SERIES-O REGULATOR
Instructions for use 522460-05-708
3807 Series O manual
8/9/08
11:04 AM
Contents
Page 2
Page
1. INTRODUCTION
2
2. INTENDED USES
3
3. GENERAL DESCRIPTION
4
4. PARTS IDENTIFICATION
5
5. SAFETY PRECAUTIONS
6
6. EQUIPMENT USE
8
7. CHECKS AND INSPECTIONS BY THE USER
11
8. SERVICE RECOMMENDATIONS
13
9. ACCESSORIES AND REPLACEMENT PARTS
14
10.SPECIFICATIONS
15
11.WARRANTY
16
SERIES-O Regulator
2
1. INTRODUCTION These instructions are intended to provide users with information on the features, use and care of the SERIES-O medical gas regulator. The instructions should be read carefully by every intended user of a SERIES-O regulator in any of its applications before attempting to use it to ensure that they are fully conversant with all the information on the features, use and care of the regulator.
Warning Read and understand the entire Instructions for Use before operating this medical device. Keep them always with the device. Failure to do so may cause clinical damage to a patient or damage to the device.
522460-05-708
3807 Series O manual
8/9/08
11:04 AM
Page 3
2. INTENDED USES
This may require clinical or medical advice.
The SERIES-O regulator is intended to be connected to high pressure medical gas cylinders in order to supply medical gases at reduced and regulated pressures similar to those from medical gas pipeline systems. It should be used only in accordance with these Instructions, and by persons instructed and familiar with basic handling of medical gases in gas cylinders.
The SERIES-O regulator may be used as a source of pressure regulated medical gas for a demand valve. However, assembly of a complete Resuscitator or Analgesic Administration System requires consideration of all other components to be used, which together with their use for the intended application should be undertaken only by persons who have received adequate training. These Instructions for Use cover only the regulator.
The gas supply from the regulator is delivered by an SIS compatible pressure outlet at reduced and regulated pressure similar to those from medical gas pipeline systems. The regulated pressure gas is supplied to medical devices intended for typical applications like Respiratory Therapy, Resuscitation and Suction.
Details of the performance of each SERIES-O model is given in Section 10 Specifications.
The regulator can only perform its pressure regulating and gas supply functions when properly connected to a downstream application. The downstream application is expected to control the supply of gas to the patient or ultimate application. The regulator alone cannot perform these functions. Use of the SERIES-O regulators should always be undertaken in accordance with the instructions for use of the medical device which will be connected to its outlet connection to deliver care to a patient.
522460-05-708
3
SERIES-O Regulator
3807 Series O manual
8/9/08
11:04 AM
Page 4
3. GENERAL DESCRIPTION
A pressure relief valve is fitted to the outlet pressure side of the regulator to give protection against most seat failures.
The SERIES-O regulator is a reliable, compact, preset single stage regulator. Pressure reduction is effected by means of an encapsulated pressure regulating valve, protected by a unique triple filtration system in the high pressure side to protect the regulator and downstream equipment from entry of foreign particles.
Gas specific inlet and outlet connections allow the regulator to be connected only to the gas for which it is intended. Colour coded pressure gauges and identification labels help for quick identification of the gas for which the regulator is intended.
The SERIES-O regulator does not require user pressure adjustment. It has been factory preset to deliver the outlet pressure specified in AS 2896 or applicable Standard. Since it is always adjusted and ready to deliver outlet pressure and does not include a shut-off valve, it is essential that it should be connected to the high pressure gas cylinder with careful attention to the operation of the valve in the gas cylinder and only when attached downstream devices are fitted with an outlet valve. A pressure gauge is fitted in the high pressure side of the regulator in order to allow indication of the gas cylinder contents. The gauge is of the safety back design to allow gas to escape away from the operator ensuring maximum safety in the unlikely event of a failure occurring. A solid metal case also provides protection to the internal mechanism.
SERIES-O Regulator
4
522460-05-708
3807 Series O manual
8/9/08
11:04 AM
Page 5
4. PARTS IDENTIFICATION Contents Gauge 4.1 Contents of Package: Item 1 2 3
Description Quantity SERIES-O regulator assembly 1 Instructions for Use 1 Outlet connector Warning label 1 (attached to the regulator)
Inlet Handwheel Inlet O-ring Inlet Filter & Retainer Pressure Relief Valve
Note: Instructions for Use and Warning label should always be kept with the regulator.
Outlet Connector Warning Label
4.2 Pictorial Description: Contents Gauge Notes: Apart from the Inlet Handwheel or Yoke Wing Screw, there are no user controls or shut-off valves on the SERIES-O Regulator. Flow control and shut-off may be done only through the gas cylinder valve or connected downstream equipment.
Bodok Seal Inlet Filter & Holder
The cylinder contents pressure gauge only provides indication of the gas cylinder pressure.
522460-05-708
Yoke Wing Screw Outlet Connector Warning Label 5
Pressure Relief Valve SERIES-O Regulator
3807 Series O manual
8/9/08
11:04 AM
Page 6
5. SAFETY PRECAUTIONS
When using the SERIES-O regulator the following precautions should be observed: - Never allow any source of ignition or naked flame near the equipment or patient during oxygen administration and for a reasonable period afterwards. - Use no oil or greases as a fire or an explosion may result. - When connected to a pressure regulator fitted to a gas cylinder, always open the cylinder valve SLOWLY. Rapid opening of the cylinder valve may damage downstream equipment. - Ensure that the cylinder valve is not opened when there is no equipment connected downstream of the regulator. This may cause damage to the regulator. - When removing downstream equipment, ensure first the cylinder valve is fully closed. Failure to do so may damage the regulator. - Turn off the cylinder valve when the regulator is not in use. - Keep gas cylinders cool. - Do not dump or drop cylinders.
WARNING The SERIES-O regulator should be used only if all its components are in correct operating condition as described in these instructions. Use the regulator only for the purpose for which it is intended (Section 2) and in accordance with clinical or medical advice. Failure to do so may result in clinical damage to a patient. Most models of the SERIES-O regulator range are intended to be connected to high pressure gas cylinders containing oxygen or oxidising gases. Materials that react normally in contact with air may react violently and ignite in an oxygenenriched atmosphere. Regulators connected to high pressure gas cylinders must be handled with care and in accordance with the recommendations from the gas supplier.
SERIES-O Regulator
6
522460-05-708
3807 Series O manual
8/9/08
11:04 AM
Page 7
Use of the regulator with medical gas cylinders Only use the SERIES-O regulator with the medical gas for which it is intended and labelled. Ensure it is connected to the correct gas cylinder by reading the cylinder label which is the primary identification of its gas. Consult safe gas cylinder handling procedures and guidelines from the medical gas suppliers to plan location and access to the gas cylinder and the connected regulator.
When the regulator is not in use, the cylinder valve should be turned off and the regulator depressurised. Care of the regulator Since SERIES-O Regulators are always preset to deliver outlet pressure and have no means to control or shut-off gas flow, they should be pressurised only when downstream equipment fitted with a control valve are attached to them.
The SERIES-O regulator has been primarily designed for use connected to a static and properly secured medical gas cylinder. Ensure the regulator is well protected at all times against accidental contact with other objects. Impact on the regulator could cause malfunction and unintended gas release. Never use the regulator as a means of lifting, carrying and steadying the cylinder.
Operation of the pressure relief valve indicates malfunction in the pressure regulating mechanism. The cylinder valve should be closed immediately and the regulator should be removed from use and subjected to servicing. Do not tamper with the setting of the pressure relief valve.
When the regulator is moved around whilst connected to a gas cylinder, the cylinder valve should always be closed first.
All repairs should be undertaken only by personnel trained in the servicing procedures for medical equipment.
For applications where a mobile supply of medical oxygen is required, e.g. during patient transportation, a safe and suitable arrangement for securing the gas cylinder and pressure regulator is required. Ensure the regulator is adequately guarded and protected against impact and knocks during transport.
Use only manufacturer approved spare parts when servicing.
522460-05-708
7
SERIES-O Regulator
3807 Series O manual
8/9/08
11:04 AM
Page 8
6. EQUIPMENT USE
Remove any protective capping from the cylinder valve and check that the cylinder valve outlet is in good condition and matches the regulator inlet connection.
6.1 Installation Connecting the Regulator to a Gas Cylinder For a new regulator, remove the regulator from its packing, checking there has been no damage in transport. Remove the plastic protection caps from the inlet and outlet connections (where fitted).
Briefly open and close the cylinder valve to remove any foreign matter. Check that the Bodok seal (Yoke inlet models) or inlet O-ring (screw connection models) are in a good condition and replace them as required. (see Section 7).
In all cases, perform the recommended Checks and Inspections before use of the regulator (see Section 7).
CAUTION
To connect a regulator to a gas cylinder, first verify that the gas cylinder label indicates the correct medical gas for the regulator and the intended application.
Refer to safe gas cylinder handling procedures and guidelines from the medical gas suppliers as required and appropriate. Connecting equipment to gas cylinders is different to connecting equipment to Terminal Units from pipeline systems. Connect the regulator to the cylinder valve. If the regulator has a pin index yoke connection align the yoke pins with the matching holes in the cylinder valve. Secure by tightening the yoke wing screw until a leak tight seal is achieved.
WARNING Supply of incorrect gas to a patient may result in clinical damage including death.
If the regulator has an inlet handwheel simply tighten the inlet connection to the cylinder valve until a leak tight seal is achieved.
SERIES-O Regulator
8
522460-05-708
3807 Series O manual
8/9/08
11:04 AM
Page 9
To connect downstream equipment (e.g. Demand valves, Flowmeters, Suction units) to a SERIES-O regulator, first ensure that the downstream equipment is turned off. The regulator is adjusted always to deliver flow at its reated pressure of 400 kPa through the SIS outlet (or other value given in Section 10 - Specifications) and has no means of adjusting or closing it off. Connect the inlet of any downstream equipment to the regulator outlet connection and tighten. Only then slowly and fully open the cylinder valve.
WARNING -
Any difficulty in connecting the regulator indicates a possible fault or mismatch in the connections which should be investigated and repaired before use of the regulator.
-
Do not open the cylinder valve for any purpose until downstream equipment is connected to the regulator outlet. The regulator alone cannot control flow and a large discharge of gas will result, possibly damaging the regulator.
Check the regulator contents pressure gauge to confirm there is sufficient gas in the cylinder for the intended use. Check all connections for leaks. If necessary, re-tighten connections until a leak tight seal is achieved. Use the downstream equipment as required and in accordance with its own Instructions for Use, and if necessary, under clinical or medical supervision.
6.2 Connecting Downstream Equipment
WARNING
When the downstream equipment is no longer in use, close the cylinder valve and depressurise the regulator.
Use only downstream equipment that is compatible with the intended gas and suitable to operate at the rated outlet pressure of the regulator. The downstream equipment should be fitted with the correct SIS inlet connection matching the regulator outlet connection.
522460-05-708
9
SERIES-O Regulator
3807 Series O manual
8/9/08
11:04 AM
Page 10
6.3 Removing downstream equipment
should be used. Care should be taken to ensure that none of the cleaning solution enters the gas passages of the regulator. Do not attempt to clean the inside of the regulator.
CAUTION When removing downstream equipment, ensure first the cylinder valve is fully closed. An open cylinder valve will result in release of gas and possible damage to the regulator.
CAUTION If internal contamination is apparent, the regulator should be removed from use and inspected by trained personnel.
To remove downstream equipment from the regulator, first close the cylinder valve. Depressurise the downstream equipment and the regulator by operating the downstream equipment until the regulator contents gauge reads zero pressure. Remove the downstream equipment.
6.6 Storage after use If the regulator is not going to be used immediately following removal from a cylinder, it should be stored in a clean, dry, secure place to prevent accidental knocks and impacts and to prevent the ingress of dust or contamination to its gas passages.
6.4 Removing the Regulator from a Cylinder To remove the regulator from the cylinder, first close the cylinder valve. Disconnect all the downstream equipment as indicated above in 6.3. With the downstream equipment now disconnected and the regulator depressurised, remove the regulator from the cylinder valve by unscrewing the yoke wing screw or inlet handwheel. 6.5 Cleaning after use A SERIES-O regulator should not require regular cleaning after use. If the exterior of the regulator needs cleaning, a mild soap solution SERIES-O Regulator
10
522460-05-708
3807 Series O manual
8/9/08
11:04 AM
Page 11
7. CHECKS AND INSPECTION BY THE USER 7.1 Operational Check and Inspection A medical pressure regulator is a critical link in the gas supply to a patient. The following checks should be performed each time the SERIES-O regulator is connected to a gas cylinder prior to using the equipment to supply gas to a downstream application and/or patient. - Check the condition of the inlet and outlet gas specific connectors. The connection fittings should be tight and secure. - Check that the Bodok seal (Yoke inlet models) or inlet Oring (screw connection models) is in a good condition. - Check the pressure gauge to ensure that the pointer reads zero when there is no inlet pressure present. - Check for any damage that may have occurred to the regulator. - Check that units subject to Preventative Maintenance programs are within test date. (Refer to Section 8 Service Recommendations).
CAUTION
If the regulator satisfies the above requirements, it may be used.
A damaged O-ring may be replaced, when a correct replacement is available, by removing it with a pointed object, taking care not to damage the inlet nipple, and inserting the replacement O-ring in place.
522460-05-708
If the unit fails any part of this checkout procedure, or any unusual conditions are evident or suspected, the regulator must be removed from service and repaired before use to supply gas to a patient. 7.2 Checking and replacing the Bodok seal The Bodok seal in the yoke inlet should have a clean, undamaged appearance, especially the black elastomer sealing element. If any amount of excessive deformation, cuts or nicks are evident, the seal should be replaced. To replace the Bodok seal, gently lever it out of position. A new seal should be inserted simply by pushing it into position. 7.3 Checking and replacing the inlet connection O-ring The O-ring on the inlet connection should be in place and in good condition (i.e. no cuts, deformation or dirt).
11
SERIES-O Regulator
3807 Series O manual
8/9/08
11:04 AM
Page 12
7.4 Replacing the inlet filter To replace the inlet filter, first remove the filter retaining screw using a correct size Allen key (3.2 mm - 1/8” for yoke inlet models, 4.8 mm - 3/16” for screw connection models). The filter should drop out of the inlet connector when the regulator is inverted. After fitting a new inlet filter its retaining screw should only be lightly tightened (“nipped up”). Do not over tighten.
equipment and should only be carried out by trained service personnel. See Section 8. Pressure Relief Valve Setting The operation of the pressure relief valve should be periodically checked. This requires special test equipment and should only be carried out by trained service personnel. Contact your nearest authorised service agent for further information.
7.5 Regulator testing The regulator should be periodically tested for leaks, outlet set pressure and pressure relief valve setting. The recommended frequency of these tests is detailed in section 8 of this manual. Leak Test To test for leaks fit the regulator to a full gas cylinder and block off the regulator outlet. Open the cylinder valve and apply an approved leak detection solution to all regulator body connections (inlet, outlet, pressure gauge, pressure relief valve and bonnet). No bubbles should be generated within 30 seconds. Approved leak detection solutions are BUBBLES™ or SNOOPY™. Outlet Set Pressure and regulator seat performance The outlet set pressure of the regulator should be periodically checked, and adjusted if required. This requires special test SERIES-O Regulator
12
522460-05-708
3807 Series O manual
8/9/08
11:04 AM
Page 13
8. SERVICE RECOMMENDATIONS A SERIES-O regulator is a critical link in the supply of medical gas to a patient. To ensure that the regulator is always in a safe, reliable, useable condition it is recommended that the unit be placed on a preventative maintenance program.
Fixed Every Three Years Servicing - Fixed Yearly servicing - Pressure relief valve check - Regulator seat capsule replacement. All repairs or service to your SERIES-O regulator should be performed by trained service personnel only. Contact your nearest Comweld Medical Products distributor for further information.
The program should include as a minimum fixed yearly and every three years servicing. Depending on usage, an additional more frequent service (e.g. six monthly, quarterly or monthly) is recommended. Frequent Inspection program: - Operational Check (Section 7.1) - Leak test (Section 7.5) - Outlet Set Pressure and regulator seat performance (Section 7.5) Fixed Yearly servicing: - Frequent Inspection - Inlet filter, Bodok seal and inlet O-ring replacement (Section 7.2, 7.3 and 7.4)
522460-05-708
13
SERIES-O Regulator
3807 Series O manual
8/9/08
11:04 AM
Page 14
9.2 User Replacement Parts Listed below are the items considered to be user replaceable parts. It is recommended that the owner keep stock of items such as Bodok seals to ensure that the SERIES-O regulator is always ready for use.
9. SPARES AND ACCESSORIES 9.1 SERIES-O Accessories Listed below are some of the accessories suitable for use with a SERIES-O regulator. It is recommended that if any other accessories are intended to be used with a SERIES-O regulator, your nearest authorised distributor should be contacted to check the compatibility of the accessory. Description EZI-FLOW 0-15 L/min oxygen flowmeter EZI-FLOW 0-2.5 L/min oxygen flowmeter EZI-FLOW 0-15 L/min air flowmeter Flowmeter Gauge type 0-14 L/min oxygen Humidifier Twin-O-Vac Oxygen Demand Valve Resuscitator Equinox Demand Valve with 2m hose
Description Bodok seal, 10 pack Inlet filter RB38, 10 pack Inlet filter/retainer RB38/RB39, 10 pack (Screw connections) Inlet filter RB38 yoke retainer, 10 pack (Yoke inlets) Inlet O-ring 110 Neoprene, 5 pack (Screw connections) Instructions for Use
Part No 515800 515824 515820 TM17 TM11 TM117G 552095 550011
Part No 551022 551024 551026 551028 301073 522460
CAUTION Use the downstream equipment as required and in accordance with its own Instructions for Use, and if necessary, under clinical or medical supervision.
SERIES-O Regulator
14
522460-05-708
3807 Series O manual
8/9/08
11:04 AM
Page 15
10. SPECIFICATIONS 10.1 Model description Part No.
Gas
Inlet Connection
Rated Inlet Pressure, P1
Outlet Pressure, P2
Standard Discharge, Q1
518800
Oxygen
AS 2473.2 Type 10 (G5/8 RH INT - Side)
20 000 kPa @ 15˚C
400 kPa
125 l/min
518804
Oxygen
AS 2473.3 (Pin Indexed Yoke)
20 000 kPa @ 15˚C
400 kPa
125 l/min
518808
Nitrous Oxide
AS 2473.3 (Pin Indexed Yoke)
21 000 kPa
400 kPa
125 l/min
518809
Air
AS 2473.3 (Pin Indexed Yoke)
20 000 kPa @ 15˚C
400 kPa
125 l/min
518810
Nitrous oxide/Oxygen 50/50 nom.
AS 2473.3 (Pin Indexed Yoke)
20 000 kPa @ 15˚C
400 kPa
125 l/min
518803
Oxygen - (Oxy-Viva)
N/A
20 000 kPa @ 15˚C
400 kPa
125 l/min
518895
Oxygen
AS 2473.2 Type 10 (G5/8 RH INT - Vert)
20 000 kPa @ 15˚C
400 kPa
125 l/min
Notes: AS 2473.3 (Pin Indexed Yoke) inlet connections for the nominated gas AS 2473.2 Type 10 inlet connection for Oxygen (Medical Oxygen cylinders currently fitted with valves with this connection will be converted to pin-indexed yoke connection over a period of time - see AS 2473.3). SIS - outlet connection compatible with Sleeve Indexed System from AS 2896 for the nominated gas
522460-05-708
15
SERIES-O Regulator
3807 Series O manual
8/9/08
11:04 AM
Page 16
10.2 Technical specifications Flow Capacity:
SERIES-O Regulator
Pressure Rise:
16
522460-05-708
3807 Series O manual
8/9/08
11:04 AM
Page 17
Pressure Relief Valve: All models are fitted with a pressure relief valve designed to vent a flow equivalent to the Standard discharge Q1 at a pressure twice the nominal outlet pressure P2 of the regulator (see AS 3840.1). Filters: All SERIES-O regulators are fitted with three inlet filters: - a sintered filter in the inlet connection, nominal particle retention 50 micron - a fibre strand filter upstream of the seat capsule, - a sintered filter fitted to the seat capsule, nominal particle retention 34 micron.
522460-05-708
17
SERIES-O Regulator
3807 Series O manual
8/9/08
11:04 AM
Page 18
11. WARRANTY CIGWELD Pty. Ltd. warrants to the purchaser that this equipment is free from defects in material and workmanship for a period of three years from the date of purchase (conditions apply).
SERIES-O Regulator
18
522460-05-708
3807 Series O manual
522460-05-708
8/9/08
11:04 AM
Page 19
19
SERIES-O Regulator
3807 Series O manual
8/9/08
11:04 AM
Page 20
CIGWELD Pty Ltd (ABN 56 007 226 815) A Division of Thermadyne Corporation, 71 Gower Street, Preston VIC 3072 Australia
Customer Service Centre and Technical Support Ph: +61 3 9474 7314 Fax: +61 3 9474 7391
In the interest of continuous improvement, CIGWELD Pty Ltd ABN 56 007 226 815 reserve the right to change the specifications or design on any of its products without prior notice.