Chattanooga Group
Hydrocollator E-2 User Manual 22180D
User Manual
22 Pages
Preview
Page 1
TABLE OF CONTENTS
Hydrocollator® Heating Units
Foreword... 2 Precautionary Definitions... 2-3 Nomenclature... 4-6 Technical Specifications... 7-8 Setup... 9 Operation... 9 Troubleshooting... 10 Replacement Parts...11-16 Maintenance... 17-18 Warranty... 19
Hydrocollator® is a registered trademark of Chattanooga Group. HotPac™ is a trademark of Chattanooga Group.
©2006 Encore Medical Corporation or its affiliates, Austin, Texas, USA. Any use of editorial, pictorial or layout composition of this publication without express written consent from Chattanooga Group of Encore Medical, L.P. is strictly prohibited. This publication was written, illustrated and prepared for print by Chattanooga Group of Encore Medical, L.P.
ISO 13485 CERTIFIED
1
FOREWORD
Hydrocollator® Heating Units
This manual contains general instructions for operation, precautionary instructions, and maintenance recommendations for the owners and operators of the Hydrocollator SS, E-2 and E-1 Heating Units. In order to obtain maximum life and efficiency of the Hydrocollator Heating Units, please read this manual thoroughly and follow all instructions prior to operating the unit. Specifications put forth in this manual were in effect at the time of publication. However, owing to Chattanooga Group's policy of continual improvement, changes to these specifications may be made at any time without obligation on the part of Chattanooga Group.
PRECAUTIONARY DEFINITIONS
Hydrocollator® Heating Units
The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definitions of these symbols are as follows: = CAUTIONS - Text with a “CAUTION” indicator explains possible safety infractions that could have the potential to cause minor to moderate injury or damage to equipment. = WARNINGS - Text with a “WARNING” indicator explains possible safety infractions that will potentially cause serious injury and equipment damage. = DANGERS - Text with a “DANGER” indicator explains possible safety infractions that are imminently hazardous situations that would result in death or serious injury. NOTE: Throughout this manual, “NOTE” may be found. These notes are helpful information to aid in the particular area or function being described.
CAUTION • Read, understand and practice the precautionary and operating instructions. Know the limitations associated with the Hydrocollator SS, E-2 and E-1 Heating Units. Observe the cautionary and operational decals installed on the unit. • The thermostat is extremely sensitive and the slightest adjustment will alter the temperature several degrees. The recommended operating temperature is 160° F to 166° F (71° to 74°C). The temperature of the water should be checked with a thermometer after every adjustment, before using the HotPacs. Always allow sufficient time for the water temperature to stabilize. • Always keep the water level over the top of the HotPacs to avoid burning or scorching of the packs. • Never allow the HotPacs to dry after use. • Check water level daily as it has a natural loss due to evaporation.
2
PRECAUTIONARY INSTRUCTIONS
Hydrocollator® Heating Units
CAUTION • Never adjust the thermostat too high. The thermostat is extremely sensitive to the slightest adjustment. • Clean the tank periodically as described in the maintenance portion of this manual. • Always replace Hydrocollator HotPacs as soon as they show signs of wear. • If the unit is to be left unattended for a period of time, unplug the unit, remove the packs, empty the water, and clean the tank. • Do not move the Hydrocollator while filled with water. Tipping over could result in burns to the user or others. • Always unplug the Hydrocollator from the power source before attempting to empty water from unit. • Always use factory authorized replacement parts. • Always destroy and replace HotPacs which are exposed to contamination: hazardous or cleaning materials, bodily fluids, mold, etc.
WARNING • Grounding the Hydrocollator SS, E-2 and E-1 Heating Unit: Make certain the unit is electrically grounded by plugging into a hospital grade electrical outlet with a ground terminal receptacle (U-ground outlet), 120 VAC only. Follow the National Electrical Code. • The water temperature in the Hydrocollator is approximately 160° F (71° C), and the water scalding temperature is approximately 120° F (49° C).
Risk of explosion if used in the presence of flammable anesthetics.
Explanation of Symbols ATTENTION: Consult accompanying documents.
Type B equipment. Denotes the degree of protection against electric shock, particularly regarding allowable leakage currents and the reliability of the protective earth connection.
3
NOMENCLATURE
Hydrocollator® Heating Units
WHITE LAMP (where applicable), when illuminated, indicates that main power is connected, the switch is on and unit is operating. RED LAMP (where applicable), when illuminated, warns of an overheat/low water condition. Action is necessary to correct the condition. GREEN LAMP (220/240 volt units), when illuminated, indicates that main power is connected, the switch is on and unit is operating.
White Lamp
Red Lamp
IMPORTANT: Some models (220-240 Volt) are fitted with a thermal cut-out thermostat which acts as a “Non self-resetting” thermal cut-out. In the unlikely event of a fault condition, this device is designed to open the power circuit between 190°F ± 14°F (88°C ± 8°C). It is not possible to alter its operating range. The operation of the unit may be restored by depressing the red button after the water/element temperatures have been allowed to cool. If the fault condition persists, please contact Chattanooga Group or an authorized agent.
4
NOMENCLATURE
Hydrocollator® Heating Units
Diagram 1. Heating Element 2. Drain 3. Overtemp Thermostat 4. Thermostat, Hydraulic 5. Bracket (Brkt. & Bulb Themo A/B Assembly) 6. Lid 7. 3 inch Caster 8. Overtemp Light
9. Junction Box (J-Box) Cover 10. IEC connector 11. Power Cord 12. Drain Valve (Not Shown) 13. Reset Overtemp Button* 14. Fuses 15. Pilot Lamp, Red 16. Pilot Lamp, Clear 1
3
4
6 2
5
ide
ns wI
Vie
R
iew V r ea
Rear View
16 15 NOTE: E-1, E-2 models shown; models will vary. * All 220-240 Volt have this button
5
0 12
lt Vo
9
11
8
10
13
14 14
220/240 V olt
NOMENCLATURE
Hydrocollator® Heating Units
Diagram (continued) 17. Hot Pac 18. Neck Pac Insert 19. Drain, Copper Tube 20. Lid Hinge nuts 21. Lid Knob/Handle (on back) 22. Lid Hinge
23. Side Handle 24. Thermostat, Hydraulic 25. Thermostat Control Knob 26. On/Off Switch 27. Internal Wiring Harness
17 21
22
18
20 19
23
-Up View Rear Close ) ht Section (Lower Rig
Rear V iew (U pside Down 27 (J-Box ) Open ) 24 Not illustrated: 28. Stainless Steel Wire Rack 29. DrainValve, Ball - 1/2 in Replacement 30 Drain, Hose, Bib 31. Replacement Tank 32. Outer Jacket 33. Fuse 10A 34. Lid Kit Handle 35. Water Thief
Right Side Up (Back View)
37. Plastic Feet 38. Pilot Lamp, Green
26
6
25
TECHNICAL SPECIFICATIONS
Hydrocollator® Heating Units
SS Model Power Supply Rated Frequency Power Consumption Tank Capacity Temperature Range Thermal Cut-out Temp Accuracy Heating Up Time Cool Down Time Safety Class Safety Tests Dimensions Weight
110-120 V~, 220-240 V~ 50/60 Hz 1000 W 31.64 Liters 160°F - 165°F (71°C - 74°C) 190°F ± 14°F (88°C ± 8°C) +/- 10% Time to 160°F (70°C) - 3 Hours Time from 160°F (70°C) - 2 Hours Type B UL 544 20 in H x 18 in L x 16 in W (51 cm H x 46 cm L x 41 cm W) 21 lbs (9.5 kg)
Standard Accessories: Includes eight Standard size HotPacs.
7
TECHNICAL SPECIFICATIONS
Hydrocollator® Heating Units
E-2 Model Power Supply Rated Frequency Power Consumption Tank Capacity Temperature Range Thermal Cut-Out Temp Accuracy Heat Up Time Cool Down Time Safety Class Safety Tests Dimensions Weight
110-120V~, 220-240 V~ 50/60 Hz 1000 W 43.23 Liters 160°F - 165°F (71°C - 74°C) 190°F ± 14°F (88°C ± 8°C) +/- 10% Time to 160°F (70°C) - 4 Hours Time from 160°F (70°C) - 2 Hours Type B UL 544 20 in H x 15 in L x 13 in W (51 cm H x 38 cm L x 33 cm W) 12 lbs (5.5 kg)
Standard Accessories: Includes two Oversize HotPacs, three Standard size HotPacs and 1 Neck Contour HotPac.
E-1 Model Power Supply Rated Frequency Power Consumption Tank Capacity Temperature Range Thermal Cut-out Temp Accuracy Heating Up Time Cool Down Time Safety Class Safety Tests Dimensions Weight
110-120V~, 220-240 V~ 50/60Hz 1000 W 15.44 Liters 160°F - 165°F (71°C - 74°C) 190°F ± 14°F (88°C ± 8°C) +/- 10% Time to 160°F (70°C) - 2 Hours Time from 160°F (70°C) - 1 Hour Type B UL 544 16 in H x 13 in L x 8 in W (41 cm H x 33 cm L x 20 cm W) 15 lbs (6.8 kg)
Standard Accessories: Includes four Standard size HotPacs.
8
SETUP
Hydrocollator® Heating Units
Remove all of the contents inside the Hydrocollator and then replace the wire rack. Make certain the drain valve is closed before filling the tank with water. The tank should be approximately 3/4 full with packs removed. The tank may be filled from any standard type faucet. Check the voltage rating on the serial decal plate which is located on the back of the unit. To turn the unit on, simply plug the unit into a 120 Volt~, 60 Hz or 220/240 Volt~, 50-60 Hz outlet as indicated on the decal plate. To turn the unit off, simply unplug from mains power supply. Do not attempt to use direct current. Follow the procedures indicated in the Precautionary Instructions. Do not attempt to use the unit if it is not properly grounded.
OPERATION
Hydrocollator® Heating Units
The Hydrocollator is simple and easy to operate. The controls are located in the rear, near the bottom of the unit. The thermostat controls the temperature of the water. It is located near the bottom of the unit on the control panel. It has been tested and preset at the factory to normal operating temperature. If further adjustment is necessary, it can be easily accomplished by turning the adjusting screw located on the thermostat. Turning the outer screw counterclockwise lowers the temperature, while turning it clockwise raises it. NOTE: Do not turn the inner stem. The Hydrocollator units operate through the use of a hydraulic capillary-type thermostat and can be left plugged in and turned on to provide a ready supply of heated HotPacs . The HotPac rack is adjustable and accommodates different sizes and types of packs. SS Unit – Holds eight Standard Size packs. Special rack holds two Neck Contour packs. E-2 Unit – The E-2 unit comes complete with two oversize, three Standard Size packs and one Neck Contour pack. E-1 Unit – With the insert in, the rack accommodates four Standard Size packs.
9
TROUBLESHOOTING
Hydrocollator® Heating Units
The following are solutions to minor performance problems. Only qualified personnel should service equipment problems.
PROBLEM
POSSIBLE CAUSE
Unit Plugged In, Water Doesn’t Get Hot
1. Breaker for electric circuit off. 2. Thermostat not functioning. 3. Heating element burned out.
1. Turn breaker ON. 2. Replace thermostat. 3. Replace element.
Cloudiness of Water
1. Filler from packs is seeping. 2. Too long a period between cleanings.
1. Replace packs. 2. Drain and clean.
Packs Too Hot
1. Thermostat set too high. 2. Thermostat failure.
1. Adjust thermostat. 2. Replace thermostat.
Packs Too Cool
1. Thermostat set too low. 2. Thermostat failure.
1. Adjust thermostat. 2. Replace thermostat.
Unit Boils
1. Thermostat failure.
1. Replace thermostat.
Rust
1. Contact with ferrous metal.
1. Remove the ferrous metal item from the unit - usually a staple, paper clip, safety pin, etc., and clean thoroughly.
2. Deposits on inside of tank.
2. Thoroughly clean tank and remove all deposits.
3. Too much chlorine in water.
3. Clean tank and rinse thoroughly. Add dechlorinator if local water has high chlorine content.
10
POSSIBLE SOLUTION
REPLACEMENT PARTS
Hydrocollator® Heating Units
Model SS 120 VOLT – Standard Cat. No. – 2202 w/packs Key* 1 4 5 6 9 11 14 15 16 19 20 21 22 26 27 29 34 35 37 26
Part Description Element, Heating 120 Volt, 1000 Watt Hydraulic Thermostat Brkt. & Bulb Thermostat A/B Assembly (2 each)** Lid Junction Box Cover Power Cord Hose, Drain Pilot Lamp, Red Pilot Lamp, Clear Drain, Copper Tube Part A Lid Hinge Rod Nuts (2 required) Lid Knob Lid Rod Thermo Low Water Kit*** Overtemp Thermostat Control Wiring (Internal) Harness DrainValve, Ball - 1/2 in Replacement Handle, Lid Kit Water Thief Feet, Plastic On/Off Switch
Part No. 10630 10631 21023 22219 22223 20568 21635 22149 22148 21288 21737 21025 22231 20548 20548 21031 21373 22171 21882 75538 23406
* Key numbers refer to Nomenclature section (pages 5 and 6) ** Used to hold the Hydraulic Thermostat and Overtemp Thermostat to the Heating element. *** Includes Gasket Thermostat, Nut 1/2-20 Brass, and Overtemp Thermostat (#26)
11
REPLACEMENT PARTS
Hydrocollator® Heating Units
Model E-2 120 VOLT – Standard Cat. No. – 2802 w/packs Key* 1 4 5 6 9 11 14 15 16 19 21 27 29 35 26
Part Description Element, Heating 120 Volt, 1000 Watt Hydraulic Thermostat Brkt. & Bulb Thermostat A/B Assembly (2 each)** Lid Junction Box Cover Power Cord Hose, Drain Pilot Lamp, Red Pilot Lamp, Clear Drain, Copper Tube Part A E-2 Lid Handle Thermo Low Water Kit*** Wiring (Internal) Harness DrainValve, Ball - 1/2 in Replacement Water Thief On/Off Switch
Part No. 10630 10631 21023 23457 23460 20568 21635 22149 22148 21288 20512 20548 21031 21373 21882 23406
* Key numbers refer to Nomenclature section (pages 5 and 6) ** Used to hold the Hydraulic Thermostat and Overtemp Thermostat to the Heating element. *** Includes Gasket Thermostat, Nut 1/2-20 Brass, and Overtemp Thermostat (#26)
12
REPLACEMENT PARTS
Hydrocollator® Heating Units
Model E-1 120 VOLT – Standard Cat. No. – 2102 w/packs Key* 1 4 5 6 9 11 15 16 20 28 21 22 23 27 26
Part Description Element, Heating 120 Volt, 1000 Watt Hydraulic Thermostat Brkt. & Bulb Thermostat A/B Assembly (2 each)** Lid Junction Box Cover Power Cord Pilot Lamp, Red Pilot Lamp, Clear Lid Hinge Rod Nuts (2 required) Rack, Stainless Steel Handle, Lid Hinge Rod, Lid Handle, Side Kit Thermo Low Water Kit*** Wiring (Internal) Harness On/Off Switch
Part No. 21007 10631 21023 21004 21005 20568 22149 22148 21737 21062 21025 22668 21994 20548 21031 23406
* Key numbers refer to Nomenclature section (pages 5 and 6) ** Used to hold the Hydraulic Thermostat and Overtemp Thermostat to the Heating element. *** Includes Gasket Thermostat, Nut 1/2-20 Brass, and Overtemp Thermostat (#26)
13
REPLACEMENT PARTS
Hydrocollator® Heating Units
Model SS 220/240 VOLT – (Low Water) Cat. No. – 2207 w/packs Cat. No. – 2209 w/o packs Key* 1 4 5 9 14 15 19 20 22 27
Part Description Element, Heating 1000 Watt Hydraulic Thermostat Brkt. & Bulb Thermostat A/B Assembly (2 each)** Junction Box Cover Hose, Drain Pilot Lamp, Red Drain, Copper Tube Part A Lid Hinge Rod Nuts (2 required) Lid Rod Thermo Low Water Kit*** Wiring (Internal) Harness
Part No. 22245 10631 21023 22223 21635 22147 21288 21737 22231 20548 61579
29
Valve, Boiler Drain, 1/2 I.D. Chrome
21373
Carrier, Fuse
72733
Holder, Fuse Handle, Lid finished Water Thief On/Off Switch Lamp, Green
72734 20531 21882 23406 22146
34 35
Cordset (Country Specific) 29
Rack, Wire
22232
Drain, Copper, Part B
21289
18
Insert, For Neck Contour Packs
21053
32
Tank, Replacement, w/Drain
23385
14
Fuse, Anti-Surge, 6.3 Amp (2)
71772
* Key numbers refer to Nomenclature section (pages 5 and 6) ** Used to hold the Hydraulic Thermostat and Overtemp Thermostat to the Heating element. *** Includes Gasket Thermostat, Nut 1/2-20 Brass, and Overtemp Thermostat (#26)
14
REPLACEMENT PARTS
Hydrocollator® Heating Units
Model E-1 220/240 VOLT – (Low Water) Cat. No. – 2107 w/packs Cat. No. – 2109 w/o packs Key* 1 4 5 6 9 15 20 28 21 22 23 27
Part Description Element, Heating 1000 Watt Hydraulic Thermostat Brkt. & Bulb Thermostat A/B Assembly (2 each)** Lid Junction Box Cover Pilot Lamp, Red Lid Hinge Rod Nuts (2 required) Rack, Stainless Steel Handle, Lid Hinge Rod, Lid Handle, Side Kit (2) Thermo Low Water Kit*** Wiring (Internal) Harness Pilot Lamp, Green
Part No. 22245 10631 21023 21004 21005 22147 21737 21062 21025 22668 21994 20548 61579 22146
11
Cordset (Country Specific)
18
Insert, For Neck Contour Packs
21053
Receptacle, Power
74723
Carrier, Fuse
72733
Holder, Fuse
72734
Fuse, Anti-Surge, 6.3 Amp (2)
71772
Washer, Nylon
21027
14
* Key numbers refer to Nomenclature section (pages 5 and 6) ** Used to hold the Hydraulic Thermostat and Overtemp Thermostat to the Heating element. *** Includes Gasket Thermostat, Nut 1/2-20 Brass, and Overtemp Thermostat (#26)
15
REPLACEMENT PARTS
Hydrocollator® Heating Units
Model E-2 220/240 VOLT – (Low Water) Cat. No. – 2806 w/packs Cat. No. – 2808 w/o packs Key*
Part Description
Part No.
1 4 5 6 9
Element, Heating 1000 Watt Hydraulic Thermostat Brkt. & Bulb Thermostat A/B Assembly (2 each)** Lid Junction Box Cover
22245 10631 21023 23457 23460
10 14 19 21 25 27 29 35 26
Cordset (Country Specific) Hose, Drain Drain, Copper Tube Part A E-2 Lid Handle Thermo Low Water Kit*** Wiring (Internal) Harness DrainValve, Ball - 1/2 in Replacement Water Thief (2) On/Off Switch
21635 21288 20512 20548 61579 21373 21882 23406
Lamp, Green, Lens
22146
Lamp, Red, Lens
22147
Carrier, Fuse (2)
72733
Holder, Fuse (2)
72734
14
Fuse, Anti-Surge, 6.3 Amp (2)
71772
22
Lid, Rod
90796
Insert, For Neck Contour Packs
21053
Rack, Wire
22259
15
* Key numbers refer to Nomenclature section (pages 5 and 6) ** Used to hold the Hydraulic Thermostat and Overtemp Thermostat to the Heating element. *** Includes Gasket Thermostat, Nut 1/2-20 Brass, and Overtemp Thermostat (#26)
16
MAINTENANCE
Hydrocollator® Heating Units
Care and Cleaning The Hydrocollator is equipped with an immersion type heating element and a hydraulic capillary type thermostat which evenly maintains the HotPac temperature in the water and provides a ready supply of heated packs. It is critical to maintain the water level over the top of the HotPac to avoid damage to the heating element, the stainless steel, or the HotPac. Water is constantly lost during operation due to evaporation. Therefore, it is essential that water be added daily. The tank should also be drained and cleaned systematically, at minimum intervals of every two weeks. Always unplug the unit from the power source when emptying or cleaning the unit. Drain the unit by the drain valve located at the bottom rear of the unit. Either move the unit to a floor drain or attach the extension hose to it when draining.
WARNING • Do not move the Hydrocollator while filled with water. Tipping over could result in a burn to the user or others. • Always unplug the Hydrocollator from the power source before attempting to empty water from unit.
Your Hydrocollator Heating Unit is made of high quality stainless steel that should last a lifetime. You must properly care for your Hydrocollator Heating Unit if you wish for it to last a lifetime. Follow these simple steps to ensure that your Hydrocollator Heating Unit lasts. 1. Chlorine means rust to stainless steel. No bleach or any cleaner with high chlorine content should be used. 2. Clean regularly for optimal service and operation. NOTE: Failure to maintain your equipment (including regular cleaning) will void your warranty. 3. Fill daily with water. Water is constantly lost during operation due to evaporation. 4. Chlorine in regular tap water may be present in high enough concentrations to damage your unit. If you suspect high levels of chlorine in your Hydrocollator Heating Unit, we recommend the addition of a dechlorinator. 5. Certain additives (such as herbal teas, essential oils, etc.) will damage the components of the heating unit including the stainless steel. The Hydrocollator Heating Unit was designed to heat water and HotPacs only.
To avoid potential for rusting: 1. 2. 3. 4. 5. 6. 7.
Do not use bleach or cleaners with high chlorine content. Do not use paper clips, safety pins, staples, coat hangers, etc., in or on the unit. Do not clean the unit with steel wool or a carbon steel brush. Do not use abrasive cleaners which tend to disturb the grain in the finish. Water level is critical: Add water daily and keep it full. Regularly clean and drain the tank (every two weeks). Remove all deposits from interior surfaces and parts (including heating element). The deposits are concentrations of chlorine that allow rusting to begin. If allowed to accumulate on the heating element, these deposits will lower the efficiency of the element and increase the power consumption of the unit. 8. If you suspect that you have “hard water”, you may need to change your water more frequently. Salt deposits are an indication of hard water that will eventually cause rust.
17
MAINTENANCE
Hydrocollator® Heating Units
Cleaning Tips 1. The interior of the unit should be cleaned, usually every two weeks, using a low abrasive bathroom cleaner. Check for low or no chlorine content in your cleaner and make sure that the residue is thoroughly rinsed away with water. 2. A strong solution of vinegar and water will usually dissolve away deposits, which then must be thoroughly rinsed away.
Care of HotPacs Always return the Hydrocollator HotPac to the hot water of the heating unit after each treatment. There it is heated, kept clean and ready for immediate use. The HotPac may be boiled; however, this hastens the deterioration of the pack. The pack may also be cleaned by scrubbing the pack with soap and water. The simplified method of keeping the pack clean is to keep it immersed in water. The Hydrocollator HotPac, under constant daily use, should have a useful life of at least one year. When the pack begins to wear out, the filler oozes and leaks through the fabric of the cloth and sections of the pack appear to be loose and empty and will not retain heat properly. The pack should be replaced when it reaches this state. Original replacement Hydrocollator HotPacs may be ordered from an authorized Chattanooga Group dealer. NOTE: If HotPacs are to be stored for extended periods, they may be placed - while wet - in plastic bags and stored in a freezer.
18
WARRANTY
Hydrocollator® Heating Units
Chattanooga Group ("Company") warrants that the Hydrocollator Models SS, E-2, and E-1 Heating Units ("Products") are free of defects in material and workmanship. This warranty shall remain in effect for one year from the date of original consumer purchase. If these Products fail to function during the one year warranty period due to a defect in material or workmanship, Company or the selling dealer will repair or replace this Product without charge within a period of thirty days from the date on which the Product is returned to the Company or the dealer. All repairs to the Product must be performed by a service center authorized by the Company. Any modifications or repairs performed by unauthorized centers or groups will void this warranty. To participate in warranty coverage, this Product's warranty registration card (included with Product) must be filled out and returned to the Company by the original owner within ten (10) business days of purchase. This Warranty Does Not Cover: Replacement parts or labor furnished by anyone other than the Company, the selling dealer or a certified Company service agent. Defects or damage caused by labor furnished by someone other than Company, the selling dealer or a certified Company service agent. Any malfunction or failure in the Product caused by product misuse, including, but not limited to, the failure to provide reasonable and necessary maintenance or any use that is inconsistent with the Product User's Manual. COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES. Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you. To obtain service from Company or the selling dealer under this warranty: 1. A written claim must be made within the warranty period to the Company or the selling dealer. Written claims made to the Company should be sent to: 4717 Adams Road P.O. Box 489 Hixson, TN 37343 USA Telephone: (423) 870-2281 USA Telephone: (800) 361-3661 Canada Outside USA: +1 (423) 870-7200 Facsimile: (423) 875-5497 2. The Product must be returned to the Company or the selling dealer by the owner. This warranty gives you specific legal rights and you may also have other rights which vary from state to state. The Company does not authorize any person or representative to create for it any other obligation or liability in connection with the sale of the Product. Any representation or agreement not contained in the warranty shall be void and of no effect. THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
19
ISO 13485 CERTIFIED 4717 Adams Road P.O. Box 489 Hixson, TN 37343 1.423.870.2281 U.S.A. 1.800.361.6661 Canada 1.423.870.7200 Outside U.S.A. 1.423.870.2046 Fax chattgroup.com © 2006 Encore Medical, L.P.
22180D