CuroCell S.A.M. Static Antidecubitus Mattress User Manual Sept 2009
User Manual
19 Pages
Preview
Page 1
Contents Importantsafetyprecautions
3
Introduction
4
Productdescription
5
InstallationandstartͲup model1
7
InstallationandstartͲup model2
10
Cleaningandreconditioning
13
Storage
14
Maintenance
15
Troubleshooting
15
Technicaldescription
16
Spareparts
17
Optionsandaccessories
18
Warranty
19
Otherinformation
19
2
IMPORTANTSAFETYINSTRUCTIONS
Readallinstructionspriortousingthemattress WARNING!Tominimizetheriskoffire,electricalshockorbodilyinjury. 1. Theproductmustonlybeusedforitsintendedpurposeasdescribedinthis manual.OnlyuseaccessoriesthathavebeenapprovedbyCareofSweden. 2. Donotusetheproductinwetroomsorwherethereisariskofcominginto contactwithwateroranotherliquid.Neverhandleaproductthathascome intocontactwithwater/liquid.Pulltheplugoutoftheelectricalsocket immediatelyandsendtheproducttoanauthorizedservicetechnicianfor servicing. 3. DONOTUSEinthevicinityoforincontactwithfiresources/hotsurfaces,suchas fire,burningcigarettesorhotlamps.Althoughtheproducthasundergonefire testing,itcouldbecomedamagedifitcomesintocontactwithfiresources. 4. Positionandusetheproductinamannerthatpreventsitfrombecomingtrapped ordamaged.NOTE:Beparticularlyawareoftrappingdamagewhenusingbed rails. 5. Neverusetheproductifthepumppowercordorplugisdefective,thepump casingisdamagedoritisnotfunctioningproperly.Contactaservicetechnicianfor examinationandrepair. 6. Regularlycheckproductfunctionalitythroughahandcheck(seesection3.4/4.4). Alsocheckthatthecomfortsettingiscorrectasindicatedinthetableonthe pump. 7. Continualmonitoringisrequiredwhentheproductisusedforindividualsneeding specialsupervision,suchaschildren. 8. Thehygienecoverdoesnotallowliquidorairtopenetrate,butisvapour permeable.Makesurethatthepatientispositionedcorrectlytoavoidtheriskof suffocation. 9. Preventdamagetothehygienecoverfromsharpobjects.
3
1.Introduction CuroCellS.A.M.®isastaticairmattressthatcanbeusedinallkindsofcaringenvironments anditisveryfunctionalandcosteffective.Itisanoverlaymattress,tobeplacedontopofan existingmattress.Itisusedpreventativeandasanaidintreatmentofpressureulcersupto degree2. Usethismanualwhenthesystemisinstalledandstartedforthefirsttime,andkeepitfor futurereference.
1.1Generalinformation Thissystemhasbeentestedandapprovedaccordingtothefollowingstandards: EN60601Ͳ1 SSͲENISO10993 SS8760004 EN597Ͳ1 SSͲENISO12182 EN597Ͳ2 SSͲENISO14971 EN14126
1.2Intendeduse CuroCellS.A.M.®isastaticairmattressandisintendedtobeusedtominimizetheriskof pressureulcers.Itisusedpreventativeandasanaidintreatmentofpressureulcersupto degree2anditcanbeusedinallkindsofcaringenvironments.Themattressisintendedfor userweightsupto200kg.Themattressmustalwaysbecleanedbetweenpatients.
NOTE! ͲWhenusedtogetherwithpositioningpillows,readtheusermanualforthepillowscarefullyin ordertoplacetheusercorrectly. ͲItisnotrecommendedtousethisproducttogetherwithevacuationsheets. ͲBecarefulwhenusingbedrailsorotherprotectionsonthebed,sothemattressdoesnotget damaged. ͲThemattressmustnotbeliftedwithauseronthemattress,orusedforanykindof transportationoftheuser. ͲKeepinmindthattheunderlyingmattressisimportantforthefunctionofthisproduct.
4
2.Productdescription
2.1Pump Front
1.Powerswitch Usedtoturnthepumpoffandon.Agreen lightindicateswhenthepumpison. 2.Frontpanel 3.Airconnection
Backside 1.Bedhooks 2.Powercord
2.2Frontpanel Comfortsettings Thecomfortsettingadjuststheairpressurein themattress.Whentheknobisturned clockwise,theairsupplyandpressurein themattressincrease.Whentheknobis turnedcounterclockwise,theairsupplyand pressureinthemattressdecrease. Appropriateairpressureissetbetween1and8. Thetableonthepumpgivesguidance.
5
2.3Mattress IMPORTANT!TherearetwoavailablemodelsofCuroCellS.A.M.®andtheyaredescribed belowasmodel1andmodel2. Model1,wherethesupportpumpalwaysisconnected. 1.Cells/airbladder.2.Tubes.3.Tubeholder.4.CPR(Rapiddeflation)
1
2
3
4
Model2,wherethesupportpump,ifrequested,canbedisconnected. 1.Cells/airbladder.2.Tubes.3.Tubeholder.4.Connectionswithcheckvalves. 5.Air/connectionstubes.
2.4Otherparts Hygienecover TransportandStorageBag(accessory)
1 5
2
3
InnercoverSoft(accessory) Pumpbag(accessory)
6
4
3.InstallationandStartupͲModel1. IMPORTANT!Thispartonlyconcernsmodel1,themodelwherethesupportpumpalwaysis connected.
3.1Installation Whenunpackingthemattress,checkthatnoneofthepartsaredamaged.Intheeventof damage,contactCareofSwedenABoryourlocaldealerbeforeusingtheproduct. 1.Placethemattressontopofanexistingmattress.Ifpossible,secureitbypullingtheelastic bandsonthebottomoverthecornersoftheunderlyingmattress. 2.Hangthepumpbythefootendofthebed,orplaceit onaflatsturdysurface.Placethepowercordsothere isnoriskofstumblingoverit,runitoverwiththebed wheels,orgettingitjammedwhenraisingorlowering thebed. NOTE! Apumpbagisavailableas anoption. 3.Connecttheairtubescomingfromthemattressto thesideofthepump. 4.Checktomakesurethatthepowerissetto’0’(off). Pluginthepowercordintoanapprovedoutlet(220V).
NOTE! ͲMakeupthebedwithsheetstoincreaseusercomfort. ͲCheckthattheCPRͲvalveisclosed.
7
3.2StartͲup
NOTE!Alwaysreadtheusermanualpriortousage. 1.Startthepumpbypressing’1’(on),onthesideofthepump.Thegreenlightislit. 2.Thepumpstartsfillingthemattresswithair.Ittakesapproximately15Ͳ20minutesbefore themattressisreadytouse.
3.3Comfortsettings Theairpressureinthemattresscanbeadjustedaccordingtouserpreference.Alwaysconsult theprescriberorotherauthorizedpersonnelforthecorrectsetting. Whenthemattressisfilled,itisreadytobeused.Placetheuseronthemattressandadjust thepressurewiththecomfortsettingknobaccordingtothetableonthepump.
NOTE! ͲWhentheheadendofthebedisraised,thepressureshouldbeincreasedatleastone(1)step. ͲForamputatedusers,whenonlypartofthemattressisunderload,thepressureshouldbe increasedby1Ͳ2steps. ͲRecommendedusersweight is 0Ͳ200kg.
3.4Handcheck Handcheckisusedtoensurethatthemattressisproperlyinflated.Thisshouldbedone regularly,oncepershift,afterchangingthepositioninthebedorchangingthecomfort settings. Openthecover.Placeaflathandundertheaircellsdirectlyunderneaththeuser’spelvic regionandmakesurethattheuserdoesnotlaydirectlyontheunderlyingmattress.There shouldbeabout1Ͳ2fingersmargin. Ifyoucanfeeltheuser´sbodyrestingonyourhandyouneedtoincreasethecomfortsettings tonexthigheststep.Wait10minutesandperformanewhandcheck. 8
3.5CPR IncaseofanemergencywhereCPR(cardiopulmonaryresuscitation)is necessary:RapidlypulltheCPRbandonthetubesbetweenthepump andthemattress.Removethetubeconnectionsonthesideofthe pump,forevenfasterdeflation. Note!MakesurethattheCPRconnectorisproperlyconnectedafteruse.
3.6Lowpressure Byanabnormallylowairpressure,seechapter8fortroubleshooting.
WARNING! Iftheairpressureremainslow,searchfordamagesonthemattressandthetubes.Repairwhen neededorcontactthesupplier.
9
4.InstallationandStartupͲModel2. IMPORTANT!Thispartonlyconcernsmodel2,themodelwherethesupportpump,if requested,canbedisconnected.
4.1Installation Whenunpackingthemattress,checkthatnopartsaredamaged.Ifdamagesaredetected, contactthesupplierbeforeusingtheproduct. 1.Placethemattressontopofanexistingmattress.Ifpossible,secureitbypullingtheelastic bandsonthebottomoverthecornersoftheunderlyingmattress 2.Connecttheconnectiontubestothemattress.Thesearetobeconnectedbythefootendof themattressandcanbefoundbyopeningthehygienecover.Whentheyareconnected,close thezipfastener. 3.Hangthepumpbythefootendofthebed,orplaceit onaflatsturdysurface.Placethepowercordsothere isnoriskofstumblingoverit,runitoverwiththebed wheels,orgettingitjammedwhenraisingorlowering thebed. NOTE! Apumpbagisavailableas anoption. 4.Connecttheairtubescomingfromthemattressto thesideofthepump. 5.Checktomakesurethatthepowerissetto’0’(off). Pluginthepowercordintoanapprovedoutlet(220V).
NOTE! ͲMakeupthebedwithsheetstoincreaseusercomfort. 10
4.2StartͲupandsettings NOTE!Alwaysreadtheusermanualpriortousage. 1.Startthepumpbypressing’1’(on),onthesideofthepump.Thegreenlightislit. 2.Thepumpstartsfillingthemattresswithair. 3.Themattressisreadytobeusedwhenitisfilled.Thistakesapproximately15Ͳ20minutes. 4.Placetheuseronthemattress. 5.Settheappropriatepressurewiththecomfortknob,seesettingstableonthepump. 6.Checkthesetpressurewiththe‘handͲcheck’,adjustifneeded.Seechapter4.3and4.4for moreinformation. 7.Whenthepressureissetaccordingtopreference,therearetwooptions: a) Disconnecttheair/connectiontubesfromthemattressandturnoffthepump.The mattressisequippedwithcheckvalves,whichpreventtheairfromleavingthe mattressandthesetpressureremains. b) Keepthepumpconnectedandrunning.
NOTE! ͲThemattressandpumpshallbecheckedregularly.Checkmattressbyusingthe‘handͲcheck’ (seechapter4.4).Ifneeded,reconnectthepumpandadjustthepressureusingthepump. ͲIfthepumpisnotconstantlyconnectedtothemattress,werecommendthatthepumpis connectedonceaweekandstep5and6inchapter4.2arerepeated.
4.3Comfortsettings Theairpressureinthemattresscanbeadjustedaccordingtouserpreference.Alwaysconsult theprescriberorotherauthorizedpersonnelforthecorrectsetting.
NOTE! ͲWhentheheadendofthebedisraised,thepressureshouldbeincreasedbyone(1). ͲForamputatedusers,whenonlypartofthemattressisunderload,thepressureshouldbe increased(by1Ͳ2steps). ͲRecommendeduserweight:0Ͳ200 kg.
11
4.4Handcheck Handcheckisusedtoensurethatthemattressisproperlyinflated.Thisshouldbedone regularly,oncepershift,afterchangingthepositioninthebedorchangingthecomfort settings. Openthecover.Placeaflathandundertheaircellsdirectlyunderneaththeuser’spelvic regionandmakesurethattheuserdoesnotlaydirectlyontheunderlyingmattress.There shouldbeabout1Ͳ2fingersmargin. Ifyoucanfeeltheuser´sbodyrestingonyourhandyouneedtoincreasethecomfortsettings tonexthigheststep.Wait10minutesandperformanewhandcheck.
4.5Deflationofair(CPR) Afterusingthemattress,itisemptiedbyconnectingtheairtubestothe mattress,withoutthetubesbeingconnectedtothepump.Thesame procedureisusedif,forexample,rapiddeflationforCPRisneeded.
4.6Lowpressure Byanabnormallylowairpressure,seechapter8forTroubleShooting.
WARNING! Iftheairpressureremainslow,searchfordamagesonthemattressandthetubes.Repairwhen neededorcontactthesupplier.
12
5.Cleaningandreconditioning Itisimportanttofollowtheinstructionsbelowbeforeusingtheproductforadifferentuser. Consultthehygieneresponsibleorthesupplierforhelpandinstructionswhenunsure.
5.1Normalcleaning Pump
WARNING! DonotopenthepumphousingͲdangerofelectricalshock.Serviceandmaintenanceshouldbe performedbyanauthorizedservicetechnician.
WARNING! Donotgetthepumpwet! Wipeoffthepumpwithadampclothandamildcleaningagent.Ifneeded,useanalcoholͲ basedagenttodisinfect.Ifotheragentsareneeded,chooseonethatwillnoteffecttheplastic casingofthepump. Mattresscover Wipeoffwithamildcleaningagentandwater,ordisinfectwithanalcoholͲbasedcleaning agent.Excrementandbloodspotsaretoberemovedimmediatelywithcoldwater.Followthe localinstructionsandthecleaningagents’instructionscarefully. Chlorideandphenolicbasedproductsshouldbeavoidedforcleaning,sincetheycaneffectthe coversurfaceinanegativeway.Ifchloridestillisused,werecommenddilutingittomaximum 1%. Thecovercanbemachinewashedinmax.95°C,tumbledorhangeddry. Innercover(accessory) Canbemachinewashedinmax.60Ͳ70°C,tumbledorhangeddry. 13
5.2Reconditioning Pump Seeinstructionchapter5.1. Mattress 1.Removethecoverandinnercover(ifpresent). 2.Disconnectthetubesfromthepumpandlettheairoutofthemattress. 3.Movethemattresstoacleanandsuitablecleaningarea. 4.Wipeoffthecells,accessibletubesandCPRͲvalvewithanalcoholͲbasedcleaningagent.Let itdry. 5.WipeoffthecleaningareawithanalcoholͲbasedcleaningagent. Mattresscover Seeinstructionchapter5.1.
NOTE! Checkthehygienecovereverytimeitiscleaned.Ifthecoverisdamaged,itmustbereplacedor repaired.Alsocheckthepumpandpowercordwhencleaning.Damagedpartsmustberepaired orreplaced.
6.Storageandpacking Itisadvisabletostorethemattressandpumpintheproductbag(accessory).Handlethe packedproductverycarefully.Donotplaceanyheavyobjectsontopofit.
14
7.Maintenance
7.1General 1.Checkthepowercordandplugfordamagesandscratches. 2.Checkthemattresscovertomakesureitisundamagedandthatcoverandcellsare correctlyassembled. 3.Checkthatairiscomingfromtheoutletsonthesideofthepump. 4.Checkthatnotubesorconnectorsaredamagedorjammed. ContactCareofSwedenABoryourlocaldealerforspareparts.
8.Troubleshooting Problem Solution Thepumpdoesnotstart Checkthatthepowercordisconnectedtoanelectricaloutlet. CheckthatallfusesareOK. Theuserliesdirectly Pressuremightbeincorrectlyset.Adjustitaccordingtothe onthebed weighttable,andwaitafewminutes.Checkagainusingthe ‘handcheck’,seechapter3.4/4.4. NOT!Ifyouusemodel2thesupportpumpmustbeconnected first. CheckthattheCPRͲvalveisclosed(onlyformodel1) Themattressmovesaround Checkthatthemattressissecuredwiththeelasticbandstothe underlyingmattress. Thepumpmakesanoise, Checkthatthepumphangscorrectlyonthebed.Resonance orvibrates inpartsofthebedcanoccur.Removethepumpandlistenfora difference.Thiscanbefixedbyplacingthepumponthefloor. Ifyourquestionsarenotansweredbytheaboveinformation,pleasecontactCareofSweden ABoryourlocaldealerforfurtherassistance.
15
9.Technicaldescription
9.1Technicalspecification PumpSpecifikation
Pump
Specification
Model
CuroCell S.A.M.
Voltage
220-240V / 50 hz
Current consumption Fuse type
0,05 A 1A, 250V
Dimensions (LxWxH) Weight
25 x 12,5 x 8,5 cm 1,4 kg
Surroundings
Temperature
Operations: 10°C ~ 40° C Storage: -15°C ~ 50°C Shipping: -15°C ~ 70°C
Humidity
Operations: 10% ~ 90% non condensing
Classification
Storage: 10% ~ 90% non condensing Class ll, type BF, IPX0
Mattress
Specification
Model Dimensions (BxLxH)
CuroCell S.A.M. 80x200/210x10, 85x200/ 210x10 90x200/210x10 cm
Weight
2,7 kg
NOTE!Themanufacturercanatanytimechangethespecificationsfortheproduct.
WARNING! Followtheoperationsandmaintenanceinstructionscarefullyinordertopreventbodilyinjuries, aswellasdamagestotheproductorotherequipment.
9.2Labeling Themattressislabeledwithproductname,articlenumber,productiondate,size,barcode,as wellascareandwashinstructions. Theproductalsodisplayswhatsideisup/downandhead/foot.
16
9.3Symbols
CEͲmarkedacc. MDD93/42/ECC Fireretardant
Readtheusermanual Notindomesticrefuse. Followinstructionsforrecycling.
Guaranteetime
MinͲmaxuserweight
TypeBFequipmentClassIIequipment (DoubleInsulated)
Userinformation
Canberotated
Placeontop
Directlyonthe bedbase
Notdirectlyonthe bedbase
Nottoberotated
Placeontopof Notontopof Canbeturned NottobeturnedPlace existingmattress existingmattress underneath
11.3Otheraccessories Toensurethebestlevelofcomfort,werecommenduseofaccessoriesfromCareofSweden sinceourproductsaredesignedtoworkwelltogether.PleasecontactCareofSwedenAB directlyorvisitourwebsite–www.careofsweden.com–foraccessoriesormoreinformation onourproducts.
17
12.Warranty
12.1Scope ThisCuroCell®systemiscoveredbya2Ͳyearwarrantyformanufacturingdefects.Thewarranty doesnotapplytonormalwearandtearorproductimpairmentordamagestemmingfrom negligenceorimproperhandling/care.
12.2Serviceandrepairs ThisCuroCell®mattressisamedicaldevice,inaccordancewithMDD93/42/EEC,thatisused asanaidinpreventingandtreatingbedsores.Forthisreason,werecommendthatthepump beregularlyservicedandinspectedtomaintainfunctionalityandperformance(readabout annualservicebelow).Likeothertechnicaldevices,apumpthatisproperlycaredforworks betterandlastslonger. DuringanannualservicingthepumpisthoroughlyinspectedandservicedbyourService department. ServiceandmaintenancemustalwaysbeperformedbyCareofSwedenABoroneofits authorisedtechnicians.OnlyusesparepartsthathavebeenapprovedbyCareofSwedenAB. UseofnonͲapprovedsparepartsinvalidatethewarranty. AwarrantyclaimmustbesubmittedtoCareofSwedenABbeforetheproductisreturned. PleasecontactCareofSwedenABoryourlocaldealerifyouhavequestionsregarding warrantyandservice.
13.Otherinformation 13.1Disassemblyandrecycling Exceptforcertainpartsofthepumps,energyrecoveryispossibleforalmostallmaterialin CuroCellproductsthroughincinerationinwasteincinerationfacilities. Pump: AusedCuroCellpumpmustnotbedisassembled.Itmustinsteadbetakentoarecycling centreandsortedaselectronicwaste. Mattress: AusedCuroCellmattressmustbetakentoarecyclingcentreandsortedascombustiblewaste. FormoreinformationcontactCareofSwedenAB 18