CLEMENTS
Portable Suction Pumps
SUC 84112 HiFlo High Vacuum High Flow Suction Pump User Manual Issue 23
User Manual
27 Pages

Preview
Page 1
CLEMENTS HiFlo High Vacuum High Flow Suction Pump SUC 84112
User Manual Manual No. SUC 91005 415 Issue 23
CLEMENTS
Safety Thank you for purchasing this Clements HiFlo High Suction Pump. For your safety it is imperative that this unit only be operated by authorised personnel in accordance with the instructions as described in this manual. Operated in this way, the HiFlo High Suction Pump will provide the standard of service specified. Due to continual improvements in product design, the HiFlo High Suction Pump may vary in detail from the descriptions in this manual. In the event of further questions please contact your local distributor or BMDi TUTA Healthcare direct.
User Manual HiFlo High Suction Pump Manual Number SUC 91005 415 Issue 23 Copyright © 2009 BMDi TUTA Healthcare Pty Ltd The information in this manual was originated by, and is the exclusive property of BMDi TUTA Healthcare Pty Ltd. It is furnished for customer information only, and is not an authorisation or licence to make this product or to furnish this information to others.
BMDi TUTA Healthcare Pty Ltd Unit 4B, 128-130 Francs Street Lidcombe NSW 2141 Australia Phone: +61 2 9466 5300 Website: www.medaust.com
2
CLEMENTS
Contents
Classifications ...
4
Specifications ...
5
Transportation and Storage ...
6
Description ...
8
Installation and Operation ...
9
Spares ...
13
Maintenance ...
15
Troubleshooting ...
17
Diagnostic Setup ...
19
Disassembly and Repair ...
21
Warranty ...
26
Contents
3
CLEMENTS
Identification SUC 84112 HiFlo High Vacuum / High Flow Suction Pump
Intended Use To provide a continuous vacuum source, within the stated operating vacuum range, for the aspiration of fluids and particulate matter in medical procedures carried out by clinically trained and authorized personnel.
Uninterrupted Availability This equipment is mains powered. If patient care requires the uninterrupted availability of a source of vacuum, the clinical facility must make provision for either a backup power supply or standby suction units which are battery or manually powered.
Classifications GMDNS Code
36777
GMDNS Term
Suction unit, electric powered
GMDNS Synonym
Aspirator
Device Class
Class IIa
Electric Shock Protection
Class I Equipment
Additional Electric Shock Protection
Defibrillation-proof Type BF Applied Part
Sterilisation
Not supplied in sterile state
Anaesthetic Rating
NOT Category AP NOT Category APG
Operation Mode
Continuous operation
4
Classifications
CLEMENTS
Specifications HiFlo SUC 84112 Power Rating
220 - 240 V 50/60 Hz 380W
Vacuum Range
0 to –93 kPa 0 to –697 mm Hg
Max Flow Rate (free air)
55 litre / min
Motor
Thermally protected PSC motor
Pump
Oil Less twin piston
Ambient Requirement
Continuous operation in ambient temperatures of 5° C to 35°C.
Filters
Inlet: disposable bacteria filter Exhaust: clean, bleached cotton wool (5g) Internal: disposable hydrophobic filter
Collection Jar
2 x 2 litre autoclavable, shatter resistant polycarbonate jars with non-static polypropylene bungs and overflow shut-off valves.
Regulator
Needle Valve
Gauge
Bourdon tube type Dual Scale, mm Hg / kPa. Accuracy ±3% FSD
Gauge Range
0 to -760 mm Hg graduated at 20 mm Hg 0 to -100 kPa graduated at 5 kPa
Unit Dimensions
Width 406mm, Depth 305mm, Height 800mm Weight 16.5kg
Standard Conditions
25°C, Sea Level, 100kPa
Environment
Indoor
Specifications
5
CLEMENTS
Transportation and Storage Environmental conditions for transportation and storage are shown in the following table. In addition, for vehicular transportation, the unit should be packed in its original packaging and kept upright.
Parameter
Minimum
Maximum
Temperature
10ºC
40ºC
Humidity
60% RH
95% RH
Barometric Pressure 700 mBar
1060 mBar
Warning Symbols Legend The warning symbols marked on the equipment and their meanings are shown as follows.
Attention, consult accompanying documents
Defibrillation-Proof Type BF applied part
Dispose of product according to requirements of WEEE directive
6
Transportation and Storage
CLEMENTS
Electromagnetic Interference This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: •
Reorient or relocate the receiving antenna.
•
Increase the separation between the equipment and the receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced radio/TV technician for his or her help.
Changes or modifications to the equipment not expressly approved by Clements could void the user’s authority to operate the equipment.
Electromagnetic Interference
7
CLEMENTS
Description The Clements HiFlo High Suction Pump is a mobile, enclosed suction pump designed for high suction and high flow. The main feature of this unit is the low maintenance piston motor and pump assembly which can generate suction strengths from 0 to -675 mm Hg (-90 kPa) with flow rates up to 55 litres of Free Air Per Minute (FAPM), depending on the model and power supply. The unit is mounted on large rubber wheels which make it easy to move to any part of a hospital. The unit also has a storage compartment at the rear to house additional tubing, handpieces and other items needed when using the pump. Main Components of Unit
8
1
Power Switch
6
Exhaust Filter
2
Vacuum Gauge
7
Handle
3
Regulator Control Knob
8
Tube Storage Hooks
4
Inlet Connection
9
Storage Compartment
5
Pump (Internal) Description
CLEMENTS
Installation and Operation Installation The Clements HiFlo High Suction Pump is supplied with: • 2 x 2 litre Collection Jar with Bung Assembly • 1 x 2.4 metre length of Suction Tubing • 1 x 0.4 metre length of Suction Tubing Important Remove strapping from motor before starting. Ensure power is disconnected before opening unit.
WARNING This unit can generate high vacuums. It must not be used for applications which require low vacuum. Always check patient requirements before applying the Clements HiFlo High Suction Pump.
For applications which require low vacuum, please contact your distributor for information about the Clements Low Suction range of products.
Connecting The Collection Jars Because the unit is fitted with two Collection Jars it can either be connected in a single-jar arrangement or a double-jar arrangement. With the single-jar arrangement, only one jar is in use at a time, with the second jar on standby. As one jar is filled it is disconnected, so that it can be emptied, and the other jar is connected. The double-jar arrangement allows both jars to be in use at the same time, thus providing longer periods between changeover.
Installation and Operation
9
CLEMENTS Single-Jar Arrangement
To connect the Collection Jars in a single-jar arrangement, use the following procedure and refer to the block diagram below. 1.
Place two clean jars in the holders, ensuring that both jars are fitted with an Overflow Cut Off Valve.
2.
Connect the SUCTION outlet of the jar to be used first to the Bacteria Filter inlet on the main unit.
3.
Connect the PATIENT inlet of the jar to be used first to the Suction Tube leading to the patient.
4.
When the first jar is full, disconnect it and use steps 2 and 3 to connect the second jar.
Hydrophobic Shut Off Valve Filter (Internal)
V Suction Pump Unit
10
To Patient
Blue Bacteria Filter (External)
Overflow Cut Off Valves in BOTH Jars
Jar in Use
Installation and Operation
V
Jar on Standby
CLEMENTS Pump Inlet
Double-Jar Arrangement
Both Jars In Use Overflow Cut Off Valve (in this Jar Only)
To connect the Collection Jars in a double-jar arrangement, use the following procedure and refer to the block diagram below. 1.
Place two clean jars in the holders, ensuring that only one jar, Jar 1, is fitted with an Overflow Cut Off Valve.
2.
Connect the SUCTION outlet (with the Overflow Cut Off Valve) of Jar 1 to the Bacteria Filter inlet on the main unit.
3.
Connect the PATIENT inlet of Jar 1 to the SUCTION outlet of the jar without an Overflow Cut Off Valve, Jar 2.
4.
Connect the PATIENT inlet of Jar 2 to the Suction Tube leading to the patient.
Hydrophobic Shut Off Valve Filter (Internal)
Blue Bacteria Filter (External)
Overflow Cut Off Valve Suction Pump Unit
To Patient
NO Overflow Cut Off Valve
V Jar 1
Installation and Operation
Jar 2
11
CLEMENTS Operation To operate the Pump, follow the steps listed below. 1.
Place the pump on a level surface with a 15cm clearance from walls or other objects.
2.
Install the Collection Jars and Suction Tubing in either the single-jar or double-jar arrangement as described in the Connecting The Collection Jars section on page 8.
3.
Connect the unit to the mains power supply and switch the power on at the mains supply outlet.
4.
Press the Power Switch down into the ‘|’ (ON) position and the Motor will run.
5.
Occlude the Suction Tubing and adjust the Regulator, by turning the knob, to obtain the required vacuum level, as indicated on the Vacuum Gauge.
6.
Check that there is no loss of vacuum in the Suction Tubing and fittings.
7.
Connect the catheter to the patient.
8.
Check the Collection Jars to ensure that suction is being applied correctly.
Please Note: • Although the Collection Jars are fitted with Overflow Cut Off Valves, the fluid levels should be monitored to prevent spill-over into the pump unit. The Hydrophobic Filter fitted to the inlet of the pump acts as a single use shut-off valve to protect the pump. It must be replaced if fluid has spilled over into the pump unit. • Always start the pump with Suction Tubing unoccluded. Always clear any blockages before re-starting the pump. • The pump motor is protected by a thermal overload mechanism that will stop the pump if the temperature exceeds a pre-set level. The mechanism will then automatically restart the Pump after the motor has cooled to the appropriate temperature. 12
Installation and Operation
CLEMENTS
If the motor thermal overload trips, switch the power off at the mains switch. This will prevent unexpected re-starting of the pump when the thermal overload automatically resets upon cooling. • If it is suspected that fluid may have entered the pump, either by an external spill or by a jar overflow, the unit must be taken out of service until the fault has been fixed and its continued electrical safety has been confirmed. The service department must also be notified of any potential biohazard.
Spares MUL 94003 702
Fuse T5.0 Amp 250V
MUL 94004 000
Mains Power Switch Green Rocker
SUC 80297 001
Kit Yellow Suction Tubing (20 Metre roll)
SUC 80301
2 litre Collection Jar for Suction Pumps
SUC 80404 001
Single Use Plastic Handpiece (pack of 10)
SUC 84100 162
Blue Bacteria Filter (Set of 12)
SUC 84200 033
Bung Assembly with Overflow Cut Off Valve for 2 litre Collection Jar
SUC 91005 415
User Manual for the HiFlo High Suction Pump
Spares
13
CLEMENTS Spares SUC 37043
Cotton Wool for Suction Pumps Exhaust Filter (Pack of 10, 5 grams each)
SUC 80330
High Vacuum Gauge 0-100kPa
SUC 84100 049
Exhaust Filter Assembly for Suction Pumps
SUC 84100121
Resilient Motor Mount (Pack of 4)
SUC 84100 400
Hydrophobic Shut-Off Valve Filter
SUC 84100 505
Footswitch Kit for Suction Pumps with (Switch, Pneumatic Actuator)
SUC 84109 002
Motor and Pump for Hi Vac/Hi Flo
SUC 84200 501
Seal for Collection Jar Bung (Set of 2)
SUC 84200 502
Suction Inlet Nipple for 2 litre Jar Bung (Set of 4)
SUC 84200 503
Kit Grey Bung Float Chamber (Set of 2)
SUC 84200 518
Washer for Overflow Valve (Pack of 4)
SUC 84200 519
Blue Male Nipple for Bacteria Filter (Pack of 10)
SUC 84200 520
Blue Female Nipple for Bacteria Filter (Pack of 10)
SUC 89107
Kit for Pump Vacuum Controller (Needle Valve Type)
SUC 89240 069
Regulator Knob Replacement Kit for Suction Pumps (Regulator Knob, Screw)
14
Spares
CLEMENTS
Maintenance IMPORTANT There are no user-serviceable components inside. Maintenance must be carried out by qualified personnel only. To keep your Clements HiFlo High Suction Pump in good condition, the following maintenance is recommended. After Each Operation • Blue Bacterial Filter The blue bacterial filter is an important device to help protect the pump from moisture and bacteria taken from patient airways. Clements recommend that the filter be changed every month or when the unit is used on infectious patients or when damp or discoloured. • Housing Wipe clean with a damp soapy cloth. Do not immerse or allow liquid to enter the housing. • Suction Tubing Suction Tubing should be sterilised with sterilants determined by local protocols and compatible with PVC tubing. Suction Tubing may be autoclaved to a maximum of 121°C for 20 minutes. At higher temperatures the Suction Tubing will discolour and lose shape. • Collection Jars Place jars upright or up-side-down in autoclave: not on their sides. Autoclave to a maximum of 136°C for 10 minutes. Do not use phenolic solutions as disinfecting agents in polycarbonate jars. • Bung Disconnect float cage and float from Bung. Place all items upright or up-side-down in autoclave: not on their sides. Autoclave all items to a maximum of 136°C for 10 minutes. Note: Do not use abrasive cleaning agents as this may damage plastic surfaces and reduce the expected performance of an item. Maintenance
15
CLEMENTS
After Every 100 Hours or 2 Months of Operation • Replace the Exhaust Filter: 1. Remove the Exhaust Filter from the rear of the unit. 2. Unscrew the caps and replace the cotton wool. The filter takes 5 gram of bleached and teased-out cotton wool. • Check all Suction Tubing and replace if it is perished, soft or discoloured. Also check the tubing nipples. • Check the seal rings on the Bungs and replace them if they are hard, cracked or perished. Check the fit of the Bung in the Collection Jar. • Check the Overflow Cut Off Valve seals and replace them if they are perished or damaged.
IMPORTANT The preceding maintenance should also be carried out immediately after use with infectious patients.
Disposal and Recycling Waste Materials The contents of the collection jars, suction tubing, bacteria filter, internal hydrophobic shut-off filter, or the cotton wool in the exhaust filter may contain biohazard wastes. Handle using safe handling procedures, which may include the use of rubber gloves and eye protection, and dispose of according to local protocols for biohazard materials.
Recycling At the end of their service life, the pump and accessories should be dismantled if necessary, and disposed of according to the WEEE directive. 16
Maintenance
CLEMENTS
Troubleshooting Fault
Check
Rectify
No Suction But
Tubing Blocked
Clear or Replace
Motor is Working
Overflow Cut Off Valve Clear, Replace or Closed Empty Jar Hydrophobic Shut –Off Replace HyValve Filter drophobic Shut-Off Valve Filter Blocked Inlet Blocked Vacuum Gauge fective
Clear De-
Replace Vacuum Gauge
Regulator Blocked
Clear or Replace
Bung Not Sealing
Adjust or Fit New Seal
Regulator at Zero
Adjust Setting
No Suction:
Mains Fuse Blown
Replace
Motor Not Working
Faulty Mains Power Lead
Replace
Power Not Turned On
Turn On
Unit Fuses Blown
Replace
Thermal Overload Acti- Wait For Unit To Cool vated Down No Suction But Humming Can Be Heard From Motor
Blockage In Pump
Dismantle And Clean Pump
Motor Windings
Replace Motor
Troubleshooting
17
CLEMENTS Fault
Check
Rectify
Not Enough Suction
Regulator Setting Too Adjust To Correct SetLow ting Leaks In Tubing or Fittings
Tighten Connections
Blockage In Tubing or Fittings
Clear or Replace
Hydrophobic ShutOff Valve Filter Blocked
Replace Hydrophobic Shut-Off Valve Filter
Excessive Noise
Pump Worn
Replace Pump
Coming From Pump
Pump Incorrectly Assembled
Assemble Correctly
Pump Runs termittently
Thermal Overload Is Cutting In and Out
Disconnect Mains Power And Wait For Unit To Cool Down
In-
Replacing the Hydrophobic Shut Off Valve Filter Transparent Filter Housing
To Patient To Pump Filter Media To ensure there is no entry of fluid into the pump the Hydrophobic Shut Off Valve Filter MUST be fitted as shown above.
18
Troubleshooting
CLEMENTS Setup for Diagnostic and Performance Testing Note that the following arrangements are not used for actual suctioning applications. They are specified to remove unnecessary variations when diagnosing faults and as a standard setup for performance measurement.
Vacuum Check Max
1. Unoccluded - zero check
Vacuum Gauge
Disconnect all items from inlet and with pump switched off, confirm that gauge reads zero. A non-zero reading indicates a faulty gauge. Set vacuum control knob to maximum, switch on pump and confirm that gauge reads zero. A non-zero reading indicates obstruction in internal tubing or connections.
Max
2. Occluded - vacuum check
Vacuum Gauge
Switch on pump and occlude inlet. Note maximum vacuum reading.
Max
3. Occluded - gauge check Calibrated Vacuum Gauge
Vacuum Gauge
Connect a calibrated vacuum gauge directly to inlet and repeat maximum vacuum reading. Confirm that pump gauge reads within the specified tolerance. Diagnostic Setup
19
CLEMENTS Flow Check 1. Occluded - leak check
Max
Jar: SUC 81099 (1 litre) or SUC 80312 (2 litre) Tubing: SUC 80297 (8x13mm)
Vacuum Gauge
Connect pump as shown in diagram. Set vacuum control knob to maximum, and switch on pump. Occlude jar inlet and confirm that pump achieves the same maximum vacuum as in the previous vacuum check setup . Any difference indicates leaks in jar or connections.
2. Unoccluded - flow check
Max Jar: SUC 81099 (1 litre) or SUC 80312 (2 litre) Tubing: SUC 80297 (8x13mm)
Vacuum Gauge Calibrated Flowmeter
Collection Jar
Connect pump as shown in diagram. Set vacuum control knob to maximum, and switch on pump. Connect jar inlet to calibrated flow meter and note flow reading. If flow is significantly below specification, check internal tubing and pump itself (or internal shutoff valve if fitted).
20
Diagnostic Setup