handles the digital X-ray image acquisition, calibration, image display, image analysis and manipulation, patient database, image archiving, and transmittal. Vision software is the central part of this system. Vision software is Digital Imaging and Communications in Medicine (DICOM) 3.0 compliant and comes with DICOM Print, Store and Modality Work List (MWL).
Service There are no serviceable parts in Faxitron equipment. Please contact your authorized Faxitron representative for servicing.
3.0.1 Service and Troubleshooting
Faxitron Bioptics, LLC 3440 E. Britannia Drive, Suite 150 Tucson, AZ 85706 USA Tel: (520) 399-8180 Fax: (520) 399-8182 E-mail: [email protected] Website: www.faxitron.com
Troubleshooting Refer to the BioVision Technical Manual, contact your Faxitron representative, or contact Faxitron Customer Service.
SCHEDULE OF MAINTENANCE Refer to the BioVision Technical Manual, contact your Faxitron representative, or contact Faxitron Customer Service.
REGULATORY REQUIREMENTS
It is the responsibility of the facility or institution operating this system to ensure that all local, regional, and federal regulations affecting the use of Ionizing Radiation Equipment are followed.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
3.0.2 Schedule of Maintenance
3.0.3 Regulatory Requirements
17
SECTION About Your BioVision System
3
3.0.4 Physical Attributes Figure 3.0.1 Dimensions & Weight
Weight: 200 lbs (90 kg)
INPUT POWER
3.0.5 Input Power
120VAC 60Hz 600W Max Or 230VAC 50Hz 720W Max
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
18
SECTION About Your BioVision System
Cabinet X-Ray Design Safety
3
3.1.0 Safety
The primary hazard of cabinet x-ray equipment is the ionizing radiation produced by this type of equipment and the potential for biological damage as a result of exposure to x-rays. The BioVision DSR has been designed to minimize this risk. The X-ray chamber is self shielding. The sides, back and top are 2mm (0.079in) type 304 Stainless Steel. The chamber floor is 3.6mm (0.14in) type 304 Stainless Steel. The X-ray chamber access door is constructed of a 22mm thick leaded acrylic (lead equivalency of 1mm), a 2mm (0.079in) thick stainless steel back, and plastic frame. There are 2 interlock switches on the front door to prevent the generation of X-rays unless the door is closed. Systems for Canada have an additional set of interlock switches for the systems back cover. The following controls and indicators are located on the systems control panel:
System Power Button
X-Ray On Button
Red Emergency Stop
Power On Indicator (LED)
System Ready indicator (LED)
X-Ray On Indicator (LED)
Error Indicator (LED)
LCD panel
There is an audible alarm to indicate X-rays are being generated. The system imaging software displays system status at the bottom of the screen. Also, messages will be displayed in the image window when; the system is calibrating, X-rays are being generated, or if there are system errors. An X-Ray warning label is installed on the front of the system. A software key (password) is required to open VISION (the imaging software). X-Rays cannot be generated independent of the software. If the software is closed, the password must be reentered.
Emission Limits X-Ray Radiation emitted from the BioVision unit does not exceed an exposure of 0.3 milliRoentgens (mR) per hour at any point 5 centimeters (cm) outside the external surface. For countries requiring emission values measured at 10 cm from the external surfaces, the radiation does not exceed 0.1 mR per hour at 10 cm.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
19
SECTION About Your BioVision System
Shielding The BioVision system is designed to generate ionizing radiation energies at or below 40keV. The stainless steel used in the manufacturer of the BioVision x-Ray chamber is a high-Z material that effectively absorbs approximately 99% of the ionizing radiation at this energy level.
3
3.2.0 Shielding and Attenuation
No additional shielding is required.
Attenuation The primary beam from the BioVision X-ray tube is directed towards the floor of the system. Any reflected or scattered beams will be absorbed by the X-ray chamber top, walls, and door. Typical background radiation measurements at the Faxitron manufacturing facility are 0.1 µSv/hr. Dose measurements are taken at 40keV energy levels with a 25mm acrylic block in the primary beam path to represent maximum system energy levels and maximum beam scatter. Measurements at accessible surfaces are typically between 0.1 and 0.5 µSv/hour.
Faxitron advocates following ALARA, As Low As Reasonably Achievable. Use energy levels no greater than necessary to perform the task. The use of AEC, Automatic Exposure Control, ensures that the lowest ionizing radiation energy levels possible will be generated to obtain optimum image quality. With AEC activated the system will take a short sample exposure at a predetermined energy level and duration. The software will evaluate the sample image and calculate best energy and time settings.
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
3.3.0 ALARA
20
SECTION About Your BioVision System
3
3.4.0 X-Ray Sub-System
X-Ray Sub-System Images
6
3.4.1 X-Ray Generating and Safety Components
1
F i b e r O p ti c C a b le s
3 4
Figure 3.4.1 Location and Identification of XRay Generating Components
5
2
X-Ray Generating Components 1 – X-Ray Tube: Oxford Instruments, Series 5000 2 – X-Ray Chamber Door, Faxitron, 304 Stainless Steel and Leaded Acrylic 3 – X-Ray Chamber: Faxitron, Type 304 Stainless Steel 4 – HVPS, High Voltage Power Supply: Spellman, MNX 50P75 5 – Interlock Switches, Chamber Door: Cherry 6 – Interlock Switches, Back Cover: Cherry (NOTE: At this time only installed on Systems for Canada)
Source-to-Imager (SID) and Source-to-Object (SOD) Distances of BioVision
BIOVISION USERS MANUAL | Faxitron Bioptics, LLC | Doc # 05-5280-04 | v3.0 | 2012
21