Addendum for DINAMAP PRO 100-400V2 Operation and Service Manuals Rev A

Addendum

8 Pages

Addendum for DINAMAP® PRO 100-400V2 Operation and Service Manuals  Addendum for DINAMAP® PRO 100-400V2 Operation and Service Manuals This addendum contains updated information for the DINAMAP® PRO 100-400V2 Monitor and the accompanying Operation and Service Manuals.  Technical Specifications BP NOTE: This change effects page 2-13 of the PRO 100-400V2 Service Manual and page 77 of the PRO 100-400V2 Operation Manual.  Cuff Pressure Range (Normal operating range)  0 to 290 mmHg (adult/ped) 0 to 145 mmHg (neonate)  Default Target: Cuff Inflation 150 ± 15 mmHg (adult/ped) 110 ± 15 mmHg (neonate) Target Cuff Inflation: Adjustment Range (in 5 mmHg increments)  100 to 250 mmHg (adult/ped) 100 to 140 mmHg (neonate)  Blood Pressure Measurement Range (mmHg) Systolic MAP Diastolic Adult 30 - 290 20 - 260 10 -220 Neonate 30 - 140 20 - 125 10 -110  Revision A  Blood Pressure Accuracy  Meets or exceeds ANSI/AAMI standard SP-10 (mean error < 5 mmHg, standard deviation < 8 mmHg)  Maximum Determination Time  120 s (adult/ped) 85 s (neonate)  Overpressure Cutoff  300 to 330 mmHg (adult/ped) 150 to 165 mmHg (neonate)  Pulse Rate Range  30 to 200 beats/min (adult/ped) 30 to 220 beats/min (neonate)  Pulse Rate Accuracy  ± 3.5%  DINAMAP® PRO Series 100-400V2 Addendum 2016959-001  1