User manual 2010-12-21
CuroCell S.A.M.® Static Antidecubitus Mattress User manual item number: 95-001074
1
Contents Important safety precautions
3
Introduction
4
Product description
5
Installation and start-up - model 1
7
Installation and start-up - model 2
10
Cleaning and reconditioning
13
Storage
14
Maintenance
15
Troubleshooting
15
Technical description
16
Spare parts
17
Options and accessories
18
Warranty
19
Other information
19
2
IMPORTANT SAFETY INSTRUCTIONS Read all instructions prior to use WARNING! To minimise the risk of fire, electric shocks or bodily injury: 1 The product must only be used for its intended purpose as described in this manual. The product may not be combined, assembled or repaired with other parts (e.g. pump and mattress), accessories or spare parts other than those described in this manual or other documentation from Care of Sweden. 2 Do not use the pump in wet rooms or where there is a risk of the pump coming into contact with water or another liquid. Never handle a pump that has come into contact with water/liquid. Pull the plug out of the electrical socket immediately and send the product to an authorised service technician for servicing. 3 DO NOT USE in the vicinity of or in contact with fire sources/hot surfaces, such as fire, burning cigarettes or hot lamps. Although the product has undergone fire testing, it could become damaged if it comes into contact with fire sources. 4 Position and use the product in a manner that prevents it from becoming trapped or damaged. NOTE: Be particularly aware of trapping damage when using bed rails. 5
Never use the product if the pump power cord or plug is defective, the pump casing is damaged or it is not functioning properly. Contact a service technician for examination and repair.
6
Regularly check product functionality through a hand check (see section 3.4/4.4). Also check that the comfort setting is correct as indicated in the table on the pump.
7
Continual monitoring is required when the product is used for individuals needing special supervision, such as children.
8
The hygiene cover does not allow liquid or air to penetrate, but is vapour permeable. Make sure that the patient is positioned correctly to avoid the risk of suffocation.
9
Prevent damage to the hygiene cover from sharp objects.
3
1. Introduction CuroCell S.A.M.® is a static air mattress that can be used in all kinds of caring environments and it is very functional and cost effective. Always read the user manual prior to use.
1.1 General information This system is a medical device with CE-marking in accordance with MDD 93/42/EEC. This system has been tested and approved according to the following standards: EN 60601-1 SS-EN ISO 10993 SS 876 00 04 EN 597-1 SS-EN ISO 12182 EN 597-2 EN 14126 SS-EN ISO 14971
1.2 Intended use CuroCell S.A.M.® can be used in all types of medical environments and is intended to assist in the prevention and treatment of pressure ulcers, up to category 2(1). The recommended user’s weight is up to 200 kg. It is available in two models. As an overlay mattress (CuroCell S.A.M.®) to be placed on top of an existing mattress or as a mattress replacement (CuroCell S.A.M.® BAS) The mattress must always be cleaned between patients.
NOTE! - When used together with positioning pillows, read the pillow user manual carefully for correct positioning of the patient. - If you are using this product with an evacuation sheet, it is the user’s responsibility to ensure that it is safe to use for evacuation. - Be careful when using bed rails or other protection on the bed so that the mattress does not become crushed or damaged. - The mattress must not be lifted with a person lying on it or used for any other type of transport other than a possible evacuation using an evacuation sheet. (1) European Pressure Ulcer Advisory Panel, Quick reference guide 2009.
4
2. Product description 2.1 Pump Front 1. Power switch On/Off (1/0). A green light indicates when the pump is on. 2. Front panel
3. Air connection
Backside 1. Bed hooks
2. Power cord
2.2 Front panel Comfort settings The comfort setting adjusts the air pressure in the mattress. When the knob is turned clockwise, the air supply and pressure in the mattress increase. When the knob is turned counter clockwise, the air supply and pressure in the mattress decrease. Appropriate air pressure is set between 1 and 8. The table on the pump gives guidance.
5
2.3 Mattress IMPORTANT! There are two available models of CuroCell S.A.M.® and they are described below as model 1 and model 2. Model 1, where the support pump always is connected. 1. Cells/air bladder. 2. Tubes. 3. Tube holder. 4. CPR (Rapid deflation)
1
2
3
4
Model 2, where the support pump, if requested, can be disconnected. 1. Cells/air bladder. 2. Tubes. 5. Air/connections tubes.
3. Tube holder.
4. Connections with check valves.
1
2
3
5
2.4 Other parts Hygiene cover Inner cover Soft (accessory) Transport and Storage Bag (accessory) Pump bag (accessory) Integrated foam core (only CuroCell S.A.M.® BAS)
6
4
3. Installation and Start up - Model 1 IMPORTANT! This part only concerns model 1, the model where the support pump always is connected.
3.1 Installation When unpacking the mattress, check that none of the parts are damaged. In the event of damage, contact Care of Sweden or your local dealer before using the product. 1. Place the mattress on top of an existing mattress. If possible, secure it by pulling the elastic bands on the bottom over the corners of the underlying mattress. Note! If the mattress have an integrated foam core (CuroCell S.A.M.® BAS), place the mattress directly on to the bed base. If requested use the straps on the cover to fasten the mattress to the bed base. 2. Hang the pump by the foot end of the bed, or place it on a flat sturdy surface. Place the power cord so there is no risk of stumbling over it, run it over with the bed wheels, or getting it jammed when raising or lowering the bed.
NOTE! A pump bag is available as an option.
3. Connect the air tubes coming from the mattress to the side of the pump.
4. Check to make sure that the power is set to ’0’ (off). Plug in the power cord into an approved outlet (220 V).
NOTE! - Make up the bed with sheets to increase user comfort. - Check that the CPR-valve is closed.
7
3.2 Start-up NOTE! Always read the user manual prior to use.
1.
1. Start the pump by pressing ’1’ (on), on the side of the pump. The green light is lit. 2. The pump starts filling the mattress with air. It takes approximately 15-20 minutes before the mattress is ready to use.
3.3 Comfort settings The air pressure in the mattress can be adjusted according to user preference. Always consult the prescriber or other authorized personnel for the correct setting. When the mattress is filled, it is ready to be used. Adjust the pressure with the comfort setting knob according to the table on the pump.
NOTE! - When the head end of the bed is raised, the pressure should be increased at least one (1) step. - For amputated users, when only part of the mattress is under load, the pressure should be increased by 1-2 steps. - Recommended users weight is up to 200 kg.
3.4 Hand check Hand check is used to ensure that the mattress is properly inflated. This should be done regularly, once per shift, after changing the position in the bed or changing the comfort settings. Open the cover. Place a flat hand under the air cells directly underneath the user’s pelvic region and make sure that the user does not lay directly on the underlying mattress. There should be about 1-2 fingers margin. If you can feel the user´s body resting on your hand you need to increase the comfort settings to next highest step. Wait 10 minutes and perform a new hand check.
8
3.5 CPR In case of an emergency where CPR (cardiopulmonary resuscitation) is necessary: Rapidly pull the CPR band on the tubes between the pump and the mattress. Remove the tube connections on the side of the pump, for even faster deflation. Note! Make sure that the CPR connector is properly connected after use.
3.6 Low pressure By an abnormally low air pressure, see chapter 8 for trouble shooting.
WARNING! If the air pressure remains low, search for damages on the mattress and the tubes. Repair when needed or contact the supplier.
9
4. Installation and Start up - Model 2 IMPORTANT! This part only concerns model 2, the model where the support pump, if requested, can be disconnected.
4.1 Installation When unpacking the mattress, check that no parts are damaged. If damages are detected, contact Care of Sweden or your local dealer before using the product. 1. Place the mattress on top of an existing mattress. If possible, secure it by pulling the elastic bands on the bottom over the corners of the underlying mattress. Note! If the mattress have an integrated foam core (CuroCell S.A.M.® BAS), place the mattress directly on to the bed base. If requested use the straps on the cover to fasten the mattress to the bed base. 2. Connect the connection tubes to the mattress. These are to be connected by the foot end of the mattress and can be found by opening the hygiene cover. When they are connected, close the zip fastener. 3. Hang the pump by the foot end of the bed, or place it on a flat sturdy surface. Place the power cord so there is no risk of stumbling over it, run it over with the bed wheels, or getting it jammed when raising or lowering the bed. NOTE! A pump bag is available as an option.
4. Connect the air tubes coming from the mattress to the side of the pump. 5. Check to make sure that the power is set to ’0’ (off). Plug in the power cord into an approved outlet (220 V).
NOTE! Make up the bed with sheets to increase user comfort.
10
4.2 Start-up and settings NOTE! Always read the user manual prior to use. 1. Start the pump by pressing ’1’ (on), on the side of the pump. The green light is lit. 2. The pump starts filling the mattress with air. 3. The mattress is ready to be used when it is filled. This takes approximately 15-20 minutes. 4. Set the appropriate pressure with the comfort knob, see settings table on the pump. 5. Check the set pressure with the ‘hand-check’, adjust if needed. See chapter 4.3 and 4.4 for more information. 6. When the pressure is set according to preference, there are two options: a) Disconnect the air/connection tubes from the mattress and turn off the pump. The mattress is equipped with check valves, which prevent the air from leaving the mattress and the set pressure remains. b) Keep the pump connected and running.
NOTE! - The mattress and pump shall be checked regularly. Check mattress by using the ‘hand- check’ (see chapter 4.4). If needed, reconnect the pump and adjust the pressure using the pump. - If the pump is not constantly connected to the mattress, we recommend that the pump is connected once a week and step 5 and 6 in chapter 4.2 are repeated.
4.3 Comfort settings The air pressure in the mattress can be adjusted according to user preference. Always consult the prescriber or other authorized personnel for the correct setting.
NOTE! - When the head end of the bed is raised, the pressure should be increased by step one (1). - For amputated users, when only part of the mattress is under load, the pressure should be increased (by 1-2 steps).
11
1.
4.4 Hand check Hand check is used to ensure that the mattress is properly inflated. This should be done regularly, once per shift, after changing the position in the bed or changing the comfort settings. Open the cover. Place a flat hand under the air cells directly underneath the user’s pelvic region and make sure that the user does not lay directly on the underlying mattress. There should be about 1-2 fingers margin. If you can feel the user´s body resting on your hand you need to increase the comfort settings to next highest step. Wait 10 minutes and perform a new hand check.
4.5 Deflation of air (CPR) At CPR (cardiopulmonary resuscitation) the mattress is emptied by connecting the air tubes to the mattress without the tubes being connected to the pump. The same procedure is used after use.
4.6 Low pressure By an abnormally low air pressure, see chapter 8 for Trouble Shooting.
WARNING! If the air pressure remains low, search for damages on the mattress and the tubes. Repair when needed or contact the supplier.
12
5. Cleaning and reconditioning It is important to follow the instructions below before using the product for another user. Consult the hygiene responsible or the supplier for help and instructions when unsure.
5.1 Normal cleaning PUMP
VARNING! Do not open the pump housing – risk of electric shock. Service and maintenance must be performed by authorised technicians. Do not get the pump wet or submerge it in any liquid. Wipe off the pump with a damp cloth and mild cleaner, such as washing up liquid. If necessary, use an alcohol-based disinfectant/cleaner intended for this purpose. Perform, Virkon or Desisoft can also be used. If using a different agent, choose one that does not affect the plastic of the pump housing. MATTRESS COVER Covers can be wiped off with a mild cleaner (e.g. washing up liquid) and water or with an alcohol-based disinfectant/cleaner intended for this purpose. Perform, Virkon or Desisoft can also be used. Excrement and blood spots must be removed immediately with cold water. Carefully follow local instructions and usage instructions for the cleaner.
Multi-layer covers should be divided before washing. Wash with like colors. Chlorine and phenol-based cleaners could have a negative impact on the PU surface and should be avoided. If chlorine is used, we recommend a mixture of max. 1%. INNER COVER SOFT (accessories)
Clean the affected area using a mild cleaner (e.g. washing up liquid) and water or with an alcohol-based disinfectant/cleaner intended for this purpose. Perform, Virkon or Desisoft can also be used. Press out the water carefully.
13
5.2 Reconditioning PUMP See section 5.1 for instructions. MATTRESS 1. Disconnect the tubes and let the air out, also open the CPR valve. 2. Remove the cover. See instructions in chapter 5.1 for cleaning covers. If the product is equipped with a foam core this should be removed and cleaned separately. 3. Move the mattress to a clean area, which is suitable for cleaning. 4. Wipe off and disinfect the cells, all tubes and the CPR valve with an alcohol-based disinfectant/cleaner intended for this purpose. Perform, Virkon or Desisoft can also be used. Let it dry. 5. Wipe off and disinfect all the work surfaces with an alcohol-based cleaning agent intended for this purpose. 6. Put the mattress together. If cells for any reason have loosened from the tubes, these must be replaced according to drawing in chapter 2.3. FOAM CORE (CuroCell S.A.M.® BAS) Clean the affected area with a mild cleaner (e.g. washing up liquid) and water or with an alcohol-based disinfectant/cleaner intended for this purpose. Perform, Virkon or Desisoft can also be used. Press out the water carefully. Let the foam core dry in a warm, ventilated area, without direct exposure of sunlight. The foam core needs to be completely dry before it is used again.
NOTE! Check the hygiene cover each time the product is cleaned. If damaged, it must be replaced or repaired. Also check the pump, tube connectors and power cord during cleaning. Damaged parts must be replaced or repaired.
14
6. Storage and packing It is advisable to store the mattress and pump in the product bag (accessory). Handle the packed product very carefully. Do not place any heavy objects on top of it.
7. Maintenance 7.1 General 1. Check the power cord and plug for damages and scratches. 2. Check the mattress cover to make sure it is undamaged and that cover and cells are correctly assembled. 3. Check that air is coming from the outlets on the side of the pump. 4. Check that no tubes or connectors are damaged or jammed. Contact the manufacturer or your local dealer for spare parts.
15
8. Trouble shooting Problem The pump does not start
Solution Check that the power cord is connected to an electrical outlet. Check that all fuses are OK.
The user lies directly on the bed
Pressure might be incorrectly set. Adjust it according to the weight table, and wait a few minutes. Check again using the ‘hand check’, see chapter 3.4/4.4. NOT! If you use model 2 the support pump must be connected first. Check that the CPR-valve is closed (only for model 1)
The mattress moves around
Check that the mattress is secured with the elastic bands to the underlying mattress. Or, if the mattress have an integrated foam core (CuroCell S.A.M.® BAS), secured by using the straps on the cover to fasten the mattress to the bed base.
The pump makes a noise, or vibrates
Check that the pump hangs correctly on the bed. Resonance in parts of the bed can occur. Remove the pump and listen for a difference. This can be fixed by placing the pump on the floor.
If your questions are not answered by the above information, please contact Care of Sweden or your local dealer for further assistance.
16
9. Technical description 9.1 Technical specification Pump Specifi kation PUMP Model Voltage Current consumption Fuse type Dimensions (LxWxH) Weight Surroundings
SPECIFICATION CuroCell S.A.M.® 220V-240 V / 50 Hz 0.05 A 1A, 250 V 25 x 12.5 x 8.5 cm 1.4 kg Operations: 10°C ˜ 40°C Storage: -15°C ˜ 50°C Shipping: -15°C ˜ 70°C Operations: 10% ˜ 90% non condensing Storage: 10% ˜ 90% non condensing Class II, type BF, IPX0
Temperature
Humidity Classification
MATTRESS
SPECIFICATION
Model Dimensions (BxWxH) Weight
CuroCell S.A.M.®, CuroCell S.A.M.® BAS 80/85/90/105/120x200/210x10/16 cm 2.7 kg
NOTE! The manufacturer can at any time change the specifications for the product.
WARNING! Follow the operations and maintenance instructions carefully in order to prevent bodily injuries, as well as damages to the product or other equipment.
9.2 Labeling The mattress is labeled with product name, article number, production date, size, bar code, as well as care and wash instructions. The product also displays what side is up/down and head/foot.
17
9.3 Symbol Definitions XXXX-XX-XX Year-Month-Day
Read the user manual
Grounded contact
CE marked acc. to MDD 93/42/ECC
Class II equipment (Double Insulated)
Type BF equipment
Fire classification
Patient information – category
Foot placement
Recommended user weight
Recommended minimum length
The mattress should only be used lying in the lengthwise direction on the mattress
Not in domestic refuse, follow recycling instructions
Recycling
Length of warranty
Directly on bed base
Place on top of existing mattress
Can be rotated
Can be turned
Not directly on bed base
Not on top of existing mattress
Do not rotate
Do not turn
Place on top
Place underneath
Wipe off
Do not iron
Machine wash at 70° C
Machine wash at 95° C
Tumble dry
Drip dry
Do not tumble dry
Do not dry clean
Chlorine
18
Do not use chlorine
10. Spare parts Art. no CC-900151 CC-900152 CC-900153 CC-900154 CC-900155 CC-900150 CC-900156 CC-900158
Description Air cells, not incl. tubes. ” ” ” ” ” ” ”
CC-908032 CC-908033
Tubes, incl. CPR. Tubes, incl. connection tube.
Size 80x200 cm 80x210 cm 85x200 cm 85x210 cm 90x200 cm 90x210 cm 105x200 cm 120x200 cm
If other spare parts are needed, please contact Care of Sweden or your local dealer for further assistance.
19
11. Options and accessories 11.1 Options Inner cover Soft Enhances the comfort further, can also be used for very sensitive users. This is included for CuroCell S.A.M.® BAS. Sturdy transportation bag Transportation and storage bag. Cover With welded seams.
11.2 Accessories CC-9091 CC-9093 58-900909 CC-9098
Transportation and Storage Bag Transportation and Storage Bag to CuroCell S.A.M.® with Inner cover Soft. Transportation and Storage Bag to CuroCell S.A.M.® BAS. Pump bag, black
To CuroCell S.A.M.® Cover Olivia (blue) Art. no Size 294-080200 80x200 cm 294-080210 80x210 cm 294-085200 85x200 cm 294-085210 85x210 cm 294-090200 90x200 cm 294-090210 90x210 cm 294-105200 105x200 cm 294-120200 120x200 cm
Inner cover Soft Art. no Size CC-928040 80x200 cm CC-928041 80x210 cm CC-928043 85x200 cm CC-928044 85x210 cm CC-928046 90x200 cm CC-928047 90x210 cm CC-928048 105x200 cm CC-928049 120x200 cm
To CuroCell S.A.M.® BAS Cover top part, grey Art. no Size 441-080200 80x200 cm 441-080210 80x210 cm 441-085200 85x200 cm 441-085210 85x210 cm 441-090200 90x200 cm 441-090210 90x210 cm 441-105200 105x200 cm 441-120200 120x200 cm
Inner cover Soft with foam pocket Art. no Size CC-928060 80x200 cm CC-928061 80x210 cm CC-928063 85x200 cm CC-928064 85x210 cm CC-928066 90x200 cm CC-928067 90x210 cm CC-928068 105x200 cm CC-928069 120x200 cm 20