® PRODISC C VIVO Cervical disc prosthesis to restore disc height and maintain segmental motion.
SURGICAL TECHNIQUE
Image intensifier control Warning This description alone does not provide sufficient background for direct use of Centinel Spine products. Instruction by a surgeon experienced in handling these products is mandatory. Processing, Reprocessing, Care and Maintenance For general guidelines, function control and dismantling of multi-part instruments, as well as processing guidelines for implants, please contact your local sales representative or refer to: http://www.centinelspine.com/prodisc_reprocessing.html For general information about reprocessing, care and maintenance of Centinel Spine reusable devices, instrument trays and cases, please consult the Important Information leaflet (SE_023827) or refer to: http://www.centinelspine.com/prodisc_reprocessing.html
TABLE OF CONTENTS
INTRODUCTION
SURGICAL TECHNIQUE
PRODUCT INFORMATION
BIBLIOGRAPHY
2
Kinematics
3
Intended Use, Indications and Contraindications
4
Surgical Technique
5
Multi-Level Cases
14
Removal Procedure
15
Implants
17
Instruments
19
Pre-Cleaning Instructions
26
28
1
PRODISC C VIVO
FEATURES AND BENEFITS prodisc C Vivo is intended to replace a diseased and/or degenerated intervertebral disc of the cervical spine in patients with symptomatic cervical disc disease. The prodisc C Vivo procedure
is intended to significantly reduce pain by allowing for the removal of the diseased disc while restoring disc height and providing the potential for motion at the affected vertebral segment.
Simple surgical technique • Simple technique with two main steps: trial and implant insertion
Anatomical design • Convex superior plate for anatomical fixation • Trapezoidal footprint design for optimal anatomical fit and maximum endplate coverage Proven materials • Superior and inferior implant plates made from titanium alloy for improved MRI compatibility. • Rough surface coating of pure titanium allows bony ongrowth • Inlay made from ultra-high molecular weight polyethylene (UHMWPE) • UHMWPE on CoCrMo alloy articulation Ball and socket articulation • Permits a physiological range of motion in flexion/extension, rotation, and lateral bending • Allows for restoration of anatomical balance • Resists shear forces
2
Centinel Spine prodisc C Vivo Surgical Technique
KINEMATICS
prodisc C Vivo has a center of rotation which is located just below the inferior endplate of the prosthesis. Pure translation movements are controlled by the ball and socket inter-face.
min. 16.9°
Flexion/extension The location of the center of rotation (COR) and the flexion radius are in accordance to the natural joint guidance in the vertebral joints. COR
COR
min. 16.9°
Lateral bending The physiological range of motion in lateral bending is restored.
COR
COR
360º
Axial rotation The axial rotation is limited by the anatomical structures and not by the prosthesis.
prodisc C Vivo Surgical Technique Centinel Spine
1
INTENDED USE, INDICATIONS AND CONTRAINDICATIONS
Intended use prodisc C Vivo implants are used to replace a cervical intervertebral disc, to restore disc height and maintain segmental motion. Successful clinical outcomes depend on a number of critical factors, including:
• Completion of a training program on the use of prodisc C, prodisc C Nova or prodisc C Vivo • Proper patient selection • Adequate bone quality (investigation to determine bone quality is recommended) • Complete and meticulous discectomy, decompression, and remobilization of the disc space • Optimal implant sizing and placement • Postoperative treatment Indications Symptomatic cervical disc disease (SCDD), which is defined as neck or arm (radicular) pain and/or a functional/neuro logical deficit with at least one of the following conditions confirmed by imaging (CT, MRI, or x-rays):
• herniated nucleus pulposus • spondylosis (defined by the presence of osteophytes) • loss of disc height Specific contraindications • Fractures, infections, tumors • Spinal stenosis by hypertrophic spondylarthrosis • Severe facet joint degeneration • Segmental instability • Ossification of posterior longitudinal ligament (OPLL) General contraindications • Osteoporosis, osteochondrosis, and severe osteopenia • Acute or chronic systemic, spinal, or localized infections • Systemic and metabolic diseases • Any medical and surgical conditions precluding the b enefits of spinal surgery • Foreign body sensitivity to the implant materials • Pregnancy • Severe obesity (Body Mass Index above 40)
4
Centinel Spine prodisc C Vivo Surgical Technique
Patient exclusion recommendations Patient selection is one of the most important factors contributing to the outcome of the total disc replacement procedure. The following may affect clinical outcomes: • A condition of senility or mental illness, alcoholism or smoking • Dependency on pharmaceutical drugs or drug abuse • The patient’s occupation or activity level • Compromised vertebral bodies at affected level due to current or past trauma (fractures) • Substantial loss of disc height, where applied segmental distraction may lead to damage of the great vessels • Involved vertebral endplate dimensionally smaller than the minimum implant footprint size in both the mediallateral and the anteriorposterior directions • Severe abnormality of the endplate (e.g. large Schmorl nodes)
SURGICAL TECHNIQUE
1 Prerequisites and patient positioning Insertion of a prodisc C Vivo is dependent on the use of anterior-posterior (AP) and lateral fluoroscopy throughout the procedure. Patient positioning should allow for circumferential use of the C-arm at the operative level. Position the patient in a supine, neutral position on a radiolucent operating table. Ensure that the neck of the patient is in a sagittally neutral position and supported by a cushion. When treating C6–C7 make sure that the shoulders do not limit the x-ray monitoring. In any case both vertebrae should be completely visible.
2 Access Use a standard anterior approach to the cervical spine. Mark the level of the surgery and expose the intervertebral disc segment and adjacent vertebrae. Determine the median line using image intensifier control and make a permanent midline mark on the superior and inferior vertebral bodies, e.g. using an osteotome or electro cauterization.
prodisc C Vivo Surgical Technique Centinel Spine
1
Surgical Technique
3 Fix retainer screw system Instruments 03.820.100
Punch
03.820.101
Screwdriver
03.820.111
Vertebral Body Retainer
03.820.102– 03.820.109
Retainer Screws
03.820.110
Locking Nut
Perforate the anterior cortex of the superior and inferior vertebra in the lateral midline and vertical center with the punch. Insert retainer screws into the perforations and place them close to the posterior cortical wall. Their trajectory should be parallel to the endplates of the treated disc. Begin with the smaller diameter screws (3.5 mm) of the longest possible length. Use a larger diameter screw (4.5 mm) when extra bone purchase is needed or a smaller screw diameter has been used unsuccessfully (“rescue” screw).
Warning: • Insert screws under image intensifier control for optimal trajectory and depth control. • Do not perforate the posterior cortex with the tip of the screw. Slide the vertebral body retainer over the screws and lock it in place with the locking nuts. This assembly secures parallelism of the retainer screws and the vertebral endplates of the operated level. Avoid using the retainer to distract the segment. For proper distraction refer to Step 4 on page 7.
6
Centinel Spine prodisc C Vivo Surgical Technique
4 Mobilize and distract the segment Instrument 03.820.112
Vertebral Distractor
Start the discectomy using standard instruments. Remove as much disc material as possible to allow the vertebral distractor tips to be placed as posteriorly as possible into the intervertebral space. Mild predistraction with the retainer can be applied to support disc removal.
7
Surgical Technique
Under fluoroscopic control, insert the tip of the vertebral distractor to the posterior margin of the vertebral bodies. Distract the intervertebral space with the vertebral distractor to restore the height and to gain access to the posterior intervertebral space. Readjust the retainer to the distracted height of the intervertebral space. This step should be repeated until maximum distraction has been achieved. Then release and withdraw the vertebral distractor. Continue the discectomy and remove the cartilageous endplate carefully. All soft tissue must be removed from the endplates. Care should be taken to avoid damage of the endplates. Continue with the canal and foraminal decompression.
Precaution: • Avoid using the vertebral body retainer as a distractor. Excess force on the vertebral body retainer can lead to bending and/or pull-out of the screws from the bone. • Expose the posterior longitudinal ligament to remobilize the segment. If required for decompression, the PLL may be resected. Warning: • Avoid overdistraction with the vertebral distractor as this can lead to nerve root tension or improper implant selection. • Avoid excessive endplate damage or removal. This increases the risk of implant subsidence. • The uncinate process should be preserved. If required for adequate decompression, the posterior third of the uncinatus process may be removed. • Ensure the cartilageous tissue is removed from the endplates. Cartilageous tissue may hinder osseo integration of the implant and reduce the fixation strength.
8
Centinel Spine prodisc C Vivo Surgical Technique
5 Define the implant size Instruments 03.670.925– 03.670.977
Trial Implants
03.670.204
Holder for Trial Implants
03.820.113
Mallet
After completing discectomy and decompression, use the trial implants to determine the appropriate disc height and size of footprint. The goal is to select a prosthesis with the best possible anatomical fit, using the largest footprint and the smallest height needed to restore the natural disc. Prosthesis center of rotation (COR) should be positioned at the midline of the vertebral body or slightly posterior. The implant should cover the majority of the vertebral body endplate. Implants with undersized footprints lead to increased risk of implant subsidence and heterotopic ossification. Insert the inner shaft of the holder for trial implants into the corresponding sleeve and push it until it snaps into place. Select the appropriate trial implant and make sure that its protrusion is captured in the stop of the shaft before assembling. Ensure that the shaft is fully screwed in before use. Align the trial implant with the midline and advance it under image intensifier control into the disc space by tapping it cautiously.
prodisc C Vivo Surgical Technique Centinel Spine
9
Surgical Technique
In the lateral fluoroscopy view the optimal position of the trial implant is given by the best possible anatomical fit. If the stop does not allow the trial implant to go deep enough, the stop can be adjusted by turning the trial shaft counter clockwise (1 rev = 0.5 mm), enabling the trial implant to be advanced slightly deeper. At the same time the trial implant should be kept centered on the midline. Now release the distraction to determine if the trial implant height is appropriate for the patient. Its height should be the smallest height to match normal adjacent discs. When the correct size for the implant is determined, the trial implant is removed (apply slight distraction with the vertebral retainer if necessary).
Precaution: • Selecting an implant that is too tall can limit the segmental range of motion. • Avoid kyphotic position of the corresponding segment. Note: • Do not unscrew stop more than 4mm or contact to trial implant may be lost. Use next size trial implant instead.
11
Centinel Spine prodisc C Vivo Surgical Technique
6 Insert implant Instruments 03.670.201– 03.670.203
Implant Holder
03.670.213
Shaft for Implant Holder
03.670.212
Stop for Implant Holder (optional)
03.670.305– 03.670.307
Spacer Clamp, sizes M/MD, heights 5–7 mm
03.670.315– 03.670.317
Spacer Clamp, sizes L/LD, heights 5–7 mm
03.670.325– 03.670.327
Spacer Clamp, sizes XL/XLD, heights 5–7 mm
03.820.113
Mallet
03.820.101
Screwdriver
03.670.207
Positioner
Preparation Assemble the shaft for the implant holder. Open the packaging of the implant and follow these steps:
• Choose spacer clamp size M/MD, L/LD or XL/XLD corresponding to the implant. • Attach the appropriate spacer clamp to the prosthesis until the arms snap into the holding features in the implant. • Attach the corresponding implant holder (M/MD, L/LD or XL/XLD) to the spacer clamp, making sure that the lateral projections of the spacer clamp are captured in the arms of the implant holder. • Tighten the implant holder to the spacer clamp turning the head of the inner shaft clockwise. • Pull the implant en-bloc out of the packaging tray. Optionally, the stop can be attached to the implant holder.
Note: The prodisc C Vivo implants are not designed to be used with bone cement.
11
Surgical Technique
Insertion Apply distraction as necessary, to facilitate the insertion of the implant. Ensure that the black midline on the superior endplate faces cranially and align it with the midline marking of the vertebral body.
Without Stop
Under lateral fluoroscopic control, advance the implant into its final position providing the best possible anatomical fit with the vertebral bodies. The center of rotation (COR) of the prosthesis should be positioned at the midline of the vertebral body or slightly posterior. Avoid excessive cranial, caudal or lateral corrections during insertion and ensure that the implant doesn’t exceed the posterior margin of the vertebral body. The spacer clamp includes two grooves that visualize the anterior margin of the implant under lateral fluoroscopy. If the correct position of the implant is confirmed using image intensifier, release the retainer and apply slight compression with the retainer. Slight compression from the retainer will help the spikes on the implant to penetrate into the vertebral bodies.
11
Centinel Spine prodisc C Vivo Surgical Technique
With Stop
Release To release the connection between spacer clamp and implant, follow the two steps below:
1. Rotate the shaft of the implant holder two full turns in the counterclockwise direction. 2. Move the implant holder sideward and pull it back until the spacer clamp disconnects from the implant. Step by step remove the locking nuts, the vertebral body retainer and the retainer screws.
Precaution: Heterotopic ossification (HO) is a possible cause for f usion of the treated segment. Copious saline lavage is recommended to remove osteogenic stimuli (blood/bone marrow). HO might be reduced when bone wax is used to close cavities in the bone (screw holes) and open bone surfaces after removal of anterior o steophytes.¹
¹ See Barbagallo 2014
11
MULTI-LEVEL CASES
The most symptomatic level should be operated first. Multi-level prodisc C Vivo surgeries should be performed sequentially level by level. If no symptomatic difference between the levels can be identified, start operating the most collapsed or caudal one. The retainer screws should be placed centrally in the vertebrae according to the surgical technique for single levels. Insert the screws under image intensifier control. Repeated image intensifier control in AP direction will be necessary to ensure proper alignment of the disc prostheses.
11
Centinel Spine prodisc C Vivo Surgical Technique
REMOVAL PROCEDURE
If a prodisc C Vivo implant must be removed, the following technique is recommended.
1 Preparation Instruments 03.670.201– 03.670.203
Implant Holder
03.670.213
Shaft for Implant Holder
03.670.400
Remover Clamp size M/MD
03.670.410
Remover Clamp size L/LD
03.670.420
Remover Clamp size XL/XLD
Start the procedure by distracting the prosthesis index level by using the vertebral distractor and the retainer system. See page 6 for details on how to use the retainer system. Attach the appropriate remover clamp to the corresponding implant holder with minimal thread engagement without tightening the head of the inner shaft of the implant holder. Note: If the thread of the shaft is fully engaged and tightened, the remover clamp cannot be attached to the implant in the next step. Attach the remover clamp assembly to the inferior endplate of the implant and securely attach it by turning the head of the implant holder shaft in clockwise direction. Note: The surface edged with “inside” must be oriented t owards the center of the disc space with respect to the implant endplate being removed.
11
Removal Procedure
2 Implant Removal Instruments 03.670.201– 03.670.203
Implant Holder
03.670.213
Shaft for Implant Holder
03.670.400
Remover Clamp size M/MD
03.670.410
Remover Clamp size L/LD
03.670.420
Remover Clamp size XL/XLD
03.820.113
Mallet
03.820.282
Slide Hammer
Remove the inferior endplate of the implant by cautiously pulling the implant holder. Alternatively, the slotted mallet or slide hammer can be used to aid implant removal. The superior endplate of the implant can be removed using the remover clamps or alternatively suitable forceps.
11
Centinel Spine prodisc C Vivo Surgical Technique
IMPLANTS
Dimensions Six different footprints are available for optimal coverage of the vertebral endplate: M, MD, L, LD, XL, XLD.
14.5–19.2 mm
16.5–22.5 mm
15 /17/19 mm
12/14/16/18 mm
Panjabi 1991
Three different heights (5, 6, and 7 mm) allow adjustment to the individual dimensions of the patient’s disc.
4.2–7.5 mm
5 /6 /7 mm
Yoganandan et al. 2001
11
Implants
The prodisc C Vivo implant consists of two titanium endplates. The superior endplate has a convex shape, while the inferior plate is flat. There are six spikes on both plates. prodisc C Vivo has always been based on the ball and socket principle with a “poly-on-metal” pairing. The polyethylene inlay (ball) is securely locked in the lower endplate while the upper endplate embraces the calotte (socket) made of CoCrMo alloy.
prodisc C Vivo, uncemented Implant M Width Depth 12 mm
15
mm
Implant MD
Implant L
Width 15 mm Depth 14 mm
Width 17 mm Depth 14 mm
Art. No.
Height
Art. No.
Height
Art. No.
Height
04.670.925S
5 mm
04.670.935S
5 mm
04.670.945S
5 mm
04.670.926S
6 mm
04.670.936S
6 mm
04.670.946S
6 mm
04.670.927S
7 mm
04.670.937S
7 mm
04.670.947S
7 mm
Implant LD
Implant XL
Implant XLD
Width 17 mm Depth 16 mm
Width 19 mm Depth 16 mm
Width 19 mm Depth 18 mm
Art. No.
Height
Art. No.
Height
Art. No.
Height
04.670.955S
5 mm
04.670.965S
5 mm
04.670.975S
5 mm
04.670.956S
6 mm
04.670.966S
6 mm
04.670.976S
6 mm
04.670.957S
7 mm
04.670.967S
7 mm
04.670.977S
7 mm
11
Centinel Spine prodisc C Vivo Surgical Technique