CareFusion

CareFusion Infant Flow SiPAP Systems

Infant Flow SiPAP Comprehensive Driver Guide Guide

Guide

3 Pages

Infant Flow SiPAP Comprehensive Driver Guide* Two-Point Oxygen Sensor Calibration A two-point oxygen sensor calibration should be performed before initial use and with each circuit change. To avoid unwanted alarms, occlude the prongs or mask and set the Low Pressure Flow to 8 LPM prior to turning Infant Flow SiPAP driver on. When the Infant Flow SiPAP driver is turned on, a power on self-check will automatically be performed. 1. Enter the calibration menu by pressing the tool button. 2. Set the Pressure Low flow meter to 8 LPM and the Pressure High flow meter to 3 LPM . Note: The Pressure High flow meter must be set during setup process to enable the manual breath button. 3. Adjust the Oxygen Control to 21%. Allow time for the reading in the oxygen display window to stabilize. 4. Press the flashing Question Mark button located under the 21% icon. The Question Mark will change to a static hourglass. When calibration is complete, a static green Check Mark icon will appear and the oxygen display window will read 21%. 5. Adjust the Oxygen Control to 100%. Allow time for the reading in the oxygen display window to stabilize. 6. Press the flashing Question Mark located under the 100% icon. The Question Mark will change to static Hourglass icon. When calibration is complete, the Hourglass icon will change to a static Check Mark. The oxygen display window will read 100%. 7. If oxygen calibration fails, a red X is displayed on the button of the screen, the alarm sounds and an Error code is displayed on the top left corner. Switch the driver off and turn back on. Repeat the procedure. Disable Oxygen Sensor The internal oxygen sensor may be disabled by pressing the Disable O2 button in the Calibration screen. This will disable oxygen monitoring and the audible oxygen alarm. An error code is displayed to indicate the oxygen monitor is inoperative. An external oxygen monitor must be used. Leak Test Set a flow of 8 LPM on the Pressure Low flow meter, while occluding the prongs or mask. A CPAP of 5 cmH2O ± 1should be generated. If unable to reach pressure, check for leaks. Release the occlusion, the pressure should fall below 2 cmH2O. If pressure does not fall, check for occlusions.  *To be used in conjunction with Departmental Policy and Procedure. This guide provides an overview of the setup procedure, please refer to Infant Flow SiPAP Operator’s Manual for more detailed instructions. CareFusion defers to the institution with respect to the actual competency of a specific user.
File Type: PDF
File Size: 117 KB
File Name: Infant Flow SiPAP Comprehensive Driver Guide Guide.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions